Published in: The International Journal of Cardiovascular Imaging 2/2011

01-02-2011 | Review

Optical coherence tomography endpoints in stent clinical investigations: strut coverage

Authors: Satoko Tahara, Daniel Chamié, Motaz Baibars, Chadi Alraies, Marco Costa

Published in: The International Journal of Cardiovascular Imaging | Issue 2/2011

Login to get access

Abstract

Late stent thrombosis (LST) and very LST (VLST) are infrequent complications after drug-eluting stent (DES) implantation, but they carry a significant risk for patients. Delayed healing, which may be represented by incomplete stent coverage, has been observed in necropsy vessel specimens treated with DES. As a result, in vivo assessment of stent coverage, as well as stent apposition using optical coherence tomography (OCT), have been recently used as surrogate safety endpoints in clinical trials testing DES platforms. By adopting strut coverage assessed by OCT, one can assess the safety profile of the new generation of DES in preregistration studies. This article focuses on stent strut coverage as a central predictor of late DES thrombosis from the histopathological point of view, discusses the limitations of the current imaging modalities and presents the technical characteristics of OCT for the detection of neointimal coverage after stent implantation. We also review the preclinical and clinical investigations using this novel imaging modality.
Literature
1.
go back to reference Morice MC, Serruys PW, Sousa JE et al (2002) Randomized study with the sirolimus-coated Bx velocity balloon-expandable stent in the treatment of patients with de novo native coronary artery lesions. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. RAVEL study group. N Engl J Med 346:1773–1780PubMedCrossRef Morice MC, Serruys PW, Sousa JE et al (2002) Randomized study with the sirolimus-coated Bx velocity balloon-expandable stent in the treatment of patients with de novo native coronary artery lesions. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. RAVEL study group. N Engl J Med 346:1773–1780PubMedCrossRef
2.
go back to reference Stone GW, Ellis SG, Cox DA et al (2004) A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. TAXUS-IV investigators. N Engl J Med 350:221–231PubMedCrossRef Stone GW, Ellis SG, Cox DA et al (2004) A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. TAXUS-IV investigators. N Engl J Med 350:221–231PubMedCrossRef
3.
go back to reference Sousa JE, Costa MA, Abizaid A et al (2003) Sirolimus-eluting stent for the treatment of in-stent restenosis: a quantitative angiography and three-dimensional intravascular ultrasound study. Circulation 107:24–27PubMedCrossRef Sousa JE, Costa MA, Abizaid A et al (2003) Sirolimus-eluting stent for the treatment of in-stent restenosis: a quantitative angiography and three-dimensional intravascular ultrasound study. Circulation 107:24–27PubMedCrossRef
4.
go back to reference Stone GW, Ellis SG, Cannon L, Mann JT, Greenberg JD, Spriggs D et al (2005) Comparison of a polymer-based paclitaxel-eluting stent with a bare metal stent in patients with complex coronary artery disease. A randomized controlled trial. JAMA 294:1215–1223PubMedCrossRef Stone GW, Ellis SG, Cannon L, Mann JT, Greenberg JD, Spriggs D et al (2005) Comparison of a polymer-based paclitaxel-eluting stent with a bare metal stent in patients with complex coronary artery disease. A randomized controlled trial. JAMA 294:1215–1223PubMedCrossRef
5.
go back to reference Thuesen L, Kelbaek H, Kløvgaard L et al (2006) Comparison of sirolimus-eluting and bare metal stents in coronary bifurcation lesions: subgroup analysis of the stenting coronary arteries in non-stress/Benestent disease trial (SCANDSTENT). Am Heart J 152:1140–1145PubMedCrossRef Thuesen L, Kelbaek H, Kløvgaard L et al (2006) Comparison of sirolimus-eluting and bare metal stents in coronary bifurcation lesions: subgroup analysis of the stenting coronary arteries in non-stress/Benestent disease trial (SCANDSTENT). Am Heart J 152:1140–1145PubMedCrossRef
6.
go back to reference Pasceri V, Patti G, Speciale G et al (2007) Meta-analysis of clinical trials on use of drug-eluting stents for treatment of acute myocardial infarction. Am Heart J 153:749–754PubMedCrossRef Pasceri V, Patti G, Speciale G et al (2007) Meta-analysis of clinical trials on use of drug-eluting stents for treatment of acute myocardial infarction. Am Heart J 153:749–754PubMedCrossRef
7.
go back to reference Baumgart D, Klauss V, Baer F et al (2007) One-year results of the SCORPIUS study. A german multicenter investigation on the effectiveness of sirolimus-eluting stents in diabetic patients. J Am Coll Cardiol 50:1627–1634PubMedCrossRef Baumgart D, Klauss V, Baer F et al (2007) One-year results of the SCORPIUS study. A german multicenter investigation on the effectiveness of sirolimus-eluting stents in diabetic patients. J Am Coll Cardiol 50:1627–1634PubMedCrossRef
8.
go back to reference Camenzind E, Steg PG, Wijns W (2007) Stent thrombosis late after implantation of first-generation drug-eluting stents: a cause for concern. Circulation 115:1440–1455PubMedCrossRef Camenzind E, Steg PG, Wijns W (2007) Stent thrombosis late after implantation of first-generation drug-eluting stents: a cause for concern. Circulation 115:1440–1455PubMedCrossRef
9.
go back to reference Daemen J, Wenawesser P, Tsuchida K, Abrecht L, Vaina S, Morger C et al (2007) Early and late coronary stent thrombosis of sirolimus-eluting and paclitaxel-eluting stents in routine clinical practice: data from a large two-institutional cohort study. Lancet 369:667–678PubMedCrossRef Daemen J, Wenawesser P, Tsuchida K, Abrecht L, Vaina S, Morger C et al (2007) Early and late coronary stent thrombosis of sirolimus-eluting and paclitaxel-eluting stents in routine clinical practice: data from a large two-institutional cohort study. Lancet 369:667–678PubMedCrossRef
10.
go back to reference Lagerqvist B, James SK, Stenestrand U, Lindbäck J, Nilsson T, Wallentin L, SCAAR Study Group (2007) Long-term outcomes with drug-eluting stents versus bare-metal stents in Sweden. N Engl J Med 356:1009–1019PubMedCrossRef Lagerqvist B, James SK, Stenestrand U, Lindbäck J, Nilsson T, Wallentin L, SCAAR Study Group (2007) Long-term outcomes with drug-eluting stents versus bare-metal stents in Sweden. N Engl J Med 356:1009–1019PubMedCrossRef
11.
go back to reference Farb A, Heller PF, Shroff S et al (2001) Pathological analysis of local delivery of paclitaxel via a polymer-coated stent. Circulation 104:473–479PubMedCrossRef Farb A, Heller PF, Shroff S et al (2001) Pathological analysis of local delivery of paclitaxel via a polymer-coated stent. Circulation 104:473–479PubMedCrossRef
12.
go back to reference Guagliumi G, Farb A, Musumeci G et al (2003) Images in cardiovascular medicine: sirolimus-eluting stent implanted in human coronary artery for 16 months: pathological findings. Circulation 107:1340–1341PubMedCrossRef Guagliumi G, Farb A, Musumeci G et al (2003) Images in cardiovascular medicine: sirolimus-eluting stent implanted in human coronary artery for 16 months: pathological findings. Circulation 107:1340–1341PubMedCrossRef
13.
go back to reference Sousa JE, Costa MA, Farb A et al (2004) Images in cardiovascular medicine: vascular healing 4 years after the implantation of sirolimus-eluting stent in humans: a histophatological examination. Circulation 110:e5–e6PubMedCrossRef Sousa JE, Costa MA, Farb A et al (2004) Images in cardiovascular medicine: vascular healing 4 years after the implantation of sirolimus-eluting stent in humans: a histophatological examination. Circulation 110:e5–e6PubMedCrossRef
14.
go back to reference Virmani R, Guagliumi G, Farb A et al (2004) Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious? Circulation 109:701–705PubMedCrossRef Virmani R, Guagliumi G, Farb A et al (2004) Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious? Circulation 109:701–705PubMedCrossRef
15.
go back to reference Joner M, Finn AV, Farb A et al (2006) Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk. J Am Coll Cardiol 48:193–202PubMedCrossRef Joner M, Finn AV, Farb A et al (2006) Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk. J Am Coll Cardiol 48:193–202PubMedCrossRef
16.
go back to reference Luscher TF, Steffel J, Eberli FR et al (2007) Drug-eluting stent and coronary thrombosis: biological mechanisms and clinical implications. Circulation 115:1051–1058PubMedCrossRef Luscher TF, Steffel J, Eberli FR et al (2007) Drug-eluting stent and coronary thrombosis: biological mechanisms and clinical implications. Circulation 115:1051–1058PubMedCrossRef
17.
go back to reference Finn AV, Joner M, Nakazawa G et al (2007) Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation 115:2435–2441PubMedCrossRef Finn AV, Joner M, Nakazawa G et al (2007) Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation 115:2435–2441PubMedCrossRef
18.
go back to reference Nakazawa G, Finn AV, Joner M et al (2008) Delayed arterial healing and increased late stent thrombosis at culprit sites after drug-eluting stent placement for acute myocardial infarction patients: an autopsy study. Circulation 118:1138–1145PubMedCrossRef Nakazawa G, Finn AV, Joner M et al (2008) Delayed arterial healing and increased late stent thrombosis at culprit sites after drug-eluting stent placement for acute myocardial infarction patients: an autopsy study. Circulation 118:1138–1145PubMedCrossRef
19.
go back to reference Guagliumi G, Sirbu V (2008) Optical coherence tomography: high resolution intravascular imaging to evaluate vascular healing after coronary stenting. Catheter Cardiovasc Interv 72:237–247PubMedCrossRef Guagliumi G, Sirbu V (2008) Optical coherence tomography: high resolution intravascular imaging to evaluate vascular healing after coronary stenting. Catheter Cardiovasc Interv 72:237–247PubMedCrossRef
20.
go back to reference Tanigawa J, Barlis P, Di Mario C et al (2007) Intravascular optical coherence tomography: optimization of image acquisition and quantitative assessment of stent strut apposition. EuroIntervention 3:128–136PubMed Tanigawa J, Barlis P, Di Mario C et al (2007) Intravascular optical coherence tomography: optimization of image acquisition and quantitative assessment of stent strut apposition. EuroIntervention 3:128–136PubMed
21.
go back to reference Bezerra HG, Costa MA, Guagliumi G et al (2009) Intracoronary optical coherence tomography: a comprehensive review: clinical and research applications. JACC Cardiovasc Interv 2:1035–1046PubMedCrossRef Bezerra HG, Costa MA, Guagliumi G et al (2009) Intracoronary optical coherence tomography: a comprehensive review: clinical and research applications. JACC Cardiovasc Interv 2:1035–1046PubMedCrossRef
22.
go back to reference Joner M, Nakazawa G, Finn AV et al (2008) Endothelial cell recovery between comparator polymer-based drug-eluting stents. J Am Coll Cardiol 52:333–342PubMedCrossRef Joner M, Nakazawa G, Finn AV et al (2008) Endothelial cell recovery between comparator polymer-based drug-eluting stents. J Am Coll Cardiol 52:333–342PubMedCrossRef
23.
go back to reference Klugherz BD, Llanos G, Lieuallen W et al (2002) Twenty-eight-day efficacy and pharmacokinetics of the sirolimus-eluting stent. Coron Artery Dis 13:183–188PubMedCrossRef Klugherz BD, Llanos G, Lieuallen W et al (2002) Twenty-eight-day efficacy and pharmacokinetics of the sirolimus-eluting stent. Coron Artery Dis 13:183–188PubMedCrossRef
24.
go back to reference Suzuki T, Kopia G, Hayashi S et al (2001) Stent-based delivery of sirolimus reduces neointimal formation in a porcine coronary model. Circulation 104:1188–1193PubMedCrossRef Suzuki T, Kopia G, Hayashi S et al (2001) Stent-based delivery of sirolimus reduces neointimal formation in a porcine coronary model. Circulation 104:1188–1193PubMedCrossRef
25.
go back to reference Carter AJ, Aggarwal M, Kopia GA et al (2004) Long-term effects of polymer-based, slow-release, sirolimus-eluting stents in porcine coronary model. Cardiovasc Res 63:617–624PubMedCrossRef Carter AJ, Aggarwal M, Kopia GA et al (2004) Long-term effects of polymer-based, slow-release, sirolimus-eluting stents in porcine coronary model. Cardiovasc Res 63:617–624PubMedCrossRef
26.
go back to reference Regar E, Werner F, Siebert U et al (2000) Reproducibility of neointima quantification with motorized intravascular ultrasound pullback in stented coronary arteries. Am Heart J 139:632–637PubMedCrossRef Regar E, Werner F, Siebert U et al (2000) Reproducibility of neointima quantification with motorized intravascular ultrasound pullback in stented coronary arteries. Am Heart J 139:632–637PubMedCrossRef
27.
go back to reference Cook S, Wenaweser P, Togni M et al (2007) Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation. Circulation 115:2426–2434PubMedCrossRef Cook S, Wenaweser P, Togni M et al (2007) Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation. Circulation 115:2426–2434PubMedCrossRef
28.
go back to reference Alfonso F, Suárez A, Pérez-Vizcayno MJ et al (2007) Intravascular ultrasound findings during episodes of drug-eluting stent thrombosis. J Am Coll Cardiol 50:2095–2097PubMedCrossRef Alfonso F, Suárez A, Pérez-Vizcayno MJ et al (2007) Intravascular ultrasound findings during episodes of drug-eluting stent thrombosis. J Am Coll Cardiol 50:2095–2097PubMedCrossRef
29.
go back to reference Lee CW, Kang SJ, Park DW et al (2010) Intravascular ultrasound findings in patients with very late stent thrombosis after either drug-eluting or bare-metal stent implantation. J Am Coll Cardiol 55:1936–1942PubMedCrossRef Lee CW, Kang SJ, Park DW et al (2010) Intravascular ultrasound findings in patients with very late stent thrombosis after either drug-eluting or bare-metal stent implantation. J Am Coll Cardiol 55:1936–1942PubMedCrossRef
30.
go back to reference Hassan AK, Bergheanu SC, Stijnen T et al (2010) Late stent malapposition risk is higher after drug-eluting stent compared with bare-metal stent implantation and associates with late stent thrombosis. Eur Heart J 31:1172–1180PubMedCrossRef Hassan AK, Bergheanu SC, Stijnen T et al (2010) Late stent malapposition risk is higher after drug-eluting stent compared with bare-metal stent implantation and associates with late stent thrombosis. Eur Heart J 31:1172–1180PubMedCrossRef
31.
go back to reference Fujii K, Carlier SG, Mintz GS et al (2005) Stent underexpansion and residual reference segment stenosis are related to stent thrombosis after drug-eluting stent implantation. An intravascular ultrasound study. J Am Coll Cardiol 45:995–998PubMedCrossRef Fujii K, Carlier SG, Mintz GS et al (2005) Stent underexpansion and residual reference segment stenosis are related to stent thrombosis after drug-eluting stent implantation. An intravascular ultrasound study. J Am Coll Cardiol 45:995–998PubMedCrossRef
32.
go back to reference Okabe T, Mintz GS, Buch AN, Roy P et al (2007) Intravascular ultrasound parameters associated with stent thrombosis after drug-eluting stent deployment. Am J Cardiol 100:615–620PubMedCrossRef Okabe T, Mintz GS, Buch AN, Roy P et al (2007) Intravascular ultrasound parameters associated with stent thrombosis after drug-eluting stent deployment. Am J Cardiol 100:615–620PubMedCrossRef
33.
go back to reference Lermusiaux P, Martinez R, Donadey A et al (2000) Intravascular ultrasound: limitations and prospects. J Mal Vasc 25:229–236PubMed Lermusiaux P, Martinez R, Donadey A et al (2000) Intravascular ultrasound: limitations and prospects. J Mal Vasc 25:229–236PubMed
34.
go back to reference Di Mario C, Gorge G, Peters R et al (1998) Clinical application and image interpretation in intracoronary ultrasound. Study group on intracoronary imaging of the working group of coronary circulation and of the subgroup on intravascular ultrasound of the working group of echocardiography of the european society of cardiology. Eur Heart J 19:207–229PubMedCrossRef Di Mario C, Gorge G, Peters R et al (1998) Clinical application and image interpretation in intracoronary ultrasound. Study group on intracoronary imaging of the working group of coronary circulation and of the subgroup on intravascular ultrasound of the working group of echocardiography of the european society of cardiology. Eur Heart J 19:207–229PubMedCrossRef
35.
go back to reference Di Mario C, Barlis P (2008) Optical coherence tomography: a new tool to detect tissue coverage in drug-eluting stents. JACC Cardiovasc Interv 1:174–175PubMedCrossRef Di Mario C, Barlis P (2008) Optical coherence tomography: a new tool to detect tissue coverage in drug-eluting stents. JACC Cardiovasc Interv 1:174–175PubMedCrossRef
36.
go back to reference Matsumoto D, Shite J, Shinke T et al (2007) Neointimal coverage of sirolimus-eluting stents at 6-month follow-up: evaluated by optical coherence tomography. Eur Heart J 28:961–967PubMedCrossRef Matsumoto D, Shite J, Shinke T et al (2007) Neointimal coverage of sirolimus-eluting stents at 6-month follow-up: evaluated by optical coherence tomography. Eur Heart J 28:961–967PubMedCrossRef
37.
go back to reference Sera F, Awata M, Uematsu M et al (2009) Optimal stent-sizing with intravascular ultrasound contributes to complete neointimal coverage after sirolimus-eluting stent implantation assessed by angioscopy. JACC Cardiovasc Interv 2:989–994PubMedCrossRef Sera F, Awata M, Uematsu M et al (2009) Optimal stent-sizing with intravascular ultrasound contributes to complete neointimal coverage after sirolimus-eluting stent implantation assessed by angioscopy. JACC Cardiovasc Interv 2:989–994PubMedCrossRef
38.
go back to reference Kotani J, Awata M, Nanto S et al (2006) Incomplete neointimal coverage of sirolimus-eluting stents: angioscopic findings. J Am Coll Cardiol 47:2108–2111PubMedCrossRef Kotani J, Awata M, Nanto S et al (2006) Incomplete neointimal coverage of sirolimus-eluting stents: angioscopic findings. J Am Coll Cardiol 47:2108–2111PubMedCrossRef
39.
go back to reference Higo T, Ueda Y, Oyabu J et al (2009) Atherosclerotic and thrombogenic neointima formed over sirolimus drug-eluting stent: an angioscopic study. JACC Cardiovasc Imaging 2:616–624PubMedCrossRef Higo T, Ueda Y, Oyabu J et al (2009) Atherosclerotic and thrombogenic neointima formed over sirolimus drug-eluting stent: an angioscopic study. JACC Cardiovasc Imaging 2:616–624PubMedCrossRef
40.
go back to reference Awata M, Kotani J, Uematsu M et al (2007) Serial angioscopic evidence of incomplete neointimal coverage after sirolimus-eluting stent implantation: comparison with bare-metal stents. Circulation 116:910–916PubMedCrossRef Awata M, Kotani J, Uematsu M et al (2007) Serial angioscopic evidence of incomplete neointimal coverage after sirolimus-eluting stent implantation: comparison with bare-metal stents. Circulation 116:910–916PubMedCrossRef
41.
go back to reference Prati F, Zimarino M, Stabile D et al (2008) Does optical coherence tomography identify arterial healing after stenting? An in vivo comparison with histology on a rabbit carotid model. Heart 94:217–294PubMedCrossRef Prati F, Zimarino M, Stabile D et al (2008) Does optical coherence tomography identify arterial healing after stenting? An in vivo comparison with histology on a rabbit carotid model. Heart 94:217–294PubMedCrossRef
42.
go back to reference Murata A, Wallace-Bradley D, Tellez A et al (2010) Accuracy of optical coherence tomography in the evaluation of neointimal coverage after stent implantation. JACC Cardiovasc Imaging 3:76–84PubMedCrossRef Murata A, Wallace-Bradley D, Tellez A et al (2010) Accuracy of optical coherence tomography in the evaluation of neointimal coverage after stent implantation. JACC Cardiovasc Imaging 3:76–84PubMedCrossRef
43.
go back to reference Suzuki Y, Ikeno F, Koizumi T et al (2008) In vivo comparison between optical coherence tomography and intravascular ultrasound for detecting small degrees of in-stent neointima after stent implantation. JACC Cardiovasc Interv 1:168–173PubMedCrossRef Suzuki Y, Ikeno F, Koizumi T et al (2008) In vivo comparison between optical coherence tomography and intravascular ultrasound for detecting small degrees of in-stent neointima after stent implantation. JACC Cardiovasc Interv 1:168–173PubMedCrossRef
44.
go back to reference Guagliumi G, Musumeci G, Sirbu V, Bezerra HG, Suzuki N, Fiocca L, Matiashvili A, Lortkipanidze N, Trivisonno A, Valsecchi O, Biondi-Zoccai G, Costa MA (2010) Optical coherence tomography assessment of in vivo vascular response following implantation of overlapping bare-metal and drug-eluting stents. JACC Cardivasc Interv 3:531–539CrossRef Guagliumi G, Musumeci G, Sirbu V, Bezerra HG, Suzuki N, Fiocca L, Matiashvili A, Lortkipanidze N, Trivisonno A, Valsecchi O, Biondi-Zoccai G, Costa MA (2010) Optical coherence tomography assessment of in vivo vascular response following implantation of overlapping bare-metal and drug-eluting stents. JACC Cardivasc Interv 3:531–539CrossRef
45.
go back to reference Guagliumi G, Sirbu V, Costa MA, Musumeci G, Trivisonno A, Matiashvili A, Lortkipanidze N, Mihalcsik L, Valsecchi O, Bezerra HG, Suzuki N, Coletta J, Mintz G, Maehara A, Parise H, Lansky A, Cristea E, Mehran R, Stone G (2011) Long-term strut coverage of paclitaxel eluting stents compared to bare-metal stents implanted during primary PCI in acute myocardial infarction: HORIZONS-OCT. Circulation 123:274–281PubMedCrossRef Guagliumi G, Sirbu V, Costa MA, Musumeci G, Trivisonno A, Matiashvili A, Lortkipanidze N, Mihalcsik L, Valsecchi O, Bezerra HG, Suzuki N, Coletta J, Mintz G, Maehara A, Parise H, Lansky A, Cristea E, Mehran R, Stone G (2011) Long-term strut coverage of paclitaxel eluting stents compared to bare-metal stents implanted during primary PCI in acute myocardial infarction: HORIZONS-OCT. Circulation 123:274–281PubMedCrossRef
46.
go back to reference Guagliumi G, Sirbu V, Bezerra HG, Zoccai GB, Fiocca L, Musumeci G, Matiashvili A, Lortkipanidze N, Tahara S, Valsecchi O, Costa MA (2010) Strut coverage and vessel wall response to zotarolimus-eluting and bare metal stents implanted in patients with ST-elevation myocardial infarction: the OCTAMI (optical coherence tomography in acute myocardial infarction) study. JACC Cardiovasc Interv 3:680–687PubMedCrossRef Guagliumi G, Sirbu V, Bezerra HG, Zoccai GB, Fiocca L, Musumeci G, Matiashvili A, Lortkipanidze N, Tahara S, Valsecchi O, Costa MA (2010) Strut coverage and vessel wall response to zotarolimus-eluting and bare metal stents implanted in patients with ST-elevation myocardial infarction: the OCTAMI (optical coherence tomography in acute myocardial infarction) study. JACC Cardiovasc Interv 3:680–687PubMedCrossRef
47.
go back to reference Guagliumi G, Sirbu V, Musumeci G, Bezerra HG, Aprile A, Kyono H, Fiocca L, Mathiasvili A, Lortkipanidze N, Vassileva A, Popma JJ, Allocco DJ, Dawkins KD, Valsecchi O, Costa MA (2010) Strut coverage and vessel wall response to a new generation paclitaxel-eluting stent with an ultrathin biodegradable abluminal polymer: optical coherence tomography drug eluting stent investigation (OCTDESI). Circ Cardiovasc Interv 3:367–375PubMedCrossRef Guagliumi G, Sirbu V, Musumeci G, Bezerra HG, Aprile A, Kyono H, Fiocca L, Mathiasvili A, Lortkipanidze N, Vassileva A, Popma JJ, Allocco DJ, Dawkins KD, Valsecchi O, Costa MA (2010) Strut coverage and vessel wall response to a new generation paclitaxel-eluting stent with an ultrathin biodegradable abluminal polymer: optical coherence tomography drug eluting stent investigation (OCTDESI). Circ Cardiovasc Interv 3:367–375PubMedCrossRef
48.
go back to reference Barlis P, Regar E, Serruys PW, Dimopoulos K, van der Giessen WJ, van Geuns RJ, Ferrante G, Wandel S, Windecker S, van Es GA, Eerdmans P, Jüni P, di Mario C (2010) An optical coherence tomography study of a biodegradable vs. durable polymer-coated limus-eluting stent: a LEADERS trial sub-study. Eur Heart J 31:165–176PubMedCrossRef Barlis P, Regar E, Serruys PW, Dimopoulos K, van der Giessen WJ, van Geuns RJ, Ferrante G, Wandel S, Windecker S, van Es GA, Eerdmans P, Jüni P, di Mario C (2010) An optical coherence tomography study of a biodegradable vs. durable polymer-coated limus-eluting stent: a LEADERS trial sub-study. Eur Heart J 31:165–176PubMedCrossRef
49.
go back to reference Kim JS, Hong MK, Fan C, Kim TH, Shim JM, Park SM, Ko YG, Choi D, Jang Y (2010) Intracoronary thrombus formation after drug-eluting stents implantation: optical coherence tomographic study. Am Heart J 159:278–283PubMedCrossRef Kim JS, Hong MK, Fan C, Kim TH, Shim JM, Park SM, Ko YG, Choi D, Jang Y (2010) Intracoronary thrombus formation after drug-eluting stents implantation: optical coherence tomographic study. Am Heart J 159:278–283PubMedCrossRef
50.
go back to reference Murakami D, Takano M, Yamamoto M, Inami S, Ohba T, Seino Y, Mizuno K (2009) Advanced neointimal growth is not associated with a low risk of in-stent thrombus. Optical coherence tomographic findings after first-generation drug-eluting stent implantation. Circ J 73:1627–1634PubMedCrossRef Murakami D, Takano M, Yamamoto M, Inami S, Ohba T, Seino Y, Mizuno K (2009) Advanced neointimal growth is not associated with a low risk of in-stent thrombus. Optical coherence tomographic findings after first-generation drug-eluting stent implantation. Circ J 73:1627–1634PubMedCrossRef
51.
go back to reference Fan C, Kim JS, Lee JM, Kim TH, Park SM, Wi J, Paik SI, Ko YG, Choi D, Hong MK, Jang Y (2010) Different vascular healing patterns with various drug-eluting stents in primary percutaneous coronary intervention for ST-segment elevation myocardial infarction: optical coherence tomographic findings. Am J Cardiol 105:972–976PubMedCrossRef Fan C, Kim JS, Lee JM, Kim TH, Park SM, Wi J, Paik SI, Ko YG, Choi D, Hong MK, Jang Y (2010) Different vascular healing patterns with various drug-eluting stents in primary percutaneous coronary intervention for ST-segment elevation myocardial infarction: optical coherence tomographic findings. Am J Cardiol 105:972–976PubMedCrossRef
52.
go back to reference Kubo T, Imanishi T, Kitabata H, Kuroi A, Ueno S, Yamano T, Tanimoto T, Matsuo Y, Masho T, Takarada S, Tanaka A, Nakamura N, Mizukoshi M, Tomobuchi Y, Akasaka T (2008) Comparison of vascular response after sirolimus-eluting stent implantation between patients with unstable and stable angina pectoris: a serial optical coherence tomography study. JACC Cardiovasc Imaging 1:475–484PubMedCrossRef Kubo T, Imanishi T, Kitabata H, Kuroi A, Ueno S, Yamano T, Tanimoto T, Matsuo Y, Masho T, Takarada S, Tanaka A, Nakamura N, Mizukoshi M, Tomobuchi Y, Akasaka T (2008) Comparison of vascular response after sirolimus-eluting stent implantation between patients with unstable and stable angina pectoris: a serial optical coherence tomography study. JACC Cardiovasc Imaging 1:475–484PubMedCrossRef
53.
go back to reference Ozaki Y, Okumura M, Ismail TF, Naruse H, Hattori K, Kan S, Ishikawa M, Kawai T, Takagi Y, Ishii J, Prati F, Serruys PW (2010) The fate of incomplete stent apposition with drug-eluting stents: an optical coherence tomography-based natural history study. Eur Heart J 31:1470–1476PubMedCrossRef Ozaki Y, Okumura M, Ismail TF, Naruse H, Hattori K, Kan S, Ishikawa M, Kawai T, Takagi Y, Ishii J, Prati F, Serruys PW (2010) The fate of incomplete stent apposition with drug-eluting stents: an optical coherence tomography-based natural history study. Eur Heart J 31:1470–1476PubMedCrossRef
54.
go back to reference Kim JS, Jang IK, Kim JS, Kim TH, Takano M, Kume T, Hur NW, Ko YG, Choi D, Hong MK, Jang Y (2009) Optical coherence tomography evaluation of zotarolimus-eluting stents at 9-month follow-up: comparison with sirolimus-eluting stents. Heart 95:1907–1912PubMedCrossRef Kim JS, Jang IK, Kim JS, Kim TH, Takano M, Kume T, Hur NW, Ko YG, Choi D, Hong MK, Jang Y (2009) Optical coherence tomography evaluation of zotarolimus-eluting stents at 9-month follow-up: comparison with sirolimus-eluting stents. Heart 95:1907–1912PubMedCrossRef
55.
go back to reference Chen BX, Ma FY, Luo W, Ruan JH, Xie WL, Zhao XZ, Sun SH, Guo XM, Wang F, Tian T, Chu XW (2008) Neointimal coverage of bare-metal and sirolimus-eluting stents evaluated with optical coherence tomography. Heart 94:566–570PubMedCrossRef Chen BX, Ma FY, Luo W, Ruan JH, Xie WL, Zhao XZ, Sun SH, Guo XM, Wang F, Tian T, Chu XW (2008) Neointimal coverage of bare-metal and sirolimus-eluting stents evaluated with optical coherence tomography. Heart 94:566–570PubMedCrossRef
56.
go back to reference Xie Y, Takano M, Murakami D, Yamamoto M, Okamatsu K, Inami S, Seimiya K, Ohba T, Seino Y, Mizuno K (2008) Comparison of neointimal coverage by optical coherence tomography of a sirolimus-eluting stent versus a bare-metal stent three months after implantation. Am J Cardiol 102:27–31PubMedCrossRef Xie Y, Takano M, Murakami D, Yamamoto M, Okamatsu K, Inami S, Seimiya K, Ohba T, Seino Y, Mizuno K (2008) Comparison of neointimal coverage by optical coherence tomography of a sirolimus-eluting stent versus a bare-metal stent three months after implantation. Am J Cardiol 102:27–31PubMedCrossRef
57.
go back to reference Katoh H, Shite J, Shinke T, Matsumoto D, Tanino Y, Ogasawara D, Sawada T, Miyoshi N, Kawamori H, Yoshino N, Hirata K (2009) Delayed neointimalization on sirolimus-eluting stents: 6-month and 12-month follow up by optical coherence tomography. Circ J 73:1033–1037PubMedCrossRef Katoh H, Shite J, Shinke T, Matsumoto D, Tanino Y, Ogasawara D, Sawada T, Miyoshi N, Kawamori H, Yoshino N, Hirata K (2009) Delayed neointimalization on sirolimus-eluting stents: 6-month and 12-month follow up by optical coherence tomography. Circ J 73:1033–1037PubMedCrossRef
58.
go back to reference Serruys PW, Ormiston JA, Onuma Y, Regar E, Gonzalo N, Garcia-Garcia HM, Nieman K, Bruining N, Dorange C, Miquel-Hébert K, Veldhof S, Webster M, Thuesen L, Dudek D (2009) A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods. Lancet 373:897–910PubMedCrossRef Serruys PW, Ormiston JA, Onuma Y, Regar E, Gonzalo N, Garcia-Garcia HM, Nieman K, Bruining N, Dorange C, Miquel-Hébert K, Veldhof S, Webster M, Thuesen L, Dudek D (2009) A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods. Lancet 373:897–910PubMedCrossRef
59.
go back to reference Ormiston JA, Serruys PW, Regar E, Dudek D, Thuesen L, Webster MW, Onuma Y, Garcia-Garcia HM, McGreevy R, Veldhof S (2008) A bioabsorbable everolimus-eluting coronary stent system for patients with single de novo coronary artery lesions (ABSORB): a prospective open-label trial. Lancet 371:899–907PubMedCrossRef Ormiston JA, Serruys PW, Regar E, Dudek D, Thuesen L, Webster MW, Onuma Y, Garcia-Garcia HM, McGreevy R, Veldhof S (2008) A bioabsorbable everolimus-eluting coronary stent system for patients with single de novo coronary artery lesions (ABSORB): a prospective open-label trial. Lancet 371:899–907PubMedCrossRef
60.
go back to reference Tamburino C, La Manna A, Di Salvo ME, Sacchetta G, Capodanno D, Mehran R, Dangas G, Corcos T, Prati F (2009) First-in-man 1-year clinical outcomes of the catania coronary stent system with nanothin polyzene-F in de novo native coronary artery lesions: the ATLANTA (assessment of the latest non-thrombogenic angioplasty stent) trial. JACC Cardiovasc Interv 2:197–204PubMedCrossRef Tamburino C, La Manna A, Di Salvo ME, Sacchetta G, Capodanno D, Mehran R, Dangas G, Corcos T, Prati F (2009) First-in-man 1-year clinical outcomes of the catania coronary stent system with nanothin polyzene-F in de novo native coronary artery lesions: the ATLANTA (assessment of the latest non-thrombogenic angioplasty stent) trial. JACC Cardiovasc Interv 2:197–204PubMedCrossRef
61.
go back to reference Takano M, Inami S, Jang IK, Yamamoto M, Murakami D, Seimiya K, Ohba T, Mizuno K (2007) Evaluation by optical coherence tomography of neointimal coverage of sirolimus-eluting stent three months after implantation. Am J Cardiol 99:1033–1038PubMedCrossRef Takano M, Inami S, Jang IK, Yamamoto M, Murakami D, Seimiya K, Ohba T, Mizuno K (2007) Evaluation by optical coherence tomography of neointimal coverage of sirolimus-eluting stent three months after implantation. Am J Cardiol 99:1033–1038PubMedCrossRef
62.
go back to reference Kim JS, Jang IK, Fan C, Kim TH, Kim JS, Park SM, Choi EY, Lee SH, Ko YG, Choi D, Hong MK, Jang Y (2009) Evaluation in 3 months duration of neointimal coverage after zotarolimus-eluting stent implantation by optical coherence tomography: the ENDEAVOR OCT trial. JACC Cardiovasc Interv 2:1240–1247PubMedCrossRef Kim JS, Jang IK, Fan C, Kim TH, Kim JS, Park SM, Choi EY, Lee SH, Ko YG, Choi D, Hong MK, Jang Y (2009) Evaluation in 3 months duration of neointimal coverage after zotarolimus-eluting stent implantation by optical coherence tomography: the ENDEAVOR OCT trial. JACC Cardiovasc Interv 2:1240–1247PubMedCrossRef
63.
go back to reference Finn AV, Nakazawa G, Joner M, Kolodgie FD, Mont EK, Gold HK, Virmani R (2007) Vascular responses to drug eluting stents: importance of delayed healing. Arterioscler Thromb Vasc Biol 27:1500–1510PubMedCrossRef Finn AV, Nakazawa G, Joner M, Kolodgie FD, Mont EK, Gold HK, Virmani R (2007) Vascular responses to drug eluting stents: importance of delayed healing. Arterioscler Thromb Vasc Biol 27:1500–1510PubMedCrossRef
64.
go back to reference Tian F, Chen YD, Sun ZJ, Chen L, Yuan F, Song XT, Lü SZ (2009) Evaluation of neointimal coverage of overlapping sirolimus-eluting stents by optical coherence tomography. Chin Med J 20(122):670–674 Tian F, Chen YD, Sun ZJ, Chen L, Yuan F, Song XT, Lü SZ (2009) Evaluation of neointimal coverage of overlapping sirolimus-eluting stents by optical coherence tomography. Chin Med J 20(122):670–674
65.
go back to reference Bezerra HG, Guagliumi G, Valescchi O, Lortkipanidze N, Rosenthal N, Tahara S, Kyono H, Simon DI, Costa MA (2009) Unraveling the lack of neointimal hyperplasia detected by IVUS using OCT: lack of spatial resolution or a true biological effect? The ODESSA sub-study. J Am Coll Cardiol 53:2523 Bezerra HG, Guagliumi G, Valescchi O, Lortkipanidze N, Rosenthal N, Tahara S, Kyono H, Simon DI, Costa MA (2009) Unraveling the lack of neointimal hyperplasia detected by IVUS using OCT: lack of spatial resolution or a true biological effect? The ODESSA sub-study. J Am Coll Cardiol 53:2523
66.
go back to reference Ishigami K, Uemura S, Morikawa Y, Soeda T, Okayama S, Nishida T, Takemoto Y, Onoue K, Somekawa S, Takeda Y, Kawata H, Horii M, Saito Y (2009) Long-term follow-up of neointimal coverage of sirolimus-eluting stents—evaluation with optical coherence tomography. Circ J 73:2300–2307PubMedCrossRef Ishigami K, Uemura S, Morikawa Y, Soeda T, Okayama S, Nishida T, Takemoto Y, Onoue K, Somekawa S, Takeda Y, Kawata H, Horii M, Saito Y (2009) Long-term follow-up of neointimal coverage of sirolimus-eluting stents—evaluation with optical coherence tomography. Circ J 73:2300–2307PubMedCrossRef
67.
go back to reference Bezerra HG, Guagliumi G, Kyono H, Tahara S, Rosenthal N, Musumeci G, Sirbu V, Valescchi O, Aprile A, Simon DI, Costa MA (2009) Determining the optimal cross-sectional analysis interval for OCT assessment of coronary stenting. Circulation 120:S1000 Bezerra HG, Guagliumi G, Kyono H, Tahara S, Rosenthal N, Musumeci G, Sirbu V, Valescchi O, Aprile A, Simon DI, Costa MA (2009) Determining the optimal cross-sectional analysis interval for OCT assessment of coronary stenting. Circulation 120:S1000
68.
go back to reference Yao ZH, Matsubara T, Inada T, Suzuki Y, Suzuki T (2008) Neointimal coverage of sirolimus-eluting stents 6 months and 12 months after implantation: evaluation by optical coherence tomography. Chin Med J 121:503–507PubMed Yao ZH, Matsubara T, Inada T, Suzuki Y, Suzuki T (2008) Neointimal coverage of sirolimus-eluting stents 6 months and 12 months after implantation: evaluation by optical coherence tomography. Chin Med J 121:503–507PubMed
Metadata
Title
Optical coherence tomography endpoints in stent clinical investigations: strut coverage
Authors
Satoko Tahara
Daniel Chamié
Motaz Baibars
Chadi Alraies
Marco Costa
Publication date
01-02-2011
Publisher
Springer Netherlands
Published in
The International Journal of Cardiovascular Imaging / Issue 2/2011
Print ISSN: 1569-5794
Electronic ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-011-9796-3

Other articles of this Issue 2/2011

The International Journal of Cardiovascular Imaging 2/2011 Go to the issue