01-06-2008 | Original Article

High Incidence of Squeaking in THAs with Alumina Ceramic-on-ceramic Bearings

Authors: J. C. Keurentjes, MD, R. M. Kuipers, MD, D. J. Wever, MD, PhD, B. W. Schreurs, MD, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 6/2008

Login to get access

Abstract

Ceramic-on-ceramic bearings in THA are a popular alternative to overcome wear concerns in traditional metal-polyethylene bearings. However, squeaking is a potentially worrisome phenomenon in ceramic-on-ceramic THAs which we observed in some of our patients. We reviewed all 42 patients who underwent 43 ceramic-on-ceramic noncemented THAs during the time of the study. Squeaking, defined as a reproducible sound of squeaking, clicking, or grating, occurred in nine of 43 implants (20.9%). Standard radiographs were normal. We used CT imaging to determine cup anteversion and inclination angles, comparing the squeaking hips with those of a randomly selected control group, but found no differences. We then hypothesized specific design features (stem size, cup size, head size, and neck length of the head) would be risk factors for squeaking. We found a difference in neck length between squeaking and nonsqueaking implants. A neck length of −4 mm or shorter resulted in a relative risk of 5.56 (95% confidence interval, 1.14–27.01) for squeaking. We found a high incidence of squeaking in our population, and we believe this phenomenon is an underreported side effect of these types of bearings. A short neck length of the femoral implant was a risk factor for squeaking in ceramic-on-ceramic THA.
Level of Evidence: Level III, therapeutic study.
Literature
1.
go back to reference Bal BS, Rahaman MN, Aleto T, Miller FS, Traina F, Toni A. The significance of metal staining on alumina femoral heads in total hip arthroplasty. J Arthroplasty. 2007;22:14–19.PubMedCrossRef Bal BS, Rahaman MN, Aleto T, Miller FS, Traina F, Toni A. The significance of metal staining on alumina femoral heads in total hip arthroplasty. J Arthroplasty. 2007;22:14–19.PubMedCrossRef
2.
go back to reference Barrack RL, Burak C, Skinner HB. Concerns about ceramics in THA. Clin Orthop Relat Res. 2004;429:73–79.PubMedCrossRef Barrack RL, Burak C, Skinner HB. Concerns about ceramics in THA. Clin Orthop Relat Res. 2004;429:73–79.PubMedCrossRef
3.
go back to reference Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH. Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes. J Arthroplasty. 2005;20:11–19.PubMedCrossRef Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH. Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes. J Arthroplasty. 2005;20:11–19.PubMedCrossRef
4.
go back to reference D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.PubMedCrossRef D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.PubMedCrossRef
5.
go back to reference D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW Jr. The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am. 2000;82:315–321.PubMed D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW Jr. The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am. 2000;82:315–321.PubMed
6.
go back to reference Eickmann T, Manaka M, Clarke I, Gustafson A. Squeaking and neck-socket impingement in a ceramic total hip arthroplasty. Bioceramics 15. 2003;240–2:849–852. Eickmann T, Manaka M, Clarke I, Gustafson A. Squeaking and neck-socket impingement in a ceramic total hip arthroplasty. Bioceramics 15. 2003;240–2:849–852.
7.
go back to reference Hannouche D, Hamadouche M, Nizard R, Bizot P, Meunier A, Sedel L. Ceramics in total hip replacement. Clin Orthop Relat Res. 2005;430:62–71.PubMedCrossRef Hannouche D, Hamadouche M, Nizard R, Bizot P, Meunier A, Sedel L. Ceramics in total hip replacement. Clin Orthop Relat Res. 2005;430:62–71.PubMedCrossRef
8.
go back to reference Huo MH, Parvizi J, Gilbert NF. What’s new in hip arthroplasty. J Bone Joint Surg Am. 2006;88:2100–2113.PubMedCrossRef Huo MH, Parvizi J, Gilbert NF. What’s new in hip arthroplasty. J Bone Joint Surg Am. 2006;88:2100–2113.PubMedCrossRef
9.
go back to reference Kalteis T, Handel M, Herold T, Perlick L, Paetzel C, Grifka J. Position of the acetabular cup: accuracy of radiographic calculation compared to CT-based measurement. Eur J Radiol. 2006;58:294–300.PubMedCrossRef Kalteis T, Handel M, Herold T, Perlick L, Paetzel C, Grifka J. Position of the acetabular cup: accuracy of radiographic calculation compared to CT-based measurement. Eur J Radiol. 2006;58:294–300.PubMedCrossRef
10.
go back to reference Kobayashi A, Freeman MA, Bonfield W, Kadoya Y, Yamac T, Al-Saffar N, Scott G, Revell PA. Number of polyethylene particles and osteolysis in total joint replacements: a quantitative study using a tissue-digestion method. J Bone Joint Surg Br. 1997;79:844–848.PubMedCrossRef Kobayashi A, Freeman MA, Bonfield W, Kadoya Y, Yamac T, Al-Saffar N, Scott G, Revell PA. Number of polyethylene particles and osteolysis in total joint replacements: a quantitative study using a tissue-digestion method. J Bone Joint Surg Br. 1997;79:844–848.PubMedCrossRef
11.
go back to reference Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed
12.
go back to reference Lombardi AV Jr, Mallory TH, Dennis DA, Komistek RD, Fada RA, Northcut EJ. An in vivo determination of total hip arthroplasty pistoning during activity. J Arthroplasty. 2000;15:702–709.PubMedCrossRef Lombardi AV Jr, Mallory TH, Dennis DA, Komistek RD, Fada RA, Northcut EJ. An in vivo determination of total hip arthroplasty pistoning during activity. J Arthroplasty. 2000;15:702–709.PubMedCrossRef
14.
go back to reference Morlock M, Nassutt R, Janssen R, Willmann G, Honl M. Mismatched wear couple zirconium oxide and aluminum oxide in total hip arthroplasty. J Arthroplasty. 2001;16:1071–1074.PubMedCrossRef Morlock M, Nassutt R, Janssen R, Willmann G, Honl M. Mismatched wear couple zirconium oxide and aluminum oxide in total hip arthroplasty. J Arthroplasty. 2001;16:1071–1074.PubMedCrossRef
15.
go back to reference Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 2000;15:793–795.PubMedCrossRef Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 2000;15:793–795.PubMedCrossRef
16.
go back to reference Nevelos JE, Prudhommeaux F, Hamadouche M, Doyle C, Ingham E, Meunier A, Nevelos AB, Sedel L, Fisher J. Comparative analysis of two different types of alumina-alumina hip prosthesis retrieved for aseptic loosening. J Bone Joint Surg Br. 2001;83:598–603.PubMed Nevelos JE, Prudhommeaux F, Hamadouche M, Doyle C, Ingham E, Meunier A, Nevelos AB, Sedel L, Fisher J. Comparative analysis of two different types of alumina-alumina hip prosthesis retrieved for aseptic loosening. J Bone Joint Surg Br. 2001;83:598–603.PubMed
17.
go back to reference Patek SN. Squeaking with a sliding joint: mechanics and motor control of sound production in palinurid lobsters. J Exp Biol. 2002;205:2375–2385.PubMed Patek SN. Squeaking with a sliding joint: mechanics and motor control of sound production in palinurid lobsters. J Exp Biol. 2002;205:2375–2385.PubMed
18.
go back to reference Schuh A, Holzwarth U, Kachler W, Goske J, Zeiler G. [Titanium deposits on the ceramic heads of dislocated total hip replacements] [in German]. Orthopade. 2004;33:1194–1200.PubMedCrossRef Schuh A, Holzwarth U, Kachler W, Goske J, Zeiler G. [Titanium deposits on the ceramic heads of dislocated total hip replacements] [in German]. Orthopade. 2004;33:1194–1200.PubMedCrossRef
19.
go back to reference Stewart T, Tipper J, Streicher R, Ingham E, Fisher J. Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions. J Mater Sci Mater Med. 2001;12:1053–1056.PubMedCrossRef Stewart T, Tipper J, Streicher R, Ingham E, Fisher J. Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions. J Mater Sci Mater Med. 2001;12:1053–1056.PubMedCrossRef
20.
go back to reference Stewart TD, Tipper JL, Insley G, Streicher RM, Ingham E, Fisher J. Long-term wear of ceramic matrix composite materials for hip prostheses under severe swing phase microseparation. J Biomed Mater Res B Appl Biomater. 2003;66:567–573.PubMedCrossRef Stewart TD, Tipper JL, Insley G, Streicher RM, Ingham E, Fisher J. Long-term wear of ceramic matrix composite materials for hip prostheses under severe swing phase microseparation. J Biomed Mater Res B Appl Biomater. 2003;66:567–573.PubMedCrossRef
21.
go back to reference Stewart TD, Tipper JL, Insley G, Streicher RM, Ingham E, Fisher J. Severe wear and fracture of zirconia heads against alumina inserts in hip simulator studies with microseparation. J Arthroplasty. 2003;18:726–734.PubMedCrossRef Stewart TD, Tipper JL, Insley G, Streicher RM, Ingham E, Fisher J. Severe wear and fracture of zirconia heads against alumina inserts in hip simulator studies with microseparation. J Arthroplasty. 2003;18:726–734.PubMedCrossRef
22.
go back to reference Suzuki K, Matsubara M, Morita S, Muneta T, Shinomiya K. Fracture of a ceramic acetabular insert after ceramic-on-ceramic THA: a case report. Acta Orthop Scand. 2003;74:101–103.PubMedCrossRef Suzuki K, Matsubara M, Morita S, Muneta T, Shinomiya K. Fracture of a ceramic acetabular insert after ceramic-on-ceramic THA: a case report. Acta Orthop Scand. 2003;74:101–103.PubMedCrossRef
23.
go back to reference Taylor S, Manley MT, Sutton K. The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings. J Arthroplasty. 2007;22:47–51.PubMedCrossRef Taylor S, Manley MT, Sutton K. The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings. J Arthroplasty. 2007;22:47–51.PubMedCrossRef
24.
go back to reference Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22:496–503.PubMedCrossRef Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22:496–503.PubMedCrossRef
25.
go back to reference White SP, Blom AW, Lee M, Smith EJ. The crescent sign: dissociation of the polyethylene liner from a modular acetabular component in total hip arthroplasty. Skeletal Radiol. 2005;34:620–624.PubMedCrossRef White SP, Blom AW, Lee M, Smith EJ. The crescent sign: dissociation of the polyethylene liner from a modular acetabular component in total hip arthroplasty. Skeletal Radiol. 2005;34:620–624.PubMedCrossRef
26.
go back to reference Yamamoto T, Saito M, Ueno M, Hananouchi T, Tokugawa Y, Yonenobu K. Wear analysis of retrieved ceramic-on-ceramic articulations in total hip arthroplasty: femoral head makes contact with the rim of the socket outside of the bearing surface. J Biomed Mater Res B Appl Biomater. 2005;73:301–307.PubMed Yamamoto T, Saito M, Ueno M, Hananouchi T, Tokugawa Y, Yonenobu K. Wear analysis of retrieved ceramic-on-ceramic articulations in total hip arthroplasty: femoral head makes contact with the rim of the socket outside of the bearing surface. J Biomed Mater Res B Appl Biomater. 2005;73:301–307.PubMed
27.
go back to reference Yoshimine F, Ginbayashi K. A mathematical formula to calculate the theoretical range of motion for total hip replacement. J Biomech. 2002;35:989–993.PubMedCrossRef Yoshimine F, Ginbayashi K. A mathematical formula to calculate the theoretical range of motion for total hip replacement. J Biomech. 2002;35:989–993.PubMedCrossRef
Metadata
Title
High Incidence of Squeaking in THAs with Alumina Ceramic-on-ceramic Bearings
Authors
J. C. Keurentjes, MD
R. M. Kuipers, MD
D. J. Wever, MD, PhD
B. W. Schreurs, MD, PhD
Publication date
01-06-2008
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 6/2008
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-008-0177-8

Other articles of this Issue 6/2008

Clinical Orthopaedics and Related Research® 6/2008 Go to the issue

Symposium: Papers Presented at the 2007 Meeting of the Musculoskeletal Infection Society

A Preoperative Decolonization Protocol for Staphylococcus aureus Prevents Orthopaedic Infections

Symposium: Selected Papers Presented at the 2007 Meeting of the Musculoskeletal Tumor Society

Predicting Distal Femur Bone Strength in a Murine Model of Tumor Osteolysis

Symposium: Papers Presented at the 2007 Meeting of the Musculoskeletal Infection Society

FDG-PET Imaging Can Diagnose Periprosthetic Infection of the Hip

Symposium: Papers Presented at the 2007 Meeting of the Musculoskeletal Infection Society

Staphylococcus aureus Nasal Decolonization in Joint Replacement Surgery Reduces Infection