01-06-2010 | Ideas and Technical Innovations

Vertebroplasty with self-locking hexagonal metal implants shows comparable primary and secondary stiffness to PMMA cement augmentation techniques in a biomechanical vertebral compression fracture model

Authors: W. Schmoelz, A. C. Disch, J. F. Huber

Published in: European Spine Journal | Issue 6/2010

Login to get access

Abstract

With the growing incidence of vertebral compression fractures in elderly patients having a fair overall health condition, minimal-invasive treatment techniques are getting in focus of surgical therapy. Cement augmentation is widely performed and its complications and mechanical limitations are well described. Implants avoiding the side effects of cement augmentation while reaching the same level of stability would be desirable. The primary and secondary stability of a new augmentation method with self-locking hexagonal metal implants were investigated and compared with the performance of established augmentation options. 18 fresh-frozen human spinal specimens (Th12–L2/L3–L5) were tested with pure moments of 7.5 Nm in a six-degree-of-freedom spine simulator to investigate primary and secondary stability of three augmentation techniques: (1) vertebroplasty, (2) PMMA filled cavity and (3) hexagonal metal implants. An increasing three-step cyclic loading model was included. Elastic displacement and height loss under loading did not show significant differences between the three test groups. Investigation of primary and secondary stability evenly demonstrated comparable results for all techniques indicating an insufficiency to stabilise the fracture with higher load cycles. The newly introduced method for augmentation with the metal implant Spine Pearls achieved comparable results to bone cement based techniques in a biomechanical in vitro study. Midterm and longterm reduction preservation and ingrowth of the implants have to be proven in further studies.
Literature
1.
go back to reference Belkoff SM, Mathis JM, Jasper LE, Deramond H (2001) The biomechanics of vertebroplasty. The effect of cement volume on mechanical behavior. Spine 26:1537–1541CrossRefPubMed Belkoff SM, Mathis JM, Jasper LE, Deramond H (2001) The biomechanics of vertebroplasty. The effect of cement volume on mechanical behavior. Spine 26:1537–1541CrossRefPubMed
2.
go back to reference Bergot C, Laval-Jeantet AM, Hutchinson K, Dautraix I, Caulin F, Genant HK (2001) A comparison of spinal quantitative computed tomography with dual energy X-ray absorptiometry in European women with vertebral and nonvertebral fractures. Calcif Tissue Int 68:74–82CrossRefPubMed Bergot C, Laval-Jeantet AM, Hutchinson K, Dautraix I, Caulin F, Genant HK (2001) A comparison of spinal quantitative computed tomography with dual energy X-ray absorptiometry in European women with vertebral and nonvertebral fractures. Calcif Tissue Int 68:74–82CrossRefPubMed
3.
go back to reference Boger A, Heini P, Windolf M, Schneider E (2007) Adjacent vertebral failure after vertebroplasty: a biomechanical study of low-modulus PMMA cement. Eur Spine J 16:2118–2125CrossRefPubMed Boger A, Heini P, Windolf M, Schneider E (2007) Adjacent vertebral failure after vertebroplasty: a biomechanical study of low-modulus PMMA cement. Eur Spine J 16:2118–2125CrossRefPubMed
4.
go back to reference Cheung KM, Lu WW, Luk KD, Wong CT, Chan D, Shen JX, Qiu GX, Zheng ZM, Li CH, Liu SL, Chan WK, Leong JC (2005) Vertebroplasty by use of a strontium-containing bioactive bone cement. Spine 30:S84–S91CrossRefPubMed Cheung KM, Lu WW, Luk KD, Wong CT, Chan D, Shen JX, Qiu GX, Zheng ZM, Li CH, Liu SL, Chan WK, Leong JC (2005) Vertebroplasty by use of a strontium-containing bioactive bone cement. Spine 30:S84–S91CrossRefPubMed
5.
go back to reference Colon-Emeric CS, Saag KG (2006) Osteoporotic fractures in older adults. Best Pract Res Clin Rheumatol 20:695–706CrossRefPubMed Colon-Emeric CS, Saag KG (2006) Osteoporotic fractures in older adults. Best Pract Res Clin Rheumatol 20:695–706CrossRefPubMed
6.
go back to reference Ferguson SJ, Winkler F, Nolte LP (2002) Anterior fixation in the osteoporotic spine: cut-out and pullout characteristics of implants. Eur Spine J 11:527–534CrossRefPubMed Ferguson SJ, Winkler F, Nolte LP (2002) Anterior fixation in the osteoporotic spine: cut-out and pullout characteristics of implants. Eur Spine J 11:527–534CrossRefPubMed
7.
go back to reference Fourney DR, Schomer DF, Nader R, Chlan-Fourney J, Suki D, Ahrar K, Rhines LD, Gokaslan ZL (2003) Percutaneous vertebroplasty and kyphoplasty for painful vertebral body fractures in cancer patients. J Neurosurg 98:21–30CrossRefPubMed Fourney DR, Schomer DF, Nader R, Chlan-Fourney J, Suki D, Ahrar K, Rhines LD, Gokaslan ZL (2003) Percutaneous vertebroplasty and kyphoplasty for painful vertebral body fractures in cancer patients. J Neurosurg 98:21–30CrossRefPubMed
8.
go back to reference Fribourg D, Tang C, Sra P, Delamarter R, Bae H (2004) Incidence of subsequent vertebral fracture after kyphoplasty. Spine 29:2270–2276 (discussion 2277)CrossRefPubMed Fribourg D, Tang C, Sra P, Delamarter R, Bae H (2004) Incidence of subsequent vertebral fracture after kyphoplasty. Spine 29:2270–2276 (discussion 2277)CrossRefPubMed
9.
go back to reference Gaitanis IN, Hadjipavlou AG, Katonis PG, Tzermiadianos MN, Pasku DS, Patwardhan AG (2005) Balloon kyphoplasty for the treatment of pathological vertebral compressive fractures. Eur Spine J 14:250–260CrossRefPubMed Gaitanis IN, Hadjipavlou AG, Katonis PG, Tzermiadianos MN, Pasku DS, Patwardhan AG (2005) Balloon kyphoplasty for the treatment of pathological vertebral compressive fractures. Eur Spine J 14:250–260CrossRefPubMed
10.
go back to reference Goldhahn J, Reinhold M, Stauber M, Knop C, Frei R, Schneider E, Linke B (2006) Improved anchorage in osteoporotic vertebrae with new implant designs. J Orthop Res 24:917–925CrossRefPubMed Goldhahn J, Reinhold M, Stauber M, Knop C, Frei R, Schneider E, Linke B (2006) Improved anchorage in osteoporotic vertebrae with new implant designs. J Orthop Res 24:917–925CrossRefPubMed
11.
go back to reference Heini PF (2005) The current treatment—a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon’s perspective. Osteoporos Int 16(Suppl 2):S85–S92CrossRefPubMed Heini PF (2005) The current treatment—a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon’s perspective. Osteoporos Int 16(Suppl 2):S85–S92CrossRefPubMed
12.
go back to reference Heini PF, Berlemann U (2001) Bone substitutes in vertebroplasty. Eur Spine J 10(Suppl 2):S205–S213PubMed Heini PF, Berlemann U (2001) Bone substitutes in vertebroplasty. Eur Spine J 10(Suppl 2):S205–S213PubMed
13.
go back to reference Heini PF, Orler R (2004) Kyphoplasty for treatment of osteoporotic vertebral fractures. Eur Spine J 13:184–192CrossRefPubMed Heini PF, Orler R (2004) Kyphoplasty for treatment of osteoporotic vertebral fractures. Eur Spine J 13:184–192CrossRefPubMed
14.
go back to reference Heo DH, Kuh SU (2007) Progressive, repeated lumbar compression fracture at the same level after vertebral kyphoplasty with calcium phosphate cement. Case report. J Neurosurg Spine 6:559–562CrossRefPubMed Heo DH, Kuh SU (2007) Progressive, repeated lumbar compression fracture at the same level after vertebral kyphoplasty with calcium phosphate cement. Case report. J Neurosurg Spine 6:559–562CrossRefPubMed
15.
go back to reference Herren T, Remagen W, Schenk R (1987) Histology of the implant-bone interface in cemented and uncemented endoprostheses. Orthopade 16:239–251PubMed Herren T, Remagen W, Schenk R (1987) Histology of the implant-bone interface in cemented and uncemented endoprostheses. Orthopade 16:239–251PubMed
16.
go back to reference Higgins KB, Harten RD, Langrana NA, Reiter MF (2003) Biomechanical effects of unipedicular vertebroplasty on intact vertebrae. Spine 28:1540–1547 (discussion 1548)CrossRefPubMed Higgins KB, Harten RD, Langrana NA, Reiter MF (2003) Biomechanical effects of unipedicular vertebroplasty on intact vertebrae. Spine 28:1540–1547 (discussion 1548)CrossRefPubMed
17.
go back to reference Kettler A, Schmoelz W, Shezifi Y, Ohana N, Ben-Arye A, Claes L, Wilke HJ (2006) Biomechanical performance of the new BeadEx implant in the treatment of osteoporotic vertebral body compression fractures: restoration and maintenance of height and stability. Clin Biomech (Bristol, Avon) 21:676–682CrossRef Kettler A, Schmoelz W, Shezifi Y, Ohana N, Ben-Arye A, Claes L, Wilke HJ (2006) Biomechanical performance of the new BeadEx implant in the treatment of osteoporotic vertebral body compression fractures: restoration and maintenance of height and stability. Clin Biomech (Bristol, Avon) 21:676–682CrossRef
18.
go back to reference Kim MJ, Lindsey DP, Hannibal M, Alamin TF (2006) Vertebroplasty versus kyphoplasty: biomechanical behavior under repetitive loading conditions. Spine 31:2079–2084CrossRefPubMed Kim MJ, Lindsey DP, Hannibal M, Alamin TF (2006) Vertebroplasty versus kyphoplasty: biomechanical behavior under repetitive loading conditions. Spine 31:2079–2084CrossRefPubMed
19.
go back to reference King AB, Saag KG, Burge RT, Pisu M, Goel N (2005) Fracture reduction affects medicare economics (FRAME): impact of increased osteoporosis diagnosis and treatment. Osteoporos Int 16:1545–1557CrossRefPubMed King AB, Saag KG, Burge RT, Pisu M, Goel N (2005) Fracture reduction affects medicare economics (FRAME): impact of increased osteoporosis diagnosis and treatment. Osteoporos Int 16:1545–1557CrossRefPubMed
20.
go back to reference Knop C, Lange U, Bastian L, Blauth M (2000) Three-dimensional motion analysis with Synex. Comparative biomechanical test series with a new vertebral body replacement for the thoracolumbar spine. Eur Spine J 9:472–485CrossRefPubMed Knop C, Lange U, Bastian L, Blauth M (2000) Three-dimensional motion analysis with Synex. Comparative biomechanical test series with a new vertebral body replacement for the thoracolumbar spine. Eur Spine J 9:472–485CrossRefPubMed
21.
22.
go back to reference Leslie-Mazwi T, Deen HG (2006) Repeated fracture of a vertebral body after treatment with balloon kyphoplasty: case illustration. J Neurosurg Spine 4:270CrossRefPubMed Leslie-Mazwi T, Deen HG (2006) Repeated fracture of a vertebral body after treatment with balloon kyphoplasty: case illustration. J Neurosurg Spine 4:270CrossRefPubMed
23.
go back to reference Lewis G (2007) Percutaneous vertebroplasty and kyphoplasty for the stand-alone augmentation of osteoporosis-induced vertebral compression fractures: present status and future directions. J Biomed Mater Res B Appl Biomater 81:371–386PubMed Lewis G (2007) Percutaneous vertebroplasty and kyphoplasty for the stand-alone augmentation of osteoporosis-induced vertebral compression fractures: present status and future directions. J Biomed Mater Res B Appl Biomater 81:371–386PubMed
24.
go back to reference Lieberman IH, Togawa D, Kayanja MM (2005) Vertebroplasty and kyphoplasty: filler materials. Spine J 5:305S–316SCrossRefPubMed Lieberman IH, Togawa D, Kayanja MM (2005) Vertebroplasty and kyphoplasty: filler materials. Spine J 5:305S–316SCrossRefPubMed
25.
go back to reference Liebschner MA, Rosenberg WS, Keaveny TM (2001) Effects of bone cement volume and distribution on vertebral stiffness after vertebroplasty. Spine 26:1547–1554CrossRefPubMed Liebschner MA, Rosenberg WS, Keaveny TM (2001) Effects of bone cement volume and distribution on vertebral stiffness after vertebroplasty. Spine 26:1547–1554CrossRefPubMed
26.
go back to reference Lim TH, Brebach GT, Renner SM, Kim WJ, Kim JG, Lee RE, Andersson GB, An HS (2002) Biomechanical evaluation of an injectable calcium phosphate cement for vertebroplasty. Spine 27:1297–1302CrossRefPubMed Lim TH, Brebach GT, Renner SM, Kim WJ, Kim JG, Lee RE, Andersson GB, An HS (2002) Biomechanical evaluation of an injectable calcium phosphate cement for vertebroplasty. Spine 27:1297–1302CrossRefPubMed
27.
go back to reference McCann H, LePine M, Glaser J (2006) Biomechanical comparison of augmentation techniques for insufficiency fractures. Spine 31:E499–E502CrossRefPubMed McCann H, LePine M, Glaser J (2006) Biomechanical comparison of augmentation techniques for insufficiency fractures. Spine 31:E499–E502CrossRefPubMed
28.
go back to reference Mehbod A, Aunoble S, Le Huec JC (2003) Vertebroplasty for osteoporotic spine fracture: prevention and treatment. Eur Spine J 12(Suppl 2):S155–S162CrossRefPubMed Mehbod A, Aunoble S, Le Huec JC (2003) Vertebroplasty for osteoporotic spine fracture: prevention and treatment. Eur Spine J 12(Suppl 2):S155–S162CrossRefPubMed
29.
go back to reference Panjabi MM (1988) Biomechanical evaluation of spinal fixation devices: I. A conceptual framework. Spine 13:1129–1134CrossRefPubMed Panjabi MM (1988) Biomechanical evaluation of spinal fixation devices: I. A conceptual framework. Spine 13:1129–1134CrossRefPubMed
30.
go back to reference Panjabi MM, Krag M, Summers D, Videman T (1985) Biomechanical time-tolerance of fresh cadaveric human spine specimens. J Orthop Res 3:292–300CrossRefPubMed Panjabi MM, Krag M, Summers D, Videman T (1985) Biomechanical time-tolerance of fresh cadaveric human spine specimens. J Orthop Res 3:292–300CrossRefPubMed
31.
go back to reference Panjabi MM, Wrathall JR (1988) Biomechanical analysis of experimental spinal cord injury and functional loss. Spine 13:1365–1370CrossRefPubMed Panjabi MM, Wrathall JR (1988) Biomechanical analysis of experimental spinal cord injury and functional loss. Spine 13:1365–1370CrossRefPubMed
32.
go back to reference Perry A, Mahar A, Massie J, Arrieta N, Garfin S, Kim C (2005) Biomechanical evaluation of kyphoplasty with calcium sulfate cement in a cadaveric osteoporotic vertebral compression fracture model. Spine J 5:489–493CrossRefPubMed Perry A, Mahar A, Massie J, Arrieta N, Garfin S, Kim C (2005) Biomechanical evaluation of kyphoplasty with calcium sulfate cement in a cadaveric osteoporotic vertebral compression fracture model. Spine J 5:489–493CrossRefPubMed
33.
go back to reference Reinhold M, Schwieger K, Goldhahn J, Linke B, Knop C, Blauth M (2006) Influence of screw positioning in a new anterior spine fixator on implant loosening in osteoporotic vertebrae. Spine 31:406–413CrossRefPubMed Reinhold M, Schwieger K, Goldhahn J, Linke B, Knop C, Blauth M (2006) Influence of screw positioning in a new anterior spine fixator on implant loosening in osteoporotic vertebrae. Spine 31:406–413CrossRefPubMed
34.
go back to reference Rotter R, Pflugmacher R, Kandziora F, Ewert A, Duda G, Mittlmeier T (2007) Biomechanical in vitro testing of human osteoporotic lumbar vertebrae following prophylactic kyphoplasty with different candidate materials. Spine 32:1400–1405CrossRefPubMed Rotter R, Pflugmacher R, Kandziora F, Ewert A, Duda G, Mittlmeier T (2007) Biomechanical in vitro testing of human osteoporotic lumbar vertebrae following prophylactic kyphoplasty with different candidate materials. Spine 32:1400–1405CrossRefPubMed
35.
go back to reference Song Y, Beaupre G, Goodman SB (1999) Osseointegration of total hip arthroplasties: studies in humans and animals. J Long Term Eff Med Implants 9:77–112PubMed Song Y, Beaupre G, Goodman SB (1999) Osseointegration of total hip arthroplasties: studies in humans and animals. J Long Term Eff Med Implants 9:77–112PubMed
36.
go back to reference Steens J, Verdonschot N, Aalsma AM, Hosman AJ (2007) The influence of endplate-to-endplate cement augmentation on vertebral strength and stiffness in vertebroplasty. Spine 32:E419–E422CrossRefPubMed Steens J, Verdonschot N, Aalsma AM, Hosman AJ (2007) The influence of endplate-to-endplate cement augmentation on vertebral strength and stiffness in vertebroplasty. Spine 32:E419–E422CrossRefPubMed
37.
go back to reference Sundaresan N, Boriani S, Rothman A, Holtzman R (2004) Tumors of the osseous spine. J Neurooncol 69:273–290CrossRefPubMed Sundaresan N, Boriani S, Rothman A, Holtzman R (2004) Tumors of the osseous spine. J Neurooncol 69:273–290CrossRefPubMed
38.
go back to reference Wilke HJ, Wenger K, Claes L (1998) Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J 7:148–154CrossRefPubMed Wilke HJ, Wenger K, Claes L (1998) Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J 7:148–154CrossRefPubMed
39.
go back to reference Willert HG, Buchhorn GH (1999) Osseointegration of cemented and noncemented implants in artificial hip replacement: long-term findings in man. J Long Term Eff Med Implants 9:113–130PubMed Willert HG, Buchhorn GH (1999) Osseointegration of cemented and noncemented implants in artificial hip replacement: long-term findings in man. J Long Term Eff Med Implants 9:113–130PubMed
40.
go back to reference Yuan HA, Brown CW, Phillips FM (2004) Osteoporotic spinal deformity: a biomechanical rationale for the clinical consequences and treatment of vertebral body compression fractures. J Spinal Disord Tech 17:236–242CrossRefPubMed Yuan HA, Brown CW, Phillips FM (2004) Osteoporotic spinal deformity: a biomechanical rationale for the clinical consequences and treatment of vertebral body compression fractures. J Spinal Disord Tech 17:236–242CrossRefPubMed
Metadata
Title
Vertebroplasty with self-locking hexagonal metal implants shows comparable primary and secondary stiffness to PMMA cement augmentation techniques in a biomechanical vertebral compression fracture model
Authors
W. Schmoelz
A. C. Disch
J. F. Huber
Publication date
01-06-2010
Publisher
Springer-Verlag
Published in
European Spine Journal / Issue 6/2010
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-010-1342-9

Other articles of this Issue 6/2010

European Spine Journal 6/2010 Go to the issue