01-08-2020 | Original Research

Evaluation of continuous flash glucose monitoring in a pediatric ICU setting

Authors: Eirini Kotzapanagiotou, Eleni Tsotridou, Eleni Volakli, Meropi Dimitriadou, Elpis Chochliourou, Serafeia Kalamitsou, Fani Kotzapanagiotou, Maria Sdougka, Athanasios Christoforidis

Published in: Journal of Clinical Monitoring and Computing | Issue 4/2020

Login to get access

Abstract

Glucose monitoring is of great importance among patients in intensive care units (ICU). The purpose of this study is to assess the performance of a new flash glucose monitoring (FGM) system in a pediatric ICU setting. Sixteen consecutive patients admitted in pediatric ICU aged > 4 years, expected length stay > 2 days and with no medication or existing diagnosis affecting glucose metabolism were enrolled. FreeStyle Libre sensor was applied to the upper arm of the patients (8 boys). FGM measurements were compared to 3 “references”: arterial blood gas analysis, capillary blood analysis and biochemical serum analysis. Mean age of patients was 8.03 ± 2.91 years. Sensors remained in situ for a median of 9.71 ± 5.35 days. Removal of the sensor was mainly attributed to the completion of the predefine life-span of the sensor or discharge of the patient from the ICU. We compared 711 pairs of measurements between the sensor and other glucose measurement methods. Glucose values from the sensor were consistently lower with mean absolute relative difference (MARD) being 28.34%, 25.11% and 18.99% compared to the blood gas analyzer, capillary blood glucose meter, and biochemical serum analysis, respectively, but a wide interindividual variability. Significant linear correlations between age and MARD values were observed. Surveillance error grid (SEG) analysis showed 92.04%, 94.67% and 95.52% of the readings in the none or slight risk zone respectively. FreeStyle Libre is well tolerated although not adequately accurate with a tendency to underestimate glucose levels in critically ill pediatric patients.
Literature
1.
go back to reference Aramendi I, Burghi G, Manzanares W. Dysglycemia in the critically ill patient: current evidence and future perspectives. Rev Bras Ter Intensiv. 2017;29:364–72.CrossRef Aramendi I, Burghi G, Manzanares W. Dysglycemia in the critically ill patient: current evidence and future perspectives. Rev Bras Ter Intensiv. 2017;29:364–72.CrossRef
2.
go back to reference Preiser J-C, Ichai C, Orban J-C, Groeneveld ABJ. Metabolic response to the stress of critical illness. Br J Anaesth. 2014;113:945–54.PubMedCrossRef Preiser J-C, Ichai C, Orban J-C, Groeneveld ABJ. Metabolic response to the stress of critical illness. Br J Anaesth. 2014;113:945–54.PubMedCrossRef
3.
go back to reference van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, et al. Intensive insulin therapy in critically ill patients. N Engl J Med. 2001;345:1359–67.PubMedCrossRef van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, et al. Intensive insulin therapy in critically ill patients. N Engl J Med. 2001;345:1359–67.PubMedCrossRef
4.
go back to reference Krinsley JS. Association between hyperglycemia and increased hospital mortality in a heterogeneous population of critically ill patients. Mayo Clin Proc. 2003;78:1471–8.PubMedCrossRef Krinsley JS. Association between hyperglycemia and increased hospital mortality in a heterogeneous population of critically ill patients. Mayo Clin Proc. 2003;78:1471–8.PubMedCrossRef
5.
go back to reference Gale SC, Sicoutris C, Reilly PM, Schwab CW, Gracias VH. Poor glycemic control is associated with increased mortality in critically ill trauma patients. Am Surg. 2007;73:454–60.PubMed Gale SC, Sicoutris C, Reilly PM, Schwab CW, Gracias VH. Poor glycemic control is associated with increased mortality in critically ill trauma patients. Am Surg. 2007;73:454–60.PubMed
6.
go back to reference van Steen SCJ, Rijkenberg S, Limpens J, van der Voort PHJ, Hermanides J, DeVries JH. The clinical benefits and accuracy of continuous glucose monitoring systems in critically ill patients—a systematic scoping review. Sensors (Basel). 2017;17:146.CrossRef van Steen SCJ, Rijkenberg S, Limpens J, van der Voort PHJ, Hermanides J, DeVries JH. The clinical benefits and accuracy of continuous glucose monitoring systems in critically ill patients—a systematic scoping review. Sensors (Basel). 2017;17:146.CrossRef
7.
go back to reference Mancini G, Berioli MG, Santi E, Rogari F, Toni G, Tascini G, et al. Flash glucose monitoring: a review of the literature with a special focus on type 1 diabetes. Nutrients. 2018;10:992.PubMedCentralCrossRef Mancini G, Berioli MG, Santi E, Rogari F, Toni G, Tascini G, et al. Flash glucose monitoring: a review of the literature with a special focus on type 1 diabetes. Nutrients. 2018;10:992.PubMedCentralCrossRef
8.
go back to reference Hoss U, Budiman ES. Factory-calibrated continuous glucose sensors: the science behind the technology. Diabetes Technol Ther. 2017;19:S44–50.PubMedCrossRef Hoss U, Budiman ES. Factory-calibrated continuous glucose sensors: the science behind the technology. Diabetes Technol Ther. 2017;19:S44–50.PubMedCrossRef
9.
go back to reference Hansen EA, Klee P, Dirlewanger M, Bouthors T, Elowe-Gruau E, Stoppa-Vaucher S, et al. Accuracy, satisfaction and usability of a flash glucose monitoring system among children and adolescents with type 1 diabetes attending a summer camp. Pediatr Diabetes. 2018;19:1276–84.PubMedCrossRef Hansen EA, Klee P, Dirlewanger M, Bouthors T, Elowe-Gruau E, Stoppa-Vaucher S, et al. Accuracy, satisfaction and usability of a flash glucose monitoring system among children and adolescents with type 1 diabetes attending a summer camp. Pediatr Diabetes. 2018;19:1276–84.PubMedCrossRef
10.
go back to reference Edge J, Acerini C, Campbell F, Hamilton-Shield J, Moudiotis C, Rahman S, et al. An alternative sensor-based method for glucose monitoring in children and young people with diabetes. Arch Dis Child. 2017;102:543–9.PubMedPubMedCentralCrossRef Edge J, Acerini C, Campbell F, Hamilton-Shield J, Moudiotis C, Rahman S, et al. An alternative sensor-based method for glucose monitoring in children and young people with diabetes. Arch Dis Child. 2017;102:543–9.PubMedPubMedCentralCrossRef
11.
go back to reference Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Mårtensson J. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit Care Resusc. 2017;19:167–74.PubMed Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Mårtensson J. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit Care Resusc. 2017;19:167–74.PubMed
12.
go back to reference Schierenbeck F, Franco-Cereceda A, Liska J. Accuracy of 2 different continuous glucose monitoring systems in patients undergoing cardiac surgery. J Diabetes Sci Technol. 2017;11:108–16.PubMedCrossRef Schierenbeck F, Franco-Cereceda A, Liska J. Accuracy of 2 different continuous glucose monitoring systems in patients undergoing cardiac surgery. J Diabetes Sci Technol. 2017;11:108–16.PubMedCrossRef
13.
go back to reference Deja G, Kleczek M, Chumiecki M, Strzala-Kleczek A, Deja R, Jarosz-Chobot P. The usefulness of the FlashStyle Libre system in glycemic control in children with type 1 diabetes during summer camp. Pediatr Endocrinol Diabetes Metab. 2018;24:11–9.PubMedCrossRef Deja G, Kleczek M, Chumiecki M, Strzala-Kleczek A, Deja R, Jarosz-Chobot P. The usefulness of the FlashStyle Libre system in glycemic control in children with type 1 diabetes during summer camp. Pediatr Endocrinol Diabetes Metab. 2018;24:11–9.PubMedCrossRef
14.
go back to reference Piona C, Dovc K, Mutlu GY, Grad K, Gregorc P, Battelino T, et al. Non-adjunctive flash glucose monitoring system use during summer-camp in children with type 1 diabetes: the free-summer study. Pediatr Diabetes. 2018;19:1285–93.PubMedCrossRef Piona C, Dovc K, Mutlu GY, Grad K, Gregorc P, Battelino T, et al. Non-adjunctive flash glucose monitoring system use during summer-camp in children with type 1 diabetes: the free-summer study. Pediatr Diabetes. 2018;19:1285–93.PubMedCrossRef
15.
go back to reference Szadkowska A, Gawrecki A, Michalak A, Zozulinska-Ziolkiewicz D, Fendler W, Mlynarski W. Flash glucose measurements in children with type 1 diabetes in real-life settings: to trust or not to trust? Diabetes Technol Ther. 2018;20:17–24.PubMedCrossRef Szadkowska A, Gawrecki A, Michalak A, Zozulinska-Ziolkiewicz D, Fendler W, Mlynarski W. Flash glucose measurements in children with type 1 diabetes in real-life settings: to trust or not to trust? Diabetes Technol Ther. 2018;20:17–24.PubMedCrossRef
16.
go back to reference Kuczmarski RJ, Ogden CL, Guo SS, Grummer-Strawn LM, Flegal KM, Mei Z, et al. CDC growth charts for the United States: methods and development. Vital Health Stat. 2000;11(2002):1–190. Kuczmarski RJ, Ogden CL, Guo SS, Grummer-Strawn LM, Flegal KM, Mei Z, et al. CDC growth charts for the United States: methods and development. Vital Health Stat. 2000;11(2002):1–190.
17.
go back to reference Basu A, Dube S, Slama M, Errazuriz I, Amezcua JC, Kudva YC, et al. Time lag of glucose from intravascular to interstitial compartment in humans. Diabetes. 2013;62:4083–7.PubMedPubMedCentralCrossRef Basu A, Dube S, Slama M, Errazuriz I, Amezcua JC, Kudva YC, et al. Time lag of glucose from intravascular to interstitial compartment in humans. Diabetes. 2013;62:4083–7.PubMedPubMedCentralCrossRef
18.
go back to reference Bonora B, Maran A, Ciciliot S, Avogaro A, Fadini GP. Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetes. J Endocrinol Invest. 2016;39:1391–9.PubMedCrossRef Bonora B, Maran A, Ciciliot S, Avogaro A, Fadini GP. Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetes. J Endocrinol Invest. 2016;39:1391–9.PubMedCrossRef
19.
go back to reference Hulse A, Rai S, Kumar KP. Evaluation of accuracy of ambulatory glucose profile in an outpatient setting in children with type 1 diabetes. Indian J Endocrinol Metab. 2016;20:643.PubMedPubMedCentralCrossRef Hulse A, Rai S, Kumar KP. Evaluation of accuracy of ambulatory glucose profile in an outpatient setting in children with type 1 diabetes. Indian J Endocrinol Metab. 2016;20:643.PubMedPubMedCentralCrossRef
20.
go back to reference Aberer F, Hajnsek M, Rumpler M, Zenz S, Baumann PM, Elsayed H, et al. Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes. Diabetes Obes Metab. 2017;19:1051–5.PubMedCrossRef Aberer F, Hajnsek M, Rumpler M, Zenz S, Baumann PM, Elsayed H, et al. Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes. Diabetes Obes Metab. 2017;19:1051–5.PubMedCrossRef
21.
go back to reference Fokkert MJ, Van Dijk PR, Edens MA, Abbes S, De Jong D, Slingerland RJ, et al. Performance of the freestyle Libre flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. BMJ Open Diabetes Res Care. 2017;5:e000320.PubMedPubMedCentralCrossRef Fokkert MJ, Van Dijk PR, Edens MA, Abbes S, De Jong D, Slingerland RJ, et al. Performance of the freestyle Libre flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. BMJ Open Diabetes Res Care. 2017;5:e000320.PubMedPubMedCentralCrossRef
22.
go back to reference Theodorsson E, Dahlqvist S, Attvall S, Sandgren U, Skrtic S, Lind M, et al. A Clinical trial of the accuracy and treatment experience of the flash glucose monitor freestyle Libre in adults with type 1 diabetes. Diabetes Technol Ther. 2017;19:164–72.PubMedPubMedCentralCrossRef Theodorsson E, Dahlqvist S, Attvall S, Sandgren U, Skrtic S, Lind M, et al. A Clinical trial of the accuracy and treatment experience of the flash glucose monitor freestyle Libre in adults with type 1 diabetes. Diabetes Technol Ther. 2017;19:164–72.PubMedPubMedCentralCrossRef
23.
go back to reference Amato AML, Avogaro A, Facchinetti A, Galasso S, Boscari F, Marescotti MC, et al. FreeStyle Libre and Dexcom G4 platinum sensors: accuracy comparisons during two weeks of home use and use during experimentally induced glucose excursions. Nutr Metab Cardiovasc Dis. 2017;28:180–6.PubMed Amato AML, Avogaro A, Facchinetti A, Galasso S, Boscari F, Marescotti MC, et al. FreeStyle Libre and Dexcom G4 platinum sensors: accuracy comparisons during two weeks of home use and use during experimentally induced glucose excursions. Nutr Metab Cardiovasc Dis. 2017;28:180–6.PubMed
24.
go back to reference Voelmle M, Garg SK, Lopez-Baca B, Smith J, Beatson C, Gottlieb PA. Comparison of accuracy and safety of the seven and the navigator continuous glucose monitoring systems. Diabetes Technol Ther. 2009;11:65–72.PubMedCrossRef Voelmle M, Garg SK, Lopez-Baca B, Smith J, Beatson C, Gottlieb PA. Comparison of accuracy and safety of the seven and the navigator continuous glucose monitoring systems. Diabetes Technol Ther. 2009;11:65–72.PubMedCrossRef
25.
go back to reference Damiano ER, McKeon K, El-Khatib FH, Zheng H, Nathan DM, Russell SJ. A comparative effectiveness analysis of three continuous glucose monitors: the navigator, G4 platinum, and enlite. J Diabetes Sci Technol. 2014;8:699–708.PubMedPubMedCentralCrossRef Damiano ER, McKeon K, El-Khatib FH, Zheng H, Nathan DM, Russell SJ. A comparative effectiveness analysis of three continuous glucose monitors: the navigator, G4 platinum, and enlite. J Diabetes Sci Technol. 2014;8:699–708.PubMedPubMedCentralCrossRef
26.
go back to reference Rijkenberg S, van Steen SC, DeVries JH, van der Voort PHJ. Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients. J Clin Monit Comput. 2018;32:953–64.PubMedCrossRef Rijkenberg S, van Steen SC, DeVries JH, van der Voort PHJ. Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients. J Clin Monit Comput. 2018;32:953–64.PubMedCrossRef
27.
go back to reference Siegelaar SE, Barwari T, Hermanides J, van der Voort PHJ, Hoekstra JBL, DeVries JH. Microcirculation and its relation to continuous subcutaneous glucose sensor accuracy in cardiac surgery patients in the intensive care unit. J Thorac Cardiovasc Surg. 2013;146:1283–9.PubMedCrossRef Siegelaar SE, Barwari T, Hermanides J, van der Voort PHJ, Hoekstra JBL, DeVries JH. Microcirculation and its relation to continuous subcutaneous glucose sensor accuracy in cardiac surgery patients in the intensive care unit. J Thorac Cardiovasc Surg. 2013;146:1283–9.PubMedCrossRef
28.
go back to reference Marics G, Koncz L, Eitler K, Vatai B, Szenasi B, Zakarias D, et al. Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit. Ital J Pediatr. 2015;41:17.PubMedPubMedCentralCrossRef Marics G, Koncz L, Eitler K, Vatai B, Szenasi B, Zakarias D, et al. Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit. Ital J Pediatr. 2015;41:17.PubMedPubMedCentralCrossRef
29.
go back to reference Massa GG, Gys I, Op’t Eyndt A, Bevilacqua E, Wijnands A, Declercq P, et al. Evaluation of the FreeStyle(R) Libre flash glucose monitoring system in children and adolescents with type 1 diabetes. Horm Res Paediatr. 2018;89:189–99.PubMedCrossRef Massa GG, Gys I, Op’t Eyndt A, Bevilacqua E, Wijnands A, Declercq P, et al. Evaluation of the FreeStyle(R) Libre flash glucose monitoring system in children and adolescents with type 1 diabetes. Horm Res Paediatr. 2018;89:189–99.PubMedCrossRef
30.
go back to reference Hulse A, Rai S, Prasanna Kumar KM. Evaluation of accuracy of ambulatory glucose profile in an outpatient setting in children with type 1 diabetes. Indian J Endocrinol Metab. 2016;20:643–7.PubMedPubMedCentralCrossRef Hulse A, Rai S, Prasanna Kumar KM. Evaluation of accuracy of ambulatory glucose profile in an outpatient setting in children with type 1 diabetes. Indian J Endocrinol Metab. 2016;20:643–7.PubMedPubMedCentralCrossRef
31.
go back to reference Rai S, Hulse A, Kumar P. Feasibility and acceptability of ambulatory glucose profile in children with type 1 diabetes mellitus: a pilot study. Indian J Endocrinol Metab. 2016;20:790–4.PubMedPubMedCentralCrossRef Rai S, Hulse A, Kumar P. Feasibility and acceptability of ambulatory glucose profile in children with type 1 diabetes mellitus: a pilot study. Indian J Endocrinol Metab. 2016;20:790–4.PubMedPubMedCentralCrossRef
32.
go back to reference Alsaffar H, Turner L, Yung Z, Didi M, Senniappan S. Continuous flash glucose monitoring in children with congenital hyperinsulinism; first report on accuracy and patient experience. Int J Pediatr Endocrinol. 2018;2018:3.PubMedPubMedCentralCrossRef Alsaffar H, Turner L, Yung Z, Didi M, Senniappan S. Continuous flash glucose monitoring in children with congenital hyperinsulinism; first report on accuracy and patient experience. Int J Pediatr Endocrinol. 2018;2018:3.PubMedPubMedCentralCrossRef
33.
go back to reference Staal OM, Hansen HMU, Christiansen SC, Fougner AL, Carlsen SM, Stavdahl O. Differences between flash glucose monitor and fingerprick measurements. Biosensors. 2018;8:93.PubMedCentralCrossRef Staal OM, Hansen HMU, Christiansen SC, Fougner AL, Carlsen SM, Stavdahl O. Differences between flash glucose monitor and fingerprick measurements. Biosensors. 2018;8:93.PubMedCentralCrossRef
34.
go back to reference Sato T, Oshima H, Nakata K, Kimura Y, Yano T, Furuhashi M, et al. Accuracy of flash glucose monitoring in insulin-treated patients with type 2 diabetes. J Diabetes Investig. 2018;10:846.PubMedPubMedCentralCrossRef Sato T, Oshima H, Nakata K, Kimura Y, Yano T, Furuhashi M, et al. Accuracy of flash glucose monitoring in insulin-treated patients with type 2 diabetes. J Diabetes Investig. 2018;10:846.PubMedPubMedCentralCrossRef
35.
go back to reference Olafsdottir AF, Attvall S, Sandgren U, Dahlqvist S, Pivodic A, Skrtic S, et al. A clinical trial of the accuracy and treatment experience of the flash glucose monitor FreeStyle Libre in adults with type 1 diabetes. Diabetes Technol Ther. 2017;19:164–72.PubMedPubMedCentralCrossRef Olafsdottir AF, Attvall S, Sandgren U, Dahlqvist S, Pivodic A, Skrtic S, et al. A clinical trial of the accuracy and treatment experience of the flash glucose monitor FreeStyle Libre in adults with type 1 diabetes. Diabetes Technol Ther. 2017;19:164–72.PubMedPubMedCentralCrossRef
36.
go back to reference Ji L, Guo X, Guo L, Ren Q, Yu N, Zhang J. A multicenter evaluation of the performance and usability of a novel glucose monitoring system in Chinese adults with diabetes. J Diabetes Sci Technol. 2017;11:290–5.PubMedPubMedCentralCrossRef Ji L, Guo X, Guo L, Ren Q, Yu N, Zhang J. A multicenter evaluation of the performance and usability of a novel glucose monitoring system in Chinese adults with diabetes. J Diabetes Sci Technol. 2017;11:290–5.PubMedPubMedCentralCrossRef
37.
go back to reference Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The performance and usability of a factory-calibrated flash glucose monitoring system. Diabetes Technol Ther. 2015;17:787–94.PubMedPubMedCentralCrossRef Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The performance and usability of a factory-calibrated flash glucose monitoring system. Diabetes Technol Ther. 2015;17:787–94.PubMedPubMedCentralCrossRef
38.
go back to reference Fokkert MJ, van Dijk PR, Edens MA, Abbes S, de Jong D, Slingerland RJ, et al. Performance of the FreeStyle Libre flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. BMJ Open Diabetes Res Care. 2017;5:e000320.PubMedPubMedCentralCrossRef Fokkert MJ, van Dijk PR, Edens MA, Abbes S, de Jong D, Slingerland RJ, et al. Performance of the FreeStyle Libre flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. BMJ Open Diabetes Res Care. 2017;5:e000320.PubMedPubMedCentralCrossRef
39.
go back to reference Klaff LJ, Christiansen MP, Bode BW, Alva S, Bailey TS, et al. Response to the comment on “the performance and usability of a factory-calibrated flash glucose monitoring system” by Bailey. Diabetes Technol Ther. 2016;18:336–7.PubMedCrossRef Klaff LJ, Christiansen MP, Bode BW, Alva S, Bailey TS, et al. Response to the comment on “the performance and usability of a factory-calibrated flash glucose monitoring system” by Bailey. Diabetes Technol Ther. 2016;18:336–7.PubMedCrossRef
Metadata
Title
Evaluation of continuous flash glucose monitoring in a pediatric ICU setting
Authors
Eirini Kotzapanagiotou
Eleni Tsotridou
Eleni Volakli
Meropi Dimitriadou
Elpis Chochliourou
Serafeia Kalamitsou
Fani Kotzapanagiotou
Maria Sdougka
Athanasios Christoforidis
Publication date
01-08-2020
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 4/2020
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00384-y

Other articles of this Issue 4/2020

Journal of Clinical Monitoring and Computing 4/2020 Go to the issue