Abstract

Aims/hypothesis

Glucose sensors often measure s.c. interstitial fluid (ISF) glucose rather than blood or plasma glucose. Putative differences between plasma and ISF glucose include a protracted delay during the recovery from hypoglycaemia and an increased gradient during hyperinsulinaemia. These have often been investigated using sensor systems that have delays due to signal smoothing, or require long equilibration times. The aim of the present study was to define these relationships during hypoglycaemia in a well-equilibrated system with no smoothing.

Methods

Hypoglycaemia was induced by i.v. insulin infusion (360 pmol·m−2·min−1) in ten non-diabetic subjects. Glucose was sequentially clamped at ∼5, 4.2 and 3.1 mmol/l and allowed to return to normoglycaemia. Subjects wore two s.c. glucose sensors (Medtronic MiniMed, Northridge, CA, USA) that had been inserted for more than 12 h. A two-compartment model was used to quantify the delay and gradient.

Results

The delay during the fall in plasma glucose was not different from the delay during recovery (8.3±0.67 vs 6.3±1.1 min; p=0.27) and no differences were observed in the ratio of sensor current to plasma glucose at basal insulin (2.7±0.25 nA·mmol−1·l) compared with any of the hyperinsulinaemic clamp phases (2.8±0.18, 2.7±0.021, 2.9±0.21; p=NS). The ratio was significantly elevated following recovery to normoglycaemia (3.1±0.2 nA·mmol−1·l; p<0.001).

Conclusions/interpretation

The elevated ratio suggests that the plasma to ISF glucose gradient was decreased following hypoglycaemia, possibly due to increased skin blood flow. Recovery from hypoglycaemia is not accompanied by a protracted delay and insulin does not increase the plasma to s.c. ISF glucose gradient.
Appendix
Available only for authorised users
Literature
1.
go back to reference DCCT (1993) The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. New Engl J Med 329:977–986 DCCT (1993) The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. New Engl J Med 329:977–986
2.
go back to reference Rebrin K, Steil GM (2000) Can interstitial glucose assessment replace blood glucose measurements? Diabetes Technol Ther 2:461–472CrossRefPubMed Rebrin K, Steil GM (2000) Can interstitial glucose assessment replace blood glucose measurements? Diabetes Technol Ther 2:461–472CrossRefPubMed
3.
go back to reference Baron AD (1994) Hemodynamic actions of insulin. Am J Physiol 267:E187–E202PubMed Baron AD (1994) Hemodynamic actions of insulin. Am J Physiol 267:E187–E202PubMed
4.
go back to reference Moberg E, Hagström-Toft E, Arner P et al (1997) Protracted glucose fall in subcutaneous adipose tissue and skeletal muscle compared with blood during insulin-induced hypoglycaemia. Diabetologia 40:1320–1326PubMed Moberg E, Hagström-Toft E, Arner P et al (1997) Protracted glucose fall in subcutaneous adipose tissue and skeletal muscle compared with blood during insulin-induced hypoglycaemia. Diabetologia 40:1320–1326PubMed
5.
go back to reference Cheyne EH, Cavan DA, Kerr D (2002) Performance of a continuous glucose monitoring system during controlled hypoglycaemia in healthy volunteers. Diabetes Technol Ther 4:607–613PubMed Cheyne EH, Cavan DA, Kerr D (2002) Performance of a continuous glucose monitoring system during controlled hypoglycaemia in healthy volunteers. Diabetes Technol Ther 4:607–613PubMed
6.
go back to reference Wientjes KJ, Schoonen AJ (2001) Determination of time delay between blood and interstitial adipose tissue glucose concentration change by microdialysis in healthy volunteers. Int J Artif Organs 24:884–889PubMed Wientjes KJ, Schoonen AJ (2001) Determination of time delay between blood and interstitial adipose tissue glucose concentration change by microdialysis in healthy volunteers. Int J Artif Organs 24:884–889PubMed
7.
go back to reference Rebrin K, Steil GM, Van Antwerp WP et al (1999) Subcutaneous glucose predicts plasma glucose independent of insulin: implications for continuous monitoring. Am J Physiol 277:E561–E571PubMed Rebrin K, Steil GM, Van Antwerp WP et al (1999) Subcutaneous glucose predicts plasma glucose independent of insulin: implications for continuous monitoring. Am J Physiol 277:E561–E571PubMed
8.
go back to reference Steil GM, Rebrin K, Mastrototaro J et al (2003) Determination of plasma glucose during rapid glucose excursions with a subcutaneous glucose sensor. Diabetes Technol Ther 5:27–31PubMed Steil GM, Rebrin K, Mastrototaro J et al (2003) Determination of plasma glucose during rapid glucose excursions with a subcutaneous glucose sensor. Diabetes Technol Ther 5:27–31PubMed
9.
go back to reference Kulcu E, Tamada JA, Reach G et al (2003) Physiological differences between interstitial glucose and blood glucose measured in human subjects. Diabetes Care 26:2405–2409PubMed Kulcu E, Tamada JA, Reach G et al (2003) Physiological differences between interstitial glucose and blood glucose measured in human subjects. Diabetes Care 26:2405–2409PubMed
10.
go back to reference Moberg E, Hagström-Toft E, Bolinder J (1997) Detection of hypoglycaemia by microdialysis measurements of glucose in subcutaneous adipose tissue. Horm Metab Res 29:440–443PubMed Moberg E, Hagström-Toft E, Bolinder J (1997) Detection of hypoglycaemia by microdialysis measurements of glucose in subcutaneous adipose tissue. Horm Metab Res 29:440–443PubMed
11.
go back to reference Sternberg F, Meyerhoff C, Mennel FJ et al (1996) Does fall in tissue glucose precede fall in blood glucose? Diabetologia 39:609–612PubMed Sternberg F, Meyerhoff C, Mennel FJ et al (1996) Does fall in tissue glucose precede fall in blood glucose? Diabetologia 39:609–612PubMed
12.
go back to reference Gross TM, Bode BW, Einhorn D et al (2000) Performance evaluation of the MiniMed continuous glucose monitoring system during patient home use. Diabetes Technol Ther 2:49–56PubMed Gross TM, Bode BW, Einhorn D et al (2000) Performance evaluation of the MiniMed continuous glucose monitoring system during patient home use. Diabetes Technol Ther 2:49–56PubMed
13.
go back to reference Clarke WL, Cox D, Gonder-Frederick LA et al (1987) Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 10:622–628PubMed Clarke WL, Cox D, Gonder-Frederick LA et al (1987) Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 10:622–628PubMed
14.
go back to reference Bolinder J, Kager L, Ostman J et al (1983) Differences at the receptor and postreceptor levels between human omental and subcutaneous adipose tissue in the action of insulin on lipolysis. Diabetes 32:117–123PubMed Bolinder J, Kager L, Ostman J et al (1983) Differences at the receptor and postreceptor levels between human omental and subcutaneous adipose tissue in the action of insulin on lipolysis. Diabetes 32:117–123PubMed
15.
go back to reference Virtanen KA, Lonnroth P, Parkkola R et al (2002) Glucose uptake and perfusion in subcutaneous and visceral adipose tissue during insulin stimulation in nonobese and obese humans. J Clin Endocrinol Metab 87:3902–3910PubMed Virtanen KA, Lonnroth P, Parkkola R et al (2002) Glucose uptake and perfusion in subcutaneous and visceral adipose tissue during insulin stimulation in nonobese and obese humans. J Clin Endocrinol Metab 87:3902–3910PubMed
16.
go back to reference Thome-Duret V, Reach G, Gangnerau MN et al (1996) Use of a subcutaneous glucose sensor to detect decreases in glucose concentration prior to observation in blood. Anal Chem 68:3822–3826PubMed Thome-Duret V, Reach G, Gangnerau MN et al (1996) Use of a subcutaneous glucose sensor to detect decreases in glucose concentration prior to observation in blood. Anal Chem 68:3822–3826PubMed
17.
go back to reference Aussedat B, Dupire-Angel M, Gifford R et al (2000) Interstitial glucose concentration and glycemia: implications for continuous subcutaneous glucose monitoring. Am J Physiol Endocrinol Metab 278:E716–E728PubMed Aussedat B, Dupire-Angel M, Gifford R et al (2000) Interstitial glucose concentration and glycemia: implications for continuous subcutaneous glucose monitoring. Am J Physiol Endocrinol Metab 278:E716–E728PubMed
18.
go back to reference Caplin NJ, O’Leary P, Bulsara M et al (2003) Subcutaneous glucose sensor values closely parallel blood glucose during insulin-induced hypoglycaemia. Diabet Med 20:238–241PubMed Caplin NJ, O’Leary P, Bulsara M et al (2003) Subcutaneous glucose sensor values closely parallel blood glucose during insulin-induced hypoglycaemia. Diabet Med 20:238–241PubMed
19.
go back to reference Monsod TP, Flanagan DE, Rife F et al (2002) Do sensor glucose levels accurately predict plasma glucose concentrations during hypoglycemia and hyperinsulinemia? Diabetes Care 25:889–893PubMed Monsod TP, Flanagan DE, Rife F et al (2002) Do sensor glucose levels accurately predict plasma glucose concentrations during hypoglycemia and hyperinsulinemia? Diabetes Care 25:889–893PubMed
20.
go back to reference Boyne MS, Silver DM, Kaplan J et al (2003) Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor. Diabetes 52:2790–2794PubMed Boyne MS, Silver DM, Kaplan J et al (2003) Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor. Diabetes 52:2790–2794PubMed
21.
go back to reference Aman J, Berne C, Ewald U et al (1992) Cutaneous blood flow during a hypoglycaemic clamp in insulin-dependent diabetic patients and healthy subjects. Clin Sci (Lond) 82:615–618 Aman J, Berne C, Ewald U et al (1992) Cutaneous blood flow during a hypoglycaemic clamp in insulin-dependent diabetic patients and healthy subjects. Clin Sci (Lond) 82:615–618
22.
go back to reference Serne EH, IJzerman RG, Gans RO et al (2002) Direct evidence for insulin-induced capillary recruitment in skin of healthy subjects during physiological hyperinsulinemia. Diabetes 51:1515–1522PubMed Serne EH, IJzerman RG, Gans RO et al (2002) Direct evidence for insulin-induced capillary recruitment in skin of healthy subjects during physiological hyperinsulinemia. Diabetes 51:1515–1522PubMed
23.
go back to reference Aman J, Berne C, Ewald U et al (1990) Lack of cutaneous hyperemia in response to insulin-induced hypoglycemia in IDDM. Diabetes Care 13:1029–1033PubMed Aman J, Berne C, Ewald U et al (1990) Lack of cutaneous hyperemia in response to insulin-induced hypoglycemia in IDDM. Diabetes Care 13:1029–1033PubMed
24.
go back to reference Schmidtke DW, Freeland AC, Heller A et al (1998) Measurement and modeling of the transient difference between blood and subcutaneous glucose concentrations in the rat after injection of insulin. Proc Natl Acad Sci U S A 95:294–299PubMed Schmidtke DW, Freeland AC, Heller A et al (1998) Measurement and modeling of the transient difference between blood and subcutaneous glucose concentrations in the rat after injection of insulin. Proc Natl Acad Sci U S A 95:294–299PubMed
Metadata
Title
Interstitial fluid glucose dynamics during insulin-induced hypoglycaemia
Authors
G. M. Steil
K. Rebrin
F. Hariri
S. Jinagonda
S. Tadros
C. Darwin
M. F. Saad
Publication date
01-09-2005
Publisher
Springer-Verlag
Published in
Diabetologia / Issue 9/2005
Print ISSN: 0012-186X
Electronic ISSN: 1432-0428
DOI
https://doi.org/10.1007/s00125-005-1852-x

Other articles of this Issue 9/2005

Diabetologia 9/2005 Go to the issue