Measurement Uncertainty Impacts Diagnosis of Diabetes Mellitus: Reliable Minimal Difference of Plasma Glucose Results

Authors: Sandra Keutmann, Stephanie Zylla, Mathilde Dahl, Nele Friedrich, Rüdiger Landgraf, Lutz Heinemann, Anders Kallner, Matthias Nauck, Astrid Petersmann

Published in: Diabetes Therapy | Issue 1/2020

Login to get access

Abstract

Introduction

The diagnosis of diabetes mellitus is based on suitable cut-off values of specific biomarkers, such as the concentration of glucose in plasma. The German Diabetes Association has very recently published a clinical practice guideline on the definition, classification and diagnosis of diabetes mellitus that recommends measurements of plasma glucose concentration have an imprecision defined as a minimal difference (MD) of at a fasting plasma glucose concentration of 7.0 mmol/L. To obtain reliable values for the MD, we investigated long-term and short-term measurement uncertainty.

Methods

The imprecision was determined by two approaches: (1) a long-term dataset with imprecision based on the Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations (Rili-BAEK), in a medical laboratory operating 24/7, using internal quality control (IQC) data for four concentrations during a 10-year period; and (2) a detailed short-term dataset with imprecision assessed by hourly measurements of control materials. These datasets were used to calculate the MD cut-off (MDcut-off) as: \( {\text{MD}}_{\text{cut-off}} = k \times {\text{SD}} \)  = 2  \( \times {\text{SD}} \), where SD is the standard deviation and k = 2 represents a confidence level of 95%.

Results

The MDcut-off of ≤ 0.7 mmol/L at a fasting plasma glucose concentration of 7.0 mmol/L (MDcut-off 7.0) for the long-term and the short-term approaches were 0.44 and 0.40 mmol/L, respectively. The MDcut-off 7.0 from both approaches was therefore below the recommended value of 0.7 mmol/L. It was noted that the variability in performance within and between instruments can be covered by reporting the long-term MDcut-off 7.0 across all connected instruments. In this study, stable results for the MDcut-off 7.0 were obtained after 1 year.

Conclusion

Imprecision as measured by IQC data is remarkably stable over many years of operation. Current imprecision assessment usually focuses on only single instruments, whereas clinicians perceive the measurement as the result of the combined analytical performance of all instruments used for a certain assay. In the clinical setting, the MD may be a more useful measure of imprecision, and we suggest deriving the MDcut-off combined from all instruments and control cycles that are used in the patient care setting for a given analyte.
Appendix
Available only for authorised users
Literature
1.
go back to reference Marathe PH, Gao HX, Close KL. American Diabetes Association standards of medical care in diabetes 2017. J Diabetes. 2017;9:320–4.CrossRef Marathe PH, Gao HX, Close KL. American Diabetes Association standards of medical care in diabetes 2017. J Diabetes. 2017;9:320–4.CrossRef
2.
go back to reference Müller-Wieland D, Petersmann A, Nauck M, et al. Definition, classification and diagnosis of diabetes mellitus. Diabetol Stoffwechs. 2016;11:S78–81.CrossRef Müller-Wieland D, Petersmann A, Nauck M, et al. Definition, classification and diagnosis of diabetes mellitus. Diabetol Stoffwechs. 2016;11:S78–81.CrossRef
3.
go back to reference World Health Organization. Use of glycated haemoglobin (HbA1c) in the diagnosis of diabetes mellitus: abbreviated report of a WHO consultation. WHO guidelines approved by the Guidelines Review Committee. Geneva: World Health Organization; 2011. World Health Organization. Use of glycated haemoglobin (HbA1c) in the diagnosis of diabetes mellitus: abbreviated report of a WHO consultation. WHO guidelines approved by the Guidelines Review Committee. Geneva: World Health Organization; 2011.
5.
go back to reference Büttner J, Broth R, Broughton PM, Bowyer RC. International Federation of Clinical Chemistry. Committee on standards. Expert panel on nomenclature and principles of quality control in clinical chemistry. IFCC document stage 2, draft 3; 1979-07 with a view to an IFCC recommendation. Quality control in clinical chemistry. Part 4. Internal quality control. J Clin Chem Clin Biochem 1980;18:535–541. Büttner J, Broth R, Broughton PM, Bowyer RC. International Federation of Clinical Chemistry. Committee on standards. Expert panel on nomenclature and principles of quality control in clinical chemistry. IFCC document stage 2, draft 3; 1979-07 with a view to an IFCC recommendation. Quality control in clinical chemistry. Part 4. Internal quality control. J Clin Chem Clin Biochem 1980;18:535–541.
6.
go back to reference Clinical and Laboratory Standards Institute (CLSI). Expression of measurement uncertainty in laboratory medicine; approved guideline. 2012:Document EP29. Clinical and Laboratory Standards Institute (CLSI). Expression of measurement uncertainty in laboratory medicine; approved guideline. 2012:Document EP29.
7.
go back to reference Parvin CA. New insight into the comparative power of quality-control rules that use control observations within a single analytical run. Clin Chem. 1993;39:440–7.CrossRef Parvin CA. New insight into the comparative power of quality-control rules that use control observations within a single analytical run. Clin Chem. 1993;39:440–7.CrossRef
8.
go back to reference Westgard JO, Barry PL, Hunt MR, Groth T. A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem. 1981;27:493–501.CrossRef Westgard JO, Barry PL, Hunt MR, Groth T. A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem. 1981;27:493–501.CrossRef
9.
go back to reference Revision of the “Guideline of the German Medical Association on quality assurance in medical lab examinations—Rili-BAEK” (unauthorized translation). J Laboratory Med. 2015;39(1):26–69. Revision of the “Guideline of the German Medical Association on quality assurance in medical lab examinations—Rili-BAEK” (unauthorized translation). J Laboratory Med. 2015;39(1):26–69.
10.
go back to reference Kunst A, Drager B, Ziegenhorn J. UV methods with hexokinase and glucose-6-phosphate dehydrogenase. In: Bergemeier H, editor. Methods of enzymatic analysis. Deerfield: Verlag Chemie; 1983. p. 163–72. Kunst A, Drager B, Ziegenhorn J. UV methods with hexokinase and glucose-6-phosphate dehydrogenase. In: Bergemeier H, editor. Methods of enzymatic analysis. Deerfield: Verlag Chemie; 1983. p. 163–72.
11.
go back to reference Badrick T, Cervinski M, Loh TP. A primer on patient-based quality control techniques. Clin Biochem. 2019;64:1–5.CrossRef Badrick T, Cervinski M, Loh TP. A primer on patient-based quality control techniques. Clin Biochem. 2019;64:1–5.CrossRef
12.
go back to reference Petersen PH, Ricos C, Stockl D, et al. Proposed guidelines for the internal quality control of analytical results in the medical laboratory. Eur J Clin Chem Clin Biochem. 1996;34:983–99.PubMed Petersen PH, Ricos C, Stockl D, et al. Proposed guidelines for the internal quality control of analytical results in the medical laboratory. Eur J Clin Chem Clin Biochem. 1996;34:983–99.PubMed
13.
go back to reference Petersmann A, Wasner C, Nauck M, Kallner A. Frequency of extreme differences and clinical performance of glucose concentration measurements judged from 21,000 duplicate measurements. Clin Chem. 2013;59:998–1000.CrossRef Petersmann A, Wasner C, Nauck M, Kallner A. Frequency of extreme differences and clinical performance of glucose concentration measurements judged from 21,000 duplicate measurements. Clin Chem. 2013;59:998–1000.CrossRef
14.
go back to reference Siemens Healthcare Diagnostics. Dimension Vista System reagent cartridge, glucose. Ref. K1039. Newark: Siemens Healthcare Diagnostics; 2017. Siemens Healthcare Diagnostics. Dimension Vista System reagent cartridge, glucose. Ref. K1039. Newark: Siemens Healthcare Diagnostics; 2017.
15.
go back to reference D’Orazio P, Burnett RW, Fogh-Andersen N, et al. Approved IFCC recommendation on reporting results for blood glucose (abbreviated). Clin Chem. 2005;51:1573–6.CrossRef D’Orazio P, Burnett RW, Fogh-Andersen N, et al. Approved IFCC recommendation on reporting results for blood glucose (abbreviated). Clin Chem. 2005;51:1573–6.CrossRef
Metadata
Title
Measurement Uncertainty Impacts Diagnosis of Diabetes Mellitus: Reliable Minimal Difference of Plasma Glucose Results
Authors
Sandra Keutmann
Stephanie Zylla
Mathilde Dahl
Nele Friedrich
Rüdiger Landgraf
Lutz Heinemann
Anders Kallner
Matthias Nauck
Astrid Petersmann
Publication date
01-01-2020
Publisher
Springer Healthcare
Keyword
Diabetes
Published in
Diabetes Therapy / Issue 1/2020
Print ISSN: 1869-6953
Electronic ISSN: 1869-6961
DOI
https://doi.org/10.1007/s13300-019-00740-w

Other articles of this Issue 1/2020

Diabetes Therapy 1/2020 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.