Experiences with continuous intra-arterial blood gas monitoring: precision and drift of a pure optode-system

Authors: Matthias Menzel, Jens Soukup, Dirk Henze, Kai Engelbrecht, Maik Senderreck, Alexander Scharf, Andreas Rieger, Stefan Grond

Published in: Intensive Care Medicine | Issue 12/2003

Login to get access

Abstract

Objective

The utility of continuous intra-arterial blood gas analysis (CBGA) with combined electrochemical and optode sensors has been demonstrated. More recently, a pure optode sensor with a changed sensing element architecture has become available. The aim was to determine the measurement accuracy and long-term stability of the new sensor.

Design

A prospective explorative study was performed. Simultaneous measurements of intermittent blood gas analyses (IBGA) (ABL 610, Radiometer, Copenhagen) and CBGA (Diametrics Medical, High Wycombe, Bucks., UK) were compared using Bland-Altman analysis.

Patients

Twenty-five patients admitted to the ICU and requiring mechanical ventilation for an expected minimum of about 96 h were included.

Results

Mean monitoring time was 106.1 (range 15–231) hours. Bias and precision for PO2 were −0.2 kPa (1%)±1.8 kPa (9.5%); PCO2: 0.03 kPa (0.6%)±0.44 kPa (9.3%); pH: −0.001 (0.01%)±0.04 (0.45%). The sensor showed no change of measurement characteristics during 4 days of measurement. However, in 69 cases continuous monitoring was interrupted (reversible sudden drops of PO2 measurement) possibly caused by thrombotic deposition and/or sensor bending and accidental sensor retraction.

Conclusions

The precision and bias of the PCO2- and pH-sensing elements were in line with the findings of the older sensor technology. The possibility that the PO2 optode could offer greater accuracy than the older technology is suggested by comparisons with results reported in previous studies. No sensor drift occurred during long-term measurement over more than 4 days.
Literature
1.
go back to reference Sasse SA, Chen PA, Mahutte CK (1994) Variability of arterial blood gas values over time in stable medical ICU patients. Chest 106:187–193CrossRef Sasse SA, Chen PA, Mahutte CK (1994) Variability of arterial blood gas values over time in stable medical ICU patients. Chest 106:187–193CrossRef
2.
go back to reference Hatherill M, Tibby SM, Durward A, Rajah V, Murdoch IA (1997) Continuous intra-arterial blood-gas monitoring in infants and children with cyanotic heart disease. Br J Anaesth 79:665–667CrossRef Hatherill M, Tibby SM, Durward A, Rajah V, Murdoch IA (1997) Continuous intra-arterial blood-gas monitoring in infants and children with cyanotic heart disease. Br J Anaesth 79:665–667CrossRef
3.
go back to reference Venkatesh B, Pigott DW, Fernandez A, Hendry SP (1996) Continuous measurement of arterial blood gas status during total hip replacement: a prospective study. Anaesth Intensive Care 24:334–341CrossRef Venkatesh B, Pigott DW, Fernandez A, Hendry SP (1996) Continuous measurement of arterial blood gas status during total hip replacement: a prospective study. Anaesth Intensive Care 24:334–341CrossRef
4.
go back to reference Green GE, Hassell KT, Mahutte CK (1987) Comparison of arterial blood gas with continuous intra-arterial and transcutaneous PO2 sensors in adult critically ill patients. Crit Care Med 15:491–494CrossRef Green GE, Hassell KT, Mahutte CK (1987) Comparison of arterial blood gas with continuous intra-arterial and transcutaneous PO2 sensors in adult critically ill patients. Crit Care Med 15:491–494CrossRef
5.
go back to reference Mahutte CK (1998) On-line arterial blood gas analysis with optodes: current status. Clin Biochem 31:119–130CrossRef Mahutte CK (1998) On-line arterial blood gas analysis with optodes: current status. Clin Biochem 31:119–130CrossRef
6.
go back to reference Mahutte CK (1994) Continuous intra-arterial blood gas monitoring. Intensive Care Med 20:85–86CrossRef Mahutte CK (1994) Continuous intra-arterial blood gas monitoring. Intensive Care Med 20:85–86CrossRef
7.
go back to reference Venkatesh B, Clutton-Brock TH, Hendry SP (1994) Continuous measurement of blood gases using a combined electrochemical and spectrophotometric sensor. J Med Eng Technol 18:165–168CrossRef Venkatesh B, Clutton-Brock TH, Hendry SP (1994) Continuous measurement of blood gases using a combined electrochemical and spectrophotometric sensor. J Med Eng Technol 18:165–168CrossRef
8.
go back to reference Venkatesh B, Clutton-Brock TH, Hendry SP (1995) Evaluation of the Paratrend 7 intravascular blood gas monitor during cardiac surgery: comparison with the C4000 in-line blood gas monitor during cardiopulmonary bypass. J Cardiothorac Vasc Anesth 9:412–419CrossRef Venkatesh B, Clutton-Brock TH, Hendry SP (1995) Evaluation of the Paratrend 7 intravascular blood gas monitor during cardiac surgery: comparison with the C4000 in-line blood gas monitor during cardiopulmonary bypass. J Cardiothorac Vasc Anesth 9:412–419CrossRef
9.
go back to reference Vretzakis G, Papaziogas B, Matsaridou E, Vasiliadou G, Papadopoulos G, Patsialas C, Kostopoulou F (2000) Continuous monitoring of arterial blood gases and pH during intraoperative rapid blood administration using a Paratrend sensor. Vox Sang 78:158–163CrossRef Vretzakis G, Papaziogas B, Matsaridou E, Vasiliadou G, Papadopoulos G, Patsialas C, Kostopoulou F (2000) Continuous monitoring of arterial blood gases and pH during intraoperative rapid blood administration using a Paratrend sensor. Vox Sang 78:158–163CrossRef
10.
go back to reference Weiss IK, Fink S, Harrison R, Feldman JD, Brill JE (1999) Clinical use of continuous arterial blood gas monitoring in the pediatric intensive care unit. Pediatrics 103:440–445CrossRef Weiss IK, Fink S, Harrison R, Feldman JD, Brill JE (1999) Clinical use of continuous arterial blood gas monitoring in the pediatric intensive care unit. Pediatrics 103:440–445CrossRef
11.
go back to reference Haller M, Kilger E, Briegel J, Forst H, Peter K (1994) Continuous intra-arterial blood gas and pH monitoring in critically ill patients with severe respiratory failure: a prospective, criterion standard study. Crit Care Med 22:580–587CrossRef Haller M, Kilger E, Briegel J, Forst H, Peter K (1994) Continuous intra-arterial blood gas and pH monitoring in critically ill patients with severe respiratory failure: a prospective, criterion standard study. Crit Care Med 22:580–587CrossRef
12.
go back to reference Larson CP Jr (1996) Continuous arterial blood gas monitoring: a technology in transition. Intensive Care Med 22:1141–1143CrossRef Larson CP Jr (1996) Continuous arterial blood gas monitoring: a technology in transition. Intensive Care Med 22:1141–1143CrossRef
13.
go back to reference Venkatesh B, Hendry SP (1996) Continuous intra-arterial blood gas monitoring. Intensive Care Med 22:818–828CrossRef Venkatesh B, Hendry SP (1996) Continuous intra-arterial blood gas monitoring. Intensive Care Med 22:818–828CrossRef
14.
go back to reference Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J (2000) Comparing methods of clinical measurement: reporting standards for bland and altman analysis. Anesth Analg 90:593–602CrossRef Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J (2000) Comparing methods of clinical measurement: reporting standards for bland and altman analysis. Anesth Analg 90:593–602CrossRef
15.
go back to reference Morgan C, Newell SJ, Ducker DA, Hodgkinson J, White DK, Morley CJ, Church JM (1999) Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor. Arch Dis Child Fetal Neonatal Ed 80:F93–98CrossRef Morgan C, Newell SJ, Ducker DA, Hodgkinson J, White DK, Morley CJ, Church JM (1999) Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor. Arch Dis Child Fetal Neonatal Ed 80:F93–98CrossRef
16.
go back to reference Pappert D, Rossaint R, Lewandowski K, Kuhlen R, Gerlach H, Falke KJ (1995) Preliminary evaluation of a new continuous intra-arterial blood gas monitoring device. Acta Anaesthesiol Scand [Suppl 107]:67–70 Pappert D, Rossaint R, Lewandowski K, Kuhlen R, Gerlach H, Falke KJ (1995) Preliminary evaluation of a new continuous intra-arterial blood gas monitoring device. Acta Anaesthesiol Scand [Suppl 107]:67–70
17.
go back to reference Zaugg M, Lucchinetti E, Zalunardo MP, Zumstein S, Spahn DR, Pasch T, Zollinger A (1998) Substantial changes in arterial blood gases during thoracoscopic surgery can be missed by conventional intermittent laboratory blood gas analyses. Anesth Analg 87:647–653PubMed Zaugg M, Lucchinetti E, Zalunardo MP, Zumstein S, Spahn DR, Pasch T, Zollinger A (1998) Substantial changes in arterial blood gases during thoracoscopic surgery can be missed by conventional intermittent laboratory blood gas analyses. Anesth Analg 87:647–653PubMed
18.
go back to reference Zollinger A, Spahn DR, Singer T, Zalunardo MP, Stoehr S, Weder W, Pasch T (1997) Accuracy and clinical performance of a continuous intra-arterial blood-gas monitoring system during thoracoscopic surgery. Br J Anaesth 79:47–52CrossRef Zollinger A, Spahn DR, Singer T, Zalunardo MP, Stoehr S, Weder W, Pasch T (1997) Accuracy and clinical performance of a continuous intra-arterial blood-gas monitoring system during thoracoscopic surgery. Br J Anaesth 79:47–52CrossRef
19.
go back to reference Ishikawa S, Makita K, Nakazawa K, Amaha K (1998) Continuous intra-arterial blood gas monitoring during oesophagectomy. Can J Anaesth 45:273–276CrossRef Ishikawa S, Makita K, Nakazawa K, Amaha K (1998) Continuous intra-arterial blood gas monitoring during oesophagectomy. Can J Anaesth 45:273–276CrossRef
20.
go back to reference Coule LW, Truemper EJ, Steinhart CM, Lutin WA (2001) Accuracy and utility of a continuous intra-arterial blood gas monitoring system in pediatric patients. Crit Care Med 29:420–426CrossRef Coule LW, Truemper EJ, Steinhart CM, Lutin WA (2001) Accuracy and utility of a continuous intra-arterial blood gas monitoring system in pediatric patients. Crit Care Med 29:420–426CrossRef
21.
go back to reference Abraham E, Gallagher TJ, Fink S (1996) Clinical evaluation of a multiparameter intra-arterial blood-gas sensor. Intensive Care Med 22:507–513CrossRef Abraham E, Gallagher TJ, Fink S (1996) Clinical evaluation of a multiparameter intra-arterial blood-gas sensor. Intensive Care Med 22:507–513CrossRef
22.
go back to reference Medicare MaCP (1992) Regulations Implementing the Clinical Laboratory Improvement Amendments of 1988 (CLIA '88). Federal Register 7002–7186 Medicare MaCP (1992) Regulations Implementing the Clinical Laboratory Improvement Amendments of 1988 (CLIA '88). Federal Register 7002–7186
23.
go back to reference Mahutte CK, Sassoon CS, Muro JR, Hansmann DR, Maxwell TP, Miller WW, Yafuso M (1990) Progress in the development of a fluorescent intravascular blood gas system in man. J Clin Monit 6:147–57CrossRef Mahutte CK, Sassoon CS, Muro JR, Hansmann DR, Maxwell TP, Miller WW, Yafuso M (1990) Progress in the development of a fluorescent intravascular blood gas system in man. J Clin Monit 6:147–57CrossRef
24.
go back to reference Venkatesh B, Clutton Brock TH, Hendry SP (1994) A multiparameter sensor for continuous intra-arterial blood gas monitoring: a prospective evaluation. Crit Care Med 22:588–594CrossRef Venkatesh B, Clutton Brock TH, Hendry SP (1994) A multiparameter sensor for continuous intra-arterial blood gas monitoring: a prospective evaluation. Crit Care Med 22:588–594CrossRef
Metadata
Title
Experiences with continuous intra-arterial blood gas monitoring: precision and drift of a pure optode-system
Authors
Matthias Menzel
Jens Soukup
Dirk Henze
Kai Engelbrecht
Maik Senderreck
Alexander Scharf
Andreas Rieger
Stefan Grond
Publication date
01-12-2003
Publisher
Springer Berlin Heidelberg
Published in
Intensive Care Medicine / Issue 12/2003
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-003-1962-1

Other articles of this Issue 12/2003

Intensive Care Medicine 12/2003 Go to the issue

Acknowledgements to referees

2003