Computed tomography quantification of bone density adjacent to cemented pegged polyethylene glenoid components in shoulder arthroplasty

Authors: Giovanni Merolla, Barbara Amore, Paolo Paladini, Enrico Cavagna, Giuseppe Porcellini

Published in: European Journal of Orthopaedic Surgery & Traumatology | Issue 5/2014

Login to get access

Abstract

Background

Cemented polyethylene devices are the prostheses implanted more frequently, but there is no agreement on the optimal glenoid component design. In this study, bone mineral density (BMD) adjacent to cemented all-polyethylene glenoid components was assessed to gain insights into the characteristics of glenoid bone as a potential risk factor implicated in the failure of shoulder arthroplasty.

Patients and methods

Twenty-two subjects were examined at an average follow-up of 31 months using the Constant–Murley score (CS) and multi-detector computed tomography. BMD was measured in 5 regions of interest (ROIs) at the sites where radiolucent lines are usually detected. BMD differences among the ROIs were tested. The relationship between BMD and a number of variables (CS subscores, age, gender, follow-up duration) was explored.

Results

There was a significant increase in CS scores (p < 0.05). Significantly different BMD (p = 0.0039) was found in the 5 ROIs, especially between ROIs 2 and 5 (p = 0.016, Bonferroni’s test) and between ROIs 3 and 5 (p = 0.005, Bonferroni’s test). BMD was lower in ROI 1 than ROI 3 and in ROI 2 than ROI 4, but the difference was not significant.

Discussion

The heterogeneous BMD distribution may be related to: (1) an interindividual variability in glenoid BMD; (2) the fixation technique; or (3) the different bone response to eccentric loading of the prosthetic head on the glenoid component.

Conclusions

BMD analysis may contribute to extent our knowledge on glenoid component loosening and encourage further techniques of glenoid fixation.
Literature
1.
go back to reference Merolla G, Campi F, Paladini P, Porcellini G (2008) Efficacy of anatomical prostheses in primary gleno-humeral osteoarthritis. Chir Organi Mov 91:109–115PubMedCrossRef Merolla G, Campi F, Paladini P, Porcellini G (2008) Efficacy of anatomical prostheses in primary gleno-humeral osteoarthritis. Chir Organi Mov 91:109–115PubMedCrossRef
2.
go back to reference Smith AM, Barnes SA, Sperling JW, Farrell CM, Cummings JD, Cofield RH (2006) Patient and physician assessed shoulder function after arthroplasty. J Bone Joint Surg Am 88:508–513PubMedCrossRef Smith AM, Barnes SA, Sperling JW, Farrell CM, Cummings JD, Cofield RH (2006) Patient and physician assessed shoulder function after arthroplasty. J Bone Joint Surg Am 88:508–513PubMedCrossRef
3.
go back to reference Trail IA, Nuttald D (2002) The results of shoulder arthroplasty in patients with rheumatoid arthritis. J Bone Joint Surg Br 84:1121–1125PubMedCrossRef Trail IA, Nuttald D (2002) The results of shoulder arthroplasty in patients with rheumatoid arthritis. J Bone Joint Surg Br 84:1121–1125PubMedCrossRef
4.
go back to reference Merolla G, Campi F, Paladini P, Lollino N, Fauci F, Porcellini G (2009) Correlation between radiographic risk for glenoid component loosening and clinical scores in shoulder arthroplasty. Musculoskelet Surg 93:S29–S34CrossRef Merolla G, Campi F, Paladini P, Lollino N, Fauci F, Porcellini G (2009) Correlation between radiographic risk for glenoid component loosening and clinical scores in shoulder arthroplasty. Musculoskelet Surg 93:S29–S34CrossRef
5.
go back to reference Franta AK, Lenters TR, Mounce D, Neradilek B, Matsen FA III (2007) The complex characteristics of 282 unsatisfactory shoulder arthroplasties. J Shoulder Elbow Surg 16:555–562PubMedCrossRef Franta AK, Lenters TR, Mounce D, Neradilek B, Matsen FA III (2007) The complex characteristics of 282 unsatisfactory shoulder arthroplasties. J Shoulder Elbow Surg 16:555–562PubMedCrossRef
6.
go back to reference Junaid S, Gupta S, Sanghavi S, Anglin C, Emery R, Amis A, Hansen U (2010) Failure mechanism of the all polyethylene glenoid implant. J Biomech 43:714–719CrossRef Junaid S, Gupta S, Sanghavi S, Anglin C, Emery R, Amis A, Hansen U (2010) Failure mechanism of the all polyethylene glenoid implant. J Biomech 43:714–719CrossRef
7.
go back to reference Nuttal D, Haines JF, Trail IA (2012) The early migration of partially cemented fluted pegged glenoid component using radiostereometric analysis. J Shoulder Elbow Surg 21:1191–1196CrossRef Nuttal D, Haines JF, Trail IA (2012) The early migration of partially cemented fluted pegged glenoid component using radiostereometric analysis. J Shoulder Elbow Surg 21:1191–1196CrossRef
8.
go back to reference Nyffeler RW, Meyer D, Sheikh R, Koller BJ, Gerber C (2006) The effect of cementing technique on structural fixation of pegged glenoid components in total shoulder arthroplasty. J Shoulder Elbow Surg 15:106–111PubMedCrossRef Nyffeler RW, Meyer D, Sheikh R, Koller BJ, Gerber C (2006) The effect of cementing technique on structural fixation of pegged glenoid components in total shoulder arthroplasty. J Shoulder Elbow Surg 15:106–111PubMedCrossRef
9.
go back to reference Raiss P, Sowa B, Bruckner T, Eck S, Woerz S, Rohr K, Rickert M, Kasten P (2012) Pressurisation leads to better cement penetration into the glenoid bone: a cadaveric study. J Bone Joint Surg Br 94:671–677PubMedCrossRef Raiss P, Sowa B, Bruckner T, Eck S, Woerz S, Rohr K, Rickert M, Kasten P (2012) Pressurisation leads to better cement penetration into the glenoid bone: a cadaveric study. J Bone Joint Surg Br 94:671–677PubMedCrossRef
10.
go back to reference Szabo I, Buscayret F, Edwards TB, Nemoz C, Boileau P, Walch G (2005) Radiographic comparison of flat-back and convex back glenoid components in total shoulder arthroplasty. J Shoulder Elbow Surg 14:636–642PubMedCrossRef Szabo I, Buscayret F, Edwards TB, Nemoz C, Boileau P, Walch G (2005) Radiographic comparison of flat-back and convex back glenoid components in total shoulder arthroplasty. J Shoulder Elbow Surg 14:636–642PubMedCrossRef
11.
go back to reference Wiater JM, Faber MH (2009) Shoulder arthroplasty: prosthetic options and indications. J Am Acad Orthop Surg 17:415–425PubMed Wiater JM, Faber MH (2009) Shoulder arthroplasty: prosthetic options and indications. J Am Acad Orthop Surg 17:415–425PubMed
12.
go back to reference Baumgarten KM, Lashgari CJ, Yamaguchi K (2004) Glenoid resurfacing in shoulder arthroplasty: indications and contraindications. Instr Course Lect 53:3–11PubMed Baumgarten KM, Lashgari CJ, Yamaguchi K (2004) Glenoid resurfacing in shoulder arthroplasty: indications and contraindications. Instr Course Lect 53:3–11PubMed
13.
go back to reference Constant CR, Murley AH (1987) A clinical method of functional assessment of the shoulder. Clin Orthop Relat Res 214:160–164PubMed Constant CR, Murley AH (1987) A clinical method of functional assessment of the shoulder. Clin Orthop Relat Res 214:160–164PubMed
14.
go back to reference Walch G, Badet R, Boulahia A, Khoury A (1999) Morphologic study of the glenoid in primary glenohumeral osteoarthritis. J Arthroplasty 14:756–760PubMedCrossRef Walch G, Badet R, Boulahia A, Khoury A (1999) Morphologic study of the glenoid in primary glenohumeral osteoarthritis. J Arthroplasty 14:756–760PubMedCrossRef
15.
go back to reference Yongpravat C, Lester JD, Saifi C, Trubelja A, Greiwe RM, Bigliani LU et al (2012) Glenoid morphology after reaming in computer-simulated total shoulder arthroplasty. J Shoulder Elbow Surg 22:122–128PubMedCrossRef Yongpravat C, Lester JD, Saifi C, Trubelja A, Greiwe RM, Bigliani LU et al (2012) Glenoid morphology after reaming in computer-simulated total shoulder arthroplasty. J Shoulder Elbow Surg 22:122–128PubMedCrossRef
16.
go back to reference Friedman RJ, Hawthorne KB, Genez BM (1992) The use of computerized tomography in the measurement of glenoid version. J Bone Joint Surg Am 74:1032–1037PubMed Friedman RJ, Hawthorne KB, Genez BM (1992) The use of computerized tomography in the measurement of glenoid version. J Bone Joint Surg Am 74:1032–1037PubMed
17.
go back to reference Nyffeler RW, Jost B, Pfirrmann CW, Gerber C (2003) Measurement of glenoid version: conventional radiography versus computed tomography scans. J Shoulder Elbow Surg 12:493–496PubMedCrossRef Nyffeler RW, Jost B, Pfirrmann CW, Gerber C (2003) Measurement of glenoid version: conventional radiography versus computed tomography scans. J Shoulder Elbow Surg 12:493–496PubMedCrossRef
18.
19.
go back to reference Throckmorton TW, Zarkadas TC, Sperling JW, Cofield RH (2010) Pegged versus keeled glenoid components in total shoulder arthroplasty. J Shoulder Elbow Surg 19:726–733PubMedCrossRef Throckmorton TW, Zarkadas TC, Sperling JW, Cofield RH (2010) Pegged versus keeled glenoid components in total shoulder arthroplasty. J Shoulder Elbow Surg 19:726–733PubMedCrossRef
20.
go back to reference Bohsali KI, Wirth MA, Rockwood CA Jr (2006) Complications of total shoulder arthroplasty. J Bone Joint Surg Am 88:2279–2292PubMedCrossRef Bohsali KI, Wirth MA, Rockwood CA Jr (2006) Complications of total shoulder arthroplasty. J Bone Joint Surg Am 88:2279–2292PubMedCrossRef
21.
go back to reference Franklin JL, Barrett WP, Jackins SE, Matsen FA (1988) Glenoid loosening in total shoulder arthroplasty. Association with rotator cuff deficiency. J Arthroplasty 193:39–46CrossRef Franklin JL, Barrett WP, Jackins SE, Matsen FA (1988) Glenoid loosening in total shoulder arthroplasty. Association with rotator cuff deficiency. J Arthroplasty 193:39–46CrossRef
22.
go back to reference Kraljević M, Zumstein V, Wirz D, Hügli R, Müller-Gerbl M (2011) Mineralisation and mechanical strength of the glenoid cavity subchondral bone plate. Int Orthop 35:1813–1819PubMedCentralPubMedCrossRef Kraljević M, Zumstein V, Wirz D, Hügli R, Müller-Gerbl M (2011) Mineralisation and mechanical strength of the glenoid cavity subchondral bone plate. Int Orthop 35:1813–1819PubMedCentralPubMedCrossRef
23.
go back to reference Wright JM, Pellicci PM, Salvati EA, Ghelman B, Roberts M, Koh JL (2001) Bone density adjacent to press-fit acetabular components. A prospective analysis with quantitative computed tomography. J Bone Joint Surg Am 83:529–536PubMedCrossRef Wright JM, Pellicci PM, Salvati EA, Ghelman B, Roberts M, Koh JL (2001) Bone density adjacent to press-fit acetabular components. A prospective analysis with quantitative computed tomography. J Bone Joint Surg Am 83:529–536PubMedCrossRef
24.
go back to reference Mueller LA, Kress A, Nowak T, Pfander D, Pitto RP, Forst R, Schmidt R (2006) Periacetabular bone changes after uncemented total hip arthroplasty evaluated by quantitative computed tomography. Acta Orthopaedica 77:380–385PubMedCrossRef Mueller LA, Kress A, Nowak T, Pfander D, Pitto RP, Forst R, Schmidt R (2006) Periacetabular bone changes after uncemented total hip arthroplasty evaluated by quantitative computed tomography. Acta Orthopaedica 77:380–385PubMedCrossRef
25.
go back to reference Schmidt R, Pitto RP, Kress A, Ehremann C, Nowak TE, Reulbach U, Forst R, Müller L (2005) Inter- and intraobserver assessment of periacetabular osteodensitometry after cemented and uncemented total hip arthroplasty using computed tomography. Arch Orthop Trauma Surg 125:291–397PubMedCrossRef Schmidt R, Pitto RP, Kress A, Ehremann C, Nowak TE, Reulbach U, Forst R, Müller L (2005) Inter- and intraobserver assessment of periacetabular osteodensitometry after cemented and uncemented total hip arthroplasty using computed tomography. Arch Orthop Trauma Surg 125:291–397PubMedCrossRef
26.
go back to reference Munro JT, Pandit S, Walker CG, Clatworthy M, Pitto RP (2010) Loss of tibial bone density in patients with rotating- or fixed-platform TKA. Clin Orthop Relat Res 468:775–781PubMedCentralPubMedCrossRef Munro JT, Pandit S, Walker CG, Clatworthy M, Pitto RP (2010) Loss of tibial bone density in patients with rotating- or fixed-platform TKA. Clin Orthop Relat Res 468:775–781PubMedCentralPubMedCrossRef
27.
go back to reference Boileau P, Avidor C, Krishnan SG, Walch G, Kempf JF, Molé D (2002) Cemented polyethylene versus uncemented metal-backed glenoid components in total shoulder arthroplasty: a prospective, double-blind, randomized study. J Shoulder Elbow Surg 11:351–359PubMedCrossRef Boileau P, Avidor C, Krishnan SG, Walch G, Kempf JF, Molé D (2002) Cemented polyethylene versus uncemented metal-backed glenoid components in total shoulder arthroplasty: a prospective, double-blind, randomized study. J Shoulder Elbow Surg 11:351–359PubMedCrossRef
28.
go back to reference Yian EH, Werner CML, Nyffeler RW, Pfirrmann CW, Ramappa A, Sukthankar A, Gerber C (2005) Radiographic and computed tomographic analysis of cemented pegged polyethylene glenoid components in total shoulder arthroplasty. J Bone Joint Surg Am 87:1928–1936PubMedCrossRef Yian EH, Werner CML, Nyffeler RW, Pfirrmann CW, Ramappa A, Sukthankar A, Gerber C (2005) Radiographic and computed tomographic analysis of cemented pegged polyethylene glenoid components in total shoulder arthroplasty. J Bone Joint Surg Am 87:1928–1936PubMedCrossRef
29.
go back to reference Walch G, Moraga C, Young A, Castellanos-Rosas J (2012) Results of anatomic nonconstrained prosthesis in primary osteoarthritis with biconcave glenoid. J Shoulder Elbow Surg 21:1526–1533PubMedCrossRef Walch G, Moraga C, Young A, Castellanos-Rosas J (2012) Results of anatomic nonconstrained prosthesis in primary osteoarthritis with biconcave glenoid. J Shoulder Elbow Surg 21:1526–1533PubMedCrossRef
30.
go back to reference Wirth MA, Loredo R, Garcia G, Rockwood CA Jr, Southworth C, Iannotti JP (2012) Total shoulder arthroplasty with an all-polyethylene pegged bone-ingrowth glenoid component: a clinical and radiographic outcome study. J Bone Joint Surg Am 94:260–267PubMedCrossRef Wirth MA, Loredo R, Garcia G, Rockwood CA Jr, Southworth C, Iannotti JP (2012) Total shoulder arthroplasty with an all-polyethylene pegged bone-ingrowth glenoid component: a clinical and radiographic outcome study. J Bone Joint Surg Am 94:260–267PubMedCrossRef
31.
go back to reference Iannotti JP, Ricchetti ET, Rodriguez EJ, Bryan JA (2013) Development and validation of a new method of 3-dimensional assessment of glenoid and humeral component position after total shoulder arthroplasty. J Shoulder Elbow Surg 22:1413–1422PubMedCrossRef Iannotti JP, Ricchetti ET, Rodriguez EJ, Bryan JA (2013) Development and validation of a new method of 3-dimensional assessment of glenoid and humeral component position after total shoulder arthroplasty. J Shoulder Elbow Surg 22:1413–1422PubMedCrossRef
32.
go back to reference Mimar R, Limb D, Hall RM (2008) Evaluation of the mechanical and architectural properties of glenoid bone. J Shoulder Elbow Surg 17:336–341PubMedCrossRef Mimar R, Limb D, Hall RM (2008) Evaluation of the mechanical and architectural properties of glenoid bone. J Shoulder Elbow Surg 17:336–341PubMedCrossRef
33.
go back to reference Vidil A, Valenti P, Guichoux F, Barthas JH (2012) CT scan evaluation of glenoid component fixation: a prospective study of 27 minimally cemented shoulder arthroplasties. Eur J Orthop Surg Traumatol 23:521–525PubMedCentralPubMedCrossRef Vidil A, Valenti P, Guichoux F, Barthas JH (2012) CT scan evaluation of glenoid component fixation: a prospective study of 27 minimally cemented shoulder arthroplasties. Eur J Orthop Surg Traumatol 23:521–525PubMedCentralPubMedCrossRef
34.
go back to reference Gross RM, High R, Apker K, Haggstrom J, Fehringer JA, Stephan J (2011) Vacuum assist glenoid fixation: does this technique lead to a more durable glenoid component? J Shoulder Elbow Surg 20:1050–1060PubMedCrossRef Gross RM, High R, Apker K, Haggstrom J, Fehringer JA, Stephan J (2011) Vacuum assist glenoid fixation: does this technique lead to a more durable glenoid component? J Shoulder Elbow Surg 20:1050–1060PubMedCrossRef
35.
go back to reference Matsen FA 3rd, Clinton J, Lynch J, Bertelsen A, Richardson ML (2008) Glenoid component failure in total shoulder arthroplasty. J Bone Joint Surg Am 90:885–896PubMedCrossRef Matsen FA 3rd, Clinton J, Lynch J, Bertelsen A, Richardson ML (2008) Glenoid component failure in total shoulder arthroplasty. J Bone Joint Surg Am 90:885–896PubMedCrossRef
Metadata
Title
Computed tomography quantification of bone density adjacent to cemented pegged polyethylene glenoid components in shoulder arthroplasty
Authors
Giovanni Merolla
Barbara Amore
Paolo Paladini
Enrico Cavagna
Giuseppe Porcellini
Publication date
01-07-2014
Publisher
Springer Paris
Published in
European Journal of Orthopaedic Surgery & Traumatology / Issue 5/2014
Print ISSN: 1633-8065
Electronic ISSN: 1432-1068
DOI
https://doi.org/10.1007/s00590-013-1368-x

Other articles of this Issue 5/2014

European Journal of Orthopaedic Surgery & Traumatology 5/2014 Go to the issue