Published in: Archives of Orthopaedic and Trauma Surgery 1/2012

01-01-2012 | Hip Arthroplasty

Cushion bearings versus large diameter head metal-on-metal bearings in total hip arthroplasty: a short-term metal ion study

Authors: Antonio Moroni, Emanuele Nocco, Martha Hoque, Elena Diremigio, Dario Buffoli, Francesco Cantù, Simona Catalani, Pietro Apostoli

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 1/2012

Login to get access

Abstract

Introduction

Metal-on-metal total hip arthroplasty (MOM THA) has the advantage of replicating the femoral head size, but the postoperative elevation of serum metal ion levels is a cause for concern. Metal-on-polycarbonate-urethane is a new cushion bearing featuring a large diameter metal head coupled with a polycarbonate-urethane liner.

Aim

The aim of this study was to assess and compare serum cobalt (Co) and chromium (Cr) levels in a group of 15 patients treated with a cushion bearing THA system (Group A) and a group of 15 patients treated with a MOM THA system (Group B) at short-term. At a mean follow-up of 27.3 months (18–35 months), in Group A the median Cr and Co serum levels were significantly lower than in Group B, measuring 0.24 μg/L (0.1–2.1 μg/L) and 0.6 μg/L (0.29–2.3 μg/L) compared to 1.3 μg/L (0.1–9 μg/L, p < 0.001) and 2.9 μg/L (0.85–13.8 μg/L, p < 0.001) respectively.

Results

All patients demonstrated an excellent clinical result, as shown by the Harris and Oxford hip scores. The cushion bearing THA studied in this paper showed clinical outcomes similar to the MOM THA bearing, with the advantage of no significant metal ion elevation in the serum.

Conclusion

These findings warrant the continued clinical study of compliant bearing options.
Literature
1.
go back to reference Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89(4):780–785PubMedCrossRef Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89(4):780–785PubMedCrossRef
2.
go back to reference Harris WH (1995) The problem is osteolysis. Clin Orthop Relat Res 311:46–53PubMed Harris WH (1995) The problem is osteolysis. Clin Orthop Relat Res 311:46–53PubMed
3.
go back to reference Weber BG (1996) Experience with the Metasul total hip bearing system. Clin Orthop (329 Suppl):S69 Weber BG (1996) Experience with the Metasul total hip bearing system. Clin Orthop (329 Suppl):S69
4.
go back to reference Smith TM, Berend KR, Lombardi AV, Emerson RH, Mallory TH (2005) Metal-on-metal hip arthroplasty with large heads may prevent early dislocation. Clin Orthop Rel Res 441:137–142CrossRef Smith TM, Berend KR, Lombardi AV, Emerson RH, Mallory TH (2005) Metal-on-metal hip arthroplasty with large heads may prevent early dislocation. Clin Orthop Rel Res 441:137–142CrossRef
5.
go back to reference Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH (2005) Range of motion and stability in total hip arthroplasty with 28, 32, 38 and 44 mm femoral head sizes. J Arthroplasty 20(1):11–19PubMedCrossRef Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH (2005) Range of motion and stability in total hip arthroplasty with 28, 32, 38 and 44 mm femoral head sizes. J Arthroplasty 20(1):11–19PubMedCrossRef
6.
go back to reference Cuckler JM, Moore KD, Lombardi AV, McPherson E, Emerson R (2004) Large versus small femoral heads in metal-on-metal total hip arthroplasty. J Arthroplasty 19(8 s3):41–44 Cuckler JM, Moore KD, Lombardi AV, McPherson E, Emerson R (2004) Large versus small femoral heads in metal-on-metal total hip arthroplasty. J Arthroplasty 19(8 s3):41–44
7.
go back to reference Berry DJ, von Klnoch M, Schleck CD, Harmsen WS (2005) Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg [Am] 87:2456–2463CrossRef Berry DJ, von Klnoch M, Schleck CD, Harmsen WS (2005) Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg [Am] 87:2456–2463CrossRef
8.
go back to reference Quesada MJ, Marker DR, Mont MA (2008) Metal-on-metal hip resurfacing: advantages and disadvantages. J Arthroplasty 23:69–73PubMedCrossRef Quesada MJ, Marker DR, Mont MA (2008) Metal-on-metal hip resurfacing: advantages and disadvantages. J Arthroplasty 23:69–73PubMedCrossRef
9.
go back to reference Mont MA, Seyler TM, Ragland PS, Starr R, Erhart J, Bhave A (2007) Gait analysis of patients with resurfacing hip arthroplasty compared with hip osteoarthritis and standard total hip arthroplasty. J Arthroplasty 22:100–108PubMedCrossRef Mont MA, Seyler TM, Ragland PS, Starr R, Erhart J, Bhave A (2007) Gait analysis of patients with resurfacing hip arthroplasty compared with hip osteoarthritis and standard total hip arthroplasty. J Arthroplasty 22:100–108PubMedCrossRef
10.
go back to reference Doorn PF, Campbell PA, Worrall J, Benya PD, McKellop HA, Amstutz HC (1998) Metal wear particle characterization from metal-on-metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles. J Biomed Mater Res 42:103–111PubMedCrossRef Doorn PF, Campbell PA, Worrall J, Benya PD, McKellop HA, Amstutz HC (1998) Metal wear particle characterization from metal-on-metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles. J Biomed Mater Res 42:103–111PubMedCrossRef
11.
12.
go back to reference Visuri T, Pukkala E, Paavolainen P, Pulkkinen P, Riska EB (1996) Cancer risk after metal-on-metal and polyethylene-on-metal total hip arthroplasty. Clin Orthop 329(Suppl):280–289 Visuri T, Pukkala E, Paavolainen P, Pulkkinen P, Riska EB (1996) Cancer risk after metal-on-metal and polyethylene-on-metal total hip arthroplasty. Clin Orthop 329(Suppl):280–289
13.
go back to reference Hart AJ, Buddhev P, Winship P, Faria N, Powell J, Skinner J (2008) Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing. Hip Int 18(3):212–219PubMed Hart AJ, Buddhev P, Winship P, Faria N, Powell J, Skinner J (2008) Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing. Hip Int 18(3):212–219PubMed
14.
go back to reference Desy NM, Bergeron SG, Petit A, Huk OL, Antoniou J (2011) Surgical variables influence metal ion levels after hip resurfacing. Clin Orthop Relat Res 469(6):1635–1641PubMedCrossRef Desy NM, Bergeron SG, Petit A, Huk OL, Antoniou J (2011) Surgical variables influence metal ion levels after hip resurfacing. Clin Orthop Relat Res 469(6):1635–1641PubMedCrossRef
15.
go back to reference Scholes SC, Unsworth A, Blamey JM, Burgess IC, Jones E, Smith N (2005) Design aspects of compliant, soft layer bearings for an experimental hip prosthesis. Proc Inst Mech Eng H 219(2):79–87PubMed Scholes SC, Unsworth A, Blamey JM, Burgess IC, Jones E, Smith N (2005) Design aspects of compliant, soft layer bearings for an experimental hip prosthesis. Proc Inst Mech Eng H 219(2):79–87PubMed
16.
go back to reference Unsworth A, Dowson D, Wright V (1975) The frictional behaviour of human synovial joints–Part 1: natural joints. Trans ASME J Lubr Technol 97(3):369–376CrossRef Unsworth A, Dowson D, Wright V (1975) The frictional behaviour of human synovial joints–Part 1: natural joints. Trans ASME J Lubr Technol 97(3):369–376CrossRef
17.
go back to reference Kretzer JP, Kleinhans JA, Jakubowitz E, Thomsen M, Heisel C (2009) A meta-analysis of design- and manufacturing-related parameters influencing the wear behaviour of metal-on-metal hip joint replacements. J Orthop Res 27(11):1473–1480PubMedCrossRef Kretzer JP, Kleinhans JA, Jakubowitz E, Thomsen M, Heisel C (2009) A meta-analysis of design- and manufacturing-related parameters influencing the wear behaviour of metal-on-metal hip joint replacements. J Orthop Res 27(11):1473–1480PubMedCrossRef
18.
go back to reference Elsner JJ, Shemesh M, Mezape Y, Levenshtein M, Hakshur K, Shterling A, Linder-Ganz E, Eliaz N (2011) Long-term evaluation of a compliant cushion form acetabular bearing for hip joint replacement: a 20 million cycles wear simulation. J Orthop Res. doi:10.1002/jor.21471 Elsner JJ, Shemesh M, Mezape Y, Levenshtein M, Hakshur K, Shterling A, Linder-Ganz E, Eliaz N (2011) Long-term evaluation of a compliant cushion form acetabular bearing for hip joint replacement: a 20 million cycles wear simulation. J Orthop Res. doi:10.​1002/​jor.​21471
19.
go back to reference Chan FW, Bobyn JD, Medley JB, Krygier JJ, Tanzer M (1999) Wear and lubrication of metal-on-metal hip implants. Clin Orthop 369:10–24PubMedCrossRef Chan FW, Bobyn JD, Medley JB, Krygier JJ, Tanzer M (1999) Wear and lubrication of metal-on-metal hip implants. Clin Orthop 369:10–24PubMedCrossRef
20.
go back to reference Heisel C, Streich N, Krachler M, Jakubowitz E, Kretzer JP (2008) Characterization of the running-in period in total hip resurfacing arthroplasty—an in vivo and in vitro metal ion analysis. J Bone Joint Surg [Am] 90(3):125–133CrossRef Heisel C, Streich N, Krachler M, Jakubowitz E, Kretzer JP (2008) Characterization of the running-in period in total hip resurfacing arthroplasty—an in vivo and in vitro metal ion analysis. J Bone Joint Surg [Am] 90(3):125–133CrossRef
21.
go back to reference Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg [Am] 51(4):737–755 Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg [Am] 51(4):737–755
22.
go back to reference Murray DW, Fitzpatrick R, Rogers K, Pandit H, Beard DJ, Carr AJ, Dawson J (2007) The use of the Oxford hip and knee scores. J Bone Joint Surg [Br] 89(8):1010–1014CrossRef Murray DW, Fitzpatrick R, Rogers K, Pandit H, Beard DJ, Carr AJ, Dawson J (2007) The use of the Oxford hip and knee scores. J Bone Joint Surg [Br] 89(8):1010–1014CrossRef
23.
go back to reference Jones PR, Taylor CJ, Hukins DW, Porter ML, Hardinge K (1989) Prosthetic hip failure: retrospective radiograph image analysis of the acetabular cup. J Biomed Eng 11:253–257PubMedCrossRef Jones PR, Taylor CJ, Hukins DW, Porter ML, Hardinge K (1989) Prosthetic hip failure: retrospective radiograph image analysis of the acetabular cup. J Biomed Eng 11:253–257PubMedCrossRef
24.
go back to reference Martell JM, Berdia S (1997) Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg [Am] 79:1635–1641 Martell JM, Berdia S (1997) Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg [Am] 79:1635–1641
25.
go back to reference Shaver SM, Brown TD, Hillis SL, Callaghan JJ (1997) Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J Bone Joint Surg [Am]. 79:690–700 Shaver SM, Brown TD, Hillis SL, Callaghan JJ (1997) Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J Bone Joint Surg [Am]. 79:690–700
26.
go back to reference MacDonald SJ, Brodner W, Jacobs JJ (2004) A consensus paper on metal ions in metal-on-metal hip arthroplasties. J Arthroplasty 19(8 Suppl 3):12–16PubMedCrossRef MacDonald SJ, Brodner W, Jacobs JJ (2004) A consensus paper on metal ions in metal-on-metal hip arthroplasties. J Arthroplasty 19(8 Suppl 3):12–16PubMedCrossRef
27.
go back to reference Clarke IC, Good IV, Williams P, Schroeder D, Anissian L, Stark A, Oonishi H, Schuldies J, Gustafson G (2000) Ultra-low wear rates for rigid-on-rigid bearings in total hip replacement. Proc Inst Mech Eng H 214:331–347PubMedCrossRef Clarke IC, Good IV, Williams P, Schroeder D, Anissian L, Stark A, Oonishi H, Schuldies J, Gustafson G (2000) Ultra-low wear rates for rigid-on-rigid bearings in total hip replacement. Proc Inst Mech Eng H 214:331–347PubMedCrossRef
28.
go back to reference Sieber HP, Rieker CB, Köttig P (1999) Analysis of 118 second-generation metal-on-metal retrieved hip implants. J Bone Joint Surg [Br] 81:46–50CrossRef Sieber HP, Rieker CB, Köttig P (1999) Analysis of 118 second-generation metal-on-metal retrieved hip implants. J Bone Joint Surg [Br] 81:46–50CrossRef
29.
go back to reference Kreibich DN, Moran CG, Delves HT, Owen TD, Pinder IM (1996) Systemic release of cobalt and chromium after total hip replacement. J Bone Joint Surg [Br] 78(1):18–21 Kreibich DN, Moran CG, Delves HT, Owen TD, Pinder IM (1996) Systemic release of cobalt and chromium after total hip replacement. J Bone Joint Surg [Br] 78(1):18–21
30.
go back to reference Jacobs JJ, Skipor AK, Patterson LM, Hallab NJ, Paprosky WG, Black J, Galante JO (1998) Metal release in patients who have had a primary total hip arthroplasty. A prospective, controlled, longitudinal study. J Bone Joint Surg [Am] 80(10):1447–1458 Jacobs JJ, Skipor AK, Patterson LM, Hallab NJ, Paprosky WG, Black J, Galante JO (1998) Metal release in patients who have had a primary total hip arthroplasty. A prospective, controlled, longitudinal study. J Bone Joint Surg [Am] 80(10):1447–1458
31.
go back to reference Savarino L, Granchi D, Ciapetti G, Cenni E, Nardi Pantoli A, Rotini R, Veronesi CA, Baldini N, Giunti A (2002) Ion release in patients with metal-on-metal hip bearings in total joint replacement: a comparison with metal-on-polyethylene bearings. J Biomed Mater Res 63(5):467–474PubMedCrossRef Savarino L, Granchi D, Ciapetti G, Cenni E, Nardi Pantoli A, Rotini R, Veronesi CA, Baldini N, Giunti A (2002) Ion release in patients with metal-on-metal hip bearings in total joint replacement: a comparison with metal-on-polyethylene bearings. J Biomed Mater Res 63(5):467–474PubMedCrossRef
32.
go back to reference Savarino L, Greco M, Cenni E, Cavasinni L, Rotini R, Baldini N, Giunti A (2006) Differences in ion release after ceramic-on-ceramic and metal-on-metal total hip replacement. Medium-term follow-up. J Bone Joint Surg [Br] 88:472–476 Savarino L, Greco M, Cenni E, Cavasinni L, Rotini R, Baldini N, Giunti A (2006) Differences in ion release after ceramic-on-ceramic and metal-on-metal total hip replacement. Medium-term follow-up. J Bone Joint Surg [Br] 88:472–476
33.
go back to reference Jacobs JJ, Urban RM, Gilbert JL, Skipor AK, Black J, Jasty M, Galante JO (1995) Local and distant products from modularity. Clin Orthop 319:94–105PubMed Jacobs JJ, Urban RM, Gilbert JL, Skipor AK, Black J, Jasty M, Galante JO (1995) Local and distant products from modularity. Clin Orthop 319:94–105PubMed
34.
go back to reference Brodner W, Bitzan P, Meisinger V, Kaider A, Gottsaumer-Wolf F, Kotz R (2003) Elevated serum cobalt with metal-on-metal articulating surfaces. J Bone Joint Surg [Am] 85:2168–2173 Brodner W, Bitzan P, Meisinger V, Kaider A, Gottsaumer-Wolf F, Kotz R (2003) Elevated serum cobalt with metal-on-metal articulating surfaces. J Bone Joint Surg [Am] 85:2168–2173
35.
go back to reference Antoniou J, Zukor DJ, Mwale F, Minarik W, Petit A, Huk OL (2008) Metal ion levels in the blood of patients after hip resurfacing: a comparison between twenty-eight and thirty-six millimeter-head metal-on-metal prostheses. J Bone Joint Surg [Am] 90(3):142–148CrossRef Antoniou J, Zukor DJ, Mwale F, Minarik W, Petit A, Huk OL (2008) Metal ion levels in the blood of patients after hip resurfacing: a comparison between twenty-eight and thirty-six millimeter-head metal-on-metal prostheses. J Bone Joint Surg [Am] 90(3):142–148CrossRef
36.
go back to reference Lavigne M, Belzile EL, Roy A, Morin F, Amzica T, Vendittoli PA (2011) Comparison of whole-blood metal ion levels in four types of metal-on-metal large-diameter femoral head total hip arthroplasty: the potential influence of the adapter sleeve. J Bone Joint Surg [Am] 93(2):128–136CrossRef Lavigne M, Belzile EL, Roy A, Morin F, Amzica T, Vendittoli PA (2011) Comparison of whole-blood metal ion levels in four types of metal-on-metal large-diameter femoral head total hip arthroplasty: the potential influence of the adapter sleeve. J Bone Joint Surg [Am] 93(2):128–136CrossRef
37.
go back to reference Kretzer JP, Jakubowitz E, Krachler M, Thomsen M, Heisel C (2009) Metal release and corrosion effects of modular neck total hip arthroplasty. Int Orthop 33(6):1531–1536PubMedCrossRef Kretzer JP, Jakubowitz E, Krachler M, Thomsen M, Heisel C (2009) Metal release and corrosion effects of modular neck total hip arthroplasty. Int Orthop 33(6):1531–1536PubMedCrossRef
38.
go back to reference Brown SA, Flemming CA, Kawalec JS, Placko HE, Vassaux C, Merritt K, Payer JH, Kraay MJ (1995) Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater 6(1):19–26PubMedCrossRef Brown SA, Flemming CA, Kawalec JS, Placko HE, Vassaux C, Merritt K, Payer JH, Kraay MJ (1995) Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater 6(1):19–26PubMedCrossRef
39.
go back to reference Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP (2010) The John Charnley Award: metal-on-metal hip resurfacing versus large-diameter head metal-on-metal hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res 468(2):318–325PubMedCrossRef Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP (2010) The John Charnley Award: metal-on-metal hip resurfacing versus large-diameter head metal-on-metal hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res 468(2):318–325PubMedCrossRef
40.
go back to reference Vendittoli PA, Mottard S, Roy AG, Dupont C, Lavigne M (2007) Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hip. J Bone Joint Surg [Br] 89:441–448CrossRef Vendittoli PA, Mottard S, Roy AG, Dupont C, Lavigne M (2007) Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hip. J Bone Joint Surg [Br] 89:441–448CrossRef
Metadata
Title
Cushion bearings versus large diameter head metal-on-metal bearings in total hip arthroplasty: a short-term metal ion study
Authors
Antonio Moroni
Emanuele Nocco
Martha Hoque
Elena Diremigio
Dario Buffoli
Francesco Cantù
Simona Catalani
Pietro Apostoli
Publication date
01-01-2012
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 1/2012
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-011-1364-8

Other articles of this Issue 1/2012

Archives of Orthopaedic and Trauma Surgery 1/2012 Go to the issue