Long-time “real-life” performance of the subcutaneous ICD in patients with electrical heart disease or idiopathic ventricular fibrillation

Authors: Gerrit Frommeyer, Dirk G. Dechering, Simon Kochhäuser, Markus Bettin, Julia Köbe, Lars Eckardt, Florian Reinke

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2016

Login to get access

Abstract

Backgrounds

The totally subcutaneous implantable defibrillator (S-ICD) has been designed as a new alternative to conventional implantable defibrillators. This system is especially attractive for young patients. However, long-term experience is not yet available. To address the question whether the S-ICD system is safe and feasible for young patients with electrical heart disease or idiopathic ventricular fibrillation (VF), the data of a standard of care prospective single center S-ICD registry were evaluated.

Methods

In the present study, 24 patients (age 34.2 ± 11.5 years) with electrical heart disease or idiopathic VF received an S-ICD for primary (n = 8) or secondary prevention (n = 16). The mean follow-up duration was 29.6 ± 15.1 months.

Results

Ventricular arrhythmias were adequately detected in four patients (17 %). In three patients (13 %) oversensing was noticed and led to at least one inappropriate shock in two patients (8 %). Further adverse events included surgical revision due to a mobile pulse generator as well as explantation of one system and switch to a transvenous ICD system due to several ineffective shocks.

Conclusions

The results of the present study suggest that S-ICD therapy may be an attractive alternative in young patients with electrical heart disease or idiopathic VF. However, episodes of oversensing as well as ineffective shocks may occur.
Literature
1.
go back to reference Alcalde M, Campuzano O, Sarquella-Brugada G, et al. Clinical interpretation of genetic variants in arrhythmogenic right ventricular cardiomyopathy. Clin Res Cardiol. 2015;104:288–303.CrossRefPubMed Alcalde M, Campuzano O, Sarquella-Brugada G, et al. Clinical interpretation of genetic variants in arrhythmogenic right ventricular cardiomyopathy. Clin Res Cardiol. 2015;104:288–303.CrossRefPubMed
2.
go back to reference Aydin A, Hartel F, Schluter M, et al. Shock efficacy of subcutaneous implantable cardioverter-defibrillator for prevention of sudden cardiac death: initial multicenter experience. Circ Arrhythm Electrophysiol. 2012;5:913–9.CrossRefPubMed Aydin A, Hartel F, Schluter M, et al. Shock efficacy of subcutaneous implantable cardioverter-defibrillator for prevention of sudden cardiac death: initial multicenter experience. Circ Arrhythm Electrophysiol. 2012;5:913–9.CrossRefPubMed
3.
go back to reference Aziz S, Leon AR, El-Chami MF. The subcutaneous defibrillator: a review of the literature. J Am Coll Cardiol. 2014;63:1473–9.CrossRefPubMed Aziz S, Leon AR, El-Chami MF. The subcutaneous defibrillator: a review of the literature. J Am Coll Cardiol. 2014;63:1473–9.CrossRefPubMed
4.
go back to reference Backhoff D, Muller M, Ruschewski W, et al. ICD therapy for primary prevention of sudden cardiac death after Mustard repair for d-transposition of the great arteries. Clin Res Cardiol. 2014;103:894–901.CrossRefPubMed Backhoff D, Muller M, Ruschewski W, et al. ICD therapy for primary prevention of sudden cardiac death after Mustard repair for d-transposition of the great arteries. Clin Res Cardiol. 2014;103:894–901.CrossRefPubMed
5.
go back to reference Bardy GH, Smith WM, Hood MA, et al. An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med. 2010;363:36–44.CrossRefPubMed Bardy GH, Smith WM, Hood MA, et al. An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med. 2010;363:36–44.CrossRefPubMed
6.
go back to reference Bobinger T, Kallmunzer B, Kopp M et al. Prevalence and impact on outcome of electrocardiographic early repolarization patterns among stroke patients: a prospective observational study. Clin Res Cardiol. 2015. Bobinger T, Kallmunzer B, Kopp M et al. Prevalence and impact on outcome of electrocardiographic early repolarization patterns among stroke patients: a prospective observational study. Clin Res Cardiol. 2015.
7.
go back to reference Brigadeau F, Kouakam C, Klug D, et al. Clinical predictors and prognostic significance of electrical storm in patients with implantable cardioverter defibrillators. Eur Heart J. 2006;27:700–7.CrossRefPubMed Brigadeau F, Kouakam C, Klug D, et al. Clinical predictors and prognostic significance of electrical storm in patients with implantable cardioverter defibrillators. Eur Heart J. 2006;27:700–7.CrossRefPubMed
8.
go back to reference Burke MC, Gold MR, Knight BP, et al. Safety and efficacy of the totally subcutaneous implantable defibrillator: 2-year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol. 2015;65:1605–15.CrossRefPubMed Burke MC, Gold MR, Knight BP, et al. Safety and efficacy of the totally subcutaneous implantable defibrillator: 2-year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol. 2015;65:1605–15.CrossRefPubMed
9.
go back to reference Connolly SJ, Hallstrom AP, Cappato R, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Antiarrhythmics vs implantable defibrillator study. Cardiac arrest study Hamburg. Canadian implantable defibrillator study. Eur Heart J. 2000;21:2071–8.CrossRefPubMed Connolly SJ, Hallstrom AP, Cappato R, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Antiarrhythmics vs implantable defibrillator study. Cardiac arrest study Hamburg. Canadian implantable defibrillator study. Eur Heart J. 2000;21:2071–8.CrossRefPubMed
10.
go back to reference Dabiri Abkenari L, Theuns DA, Valk SD, et al. Clinical experience with a novel subcutaneous implantable defibrillator system in a single center. Clin Res Cardiol. 2011;100:737–44.CrossRefPubMedPubMedCentral Dabiri Abkenari L, Theuns DA, Valk SD, et al. Clinical experience with a novel subcutaneous implantable defibrillator system in a single center. Clin Res Cardiol. 2011;100:737–44.CrossRefPubMedPubMedCentral
11.
go back to reference Frommeyer G, Dechering DG, Zumhagen S, et al. Limitations in S-ICD therapy: reasons for system explantation. Clin Res Cardiol. 2015;104:902–7.CrossRefPubMed Frommeyer G, Dechering DG, Zumhagen S, et al. Limitations in S-ICD therapy: reasons for system explantation. Clin Res Cardiol. 2015;104:902–7.CrossRefPubMed
12.
go back to reference Frommeyer G, Reinke F, Eckardt L, et al. Inappropriate shock in a subcutaneous ICD due to interference with a street lantern. Int J Cardiol. 2015;198:6–8.CrossRefPubMed Frommeyer G, Reinke F, Eckardt L, et al. Inappropriate shock in a subcutaneous ICD due to interference with a street lantern. Int J Cardiol. 2015;198:6–8.CrossRefPubMed
13.
go back to reference Gasparini M, Proclemer A, Klersy C, et al. Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA. 2013;309:1903–11.CrossRefPubMed Gasparini M, Proclemer A, Klersy C, et al. Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA. 2013;309:1903–11.CrossRefPubMed
14.
go back to reference Jarman JW, Todd DM. United Kingdom national experience of entirely subcutaneous implantable cardioverter-defibrillator technology: important lessons to learn. Europace. 2013;15:1158–65.CrossRefPubMed Jarman JW, Todd DM. United Kingdom national experience of entirely subcutaneous implantable cardioverter-defibrillator technology: important lessons to learn. Europace. 2013;15:1158–65.CrossRefPubMed
15.
go back to reference Kolb C, Lennerz C, Semmler V, et al. Primary prevention of sudden cardiac death with an entirely subcutaneous defibrillator in a patient with a large right atrial thrombus. Clin Res Cardiol. 2013;102:169–70.CrossRefPubMed Kolb C, Lennerz C, Semmler V, et al. Primary prevention of sudden cardiac death with an entirely subcutaneous defibrillator in a patient with a large right atrial thrombus. Clin Res Cardiol. 2013;102:169–70.CrossRefPubMed
16.
go back to reference Lambiase PD, Barr C, Theuns DA, et al. Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry. Eur Heart J. 2014;35:1657–65.CrossRefPubMedPubMedCentral Lambiase PD, Barr C, Theuns DA, et al. Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry. Eur Heart J. 2014;35:1657–65.CrossRefPubMedPubMedCentral
17.
go back to reference Maagh P, Trappe HJ, Meissner A. Non-determinable defibrillation threshold and inefficacy of implantable cardioverter/defibrillator shocks due to defective connections of the defibrillator lead terminals in the device header port. Europace. 2007;9:1161–2.CrossRefPubMed Maagh P, Trappe HJ, Meissner A. Non-determinable defibrillation threshold and inefficacy of implantable cardioverter/defibrillator shocks due to defective connections of the defibrillator lead terminals in the device header port. Europace. 2007;9:1161–2.CrossRefPubMed
18.
go back to reference Maron BJ, Shen WK, Link MS, et al. Efficacy of implantable cardioverter-defibrillators for the prevention of sudden death in patients with hypertrophic cardiomyopathy. N Engl J Med. 2000;342:365–73.CrossRefPubMed Maron BJ, Shen WK, Link MS, et al. Efficacy of implantable cardioverter-defibrillators for the prevention of sudden death in patients with hypertrophic cardiomyopathy. N Engl J Med. 2000;342:365–73.CrossRefPubMed
19.
go back to reference Moss AJ, Schuger C, Beck CA, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83.CrossRefPubMed Moss AJ, Schuger C, Beck CA, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83.CrossRefPubMed
20.
go back to reference Nanthakumar K, Epstein AE, Kay GN, et al. Prophylactic implantable cardioverter-defibrillator therapy in patients with left ventricular systolic dysfunction: a pooled analysis of 10 primary prevention trials. J Am Coll Cardiol. 2004;44:2166–72.CrossRefPubMed Nanthakumar K, Epstein AE, Kay GN, et al. Prophylactic implantable cardioverter-defibrillator therapy in patients with left ventricular systolic dysfunction: a pooled analysis of 10 primary prevention trials. J Am Coll Cardiol. 2004;44:2166–72.CrossRefPubMed
21.
go back to reference Olde Nordkamp LR, Brouwer TF, Barr C, et al. Inappropriate shocks in the subcutaneous ICD: incidence, predictors and management. Int J Cardiol. 2015;195:126–33.CrossRefPubMed Olde Nordkamp LR, Brouwer TF, Barr C, et al. Inappropriate shocks in the subcutaneous ICD: incidence, predictors and management. Int J Cardiol. 2015;195:126–33.CrossRefPubMed
22.
go back to reference Olde Nordkamp LR, Dabiri Abkenari L, Boersma LV, et al. The entirely subcutaneous implantable cardioverter-defibrillator: initial clinical experience in a large Dutch cohort. J Am Coll Cardiol. 2012;60:1933–9.CrossRefPubMed Olde Nordkamp LR, Dabiri Abkenari L, Boersma LV, et al. The entirely subcutaneous implantable cardioverter-defibrillator: initial clinical experience in a large Dutch cohort. J Am Coll Cardiol. 2012;60:1933–9.CrossRefPubMed
23.
go back to reference Tan VH, Wilton SB, Kuriachan V, et al. Impact of programming strategies aimed at reducing nonessential implantable cardioverter defibrillator therapies on mortality: a systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2014;7:164–70.CrossRefPubMed Tan VH, Wilton SB, Kuriachan V, et al. Impact of programming strategies aimed at reducing nonessential implantable cardioverter defibrillator therapies on mortality: a systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2014;7:164–70.CrossRefPubMed
24.
go back to reference Veltmann C, Borggrefe M, Schimpf R, et al. Fatal inappropriate ICD shock. J Cardiovasc Electrophysiol. 2007;18:326–8.CrossRefPubMed Veltmann C, Borggrefe M, Schimpf R, et al. Fatal inappropriate ICD shock. J Cardiovasc Electrophysiol. 2007;18:326–8.CrossRefPubMed
25.
go back to reference Watanabe H, Chinushi M, Sugiura H, et al. Unsuccessful internal defibrillation in Brugada syndrome: focus on refractoriness and ventricular fibrillation cycle length. J Cardiovasc Electrophysiol. 2005;16:262–6.CrossRefPubMed Watanabe H, Chinushi M, Sugiura H, et al. Unsuccessful internal defibrillation in Brugada syndrome: focus on refractoriness and ventricular fibrillation cycle length. J Cardiovasc Electrophysiol. 2005;16:262–6.CrossRefPubMed
26.
go back to reference Weiss R, Knight BP, Gold MR, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. 2013;128:944–53.CrossRefPubMed Weiss R, Knight BP, Gold MR, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. 2013;128:944–53.CrossRefPubMed
27.
go back to reference Yasuda M, Nakazato Y, Sasaki A, et al. Ventricular fibrillation refractory to ICD therapy. Intern Med. 2005;44:829–31.CrossRefPubMed Yasuda M, Nakazato Y, Sasaki A, et al. Ventricular fibrillation refractory to ICD therapy. Intern Med. 2005;44:829–31.CrossRefPubMed
Metadata
Title
Long-time “real-life” performance of the subcutaneous ICD in patients with electrical heart disease or idiopathic ventricular fibrillation
Authors
Gerrit Frommeyer
Dirk G. Dechering
Simon Kochhäuser
Markus Bettin
Julia Köbe
Lars Eckardt
Florian Reinke
Publication date
01-11-2016
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2016
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-016-0143-4

Other articles of this Issue 2/2016

Journal of Interventional Cardiac Electrophysiology 2/2016 Go to the issue