Single and dual coil shock efficacy and predictors of shock failure in patients with modern implantable cardioverter defibrillators—a single-center paired randomized study

Authors: Jacob Moesgaard Larsen, Finn P. Heath, Sam Riahi, Katja Holm, Martin B. Johansen, Søren P. Hjortshøj, Anna M. Thøgersen

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 1/2019

Login to get access

Abstract

Purpose

Implantable cardioverter defibrillators (ICDs) can treat life-threatening tachyarrhythmia with high-voltage shocks. The aims were to compare the efficacy of single and dual coil shock vectors in modern ICDs and to identify predictors of shock failure.

Methods

This is a single-center paired randomized study including 216 patients with mixed indications and ICDs from four manufacturers. All patients underwent two implant defibrillation tests using single and dual coil vectors with the test order randomized. Tested shock energy differed slightly between manufacturers because of differences in device programmability: first shock approximately 15 J below maximal output—if failed, second shock approximately 10 J below maximal output—if failed, third shock at maximal output.

Results

First shock success rate was 399/432 (92.4%). Comparing single and dual coil vectors, no differences were seen in first shock efficacy (91.7% vs. 93.1%, P = 0.629) or lowest tested succesfully stored energy (27.2 J vs. 27.1 J, P = 0.620). All successive internal shocks failed in 4/432 (0.9%) of inductions requiring external rescue shocks to restore circulation. Multivariate predictors of first shock failure were QRS duration (relative risk 0.81 per 10 ms, P = 0.001), amiodarone treatment (relative risk 3.30, P = 0.003), and body height (relative risk 1.70 per 10 cm, P = 0.019).

Conclusions

Implant defibrillation testing of modern intravenous ICD systems demonstrates high shock efficacy with no difference between single and dual coil vectors.
Literature
1.
go back to reference Nanthakumar K, Epstein AE, Kay GN, Plumb VJ, Lee DS. Prophylactic implantable cardioverter-defibrillator therapy in patients with left ventricular systolic dysfunction: a pooled analysis of 10 primary prevention trials. J Am Coll Cardiol. 2004;44:2166–72.CrossRefPubMed Nanthakumar K, Epstein AE, Kay GN, Plumb VJ, Lee DS. Prophylactic implantable cardioverter-defibrillator therapy in patients with left ventricular systolic dysfunction: a pooled analysis of 10 primary prevention trials. J Am Coll Cardiol. 2004;44:2166–72.CrossRefPubMed
2.
go back to reference Connolly SJ, Hallstrom AP, Cappato R, Schron EB, Kuck KH, Zipes DP, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Eur Heart J. 2000;21:2071–8.CrossRefPubMed Connolly SJ, Hallstrom AP, Cappato R, Schron EB, Kuck KH, Zipes DP, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Eur Heart J. 2000;21:2071–8.CrossRefPubMed
3.
go back to reference Gold MR, et al. Safety of a single successful conversion of ventricular fibrillation before the implantation of cardioverter defibrillators. Pacing Clin Electrophysiol. 2003;26:483–6.CrossRefPubMed Gold MR, et al. Safety of a single successful conversion of ventricular fibrillation before the implantation of cardioverter defibrillators. Pacing Clin Electrophysiol. 2003;26:483–6.CrossRefPubMed
4.
go back to reference Bardy GH, Dolack GL, Kudenchuk PJ, Poole JE, Mehra R, Johnson G. Prospective, randomized comparison in humans of a unipolar defibrillation system with that using an additional superior vena cava electrode. Circulation. 1994;89:1090–3.CrossRefPubMed Bardy GH, Dolack GL, Kudenchuk PJ, Poole JE, Mehra R, Johnson G. Prospective, randomized comparison in humans of a unipolar defibrillation system with that using an additional superior vena cava electrode. Circulation. 1994;89:1090–3.CrossRefPubMed
5.
go back to reference Schulte B, Sperzel J, Carlsson J, Schwarz T, Ehrlich W, Pitschner HF, et al. Dual-coil vs single-coil active pectoral implantable defibrillator lead systems: defibrillation energy requirements and probability of defibrillation success at multiples of the defibrillation energy requirements. Europace. 2001;3:177–80.CrossRefPubMed Schulte B, Sperzel J, Carlsson J, Schwarz T, Ehrlich W, Pitschner HF, et al. Dual-coil vs single-coil active pectoral implantable defibrillator lead systems: defibrillation energy requirements and probability of defibrillation success at multiples of the defibrillation energy requirements. Europace. 2001;3:177–80.CrossRefPubMed
6.
go back to reference Rinaldi CA, et al. A randomized prospective study of single coil versus dual coil defibrillation in patients with ventricular arrhythmias undergoing implantable cardioverter defibrillator therapy. Pacing Clin Electrophysiol. 2003;26:1684–90.CrossRefPubMed Rinaldi CA, et al. A randomized prospective study of single coil versus dual coil defibrillation in patients with ventricular arrhythmias undergoing implantable cardioverter defibrillator therapy. Pacing Clin Electrophysiol. 2003;26:1684–90.CrossRefPubMed
7.
go back to reference Gold MR, Foster AH, Shorofsky SR. Lead system optimization for transvenous defibrillation. Am J Cardiol. 1997;80:1163–7.CrossRefPubMed Gold MR, Foster AH, Shorofsky SR. Lead system optimization for transvenous defibrillation. Am J Cardiol. 1997;80:1163–7.CrossRefPubMed
8.
go back to reference Gold MR, Olsovsky MR, Pelini MA, Peters RW, Shorofsky SR. Comparison of single- and dual-coil active pectoral defibrillation lead systems. J Am Coll Cardiol. 1998;31:1391–4.CrossRefPubMed Gold MR, Olsovsky MR, Pelini MA, Peters RW, Shorofsky SR. Comparison of single- and dual-coil active pectoral defibrillation lead systems. J Am Coll Cardiol. 1998;31:1391–4.CrossRefPubMed
9.
go back to reference Gold MR, et al. Optimization of transvenous coil position for active can defibrillation thresholds. J Cardiovasc Electrophysiol. 2000;11:25–9.CrossRefPubMed Gold MR, et al. Optimization of transvenous coil position for active can defibrillation thresholds. J Cardiovasc Electrophysiol. 2000;11:25–9.CrossRefPubMed
10.
go back to reference Libero L, Lozano IF, Bocchiardo M, Marcolongo M, Sallusti L, Madrid A, et al. Comparison of defibrillation thresholds using monodirectional electrical vector versus bidirectional electrical vector. Ital Heart J. 2001;2:449–55.PubMed Libero L, Lozano IF, Bocchiardo M, Marcolongo M, Sallusti L, Madrid A, et al. Comparison of defibrillation thresholds using monodirectional electrical vector versus bidirectional electrical vector. Ital Heart J. 2001;2:449–55.PubMed
11.
go back to reference Gold M, Val-Mejias J, Leman RB, Tummala R, Goyal S, Kluger J, et al. Optimization of superior vena cava coil position and usage for transvenous defibrillation. Heart Rhythm. 2008;5:394–9.CrossRefPubMed Gold M, Val-Mejias J, Leman RB, Tummala R, Goyal S, Kluger J, et al. Optimization of superior vena cava coil position and usage for transvenous defibrillation. Heart Rhythm. 2008;5:394–9.CrossRefPubMed
12.
go back to reference Bongiorni MG, Proclemer A, Dobreanu D, Marinskis G, Pison L, Blomstrom-Lundqvist C, et al. Preferred tools and techniques for implantation of cardiac electronic devices in Europe: results of the European Heart Rhythm Association survey. Europace. 2013;15:1664–8.CrossRefPubMed Bongiorni MG, Proclemer A, Dobreanu D, Marinskis G, Pison L, Blomstrom-Lundqvist C, et al. Preferred tools and techniques for implantation of cardiac electronic devices in Europe: results of the European Heart Rhythm Association survey. Europace. 2013;15:1664–8.CrossRefPubMed
13.
go back to reference Baccillieri MS, Gasparini G, Benacchio L, Zorzi A, Marras E, Zerbo F, et al. Multicentre comparison of shock efficacy using single-vs. dual-coil lead systems and anodal vs. cathodaL polarITY defibrillation in patients undergoing transvenous cardioverter-defibrillator implantation. The MODALITY study. J Interv Card Electrophysiol. 2015;43:45–54.CrossRefPubMed Baccillieri MS, Gasparini G, Benacchio L, Zorzi A, Marras E, Zerbo F, et al. Multicentre comparison of shock efficacy using single-vs. dual-coil lead systems and anodal vs. cathodaL polarITY defibrillation in patients undergoing transvenous cardioverter-defibrillator implantation. The MODALITY study. J Interv Card Electrophysiol. 2015;43:45–54.CrossRefPubMed
14.
go back to reference Hsu JC, Saxon LA, Jones PW, Wehrenberg S, Marcus GM. Utilization trends and clinical outcomes in patients implanted with a single- vs a dual-coil implantable cardioverter-defibrillator lead: insights from the ALTITUDE study. Heart Rhythm. 2015;12:1770–5.CrossRefPubMed Hsu JC, Saxon LA, Jones PW, Wehrenberg S, Marcus GM. Utilization trends and clinical outcomes in patients implanted with a single- vs a dual-coil implantable cardioverter-defibrillator lead: insights from the ALTITUDE study. Heart Rhythm. 2015;12:1770–5.CrossRefPubMed
15.
go back to reference Larsen JM, Hjortshøj SP, Nielsen JC, Johansen JB, Petersen HH, Haarbo J, et al. Single-coil and dual-coil defibrillator leads and association with clinical outcomes in a complete Danish nationwide ICD cohort. Heart Rhythm. 2016;13:706–12.CrossRefPubMed Larsen JM, Hjortshøj SP, Nielsen JC, Johansen JB, Petersen HH, Haarbo J, et al. Single-coil and dual-coil defibrillator leads and association with clinical outcomes in a complete Danish nationwide ICD cohort. Heart Rhythm. 2016;13:706–12.CrossRefPubMed
16.
go back to reference Bansch D, et al. Shock efficacy of single and dual coil electrodes-new insights from the NORDIC ICD trial. Europace. 2017;20:971–8.CrossRef Bansch D, et al. Shock efficacy of single and dual coil electrodes-new insights from the NORDIC ICD trial. Europace. 2017;20:971–8.CrossRef
18.
go back to reference Epstein LM, Love CJ, Wilkoff BL, Chung MK, Hackler JW, Bongiorni MG, et al. Superior vena cava defibrillator coils make transvenous lead extraction more challenging and riskier. J Am Coll Cardiol. 2013;61:987–9.CrossRefPubMed Epstein LM, Love CJ, Wilkoff BL, Chung MK, Hackler JW, Bongiorni MG, et al. Superior vena cava defibrillator coils make transvenous lead extraction more challenging and riskier. J Am Coll Cardiol. 2013;61:987–9.CrossRefPubMed
19.
go back to reference Healey JS, Hohnloser SH, Glikson M, Neuzner J, Mabo P, Vinolas X, et al. Cardioverter defibrillator implantation without induction of ventricular fibrillation: a single-blind, non-inferiority, randomised controlled trial (SIMPLE). Lancet. 2015;385:785–91.CrossRefPubMed Healey JS, Hohnloser SH, Glikson M, Neuzner J, Mabo P, Vinolas X, et al. Cardioverter defibrillator implantation without induction of ventricular fibrillation: a single-blind, non-inferiority, randomised controlled trial (SIMPLE). Lancet. 2015;385:785–91.CrossRefPubMed
20.
go back to reference Bansch D, et al. Intra-operative defibrillation testing and clinical shock efficacy in patients with implantable cardioverter-defibrillators: the NORDIC ICD randomized clinical trial. Eur Heart J. 2015;36:2500–7.CrossRefPubMedPubMedCentral Bansch D, et al. Intra-operative defibrillation testing and clinical shock efficacy in patients with implantable cardioverter-defibrillators: the NORDIC ICD randomized clinical trial. Eur Heart J. 2015;36:2500–7.CrossRefPubMedPubMedCentral
21.
go back to reference Zou G. A modified poisson regression approach to prospective studies with binary data. Am J Epidemiol. 2004;159:702–6.CrossRefPubMed Zou G. A modified poisson regression approach to prospective studies with binary data. Am J Epidemiol. 2004;159:702–6.CrossRefPubMed
22.
go back to reference Birnie D, Tung S, Simpson C, Crystal E, Exner D, Ayala Paredes FA, et al. Complications associated with defibrillation threshold testing: the Canadian experience. Heart Rhythm. 2008;5:387–90.CrossRefPubMed Birnie D, Tung S, Simpson C, Crystal E, Exner D, Ayala Paredes FA, et al. Complications associated with defibrillation threshold testing: the Canadian experience. Heart Rhythm. 2008;5:387–90.CrossRefPubMed
23.
go back to reference Segreti L, di Cori A, Soldati E, Zucchelli G, Viani S, Paperini L, et al. Major predictors of fibrous adherences in transvenous implantable cardioverter-defibrillator lead extraction. Heart Rhythm. 2014;11:2196–201.CrossRefPubMed Segreti L, di Cori A, Soldati E, Zucchelli G, Viani S, Paperini L, et al. Major predictors of fibrous adherences in transvenous implantable cardioverter-defibrillator lead extraction. Heart Rhythm. 2014;11:2196–201.CrossRefPubMed
24.
go back to reference Di Cori A, et al. Transvenous extraction performance of expanded polytetrafluoroethylene covered ICD leads in comparison to traditional ICD leads in humans. Pacing Clin Electrophysiol. 2010;33:1376–81.CrossRefPubMed Di Cori A, et al. Transvenous extraction performance of expanded polytetrafluoroethylene covered ICD leads in comparison to traditional ICD leads in humans. Pacing Clin Electrophysiol. 2010;33:1376–81.CrossRefPubMed
25.
go back to reference Shukla HH, et al. High defibrillation thresholds in transvenous biphasic implantable defibrillators: clinical predictors and prognostic implications. Pacing Clin Electrophysiol. 2003;26:44–8.CrossRefPubMed Shukla HH, et al. High defibrillation thresholds in transvenous biphasic implantable defibrillators: clinical predictors and prognostic implications. Pacing Clin Electrophysiol. 2003;26:44–8.CrossRefPubMed
26.
go back to reference Russo AM, Sauer W, Gerstenfeld EP, Hsia HH, Lin D, Cooper JM, et al. Defibrillation threshold testing: is it really necessary at the time of implantable cardioverter-defibrillator insertion? Heart Rhythm. 2005;2:456–61.CrossRefPubMed Russo AM, Sauer W, Gerstenfeld EP, Hsia HH, Lin D, Cooper JM, et al. Defibrillation threshold testing: is it really necessary at the time of implantable cardioverter-defibrillator insertion? Heart Rhythm. 2005;2:456–61.CrossRefPubMed
27.
go back to reference Varma N, Efimov I. Right pectoral implantable cardioverter defibrillators: role of the proximal (SVC) coil. Pacing Clin Electrophysiol. 2008;31:1025–35.CrossRefPubMed Varma N, Efimov I. Right pectoral implantable cardioverter defibrillators: role of the proximal (SVC) coil. Pacing Clin Electrophysiol. 2008;31:1025–35.CrossRefPubMed
28.
go back to reference Verma A, Kaplan AJ, Sarak B, Oosthuizen R, Beardsall M, Higgenbottam J, et al. Incidence of very high defibrillation thresholds (DFT) and efficacy of subcutaneous (SQ) array insertion during implantable cardioverter defibrillator (ICD) implantation. J Interv Card Electrophysiol. 2010;29:127–33.CrossRefPubMed Verma A, Kaplan AJ, Sarak B, Oosthuizen R, Beardsall M, Higgenbottam J, et al. Incidence of very high defibrillation thresholds (DFT) and efficacy of subcutaneous (SQ) array insertion during implantable cardioverter defibrillator (ICD) implantation. J Interv Card Electrophysiol. 2010;29:127–33.CrossRefPubMed
29.
go back to reference Wilkoff BL, et al. HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Europace 2016. 2015;18:159–83. Wilkoff BL, et al. HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Europace 2016. 2015;18:159–83.
30.
go back to reference Osswald BR, de Simone R, Most S, Tochtermann U, Tanzeem A, Karck M. High defibrillation threshold in patients with implantable defibrillator: how effective is the subcutaneous finger lead? Eur J Cardiothorac Surg. 2009;35:489–92.CrossRefPubMed Osswald BR, de Simone R, Most S, Tochtermann U, Tanzeem A, Karck M. High defibrillation threshold in patients with implantable defibrillator: how effective is the subcutaneous finger lead? Eur J Cardiothorac Surg. 2009;35:489–92.CrossRefPubMed
Metadata
Title
Single and dual coil shock efficacy and predictors of shock failure in patients with modern implantable cardioverter defibrillators—a single-center paired randomized study
Authors
Jacob Moesgaard Larsen
Finn P. Heath
Sam Riahi
Katja Holm
Martin B. Johansen
Søren P. Hjortshøj
Anna M. Thøgersen
Publication date
01-01-2019
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2019
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-018-0443-y

Other articles of this Issue 1/2019

Journal of Interventional Cardiac Electrophysiology 1/2019 Go to the issue