Published in: Annals of Nuclear Medicine 3/2010

01-04-2010 | Original Article

The reproducibility of deep-inspiration breath-hold 18F-FDG PET/CT technique in diagnosing various cancers affected by respiratory motion

Authors: Shigeki Nagamachi, Hideyuki Wakamatsu, Shogo Kiyohara, Seigo Fujita, Shigemi Futami, Hideo Arita, Shozo Tamura, Keiichi Kawai

Published in: Annals of Nuclear Medicine | Issue 3/2010

Login to get access

Abstract

Background

The deep-inspiration breath-hold PET/CT (DIBH PET/CT) technique improves the limitations of diagnosing a lesion located in an area influenced by respiratory motion that results in spatial misregistrations caused by respiration between PET and CT. However, its reproducibility with regard to calculating the maximal standardized uptake value (SUVmax) and metabolic volume (MV) in DIBH PET/CT has not been elucidated.

Objective

The purpose of this study was to investigate the reproducibility of the DIBH PET/CT technique including calculating the SUVmax and the MV.

Methods

Sixty patients with various cancers were enrolled. The subjects had 47 abdominal lesions and 28 chest lesions. All patients demonstrated a misregistered image in the early whole-body image taken 50 min after FDG intravenous infusions. We added the delayed spot images 40 min after the first image. On the delayed image, we performed both conventional techniques with non-breath-hold (NBH) and the DIBH technique. In the four times DIBH technique, we obtained the coefficient of variance (CV) in calculating these indices for evaluating reproducibility.

Results

The SUVmax value with DIBH showed an increase of 16.1–60.1% compared with that measured by NBH. The mean value of CV was 5.5 in thoracic lesions and 6.3 in abdominal lesions. The values of MV with DIBH showed a decrease of 14.0–20.1% compared with those measured by NBH. Regarding reproducibility, mean value of CV was 7.1 in thoracic lesions and 11.9 in abdominal lesions.

Conclusion

The DIBH technique improves the inaccurate quantification of both SUVmax and MV. Although the CV value of SUVmax with DIBH technique is better in thoracic lesions compared with that in abdominal lesions, the reproducibility was acceptable.
Literature
1.
go back to reference Kim SK, Allen-Auerbach M, Goldin J, Fueger BJ, Dahlbom M, Brown M, et al. Accuracy of PET/CT in characterization of solitary pulmonary lesions. J Nucl Med. 2007;48:214–20.PubMed Kim SK, Allen-Auerbach M, Goldin J, Fueger BJ, Dahlbom M, Brown M, et al. Accuracy of PET/CT in characterization of solitary pulmonary lesions. J Nucl Med. 2007;48:214–20.PubMed
2.
go back to reference Halpern BS, Yeom K, Fueger BJ, Lufkin RB, Czernin J, Allen-Auerbach M. Evaluation of suspected local recurrence in head and neck cancer: a comparison between PET and PET/CT for biopsy-proven lesions. Eur J Radiol. 2007;62:199–204.CrossRefPubMed Halpern BS, Yeom K, Fueger BJ, Lufkin RB, Czernin J, Allen-Auerbach M. Evaluation of suspected local recurrence in head and neck cancer: a comparison between PET and PET/CT for biopsy-proven lesions. Eur J Radiol. 2007;62:199–204.CrossRefPubMed
3.
go back to reference Mottaghy FM, Sunderkötter C, Schubert R, Wohlfart P, Blumstein NM, Neumaier B, et al. Direct comparison of [18F] FDG PET/CT with PET alone and with side-by-side PET and CT in patients with malignant melanoma. Eur J Nucl Med Mol Imaging. 2007;34:1355–64.CrossRefPubMed Mottaghy FM, Sunderkötter C, Schubert R, Wohlfart P, Blumstein NM, Neumaier B, et al. Direct comparison of [18F] FDG PET/CT with PET alone and with side-by-side PET and CT in patients with malignant melanoma. Eur J Nucl Med Mol Imaging. 2007;34:1355–64.CrossRefPubMed
4.
go back to reference Nehmeh SA, Erdi YE. Respiratory motion in positron emission tomography/computed tomography: a review. Semin Nucl Med. 2008;38:167–76.CrossRefPubMed Nehmeh SA, Erdi YE. Respiratory motion in positron emission tomography/computed tomography: a review. Semin Nucl Med. 2008;38:167–76.CrossRefPubMed
5.
go back to reference Cohade C, Osman M, Marshall LN, Wahl RN. PET-CT: accuracy of PET and CT spatial registration of lung lesions. Eur J Nucl Med Imaging. 2003;30:721–6. Cohade C, Osman M, Marshall LN, Wahl RN. PET-CT: accuracy of PET and CT spatial registration of lung lesions. Eur J Nucl Med Imaging. 2003;30:721–6.
6.
go back to reference Francis DL, Visvikis D, Costa DC, Arulampalam TH, Townsend C, Luthra SK, et al. Potential impact of [18F] 3′-deoxy-3′-fluorothymidine versus [18F]fluoro-2-deoxy-d-glucose in positron emission tomography for colorectal cancer. Eur J Nucl Med Imaging. 2003;30:344–53. Francis DL, Visvikis D, Costa DC, Arulampalam TH, Townsend C, Luthra SK, et al. Potential impact of [18F] 3′-deoxy-3′-fluorothymidine versus [18F]fluoro-2-deoxy-d-glucose in positron emission tomography for colorectal cancer. Eur J Nucl Med Imaging. 2003;30:344–53.
7.
go back to reference Dawood M, Lang N, Jiang X, Schäfers KP. Lung motion correction on respiratory gated 3-D PET/CT images. IEEE Trans Med Imaging. 2006;25:476–85.CrossRefPubMed Dawood M, Lang N, Jiang X, Schäfers KP. Lung motion correction on respiratory gated 3-D PET/CT images. IEEE Trans Med Imaging. 2006;25:476–85.CrossRefPubMed
8.
9.
go back to reference Vogel WV, van Dalen JA, Wiering B, Huisman H, Corstens FH, Ruers TJ, et al. Evaluation of image registration in PET/CT of the liver and recommendations for optimized imaging. J Nucl Med. 2007;48:910–9.CrossRefPubMed Vogel WV, van Dalen JA, Wiering B, Huisman H, Corstens FH, Ruers TJ, et al. Evaluation of image registration in PET/CT of the liver and recommendations for optimized imaging. J Nucl Med. 2007;48:910–9.CrossRefPubMed
10.
go back to reference Larson SM, Nehmeh SA, Erdi YE, Humm JL. PET/CT in non-small-cell lung cancer: value of respiratory-gated PET. Chang Gung Med J. 2005;28:306–14.PubMed Larson SM, Nehmeh SA, Erdi YE, Humm JL. PET/CT in non-small-cell lung cancer: value of respiratory-gated PET. Chang Gung Med J. 2005;28:306–14.PubMed
11.
go back to reference Boucher L, Rodrigue S, Lecomte R, Bénard F. Respiratory gating for 3-dimensional PET of the thorax: feasibility and initial results. J Nucl Med. 2004;45:214–9.PubMed Boucher L, Rodrigue S, Lecomte R, Bénard F. Respiratory gating for 3-dimensional PET of the thorax: feasibility and initial results. J Nucl Med. 2004;45:214–9.PubMed
12.
go back to reference Hamill JJ, Bosmans G, Dekker A. Respiratory-gated CT as a tool for the simulation of breathing artifacts in PET and PET/CT. Med Phys. 2008;35:576–85.CrossRefPubMed Hamill JJ, Bosmans G, Dekker A. Respiratory-gated CT as a tool for the simulation of breathing artifacts in PET and PET/CT. Med Phys. 2008;35:576–85.CrossRefPubMed
13.
go back to reference Daouk J, Fin L, Bailly P, Meyer ME. Respiratory-gated positron emission tomography and breath-hold computed tomography coupling to reduce the influence of respiratory motion: methodology and feasibility. Acta Radiol. 2009;50:144–55.CrossRefPubMed Daouk J, Fin L, Bailly P, Meyer ME. Respiratory-gated positron emission tomography and breath-hold computed tomography coupling to reduce the influence of respiratory motion: methodology and feasibility. Acta Radiol. 2009;50:144–55.CrossRefPubMed
14.
go back to reference Meirelles GSP, Erdi YE, Nehmeh SA, Squire OD, Larson SM, Humm JL, et al. Deep-inspiration breath-hold PET/CT: clinical findings with a new technique for detection and characterization of thoracic lesions. J Nucl Med. 2007;48:712–9.CrossRefPubMed Meirelles GSP, Erdi YE, Nehmeh SA, Squire OD, Larson SM, Humm JL, et al. Deep-inspiration breath-hold PET/CT: clinical findings with a new technique for detection and characterization of thoracic lesions. J Nucl Med. 2007;48:712–9.CrossRefPubMed
15.
go back to reference Nehmeh SA, Erdi YE, Meirelles GSP, Squire O, Larson SM, Humm JL, et al. Deep-inspiration breath-hold PET/CT of the thorax. J Nucl Med. 2007;48:22–6.PubMed Nehmeh SA, Erdi YE, Meirelles GSP, Squire O, Larson SM, Humm JL, et al. Deep-inspiration breath-hold PET/CT of the thorax. J Nucl Med. 2007;48:22–6.PubMed
16.
go back to reference Kawano T, Ohtake E, Inoue T. Deep-inspiration breath-hold PET/CT of lung cancer: maximum standardized uptake value analysis of 108 patients. J Nucl Med. 2008;49:1223–31.CrossRefPubMed Kawano T, Ohtake E, Inoue T. Deep-inspiration breath-hold PET/CT of lung cancer: maximum standardized uptake value analysis of 108 patients. J Nucl Med. 2008;49:1223–31.CrossRefPubMed
17.
go back to reference Yamaguchi T, Ueda O, Hara H, Sakai H, Kida T, Suzuki K, et al. Usefulness of a breath-holding acquisition method in PET/CT for pulmonary lesions. Ann Nucl Med. 2009;23:65–71.CrossRefPubMed Yamaguchi T, Ueda O, Hara H, Sakai H, Kida T, Suzuki K, et al. Usefulness of a breath-holding acquisition method in PET/CT for pulmonary lesions. Ann Nucl Med. 2009;23:65–71.CrossRefPubMed
18.
go back to reference Erdi YE, Nehmeh SA, Pan T, Pevsner A, Rosenzweig KE, Mageras G, et al. The CT motion quantitation of lung lesions and its impact on PET-measured SUVs. J Nucl Med. 2004;45:1287–92.PubMed Erdi YE, Nehmeh SA, Pan T, Pevsner A, Rosenzweig KE, Mageras G, et al. The CT motion quantitation of lung lesions and its impact on PET-measured SUVs. J Nucl Med. 2004;45:1287–92.PubMed
19.
go back to reference Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Schoder H, Larson SM, et al. Effect of respiratory gating on quantifying PET images of lung cancer. J Nucl Med. 2002;43:876–81.PubMed Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Schoder H, Larson SM, et al. Effect of respiratory gating on quantifying PET images of lung cancer. J Nucl Med. 2002;43:876–81.PubMed
20.
go back to reference Nehmeh SA, Erdi YE, Pan T, Pevsner A, Rosenzweig KE, Yorke E, et al. Four-dimensional (4D) PET/CT imaging of the thorax. Med Phys. 2004;31:3179–86.CrossRefPubMed Nehmeh SA, Erdi YE, Pan T, Pevsner A, Rosenzweig KE, Yorke E, et al. Four-dimensional (4D) PET/CT imaging of the thorax. Med Phys. 2004;31:3179–86.CrossRefPubMed
21.
go back to reference Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Squire OD, Braban LE, et al. Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer. Med Phys. 2002;29:366–71.CrossRefPubMed Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Squire OD, Braban LE, et al. Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer. Med Phys. 2002;29:366–71.CrossRefPubMed
22.
go back to reference Kiyohara S, Nagamachi S, Wakamatsu H, Fujita S, Nishii R, Kamimura M, et al. The usefulness of metabolic volume in diagnosing malignant tumor in FDG-PET/CT. Jpn J Clin Radiol. 2008;53:997–1003. Kiyohara S, Nagamachi S, Wakamatsu H, Fujita S, Nishii R, Kamimura M, et al. The usefulness of metabolic volume in diagnosing malignant tumor in FDG-PET/CT. Jpn J Clin Radiol. 2008;53:997–1003.
23.
go back to reference Zeng ZC, Tang ZY, Fan J, Zhou J, Qin LX, Ye SL, et al. Consideration of the role of radiotherapy for unresectable intra-hepatic cholangiocarcinoma: a retrospective analysis of 75 patients. Cancer J. 2006;12:113–22.PubMed Zeng ZC, Tang ZY, Fan J, Zhou J, Qin LX, Ye SL, et al. Consideration of the role of radiotherapy for unresectable intra-hepatic cholangiocarcinoma: a retrospective analysis of 75 patients. Cancer J. 2006;12:113–22.PubMed
24.
go back to reference Ben-David MA, Griffith KA, Abu-Isa E, Lawrence TS, Knol J, Zalupski M, et al. External-beam radiotherapy for localized extra-hepatic cholangiocarcinoma. Int J Radiat Oncol Biol Phys. 2006;66:772–9.PubMed Ben-David MA, Griffith KA, Abu-Isa E, Lawrence TS, Knol J, Zalupski M, et al. External-beam radiotherapy for localized extra-hepatic cholangiocarcinoma. Int J Radiat Oncol Biol Phys. 2006;66:772–9.PubMed
25.
go back to reference Takeuchi S, Saito H, Hokodate H, Yuasa N, Takei T, Takamura A. Criteria for Indications for IVR Therapy from the standpoint of MDCT, MRI and Technical Points. Jpn J Diagn Imaging. 2006;26:585–94. Takeuchi S, Saito H, Hokodate H, Yuasa N, Takei T, Takamura A. Criteria for Indications for IVR Therapy from the standpoint of MDCT, MRI and Technical Points. Jpn J Diagn Imaging. 2006;26:585–94.
Metadata
Title
The reproducibility of deep-inspiration breath-hold 18F-FDG PET/CT technique in diagnosing various cancers affected by respiratory motion
Authors
Shigeki Nagamachi
Hideyuki Wakamatsu
Shogo Kiyohara
Seigo Fujita
Shigemi Futami
Hideo Arita
Shozo Tamura
Keiichi Kawai
Publication date
01-04-2010
Publisher
Springer Japan
Published in
Annals of Nuclear Medicine / Issue 3/2010
Print ISSN: 0914-7187
Electronic ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-010-0352-3

Other articles of this Issue 3/2010

Annals of Nuclear Medicine 3/2010 Go to the issue