Abstract

Background

This study evaluated the clinical results of a new implant in the internal fixation of undisplaced femoral neck fractures.

Method

Irrespective of their age, 149 patients with undisplaced (Garden I and II) femoral neck fractures were included in a prospective multicentre clinical cohort study and were treated by internal fixation by means of the Dynamic Locking Blade Plate (DLBP). The mean age was 69 years and the follow-up at least one year.

Results

The DLBP fixation resulted in 6 out of 149 failures caused by AVN (2x), non-union (2x), loss of fixation (3x) or combination of these.

Conclusion

The fixation of undisplaced femoral neck fractures by the DLBP resulted in a low failure rate of 4 %.
Literature
1.
go back to reference Loizou CL, Parker MJ. Avascular necrosis after internal fixation of intracapsular hip fractures; a study of the outcome for 1023 patients. Injury. 2009;40:1143–6.CrossRefPubMed Loizou CL, Parker MJ. Avascular necrosis after internal fixation of intracapsular hip fractures; a study of the outcome for 1023 patients. Injury. 2009;40:1143–6.CrossRefPubMed
2.
go back to reference Sevitt S. Avascular necrosis and revascularisation of the femoral head after intracapsular fractures; a combined arteriographic and histological necropsy study. J Bone Joint Surg Br. 1964;46:270–96.PubMed Sevitt S. Avascular necrosis and revascularisation of the femoral head after intracapsular fractures; a combined arteriographic and histological necropsy study. J Bone Joint Surg Br. 1964;46:270–96.PubMed
3.
go back to reference Stromqvist B, Hansson LI, Palmer J, Ceder C, Thomgren KG. Scintimetric evaluation of nailed femoral neck fracture with special reference to type of osteosynthesis. Acta Orthop Scand. 1983;54:340–7.CrossRefPubMed Stromqvist B, Hansson LI, Palmer J, Ceder C, Thomgren KG. Scintimetric evaluation of nailed femoral neck fracture with special reference to type of osteosynthesis. Acta Orthop Scand. 1983;54:340–7.CrossRefPubMed
4.
go back to reference Linde F, Anderson E, Hvass J, Madson F, Pallesen R. Avascular femoral head necrosis following fracture fixation. Injury. 1986;17:159–63.CrossRefPubMed Linde F, Anderson E, Hvass J, Madson F, Pallesen R. Avascular femoral head necrosis following fracture fixation. Injury. 1986;17:159–63.CrossRefPubMed
5.
go back to reference Perren SM. Physical and biological aspects of fracture healing with special reference to Internal fixation. Clin Ortop Relat R. 1979;138:175–96. Perren SM. Physical and biological aspects of fracture healing with special reference to Internal fixation. Clin Ortop Relat R. 1979;138:175–96.
6.
go back to reference Claes LE, Heigele CA. Magnitudes of local stress and strain along bony surfaces predict the course and type of fracture healing. J Biomech. 1999;32:255–66.CrossRefPubMed Claes LE, Heigele CA. Magnitudes of local stress and strain along bony surfaces predict the course and type of fracture healing. J Biomech. 1999;32:255–66.CrossRefPubMed
7.
go back to reference Rockwood CA, Green DP, Heckman JD, Bucholz RW. Fractures in adults. 4th ed. Philadelphia: Lippincot-Raven; 2006 Rockwood CA, Green DP, Heckman JD, Bucholz RW. Fractures in adults. 4th ed. Philadelphia: Lippincot-Raven; 2006
8.
go back to reference Rogmark C, Flensburg L, Fredin H. Undisplaced femoral neck fractures—no problems? A consecutive study of 224 patients treated with internal fixation. Injury. 2009;40:274–6.CrossRefPubMed Rogmark C, Flensburg L, Fredin H. Undisplaced femoral neck fractures—no problems? A consecutive study of 224 patients treated with internal fixation. Injury. 2009;40:274–6.CrossRefPubMed
9.
go back to reference Bjørgul K, Reikerås O. Outcome of undisplaced and moderately displaced femoral neck fractures. Acta Ortop. 2007;78:498–504.CrossRef Bjørgul K, Reikerås O. Outcome of undisplaced and moderately displaced femoral neck fractures. Acta Ortop. 2007;78:498–504.CrossRef
10.
go back to reference Parker MJ, Raghavan R, Gurusamy K. Incidence of fracture-healing complications after femoral neck fractures. Clin Orthop Relat R. 2007;458:175–9. Parker MJ, Raghavan R, Gurusamy K. Incidence of fracture-healing complications after femoral neck fractures. Clin Orthop Relat R. 2007;458:175–9.
11.
go back to reference Conn KS, Parker MJ. Undisplaced intracapsular hip fracture: results of internal fixation in 375 patients. Clin Orthop Relat R. 2004;421:249–54.CrossRef Conn KS, Parker MJ. Undisplaced intracapsular hip fracture: results of internal fixation in 375 patients. Clin Orthop Relat R. 2004;421:249–54.CrossRef
12.
go back to reference Parker MJ, White A, Boyle A. Fixation versus hemiarthroplasty for undisplaced intracapsular hip fractures. Injury. 2008;39:791–5.CrossRefPubMed Parker MJ, White A, Boyle A. Fixation versus hemiarthroplasty for undisplaced intracapsular hip fractures. Injury. 2008;39:791–5.CrossRefPubMed
13.
go back to reference van Embden D, Krijnen P, Schipper IB. De mediale collumfractuur. Is er nog plaats voor conservatieve behandeling? Ned Tijdschr Genees. 2014;158:A8105. van Embden D, Krijnen P, Schipper IB. De mediale collumfractuur. Is er nog plaats voor conservatieve behandeling? Ned Tijdschr Genees. 2014;158:A8105.
14.
go back to reference Roerdink WH, Aalsma AMM, Nijenbanning G, van Walsum ADP. The dynamic locking blade plate, a new implant for intracapsular hip fractures: biomechanical comparison with the sliding hip screw and the Twin Hook. Injury. 2009;40:283–7.CrossRefPubMed Roerdink WH, Aalsma AMM, Nijenbanning G, van Walsum ADP. The dynamic locking blade plate, a new implant for intracapsular hip fractures: biomechanical comparison with the sliding hip screw and the Twin Hook. Injury. 2009;40:283–7.CrossRefPubMed
15.
go back to reference Steinberg ME, Hayken GD, Steinberg DR. A quantitative system for staging avascular necrosis. J Bone Joint Surg Br. 1995;77(1):34–41.PubMed Steinberg ME, Hayken GD, Steinberg DR. A quantitative system for staging avascular necrosis. J Bone Joint Surg Br. 1995;77(1):34–41.PubMed
16.
go back to reference Baumgaertner MR, Curtin SL, Lindskogg DM, Keggi JM. The value of the tip-apex distance in predicting failure of peritrochanteric Fractures of the hip. J Bone Joint Surg Am. 1995;77(7):1058–64.CrossRefPubMed Baumgaertner MR, Curtin SL, Lindskogg DM, Keggi JM. The value of the tip-apex distance in predicting failure of peritrochanteric Fractures of the hip. J Bone Joint Surg Am. 1995;77(7):1058–64.CrossRefPubMed
17.
go back to reference Hui ACW, Anderson GH, Choudhry R, et al. Internal fixation or hemiarthroplasty for undisplaced fractures of the femoral neck in octogenarians. J Bone Joint Surg Br. 1994;76(6):891–4.PubMed Hui ACW, Anderson GH, Choudhry R, et al. Internal fixation or hemiarthroplasty for undisplaced fractures of the femoral neck in octogenarians. J Bone Joint Surg Br. 1994;76(6):891–4.PubMed
18.
go back to reference Lenich A, Bachmeier S, Prantl L, Nerlich M, Hammer J, Mayr E, et al. Is the rotation of the femoral head a potential initiation for cutting out? A theoretical and experimental approach. BMC Musculoskelet Disord. 2011;12(79):1–6. Lenich A, Bachmeier S, Prantl L, Nerlich M, Hammer J, Mayr E, et al. Is the rotation of the femoral head a potential initiation for cutting out? A theoretical and experimental approach. BMC Musculoskelet Disord. 2011;12(79):1–6.
19.
go back to reference Olsson O, Tanner KE, Ceder L, Ryd L. A biomechanical study on fixation stability with twin hook or lag screw in artificial cancellous bone. Int Orthop. 2002;26(6):349–55.CrossRefPubMedPubMedCentral Olsson O, Tanner KE, Ceder L, Ryd L. A biomechanical study on fixation stability with twin hook or lag screw in artificial cancellous bone. Int Orthop. 2002;26(6):349–55.CrossRefPubMedPubMedCentral
20.
go back to reference Windolf M, Braunstein V, Dutoit C, Schwieger K. Is a helical shaped implant a superior alternative to the dynamic hip screw for unstable femoral neck fractures? A biomechanical investigation. Clin Biomech. 2009;24:59–64.CrossRef Windolf M, Braunstein V, Dutoit C, Schwieger K. Is a helical shaped implant a superior alternative to the dynamic hip screw for unstable femoral neck fractures? A biomechanical investigation. Clin Biomech. 2009;24:59–64.CrossRef
21.
go back to reference Linde F, Anderson E, Hvass J, Madson F, Pallesen R. Avascular femoral head necrosis following fracture fixation. Injury. 1986;17(3):159–63.CrossRefPubMed Linde F, Anderson E, Hvass J, Madson F, Pallesen R. Avascular femoral head necrosis following fracture fixation. Injury. 1986;17(3):159–63.CrossRefPubMed
Metadata
Title
Low failure rate by means of DLBP fixation of undisplaced femoral neck fractures
Authors
A. D. P. van Walsum
J. Vroemen
H. M. J. Janzing
T. Winkelhorst
J. Kalsbeek
W. H. Roerdink
Publication date
01-08-2017
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Trauma and Emergency Surgery / Issue 4/2017
Print ISSN: 1863-9933
Electronic ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-016-0659-4

Other articles of this Issue 4/2017

European Journal of Trauma and Emergency Surgery 4/2017 Go to the issue