Correlation between resonance frequency, insertion torque and bone-implant contact in self-cutting threaded implants

Authors: Yahya Açil, Jan Sievers, Aydin Gülses, Mustafa Ayna, Jörg Wiltfang, Hendrik Terheyden

Published in: Odontology | Issue 3/2017

Login to get access

Abstract

The aim of this study was to evaluate the correlation between resonance frequency analysis (RFA) values and the histomorphometric bone-implant contact (BIC) immediately after insertion of the implant. Additionally, it was examined to define which extent peak insertion torque (PIT) was correlated with the latter measurements. 15 Xive® S plus root from dental implants were inserted in fresh porcine frontal bones. The insertion torque was measured with the Kavo Intrasurg 300 surgical unit. RFA connector was coupled to the implants and the mean value of 20 consecutive RFA measurements was calculated via Osstell® ISQ device. The implants were removed with the adjacent bone with a band saw. The blocks were processed for undecalcified histology. Two perpendicular longitudinal middle sections of the implant were made and stained with toluidine blue and the BIC was assessed by histomorphometry. The correlation coefficient (Spearman) between RFA and BIC was R = 0.579 (p = 0.026, F test). The correlation between PIT and BIC (0.33, p > 0.05) and PIT and RFA (0.153, p > 0.05) was not statistically significant. The present data confirmed a moderate and statistically positive correlation between RFA and BIC. No correlation between BIC and PIT and PIT and RFA was observed. Further studies considering different bone qualities would be beneficial in understanding the relation between RFA and BIC.
Literature
1.
go back to reference Manresa C, Bosch M, Echeverría JJ. The comparison between implant stability quotient and bone-implant contact revisited: an experiment in Beagle dog. Clin Oral Implants Res. 2014;25:1213–21.CrossRefPubMed Manresa C, Bosch M, Echeverría JJ. The comparison between implant stability quotient and bone-implant contact revisited: an experiment in Beagle dog. Clin Oral Implants Res. 2014;25:1213–21.CrossRefPubMed
2.
go back to reference Quesada-García MP, Prados-Sánchez E, Olmedo-Gaya MV, Muñoz-Soto E, González-Rodríguez MP, Valllecillo-Capilla M. Measurement of dental implant stability by resonance frequency analysis: a review of the literature. Med Oral Patol Oral Cir Bucal. 2009;14:538–46.CrossRef Quesada-García MP, Prados-Sánchez E, Olmedo-Gaya MV, Muñoz-Soto E, González-Rodríguez MP, Valllecillo-Capilla M. Measurement of dental implant stability by resonance frequency analysis: a review of the literature. Med Oral Patol Oral Cir Bucal. 2009;14:538–46.CrossRef
3.
go back to reference Akça K, Kökat AM, Cömert A, Akkocaoğlu M, Tekdemir I, Cehreli MC. Torque-fitting and resonance frequency analyses of implants in conventional sockets versus controlled bone defects in vitro. Int J Oral Maxillofac Surg. 2010;39:169–73.CrossRefPubMed Akça K, Kökat AM, Cömert A, Akkocaoğlu M, Tekdemir I, Cehreli MC. Torque-fitting and resonance frequency analyses of implants in conventional sockets versus controlled bone defects in vitro. Int J Oral Maxillofac Surg. 2010;39:169–73.CrossRefPubMed
4.
go back to reference Ohta K, Takechi M, Minami M, Shigeishi H, Hiraoka M, Nishimura M, Kamata N. Influence of factors related to implant stability detected by wireless resonance frequency analysis device. J Oral Rehabil. 2010;37:131–7.CrossRefPubMed Ohta K, Takechi M, Minami M, Shigeishi H, Hiraoka M, Nishimura M, Kamata N. Influence of factors related to implant stability detected by wireless resonance frequency analysis device. J Oral Rehabil. 2010;37:131–7.CrossRefPubMed
5.
go back to reference Turkyilmaz I, Sennerby L, Yilmaz B, Bilecenoglu B, Ozbek EN. Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: a human cadaver study. Clin Implant Dent Relat Res. 2009;11:52–8.CrossRefPubMed Turkyilmaz I, Sennerby L, Yilmaz B, Bilecenoglu B, Ozbek EN. Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: a human cadaver study. Clin Implant Dent Relat Res. 2009;11:52–8.CrossRefPubMed
6.
go back to reference Nkenke E, Hahn M, Weinzierl K, Radespiel-Tröger M, Neukam FW, Engelke K. Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants. Clin Oral Implants Res. 2003;14:601–9.CrossRefPubMed Nkenke E, Hahn M, Weinzierl K, Radespiel-Tröger M, Neukam FW, Engelke K. Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants. Clin Oral Implants Res. 2003;14:601–9.CrossRefPubMed
7.
go back to reference Schliephake H, Sewing A, Aref A. Resonance frequency measurements of implant stability in the dog mandible: experimental comparison with histomor-phometric data. Int J Oral Maxillofac Surg. 2006;35:941–6.CrossRefPubMed Schliephake H, Sewing A, Aref A. Resonance frequency measurements of implant stability in the dog mandible: experimental comparison with histomor-phometric data. Int J Oral Maxillofac Surg. 2006;35:941–6.CrossRefPubMed
8.
go back to reference Donath K, Breuner G. A method for the study of undecalcified bones and teeth with attaced soft tissues: the Säge-Schliff (sawing and grinding) technique. J Oral Pathol. 1982;II:318–26.CrossRef Donath K, Breuner G. A method for the study of undecalcified bones and teeth with attaced soft tissues: the Säge-Schliff (sawing and grinding) technique. J Oral Pathol. 1982;II:318–26.CrossRef
9.
go back to reference Kim YS, Lim YJ. Primary stability and self-tapping blades: biomechanical assessment of dental implants in medium-density bone. Clin Oral Implants Res. 2011;22:1179–84.CrossRefPubMed Kim YS, Lim YJ. Primary stability and self-tapping blades: biomechanical assessment of dental implants in medium-density bone. Clin Oral Implants Res. 2011;22:1179–84.CrossRefPubMed
10.
go back to reference Kim DR, Lim YJ, Kim MJ, Kwon HB, Kim SH. Self-cutting blades and their influence on primary stability of tapered dental implants in a simulated low-density bone model: a laboratory study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2011;112:573–80.CrossRefPubMed Kim DR, Lim YJ, Kim MJ, Kwon HB, Kim SH. Self-cutting blades and their influence on primary stability of tapered dental implants in a simulated low-density bone model: a laboratory study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2011;112:573–80.CrossRefPubMed
11.
go back to reference Chowdary R. Evaluation of a self cutting thread designed and non self cutting threaded implants placed in extraction sites with immediate temporization: a 5-year clinical study in: abstracts of the EAO 23rd Annual Scientific Meeting, September 25–27, 2014, Rome, Italy. Clin Oral Implants Res. 2014;25(Suppl 10):1–694. Chowdary R. Evaluation of a self cutting thread designed and non self cutting threaded implants placed in extraction sites with immediate temporization: a 5-year clinical study in: abstracts of the EAO 23rd Annual Scientific Meeting, September 25–27, 2014, Rome, Italy. Clin Oral Implants Res. 2014;25(Suppl 10):1–694.
12.
go back to reference Buser D, Nydegger T, Hirt HP, Cochran DL, Nolte LP. Removal torque values of titanium implants in the maxilla of miniature pigs. Int J Oral Maxillofac Implants. 1998;13:611–9.PubMed Buser D, Nydegger T, Hirt HP, Cochran DL, Nolte LP. Removal torque values of titanium implants in the maxilla of miniature pigs. Int J Oral Maxillofac Implants. 1998;13:611–9.PubMed
13.
go back to reference Büchter A, Kleinheinz J, Wiesmann HP, Seper L, Joos U, Meyer U. Peri-implant bone formation around cylindrical and conical implant systems. Mund Kiefer Gesichtschir. 2004;8:282–8.CrossRefPubMed Büchter A, Kleinheinz J, Wiesmann HP, Seper L, Joos U, Meyer U. Peri-implant bone formation around cylindrical and conical implant systems. Mund Kiefer Gesichtschir. 2004;8:282–8.CrossRefPubMed
14.
go back to reference Bashutski JD, D’Silva NJ, Wang HL. Implant compression necrosis: current understanding and case report. J Periodontol. 2009;80:700–4.CrossRefPubMed Bashutski JD, D’Silva NJ, Wang HL. Implant compression necrosis: current understanding and case report. J Periodontol. 2009;80:700–4.CrossRefPubMed
15.
go back to reference Nevins M, Nevins ML, Schupbach P, Fiorellini J, Lin Z, Kim DM. The impact of bone compression on bone-to-implant contact of an osseointegrated implant: a canine study. Int J Periodontics Restor Dent. 2012;32:637–45. Nevins M, Nevins ML, Schupbach P, Fiorellini J, Lin Z, Kim DM. The impact of bone compression on bone-to-implant contact of an osseointegrated implant: a canine study. Int J Periodontics Restor Dent. 2012;32:637–45.
16.
go back to reference Halldin A, Jimbo R, Johansson CB, Wennerberg A, Jacobsson M, Albrektsson T, Hansson S. The effect of static bone strain on implant stability and bone remodeling. Bone. 2011;49(4):783–9.CrossRefPubMed Halldin A, Jimbo R, Johansson CB, Wennerberg A, Jacobsson M, Albrektsson T, Hansson S. The effect of static bone strain on implant stability and bone remodeling. Bone. 2011;49(4):783–9.CrossRefPubMed
17.
go back to reference Jimbo R, Tovar N, Marin C, Teixeira HS, Anchieta RB, Silveira LM, Janal MN, Shibli JA, Coelho PG. The impact of a modified cutting flute implant design on osseointegration. Int J Oral Maxillofac Surg. 2014;43:883–8.CrossRefPubMed Jimbo R, Tovar N, Marin C, Teixeira HS, Anchieta RB, Silveira LM, Janal MN, Shibli JA, Coelho PG. The impact of a modified cutting flute implant design on osseointegration. Int J Oral Maxillofac Surg. 2014;43:883–8.CrossRefPubMed
18.
go back to reference Dagher M, Mokbel N, Jabbour G, Naaman N. Resonance frequency analysis, insertion torque, and bone to implant contact of 4 implant surfaces: comparison and correlation study in sheep. Implant Dent. 2014;23:672–8.PubMed Dagher M, Mokbel N, Jabbour G, Naaman N. Resonance frequency analysis, insertion torque, and bone to implant contact of 4 implant surfaces: comparison and correlation study in sheep. Implant Dent. 2014;23:672–8.PubMed
19.
go back to reference Abrahamsson I, Linder E, Lang NP. Implant stability in relation to osseointe-gration: an experimental study in the Labrador dog. Clin Oral Implants Res. 2009;20:313–8.CrossRefPubMed Abrahamsson I, Linder E, Lang NP. Implant stability in relation to osseointe-gration: an experimental study in the Labrador dog. Clin Oral Implants Res. 2009;20:313–8.CrossRefPubMed
20.
go back to reference Ito Y, Sato D, Yoneda S, Ito D, Kondo H, Kasugai S. Relevance of resonance frequency analysis to evaluate dental implant stability: simulation and histo-morphometrical animal experiments. Clin Oral Implants Res. 2007;19:9–14.PubMed Ito Y, Sato D, Yoneda S, Ito D, Kondo H, Kasugai S. Relevance of resonance frequency analysis to evaluate dental implant stability: simulation and histo-morphometrical animal experiments. Clin Oral Implants Res. 2007;19:9–14.PubMed
21.
go back to reference Scarano A, Degidi M, Iezzi G, Petrone G, Piattelli A. Correlation between im-plant stability quotient and bone-implant contact: a retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans. Clin Implant Dent Relat Res. 2006;8:218–22.CrossRefPubMed Scarano A, Degidi M, Iezzi G, Petrone G, Piattelli A. Correlation between im-plant stability quotient and bone-implant contact: a retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans. Clin Implant Dent Relat Res. 2006;8:218–22.CrossRefPubMed
22.
go back to reference Ottoni JM, Oliveira ZF, Mansini R, Cabral AM. Correlation between placement torque and survival of single-tooth implants. Int J Oral Maxillofac Implants. 2005;20:769–76.PubMed Ottoni JM, Oliveira ZF, Mansini R, Cabral AM. Correlation between placement torque and survival of single-tooth implants. Int J Oral Maxillofac Implants. 2005;20:769–76.PubMed
23.
go back to reference Eliyas S, Al-Khayatt AS. No difference between failure rates of early and con-ventionally loaded implants. Evid Based Dent. 2008;9:50.CrossRefPubMed Eliyas S, Al-Khayatt AS. No difference between failure rates of early and con-ventionally loaded implants. Evid Based Dent. 2008;9:50.CrossRefPubMed
24.
go back to reference Ersanli S, Karabuda C, Beck F, Leblebicioglu B. Resonance frequency analy-sis of one-stage dental implant stability during the osseointegration period. J Periodontol. 2005;76:1066–71.CrossRefPubMed Ersanli S, Karabuda C, Beck F, Leblebicioglu B. Resonance frequency analy-sis of one-stage dental implant stability during the osseointegration period. J Periodontol. 2005;76:1066–71.CrossRefPubMed
25.
go back to reference Esposito M, Grusovin MG, Willings M, Coulthard P, Worthington HV. The effectiveness of immediate, early, and conventional loading of dental implants: a Cochrane systematic review of randomized controlled clinical trials. Int J Oral Maxillofac Implants. 2007;22:893–904.PubMed Esposito M, Grusovin MG, Willings M, Coulthard P, Worthington HV. The effectiveness of immediate, early, and conventional loading of dental implants: a Cochrane systematic review of randomized controlled clinical trials. Int J Oral Maxillofac Implants. 2007;22:893–904.PubMed
26.
go back to reference Degidi M, Perrotti V, Strocchi R, Piattelli A, Iezzi G. Is insertion torque corre-lated to bone-implant contact percentage in the early healing period? A histological and histomorphometrical evaluation of 17 human-retrieved dental im-plants. Clin Oral Implants Res. 2009;26:330–5. Degidi M, Perrotti V, Strocchi R, Piattelli A, Iezzi G. Is insertion torque corre-lated to bone-implant contact percentage in the early healing period? A histological and histomorphometrical evaluation of 17 human-retrieved dental im-plants. Clin Oral Implants Res. 2009;26:330–5.
27.
go back to reference Tabassum A, Meijer GJ, Wolke JG, Jansen JA. Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study. Clin Oral Implants Res. 2010;21:213–20.CrossRefPubMed Tabassum A, Meijer GJ, Wolke JG, Jansen JA. Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study. Clin Oral Implants Res. 2010;21:213–20.CrossRefPubMed
28.
go back to reference Cehreli MC, Kökat AM, Comert A, Akkocaoğlu M, Tekdemir I, Akça K. Implant stability and bone density: assessment of correlation in fresh cadavers using conventional and osteotome implant sockets. Clin Oral Implants Res. 2009;20:1163–9.CrossRefPubMed Cehreli MC, Kökat AM, Comert A, Akkocaoğlu M, Tekdemir I, Akça K. Implant stability and bone density: assessment of correlation in fresh cadavers using conventional and osteotome implant sockets. Clin Oral Implants Res. 2009;20:1163–9.CrossRefPubMed
29.
go back to reference Kunnekel AT, Nair KC, Naidu EM, Sivagami G. Validation of resonance frequency analysis by comparing implant stability quotient values with histomorphometric data. J Oral Implantol. 2011;37:301–8.CrossRefPubMed Kunnekel AT, Nair KC, Naidu EM, Sivagami G. Validation of resonance frequency analysis by comparing implant stability quotient values with histomorphometric data. J Oral Implantol. 2011;37:301–8.CrossRefPubMed
30.
go back to reference Metzler P, von Wilmowsky C, Stadlinger B, Zemann W, Schlegel KA, Rosiwal S, Rupprecht S. Nano-crystalline diamond-coated titanium dental implants—a histomorphometric study in adult domestic pigs. J Cranio-Maxillofac Surg. 2013;41:532–8.CrossRef Metzler P, von Wilmowsky C, Stadlinger B, Zemann W, Schlegel KA, Rosiwal S, Rupprecht S. Nano-crystalline diamond-coated titanium dental implants—a histomorphometric study in adult domestic pigs. J Cranio-Maxillofac Surg. 2013;41:532–8.CrossRef
31.
go back to reference Laiblin C, Jaeschke G. Clinical chemistry examinations of bone and muscle metabolism under stress in the Gottingen miniature pig—an experimental study. Berl Munch Tierarztl Wochenschr. 1979;92:124–8.PubMed Laiblin C, Jaeschke G. Clinical chemistry examinations of bone and muscle metabolism under stress in the Gottingen miniature pig—an experimental study. Berl Munch Tierarztl Wochenschr. 1979;92:124–8.PubMed
Metadata
Title
Correlation between resonance frequency, insertion torque and bone-implant contact in self-cutting threaded implants
Authors
Yahya Açil
Jan Sievers
Aydin Gülses
Mustafa Ayna
Jörg Wiltfang
Hendrik Terheyden
Publication date
01-07-2017
Publisher
Springer Japan
Published in
Odontology / Issue 3/2017
Print ISSN: 1618-1247
Electronic ISSN: 1618-1255
DOI
https://doi.org/10.1007/s10266-016-0265-2

Other articles of this Issue 3/2017

Odontology 3/2017 Go to the issue