01-02-2013 | Review Article

Physical and organizational provision for installation, regulatory requirements and implementation of a simultaneous hybrid PET/MR-imaging system in an integrated research and clinical setting

Authors: Bernhard Sattler, Thies Jochimsen, Henryk Barthel, Kerstin Sommerfeld, Patrick Stumpp, Karl-Titus Hoffmann, Matthias Gutberlet, Arno Villringer, Thomas Kahn, Osama Sabri

Published in: Magnetic Resonance Materials in Physics, Biology and Medicine | Issue 1/2013

Login to get access

Abstract

The implementation of hybrid imaging systems requires thorough and anticipatory planning at local and regional levels. For installation of combined positron emission and magnetic resonance imaging systems (PET/MRI), a number of physical and constructional provisions concerning shielding of electromagnetic fields (RF- and high-field) as well as handling of radionuclides have to be met, the latter of which includes shielding for the emitted 511 keV gamma rays. Based on our experiences with a SIEMENS Biograph mMR system, a step-by-step approach is required to allow a trouble-free installation. In this article, we present a proposal for a standardized step-by-step plan to accomplish the installation of a combined PET/MRI system. Moreover, guidelines for the smooth operation of combined PET/MRI in an integrated research and clinical setting will be proposed. Overall, the most important preconditions for the successful implementation of PET/MRI in an integrated research and clinical setting is the interdisciplinary target-oriented cooperation between nuclear medicine, radiology, and all referring and collaborating institutions at all levels of interaction (personnel, imaging protocols, reporting, selection of the data transfer and communication methods).
Literature
1.
go back to reference European Parliament and Council (1996) Directive 96/29/EURATOM Laying down basic safety standards for the protection of health of workers and the general public against the dangers arising from ionizing radiation. Off J Eur Un L 159:1 European Parliament and Council (1996) Directive 96/29/EURATOM Laying down basic safety standards for the protection of health of workers and the general public against the dangers arising from ionizing radiation. Off J Eur Un L 159:1
2.
go back to reference European Parliament and Council (1997) Directive 97/43/EURATOM On health protection of individuals against dangers of ionizing radiation in relation to medical exposure. Off J Eur Un L 180:22 European Parliament and Council (1997) Directive 97/43/EURATOM On health protection of individuals against dangers of ionizing radiation in relation to medical exposure. Off J Eur Un L 180:22
3.
go back to reference Strahlenschutzverordnung vom 20. Juli 2001 (BGBl. I S. 1714; 2002 I S. 1459), die zuletzt durch Artikel 5 Absatz 7 des Gesetzes vom 24. Februar 2012 (BGBl. I S. 212) geändert worden ist Strahlenschutzverordnung vom 20. Juli 2001 (BGBl. I S. 1714; 2002 I S. 1459), die zuletzt durch Artikel 5 Absatz 7 des Gesetzes vom 24. Februar 2012 (BGBl. I S. 212) geändert worden ist
4.
go back to reference European Commission Public Health (2009) Manufacture of Radiopharmaceuticals, The Rules Governing Medicinal Products in the European Union, EU Guidelines to Good Manufacturing Practice, Medicinal Products for Human and Veterinary Use. EudraLex, vol 4, Annex 3 European Commission Public Health (2009) Manufacture of Radiopharmaceuticals, The Rules Governing Medicinal Products in the European Union, EU Guidelines to Good Manufacturing Practice, Medicinal Products for Human and Veterinary Use. EudraLex, vol 4, Annex 3
5.
go back to reference European Commission Public Health (2010) Investigational Medicinal Products, The Rules Governing Medicinal Products in the European Union; EU Guidelines to Good Manufacturing Practice, Medicinal Products for Human and Veterinary Use. EudraLex, vol 4, Annex 13 European Commission Public Health (2010) Investigational Medicinal Products, The Rules Governing Medicinal Products in the European Union; EU Guidelines to Good Manufacturing Practice, Medicinal Products for Human and Veterinary Use. EudraLex, vol 4, Annex 13
6.
go back to reference Arzneimittelgesetz in der Fassung der Bekanntmachung v. 12. Dezember 2005 (BGBl I S. 3394), das zuletzt durch Artikel 13 des Gesetzes vom 22. Dezember 2011 (BGBl I S. 2983) geändert worden ist Arzneimittelgesetz in der Fassung der Bekanntmachung v. 12. Dezember 2005 (BGBl I S. 3394), das zuletzt durch Artikel 13 des Gesetzes vom 22. Dezember 2011 (BGBl I S. 2983) geändert worden ist
7.
go back to reference European Parliament and Council (1993) Directive 93/42/EEC Concerning medical devices. Off J Eur Un L 169:1 European Parliament and Council (1993) Directive 93/42/EEC Concerning medical devices. Off J Eur Un L 169:1
8.
go back to reference European Parliament and Council (2004) Directive 2004/40/EC On the minimum health and safety requirements regarding the exposure of workers to the risks arising from physical agents (electromagnetic fields) Off Journ Eur Un L 159:1 European Parliament and Council (2004) Directive 2004/40/EC On the minimum health and safety requirements regarding the exposure of workers to the risks arising from physical agents (electromagnetic fields) Off Journ Eur Un L 159:1
9.
go back to reference Medizinproduktegesetz in der Fassung der Bekanntmachung vom 7. August 2002 (BGBl. I S. 3146), das zuletzt durch Artikel 13 des Gesetzes vom 8. November 2011 (BGBl. I S. 2178) geändert worden ist Medizinproduktegesetz in der Fassung der Bekanntmachung vom 7. August 2002 (BGBl. I S. 3146), das zuletzt durch Artikel 13 des Gesetzes vom 8. November 2011 (BGBl. I S. 2178) geändert worden ist
15.
go back to reference Busemann-Sokole E, Plachcinska A, Britten A, EANM Physics Committee (2010) Acceptance testing for nuclear medicine instrumentation. Eur J Nucl Med Mol Imaging 37(3):672–681PubMedCrossRef Busemann-Sokole E, Plachcinska A, Britten A, EANM Physics Committee (2010) Acceptance testing for nuclear medicine instrumentation. Eur J Nucl Med Mol Imaging 37(3):672–681PubMedCrossRef
16.
go back to reference National Electrical Manufacturers Association (2007) Performance measurements of positron emission tomographs. NEMA Standards Publication NU 2-2007, Rosslyn, VA National Electrical Manufacturers Association (2007) Performance measurements of positron emission tomographs. NEMA Standards Publication NU 2-2007, Rosslyn, VA
17.
go back to reference Busemann-Sokole E, Plachcinska A, Britten A, EANM Physics Committee (2010) Routine quality control recommendations for nuclear medicine instrumentation. Eur J Nucl Med Mol Imaging 37(3):662–671PubMedCrossRef Busemann-Sokole E, Plachcinska A, Britten A, EANM Physics Committee (2010) Routine quality control recommendations for nuclear medicine instrumentation. Eur J Nucl Med Mol Imaging 37(3):662–671PubMedCrossRef
18.
go back to reference Drzezga A, Souvatzoglou M, Eiber M, Beer A, Fürst S, Martinez-Möller A, Nekolla SG, Ziegler SI, Ganter C, Rummeny E, Schwaiger M (2012) First clinical experience with integrated whole-body PET/MR: comparison to PET/CT in patients with oncologic diagnoses. J Nucl Med 53:1–11CrossRef Drzezga A, Souvatzoglou M, Eiber M, Beer A, Fürst S, Martinez-Möller A, Nekolla SG, Ziegler SI, Ganter C, Rummeny E, Schwaiger M (2012) First clinical experience with integrated whole-body PET/MR: comparison to PET/CT in patients with oncologic diagnoses. J Nucl Med 53:1–11CrossRef
19.
go back to reference Schwenzer NF, Schmidt H, Claussen CD, Whole-body MR/PET (2012) Applications in abdominal imaging. Abdom Imaging 37(1):20–28PubMedCrossRef Schwenzer NF, Schmidt H, Claussen CD, Whole-body MR/PET (2012) Applications in abdominal imaging. Abdom Imaging 37(1):20–28PubMedCrossRef
20.
go back to reference Ratib O, Beyer T (2011) Whole-body hybrid PET/MRI: ready for clinical use? Eur J Nucl Med Mol Imaging 38:992–995PubMedCrossRef Ratib O, Beyer T (2011) Whole-body hybrid PET/MRI: ready for clinical use? Eur J Nucl Med Mol Imaging 38:992–995PubMedCrossRef
21.
go back to reference Schwenzer NF, Stegger L, Bisdas S, Schraml C, Kolb A, Boss A, Müller M, Reimold M, Ernemann U, Claussen CD, Pfannenberg C, Schmidt H (2012) Simultaneous PET/MR imaging in a human brain PET/MR system in 50 patients-current state of image quality. Eur J Radiol. doi:10.1016/j.ejrad.2011.12.027 Schwenzer NF, Stegger L, Bisdas S, Schraml C, Kolb A, Boss A, Müller M, Reimold M, Ernemann U, Claussen CD, Pfannenberg C, Schmidt H (2012) Simultaneous PET/MR imaging in a human brain PET/MR system in 50 patients-current state of image quality. Eur J Radiol. doi:10.​1016/​j.​ejrad.​2011.​12.​027
22.
go back to reference Gutberlet M, Noeske R, Schwinge K, Freyhardt P, Felix R, Niendorf T (2006) Comprehensive cardiac magnetic resonance imaging at 3.0 Tesla: feasibility and implications for clinical applications. Invest Radiol 41(2):154–167PubMedCrossRef Gutberlet M, Noeske R, Schwinge K, Freyhardt P, Felix R, Niendorf T (2006) Comprehensive cardiac magnetic resonance imaging at 3.0 Tesla: feasibility and implications for clinical applications. Invest Radiol 41(2):154–167PubMedCrossRef
23.
go back to reference Gutberlet M, Schwinge K, Freyhardt P, Spors B, Grothoff M, Denecke T, Lüdemann L, Noeske R, Niendorf T, Felix R (2005) Influence of high magnetic field strengths and parallel acquisition strategies on image quality in cardiac 2D CINE magnetic resonance imaging: comparison of 1.5 T vs. 3.0 T. Eur Radiol 15(8):1586–1597PubMedCrossRef Gutberlet M, Schwinge K, Freyhardt P, Spors B, Grothoff M, Denecke T, Lüdemann L, Noeske R, Niendorf T, Felix R (2005) Influence of high magnetic field strengths and parallel acquisition strategies on image quality in cardiac 2D CINE magnetic resonance imaging: comparison of 1.5 T vs. 3.0 T. Eur Radiol 15(8):1586–1597PubMedCrossRef
24.
go back to reference Kellman P, Larson AC, Hsu LY, Chung YC, Simonetti OP, McVeigh ER, Arai AE (2005) Motion-corrected free-breathing delayed enhancement imaging of myocardial infarction. Magn Reson Med 53(1):194–200PubMedCrossRef Kellman P, Larson AC, Hsu LY, Chung YC, Simonetti OP, McVeigh ER, Arai AE (2005) Motion-corrected free-breathing delayed enhancement imaging of myocardial infarction. Magn Reson Med 53(1):194–200PubMedCrossRef
25.
go back to reference Eiber M, Martinez-Möller A, Souvatzoglou M, Holzapfel K, Pickhard A, Löffelbein D, Santi I, Rummeny EJ, Ziegler S, Schwaiger M, Nekolla SG, Beer AJ (2011) Value of a Dixon-based MR/PET attenuation correction sequence for the localization and evaluation of PET-positive lesions. Eur J Nucl Med Mol Imaging 38(9):1691–1701PubMedCrossRef Eiber M, Martinez-Möller A, Souvatzoglou M, Holzapfel K, Pickhard A, Löffelbein D, Santi I, Rummeny EJ, Ziegler S, Schwaiger M, Nekolla SG, Beer AJ (2011) Value of a Dixon-based MR/PET attenuation correction sequence for the localization and evaluation of PET-positive lesions. Eur J Nucl Med Mol Imaging 38(9):1691–1701PubMedCrossRef
26.
go back to reference Delso G, Furst S, Jakoby B, Ladebeck R, Ganter C, Nekolla SG, Schwaiger M, Ziegler SI (2012) Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner. J Nucl Med 52:1–9 Delso G, Furst S, Jakoby B, Ladebeck R, Ganter C, Nekolla SG, Schwaiger M, Ziegler SI (2012) Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner. J Nucl Med 52:1–9
27.
go back to reference Keller SH, Holm S, Hansen AE, Sattler B, Andersen A, Klausen TL, Højgaard L, A Kjær A, Beyer T (2012) Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI. Magn Reson Mater Phy. doi:10.1007/s10334-012-0345-4 Keller SH, Holm S, Hansen AE, Sattler B, Andersen A, Klausen TL, Højgaard L, A Kjær A, Beyer T (2012) Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI. Magn Reson Mater Phy. doi:10.​1007/​s10334-012-0345-4
28.
go back to reference Hofmann M, Steinke F, Scheel V, Charpiat G, Farquhar J, Aschoff P, Brady M, Schölkopf B, Pichler BJ (2008) MRI-based attenuation correction for PET/MRI: a novel approach combining pattern recognition and atlas registration. J Nucl Med 49:1875–1883PubMedCrossRef Hofmann M, Steinke F, Scheel V, Charpiat G, Farquhar J, Aschoff P, Brady M, Schölkopf B, Pichler BJ (2008) MRI-based attenuation correction for PET/MRI: a novel approach combining pattern recognition and atlas registration. J Nucl Med 49:1875–1883PubMedCrossRef
29.
go back to reference Hofmann M, Pichler BJ, Schölkopf B, Beyer T (2009) Towards quantitative PET/MRI a review of MR-based attenuation correction techniques. Eur J Nucl Med Mol Imaging 36(Suppl 1):S93–S104PubMedCrossRef Hofmann M, Pichler BJ, Schölkopf B, Beyer T (2009) Towards quantitative PET/MRI a review of MR-based attenuation correction techniques. Eur J Nucl Med Mol Imaging 36(Suppl 1):S93–S104PubMedCrossRef
30.
go back to reference Hofmann M, Bezrukov I, Mantlik F, Aschoff P, Steinke F, Beyer T, Pichler BJ, Schölkopf B (2011) MRI-based attenuation correction for whole-body PET/MRI: quantitative evaluation of segmentation- and atlas-based methods. J Nucl Med 52:1392–1399PubMedCrossRef Hofmann M, Bezrukov I, Mantlik F, Aschoff P, Steinke F, Beyer T, Pichler BJ, Schölkopf B (2011) MRI-based attenuation correction for whole-body PET/MRI: quantitative evaluation of segmentation- and atlas-based methods. J Nucl Med 52:1392–1399PubMedCrossRef
31.
go back to reference Martinez-Möller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd’hotel C, Ziegler SI, Navab N, Schwaiger M, Nekolla SG (2009) Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med 50:520–526PubMedCrossRef Martinez-Möller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd’hotel C, Ziegler SI, Navab N, Schwaiger M, Nekolla SG (2009) Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med 50:520–526PubMedCrossRef
32.
go back to reference Schulz V, Torres-Espallardo I, Renisch S, Hu Z, Ojha N, Börnert P, Perkuhn M, Niendorf T, Schäfer WM, Brockmann H, Krohn T, Buhl A, Günther RW, Mottaghy FM, Krombach GA (2011) Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging 38(1):138–152PubMedCrossRef Schulz V, Torres-Espallardo I, Renisch S, Hu Z, Ojha N, Börnert P, Perkuhn M, Niendorf T, Schäfer WM, Brockmann H, Krohn T, Buhl A, Günther RW, Mottaghy FM, Krombach GA (2011) Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging 38(1):138–152PubMedCrossRef
33.
go back to reference Berker Y, Franke J, Salomon A, Palmowski M, Donker HC, Temur Y, Mottaghy FM, Kuhl C, Izquierdo-Garcia D, Fayad ZA, Kiessling F, Schulz V (2012) MRI-based attenuation correction for hybrid PET/MRI systems: a 4-class tissue segmentation technique using a combined ultrashort-echo-time/Dixon MRI sequence. J Nucl Med 53(5):796–804PubMedCrossRef Berker Y, Franke J, Salomon A, Palmowski M, Donker HC, Temur Y, Mottaghy FM, Kuhl C, Izquierdo-Garcia D, Fayad ZA, Kiessling F, Schulz V (2012) MRI-based attenuation correction for hybrid PET/MRI systems: a 4-class tissue segmentation technique using a combined ultrashort-echo-time/Dixon MRI sequence. J Nucl Med 53(5):796–804PubMedCrossRef
34.
go back to reference Zanotti-Fregonara P, Maroy R, Comtat C, Jan S, Gaura V, Bar-Hen A, Ribeiro MJ, Trébossen R (2009) Comparison of 3 methods of automated internal carotid segmentation in human brain PET studies: application to the estimation of arterial input function. J Nucl Med 50(3):461–467PubMedCrossRef Zanotti-Fregonara P, Maroy R, Comtat C, Jan S, Gaura V, Bar-Hen A, Ribeiro MJ, Trébossen R (2009) Comparison of 3 methods of automated internal carotid segmentation in human brain PET studies: application to the estimation of arterial input function. J Nucl Med 50(3):461–467PubMedCrossRef
35.
go back to reference Rousset OG, Yilong M, Evans AC (1998) Correction for partial volume effects in PET: principle and validation. J Nucl Med 39:904–911PubMed Rousset OG, Yilong M, Evans AC (1998) Correction for partial volume effects in PET: principle and validation. J Nucl Med 39:904–911PubMed
36.
go back to reference Feng D, Huang S-C, Wang X (1993) Models for computer simulation studies of input functions for tracer kinetic modeling with positron emission tomography. Int J Biomed Comput 32:95–110PubMedCrossRef Feng D, Huang S-C, Wang X (1993) Models for computer simulation studies of input functions for tracer kinetic modeling with positron emission tomography. Int J Biomed Comput 32:95–110PubMedCrossRef
38.
go back to reference Friedman L, Glover GH (2006) Report on a multicenter fMRI quality assurance protocol. J Magn Reson Imaging 23:827–839PubMedCrossRef Friedman L, Glover GH (2006) Report on a multicenter fMRI quality assurance protocol. J Magn Reson Imaging 23:827–839PubMedCrossRef
Metadata
Title
Physical and organizational provision for installation, regulatory requirements and implementation of a simultaneous hybrid PET/MR-imaging system in an integrated research and clinical setting
Authors
Bernhard Sattler
Thies Jochimsen
Henryk Barthel
Kerstin Sommerfeld
Patrick Stumpp
Karl-Titus Hoffmann
Matthias Gutberlet
Arno Villringer
Thomas Kahn
Osama Sabri
Publication date
01-02-2013
Publisher
Springer-Verlag
Published in
Magnetic Resonance Materials in Physics, Biology and Medicine / Issue 1/2013
Print ISSN: 0968-5243
Electronic ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-012-0347-2

Other articles of this Issue 1/2013

Magnetic Resonance Materials in Physics, Biology and Medicine 1/2013 Go to the issue