Published in: Intensive Care Medicine 12/2003

01-12-2003 | Neonatal And Pediatric Intensive Care

Comparison of the bispectral index monitor with the Comfort score in assessing level of sedation of critically ill children

Authors: Simon P. Courtman, Allan Wardurgh, Andy J. Petros

Published in: Intensive Care Medicine | Issue 12/2003

Login to get access

Abstract

Objective

To compare the value of bispectral index as a monitor of sedation in critically ill children with a validated sedation scoring system.

Design

Prospective convenience sample.

Setting

Paediatric intensive care unit in a tertiary paediatric centre.

Patients and participants

Forty-three critically ill children receiving sedation and mechanical ventilation.

Measurements and results

Simultaneous recording of bispectral index (BIS) and assessment of depth of sedation using the Comfort score were performed at regular intervals. To determine if BIS could detect episodes of arousal, times of endotracheal suctioning and the corresponding BIS score were recorded. There was an overall moderate correlation between BIS scores and Comfort scores (r=0.50, r 2=0.25, p<0.0001). Children who had a neurological reason for their current admission (n=25) showed a weaker correlation (r=0.26, r 2=0.06, p<0.007) than those (n=15) with normal neurology (r=0.51, r 2=0.26, p<0.0001). There were no significant differences in the rise in BIS following endotracheal suctioning among any of the predefined depths of sedation. There was a correlation of r=0.84 (r 2=0.71) (SE of slope 0.49, CI95 1.79–3.88) for mean BIS values for each individual Comfort score from 8–23. Using Spearman's rank correlation of Comfort versus mean BIS, the correlation coefficient was r=0.92.

Conclusions

Bispectral index scores correlate with Comfort scores to a moderate degree. BIS is able to discriminate between light and deep levels of sedation, but not between deep and very deep levels of sedation. The BIS monitor may provide a useful method for assessing sedation in critically ill children, especially those receiving neuromuscular blockers.
Appendix
Available only for authorised users
Literature
1.
go back to reference Playfor SD, Thomas DA, Choonara II (2000) Recall following paediatric intensive care. Paediatr Anaesth 10:703–704PubMed Playfor SD, Thomas DA, Choonara II (2000) Recall following paediatric intensive care. Paediatr Anaesth 10:703–704PubMed
2.
go back to reference Kollef M, Levy NT, Ahrens TS, Schaiff R, Prentice D, Sherman G (1998) The use of continuous i.v. sedation is associated with prolongation of mechanical ventilation. Chest 114:541–548PubMed Kollef M, Levy NT, Ahrens TS, Schaiff R, Prentice D, Sherman G (1998) The use of continuous i.v. sedation is associated with prolongation of mechanical ventilation. Chest 114:541–548PubMed
3.
go back to reference Cook D, Walter SD, Cook RJ, Griffith LE, Guyatt GH, Leasa D, Jaeschke RZ, Brun-Buisson C (1998) Incidence of risk factors for ventilator-associated pneumonia in critically ill patients. Ann Intern Med 129:433–440PubMed Cook D, Walter SD, Cook RJ, Griffith LE, Guyatt GH, Leasa D, Jaeschke RZ, Brun-Buisson C (1998) Incidence of risk factors for ventilator-associated pneumonia in critically ill patients. Ann Intern Med 129:433–440PubMed
4.
go back to reference Cammarano W, Pittet JF, Weitz S, Schlobohm RM, Marks JD (1998) Acute withdrawal syndrome related to the administration of analgesic and sedative medications in adult intensive care units. Crit Care Med 26:676–684PubMed Cammarano W, Pittet JF, Weitz S, Schlobohm RM, Marks JD (1998) Acute withdrawal syndrome related to the administration of analgesic and sedative medications in adult intensive care units. Crit Care Med 26:676–684PubMed
5.
go back to reference Tobias JD (2000) Tolerance, withdrawal and physical dependency after long-term sedation and analgesia of children in the pediatric intensive care unit. Crit Care Med 28:2122–2132PubMed Tobias JD (2000) Tolerance, withdrawal and physical dependency after long-term sedation and analgesia of children in the pediatric intensive care unit. Crit Care Med 28:2122–2132PubMed
6.
go back to reference Anand KJ, Sippell WG, Aynsley-Green A (1987) Pain, anaesthesia and babies. Lancet 2:1210–1212PubMed Anand KJ, Sippell WG, Aynsley-Green A (1987) Pain, anaesthesia and babies. Lancet 2:1210–1212PubMed
7.
go back to reference Kress JP, Pohlman AS, Hall JB (2002) Sedation and analgesia in the intensive care unit. Am J Respir Crit Care Med 166:1024–1028PubMed Kress JP, Pohlman AS, Hall JB (2002) Sedation and analgesia in the intensive care unit. Am J Respir Crit Care Med 166:1024–1028PubMed
8.
go back to reference Playfor SD, Thomas DA, Choonara I (2003) Sedation and neuromuscular blockade in paediatric intensive care: a review of current practice in the UK. Paediatr Anaesth 13:147–151PubMed Playfor SD, Thomas DA, Choonara I (2003) Sedation and neuromuscular blockade in paediatric intensive care: a review of current practice in the UK. Paediatr Anaesth 13:147–151PubMed
9.
go back to reference Bion J, Ledingham IM (1987) Sedation in intensive care—a postal survey. Intensive Care Med 13:215–216PubMed Bion J, Ledingham IM (1987) Sedation in intensive care—a postal survey. Intensive Care Med 13:215–216PubMed
10.
go back to reference De Jonghe B, Cook D, Appere-de-Vecchi C, Guyatt G, Meade M, Outin H (2000) Using and understanding sedation scoring systems: a systematic review. Intensive Care Med 26:275–285 De Jonghe B, Cook D, Appere-de-Vecchi C, Guyatt G, Meade M, Outin H (2000) Using and understanding sedation scoring systems: a systematic review. Intensive Care Med 26:275–285
11.
go back to reference Ambuel B, Hamlett KW, Marx CM, Blumer JL (1992) Assessing distress in pediatric intensive care environments: the COMFORT scale. J Pediatr Psychol 17:95–109PubMed Ambuel B, Hamlett KW, Marx CM, Blumer JL (1992) Assessing distress in pediatric intensive care environments: the COMFORT scale. J Pediatr Psychol 17:95–109PubMed
12.
go back to reference Marx CM, Smith PG, Lowrie LH, Hamlett KW, Ambuel B, Yamashita TS, Blumer JL (1994) Optimal sedation of mechanically ventilated pediatric critical care patients. Crit Care Med 22:163–170PubMed Marx CM, Smith PG, Lowrie LH, Hamlett KW, Ambuel B, Yamashita TS, Blumer JL (1994) Optimal sedation of mechanically ventilated pediatric critical care patients. Crit Care Med 22:163–170PubMed
13.
go back to reference Denman WT, Swanson EL, Rosow D, Ezbicki K, Connors PD, Rosow CE (2000) Pediatric evaluation of the bispectral index (BIS) monitor and correlation of BIS with end-tidal sevoflurane concentration in infants and children. Anesth Analg 90:872–877PubMed Denman WT, Swanson EL, Rosow D, Ezbicki K, Connors PD, Rosow CE (2000) Pediatric evaluation of the bispectral index (BIS) monitor and correlation of BIS with end-tidal sevoflurane concentration in infants and children. Anesth Analg 90:872–877PubMed
14.
go back to reference Glass PS, Bloom M, Kearse L, Rosow C, Sebel P, Manberg P (1997) Bispectral analysis measures sedation and memory effects of propofol, midazolam, isoflurane and alfentanil in healthy volunteers. Anesthesiology 86:836–847PubMed Glass PS, Bloom M, Kearse L, Rosow C, Sebel P, Manberg P (1997) Bispectral analysis measures sedation and memory effects of propofol, midazolam, isoflurane and alfentanil in healthy volunteers. Anesthesiology 86:836–847PubMed
15.
go back to reference Suzuki M, Edmonds HL Jr, Tsueda K, Malkani AL, Roberts CS (1998) Effect of ketamine on bispectral index and levels of sedation. J Clin Monit Comput 14:373PubMed Suzuki M, Edmonds HL Jr, Tsueda K, Malkani AL, Roberts CS (1998) Effect of ketamine on bispectral index and levels of sedation. J Clin Monit Comput 14:373PubMed
16.
go back to reference Degoute CS, Macabeo C, Dubreuil C, Duclaux R, Banssillon V (2001) EEG bispectral index and hypnotic component of anaesthesia induced by sevoflurane: comparison between children and adults. Br J Anaesth 86:209–212PubMed Degoute CS, Macabeo C, Dubreuil C, Duclaux R, Banssillon V (2001) EEG bispectral index and hypnotic component of anaesthesia induced by sevoflurane: comparison between children and adults. Br J Anaesth 86:209–212PubMed
17.
go back to reference Liu J, Singh H, White PF (1997) Electroencephalographic bispectral index correlates with intraoperative recall and depth of propofol-induced sedation. Anesth Analg 84:185–189PubMed Liu J, Singh H, White PF (1997) Electroencephalographic bispectral index correlates with intraoperative recall and depth of propofol-induced sedation. Anesth Analg 84:185–189PubMed
18.
go back to reference Aneja R, Heard AM, Fletcher JE, Heard CM (2003) Sedation monitoring of children by the Bispectral Index in the pediatric intensive care unit. Pediatr Crit Care Med 4:60–64PubMed Aneja R, Heard AM, Fletcher JE, Heard CM (2003) Sedation monitoring of children by the Bispectral Index in the pediatric intensive care unit. Pediatr Crit Care Med 4:60–64PubMed
19.
go back to reference Bland JM, Altman DG (1995) Calculating correlation coefficients with repeated observations: Part 1—Correlation within subjects. BMJ 310:446PubMedPubMedCentral Bland JM, Altman DG (1995) Calculating correlation coefficients with repeated observations: Part 1—Correlation within subjects. BMJ 310:446PubMedPubMedCentral
20.
go back to reference Bland JM, Altman DG (1995) Calculating correlation coefficients with repeated observations: Part 2—Correlation between subjects. BMJ 310:633PubMedPubMedCentral Bland JM, Altman DG (1995) Calculating correlation coefficients with repeated observations: Part 2—Correlation between subjects. BMJ 310:633PubMedPubMedCentral
21.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
22.
go back to reference Playfor SD, Thomas DA, Choonara I, Collier J, Jarvis A (2001) Parental perceptions of comfort during mechanical ventilation. Paediatr Anaesth 11:99–103PubMed Playfor SD, Thomas DA, Choonara I, Collier J, Jarvis A (2001) Parental perceptions of comfort during mechanical ventilation. Paediatr Anaesth 11:99–103PubMed
23.
go back to reference Crain N, Slonim, A, Pollack MM (2002) Assessing sedation in the paediatric intensive care unit by using BIS and the COMFORT scale. Pedatric Crit Care Med 3:11–14 Crain N, Slonim, A, Pollack MM (2002) Assessing sedation in the paediatric intensive care unit by using BIS and the COMFORT scale. Pedatric Crit Care Med 3:11–14
24.
go back to reference Berkenbosch JW, Fichter CR, Tobias JD (2002) The correlation of the bispectral index monitor with clinical sedation scores during mechanical ventilation in the pediatric intensive care unit. Anesth Analg 94:506–511PubMed Berkenbosch JW, Fichter CR, Tobias JD (2002) The correlation of the bispectral index monitor with clinical sedation scores during mechanical ventilation in the pediatric intensive care unit. Anesth Analg 94:506–511PubMed
25.
go back to reference Witte H, Putsche P, Eiselt M, Hoffmann K, Arnold M, Jager H, Leistritz L (1998) Multimodal time-variant signal analysis of neonatal EEG burst patterns. Medinfo 9:1250–1254 Witte H, Putsche P, Eiselt M, Hoffmann K, Arnold M, Jager H, Leistritz L (1998) Multimodal time-variant signal analysis of neonatal EEG burst patterns. Medinfo 9:1250–1254
26.
go back to reference Eeg-Olofsson O (1980) Longitudinal developmental course of electrical activity of brain. Brain Dev 2:33–44PubMed Eeg-Olofsson O (1980) Longitudinal developmental course of electrical activity of brain. Brain Dev 2:33–44PubMed
27.
go back to reference Davidson AJ, McCann ME, Devavaram P, Auble SA, Sullivan LJ, Gillis JM, Laussen PC (2001) The differences in the bispectral index between infants and children during emergence from anesthesia after circumcision surgery. Anesth Analg 93:326–330PubMed Davidson AJ, McCann ME, Devavaram P, Auble SA, Sullivan LJ, Gillis JM, Laussen PC (2001) The differences in the bispectral index between infants and children during emergence from anesthesia after circumcision surgery. Anesth Analg 93:326–330PubMed
28.
go back to reference Scher M (1999) Pediatric neurophysiologic evaluation. In: Swaiman KF, Ashwal S (ed) Pediatric neurology. Mosby, St Louis, pp 142–181 Scher M (1999) Pediatric neurophysiologic evaluation. In: Swaiman KF, Ashwal S (ed) Pediatric neurology. Mosby, St Louis, pp 142–181
29.
go back to reference McCann ME, Bacsik J, Davidson A, Auble S, Sullivan L, Laussen P (2002) The correlation of bispectral index with end tidal sevoflurane concentration and haemodynamic parameters in preschoolers. Paediatr Anaesth 12:519–525PubMed McCann ME, Bacsik J, Davidson A, Auble S, Sullivan L, Laussen P (2002) The correlation of bispectral index with end tidal sevoflurane concentration and haemodynamic parameters in preschoolers. Paediatr Anaesth 12:519–525PubMed
30.
go back to reference Bannister CF, Brosius KK, Sigl JC, Meyer BJ, Sebel PS (2001) The effect of bispectral index monitoring on anesthetic use and recovery in children anesthetized with sevoflurane in nitrous oxide. Anesth Analg 92:877–881PubMed Bannister CF, Brosius KK, Sigl JC, Meyer BJ, Sebel PS (2001) The effect of bispectral index monitoring on anesthetic use and recovery in children anesthetized with sevoflurane in nitrous oxide. Anesth Analg 92:877–881PubMed
31.
go back to reference Chawathe MS, Hall JE, Mecklenburgh JS, Jones RM (2002) BIS-cautionary interpretation of data in children. Br J Anaesth 89:672P Chawathe MS, Hall JE, Mecklenburgh JS, Jones RM (2002) BIS-cautionary interpretation of data in children. Br J Anaesth 89:672P
32.
go back to reference Simmons LE, Riker RR, Prato BS, Fraser GL (1999) Assessing sedation during intensive care unit mechanical ventilation with the Bispectral Index and the Sedation-Agitation Scale. Crit Care Med 27:1499–1504PubMed Simmons LE, Riker RR, Prato BS, Fraser GL (1999) Assessing sedation during intensive care unit mechanical ventilation with the Bispectral Index and the Sedation-Agitation Scale. Crit Care Med 27:1499–1504PubMed
33.
go back to reference Nasraway SS Jr, Wu EC, Kelleher RM, Yasuda CM, Donnelly AM (2002) How reliable is the Bispectral Index in critically ill patients? A prospective, comparative, single-blinded observer study. Crit Care Med 30:1483–1487 Nasraway SS Jr, Wu EC, Kelleher RM, Yasuda CM, Donnelly AM (2002) How reliable is the Bispectral Index in critically ill patients? A prospective, comparative, single-blinded observer study. Crit Care Med 30:1483–1487
34.
go back to reference Bruhn J, Bouillon TW, Shafer SL (2002) Electromyographic activity falsely elevates the bispectral index. Anesthesiology 92:1485–1487 Bruhn J, Bouillon TW, Shafer SL (2002) Electromyographic activity falsely elevates the bispectral index. Anesthesiology 92:1485–1487
35.
go back to reference Riker RR, Fraser GL (2002) Sedation in the intensive care unit: refining the models and defining the questions. Crit Care Med 30:1661–1663PubMed Riker RR, Fraser GL (2002) Sedation in the intensive care unit: refining the models and defining the questions. Crit Care Med 30:1661–1663PubMed
36.
go back to reference Gallagher JD (1999) Pacer-induced artifact in the bispectral index during cardiac surgery. Anesthesiology 90:636PubMed Gallagher JD (1999) Pacer-induced artifact in the bispectral index during cardiac surgery. Anesthesiology 90:636PubMed
37.
go back to reference Guignard B, Chauvin M (2000) Bispectral index increases and decreases are not always signs of inadequate anesthesia. Anesthesiology 92:903PubMed Guignard B, Chauvin M (2000) Bispectral index increases and decreases are not always signs of inadequate anesthesia. Anesthesiology 92:903PubMed
38.
go back to reference Sakai T, Singh H, Mi WD, Kudo T, Matsuki A (1999) The effect of ketamine on clinical endpoints of hypnosis and EEG variables during propofol infusion. Acta Anaesthesiol Scand 43:212–216PubMed Sakai T, Singh H, Mi WD, Kudo T, Matsuki A (1999) The effect of ketamine on clinical endpoints of hypnosis and EEG variables during propofol infusion. Acta Anaesthesiol Scand 43:212–216PubMed
39.
go back to reference Andrzejowski J, Sleigh JW, Johnson IA, Sikiotis L (2000) The effect of intravenous epinephrine on the bispectral index and sedation. Anaesthesia 55:761–763PubMed Andrzejowski J, Sleigh JW, Johnson IA, Sikiotis L (2000) The effect of intravenous epinephrine on the bispectral index and sedation. Anaesthesia 55:761–763PubMed
40.
go back to reference Riess ML, Graefe UA, Goeters C, Van AH, Bone HG (2000) Sedation assessment in critically ill patients with bispectral index. Eur J Anaesthesiol 19:18–22 Riess ML, Graefe UA, Goeters C, Van AH, Bone HG (2000) Sedation assessment in critically ill patients with bispectral index. Eur J Anaesthesiol 19:18–22
41.
go back to reference Kim DW, Kil HY, White PF (2001) The effect of noise on the bispectral index during propofol sedation. Anesth Analg 93:1170–1173PubMed Kim DW, Kil HY, White PF (2001) The effect of noise on the bispectral index during propofol sedation. Anesth Analg 93:1170–1173PubMed
42.
go back to reference Iselin-Chaves I, Flaishon R, Sebel PS, Howell S, Gan TJ, Sigl J, Ginsberg B, Glass PS (1998) The effect of the interaction of propofol and alfentanil on recall, loss of consciousness and the bispectral index. Anesth Analg 87:949–955PubMed Iselin-Chaves I, Flaishon R, Sebel PS, Howell S, Gan TJ, Sigl J, Ginsberg B, Glass PS (1998) The effect of the interaction of propofol and alfentanil on recall, loss of consciousness and the bispectral index. Anesth Analg 87:949–955PubMed
43.
go back to reference Kussman BD, Gruber EM, Zurakowski D, Hansen DD, Sullivan LJ, Laussen PC (2001) Bispectral index monitoring during infant cardiac surgery: relationship of BIS to the stress response and plasma fentanyl levels. Paediatr Anaesth 11:663–669PubMed Kussman BD, Gruber EM, Zurakowski D, Hansen DD, Sullivan LJ, Laussen PC (2001) Bispectral index monitoring during infant cardiac surgery: relationship of BIS to the stress response and plasma fentanyl levels. Paediatr Anaesth 11:663–669PubMed
44.
go back to reference Theilen HJ, Ragaller M, Tscho U, May SA, Schackert G, Albrecht MD (2000) Electroencephalogram silence ratio for early outcome prognosis in severe head trauma. Crit Care Med 28:3522–3529PubMed Theilen HJ, Ragaller M, Tscho U, May SA, Schackert G, Albrecht MD (2000) Electroencephalogram silence ratio for early outcome prognosis in severe head trauma. Crit Care Med 28:3522–3529PubMed
45.
go back to reference Gill M, Green SM, Krauss B (2003) Can the bispectral index monitor quantify altered level of consciousness in emergency department patients? Acad Emerg Med 10:175–179PubMed Gill M, Green SM, Krauss B (2003) Can the bispectral index monitor quantify altered level of consciousness in emergency department patients? Acad Emerg Med 10:175–179PubMed
46.
go back to reference Gilbert TT, Wagner MR, Halukurike V, Paz HL, Garland A (2001) Use of bispectral electroencephalogram monitoring to assess neurologic status in unsedated, critically ill patients. Crit Care Med 29:1996–2000PubMed Gilbert TT, Wagner MR, Halukurike V, Paz HL, Garland A (2001) Use of bispectral electroencephalogram monitoring to assess neurologic status in unsedated, critically ill patients. Crit Care Med 29:1996–2000PubMed
47.
go back to reference Nieuwenhuijs D, Coleman EL, Douglas NJ, Drummond GB, Dahan A (2002) Bispectral index values and spectral edge frequency at different stages of physiologic sleep. Anesth Analg 94:125–129PubMed Nieuwenhuijs D, Coleman EL, Douglas NJ, Drummond GB, Dahan A (2002) Bispectral index values and spectral edge frequency at different stages of physiologic sleep. Anesth Analg 94:125–129PubMed
48.
go back to reference Sleigh J, Andrzejowski J, Steyn-Ross A, Steyn-Ross M (1999) The bispectral index: a measure of depth of sleep? Anesth Analg 88:659–661PubMed Sleigh J, Andrzejowski J, Steyn-Ross A, Steyn-Ross M (1999) The bispectral index: a measure of depth of sleep? Anesth Analg 88:659–661PubMed
49.
go back to reference Wagner BK, Zavotsky KE, Sweeney JB, Palmeri BA, Hammond JS (1998) Patient recall of therapeutic paralysis in a surgical critical care unit. Pharmacotherapy 18:358–363PubMed Wagner BK, Zavotsky KE, Sweeney JB, Palmeri BA, Hammond JS (1998) Patient recall of therapeutic paralysis in a surgical critical care unit. Pharmacotherapy 18:358–363PubMed
50.
go back to reference Triltsch AE, Welte M, von Homeyer P, Welte M, von Homeyer P, Grosse J, Genahr A, Moshirzadeh M, Sidiropoulos A, Konertz W, Kox WJ, Spies CD (2002) Bispectral index-guided sedation with dexmedetomidine in intensive care: a prospective, randomized, double blind, placebo-controlled phase II study. Crit Care Med 30:1007–1014PubMed Triltsch AE, Welte M, von Homeyer P, Welte M, von Homeyer P, Grosse J, Genahr A, Moshirzadeh M, Sidiropoulos A, Konertz W, Kox WJ, Spies CD (2002) Bispectral index-guided sedation with dexmedetomidine in intensive care: a prospective, randomized, double blind, placebo-controlled phase II study. Crit Care Med 30:1007–1014PubMed
Metadata
Title
Comparison of the bispectral index monitor with the Comfort score in assessing level of sedation of critically ill children
Authors
Simon P. Courtman
Allan Wardurgh
Andy J. Petros
Publication date
01-12-2003
Publisher
Springer Berlin Heidelberg
Published in
Intensive Care Medicine / Issue 12/2003
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-003-1997-3

Other articles of this Issue 12/2003

Intensive Care Medicine 12/2003 Go to the issue