01-09-2011 | Clinical Research

Is Metal-on-Metal Squeaking Related to Acetabular Angle of Inclination?

Authors: Thomas Bernasek, MD, David Fisher, MD, David Dalury, MD, Melissa Levering, BS, Kirk Dimitris, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 9/2011

Login to get access

Abstract

Background

Postoperative audible squeaking has been well documented in ceramic-on-ceramic hip prostheses, and several metal-on-metal (MOM) THA designs, specifically those used for large-head resurfacing and MOM polyethylene sandwich designs, and are attributed to different implant- and patient-specific factors. Current literature does not identify the incidence of squeaking in modular MOM THA or possible etiologic factors.

Questions/purposes

Our purposes were to (1) identify the incidence of squeaking in modular MOM prostheses in THA; (2) determine whether males or females were more likely to have squeaking; and (3) determine whether the incidence of squeaking relates to acetabular inclination angle.

Methods

We retrospectively reviewed the patient records and radiographs of 539 patients (542 hips) from three independent centers who underwent a MOM THA between February 2001 and December 2005. Demographic and implant factors were evaluated, including measurement of cup inclination angles. The minimum followup was 36 months (mean, 76 months; range, 36–119 months).

Results

We identified squeaking in eight of the 542 hips (1.5%); five were in women and two were in men (one patient had bilateral squeaking). The time to onset of patient-reported audible squeaking averaged 23 months (range, 6–84 months). Squeaking was more likely to occur in women (six of eight hips). No hips with 45º or less acetabular inclination squeaked (291 hips); eight of 251 hips (3.2%) with inclination angles greater than 45º squeaked. Patients who reported squeaking had higher inclination angles than those who did not report squeaking.

Conclusions

Our observations suggest an increased frequency of squeaking in female patients and in patients with greater inclination angles with this MOM implant design.
Literature
1.
go back to reference Back DL, Dalziel R, Young D, Shimmin A. Early results of primary Birmingham hip resurfacings: an independent prospective study of the first 230 hips. J Bone Joint Surg Br. 2005;87:324–329.PubMedCrossRef Back DL, Dalziel R, Young D, Shimmin A. Early results of primary Birmingham hip resurfacings: an independent prospective study of the first 230 hips. J Bone Joint Surg Br. 2005;87:324–329.PubMedCrossRef
2.
go back to reference Brockett CL, Harper P, Williams S, Isaac GH, Dwyer-Joyce RS, Jin Z, Fisher J. The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements. J Mater Sci Mater Med. 2008;19:1575–1579.PubMedCrossRef Brockett CL, Harper P, Williams S, Isaac GH, Dwyer-Joyce RS, Jin Z, Fisher J. The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements. J Mater Sci Mater Med. 2008;19:1575–1579.PubMedCrossRef
3.
go back to reference Clarke IC, Manley MT; Implant Wear Symposium 2007 Engineering Work Group. How do alternative bearing surfaces influence wear behavior? J Am Acad Orthop Surg. 2008;16(suppl 1):S86–93. Clarke IC, Manley MT; Implant Wear Symposium 2007 Engineering Work Group. How do alternative bearing surfaces influence wear behavior? J Am Acad Orthop Surg. 2008;16(suppl 1):S86–93.
4.
go back to reference Glaser D, Dennis DA, Komistek RD, Miner TM. In vivo comparison of hip mechanics for minimally invasive versus traditional total hip arthroplasty. Clin Biomech (Bristol, Avon). 2008;23:127–134.CrossRef Glaser D, Dennis DA, Komistek RD, Miner TM. In vivo comparison of hip mechanics for minimally invasive versus traditional total hip arthroplasty. Clin Biomech (Bristol, Avon). 2008;23:127–134.CrossRef
5.
go back to reference Holloway I, Walter WL, Zicat B, Walter WK. Osteolysis with a cementless second generation metal-on-metal cup in total hip replacement. Int Orthop. 2009;33:1537–1542.PubMedCrossRef Holloway I, Walter WL, Zicat B, Walter WK. Osteolysis with a cementless second generation metal-on-metal cup in total hip replacement. Int Orthop. 2009;33:1537–1542.PubMedCrossRef
6.
go back to reference Jarrett CA, Ranawat AS, Bruzzone M, Blum YC, Rodriguez JA, Ranawat CS. The squeaking hip: a phenomenon of ceramic-on-ceramic total hip arthroplasty. J Bone Joint Surg Am. 2009;91:1344–1349.PubMedCrossRef Jarrett CA, Ranawat AS, Bruzzone M, Blum YC, Rodriguez JA, Ranawat CS. The squeaking hip: a phenomenon of ceramic-on-ceramic total hip arthroplasty. J Bone Joint Surg Am. 2009;91:1344–1349.PubMedCrossRef
7.
go back to reference Keurentjes JC, Kuipers RM, Wever DJ, Schreurs BW. High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings. Clin Orthop Relat Res. 2008;466:1438–1443.PubMedCrossRef Keurentjes JC, Kuipers RM, Wever DJ, Schreurs BW. High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings. Clin Orthop Relat Res. 2008;466:1438–1443.PubMedCrossRef
8.
go back to reference Manley MT, Sutton K. Bearings of the future for total hip arthroplasty. J Arthroplasty. 2008;23(7 suppl):47–50.PubMedCrossRef Manley MT, Sutton K. Bearings of the future for total hip arthroplasty. J Arthroplasty. 2008;23(7 suppl):47–50.PubMedCrossRef
9.
go back to reference Mendenhall Associates Inc. Hospital resources and implant cost management: a 2009 update. Orthopedic Network News. Vol. 21. Birmingham, AL: Mendenhall Associates Inc; 2010:13–19. Mendenhall Associates Inc. Hospital resources and implant cost management: a 2009 update. Orthopedic Network News. Vol. 21. Birmingham, AL: Mendenhall Associates Inc; 2010:13–19.
10.
go back to reference Ranawat AS, Ranawat CS. The squeaking hip: a cause for concern-agrees. Orthopedics. 2007;30:738, 743.PubMed Ranawat AS, Ranawat CS. The squeaking hip: a cause for concern-agrees. Orthopedics. 2007;30:738, 743.PubMed
11.
go back to reference Restrepo C, Parvizi J, Kurtz SM, Sharkey PF, Hozack WJ, Rothman RH. The noisy ceramic hip: is component malpositioning the cause? J Arthroplasty. 2008;23:643–649.PubMedCrossRef Restrepo C, Parvizi J, Kurtz SM, Sharkey PF, Hozack WJ, Rothman RH. The noisy ceramic hip: is component malpositioning the cause? J Arthroplasty. 2008;23:643–649.PubMedCrossRef
12.
go back to reference Taylor S, Manley MT, Sutton K. The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings. J Arthroplasty. 2007;22(7 suppl 3):47–51.PubMedCrossRef Taylor S, Manley MT, Sutton K. The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings. J Arthroplasty. 2007;22(7 suppl 3):47–51.PubMedCrossRef
13.
go back to reference Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22:496–503.PubMedCrossRef Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22:496–503.PubMedCrossRef
14.
go back to reference Walter WL, Waters TS, Gillies M, Donohoo S, Kurtz SM, Ranawat AS, Hozack WJ, Tuke MA. Squeaking hips. J Bone Joint Surg Am. 2008;90(suppl 4):102–111.PubMedCrossRef Walter WL, Waters TS, Gillies M, Donohoo S, Kurtz SM, Ranawat AS, Hozack WJ, Tuke MA. Squeaking hips. J Bone Joint Surg Am. 2008;90(suppl 4):102–111.PubMedCrossRef
15.
go back to reference Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints: a clinical and histomorphological study. J Bone Joint Surg Am. 2005;87:28–36.PubMedCrossRef Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints: a clinical and histomorphological study. J Bone Joint Surg Am. 2005;87:28–36.PubMedCrossRef
16.
go back to reference Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed
Metadata
Title
Is Metal-on-Metal Squeaking Related to Acetabular Angle of Inclination?
Authors
Thomas Bernasek, MD
David Fisher, MD
David Dalury, MD
Melissa Levering, BS
Kirk Dimitris, MD
Publication date
01-09-2011
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 9/2011
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-011-1900-4

Other articles of this Issue 9/2011

Clinical Orthopaedics and Related Research® 9/2011 Go to the issue

Symposium: Reverse Total Shoulder Arthroplasty

The Classic: Delta Shoulder Prosthesis for Rotator Cuff Rupture

Symposium: Reverse Total Shoulder Arthroplasty

How Reverse Shoulder Arthroplasty Works