Published in: Archives of Orthopaedic and Trauma Surgery 1/2012

01-01-2012 | Trauma Surgery

Mismatch of anatomically pre-shaped locking plate on asian femurs could lead to malalignment in the minimally invasive plating of distal femoral fractures: a cadaveric study

Authors: Jin-Ho Hwang, Jong-Keon Oh, Chang-Wug Oh, Yong-Cheol Yoon, Hyeuk Woo Choi

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 1/2012

Login to get access

Abstract

Background

LCP-DF (locking compression plate-distal femur) has been introduced as an anatomically pre-shaped plate that does not require further contouring. However, the LCP-DF was developed based on skeletal measurements in Caucasians. It is unknown whether the LCP-DF fits Asians.

Objective

The purpose of this study was to assess the conformity of the LCP-DF in Asian adult femurs and also to determine the matching pattern of this anatomically pre-shaped LCP-DF in normal Asian adult femurs and the clinical implications of the matching pattern.

Methods

Sixty adult cadaver femurs were obtained from 41 male and 19 female cadavers with an average age of 63 years (31–95). An 11-hole LCP-DF was applied to the lateral surface of the distal femur according to the contour. Any mismatches between the bone and the plate were recorded. Then the distance from the inner surface of the plate to the cortex was measured at the sites of mismatch. C-arm AP image of the distal femur was taken and the angle between the distal most locking screw and the joint line was measured, and it was named as joint screw angle (JSA).

Results

Mismatch was found at the level of proximal 4–5 holes of the plate with an average distance of 11.36 mm (range 0–32 mm) at the tip of the plate. Otherwise, the overall conformity of the LCP-DF was good. The average JSA was 2.15 ± 1.78° (ranged from 0° to 5°).

Conclusion

A rather consistent pattern of mismatch was found at the proximal part of the 11-hole LCP-DF. An attempt to fit the plate to the bone at this level may cause valgus malalignment at the fracture site.
Literature
1.
go back to reference Cole PA, Zlowodzki M, Kregor PJ (2004) Treatment of proximal tibia fractures using the less invasive stabilization system: surgical experience and early clinical results in 77 fractures. J Orthop Trauma 18:528–535PubMedCrossRef Cole PA, Zlowodzki M, Kregor PJ (2004) Treatment of proximal tibia fractures using the less invasive stabilization system: surgical experience and early clinical results in 77 fractures. J Orthop Trauma 18:528–535PubMedCrossRef
2.
go back to reference Ricci AR, Yue JJ, Taffet R, Catalano JB, DeFalco RA, Wilkens KJ (2004) Less invasive stabilization system for treatment of distal femur fractures. Am J Orthop (Belle Mead NJ) 33:250–255 Ricci AR, Yue JJ, Taffet R, Catalano JB, DeFalco RA, Wilkens KJ (2004) Less invasive stabilization system for treatment of distal femur fractures. Am J Orthop (Belle Mead NJ) 33:250–255
3.
go back to reference Kregor PJ, Stannard J, Zlowodzki M, Cole PA, Alonso J (2001) Distal femoral fracture fixation utilizing the Less Invasive Stabilization System (L.I.S.S.): the technique and early results. Injury 32(Suppl 3):SC32–SC47PubMed Kregor PJ, Stannard J, Zlowodzki M, Cole PA, Alonso J (2001) Distal femoral fracture fixation utilizing the Less Invasive Stabilization System (L.I.S.S.): the technique and early results. Injury 32(Suppl 3):SC32–SC47PubMed
4.
go back to reference Kregor PJ, Stannard JA, Zlowodzki M, Cole PA (2004) Treatment of distal femur fractures using the less invasive stabilization system: surgical experience and early clinical results in 103 fractures. J Orthop Trauma 18:509–520PubMedCrossRef Kregor PJ, Stannard JA, Zlowodzki M, Cole PA (2004) Treatment of distal femur fractures using the less invasive stabilization system: surgical experience and early clinical results in 103 fractures. J Orthop Trauma 18:509–520PubMedCrossRef
5.
go back to reference Schutz M, Muller M, Regazzoni P, Hontzsch D, Krettek C, Van der Werken C, Haas N (2005) Use of the less invasive stabilization system (LISS) in patients with distal femoral (AO33) fractures: a prospective multicenter study. Arch Orthop Trauma Surg 125:102–108PubMedCrossRef Schutz M, Muller M, Regazzoni P, Hontzsch D, Krettek C, Van der Werken C, Haas N (2005) Use of the less invasive stabilization system (LISS) in patients with distal femoral (AO33) fractures: a prospective multicenter study. Arch Orthop Trauma Surg 125:102–108PubMedCrossRef
6.
go back to reference Weight M, Collinge C (2004) Early results of the less invasive stabilization system for mechanically unstable fractures of the distal femur (AO/OTA types A2, A3, C2, and C3). J Orthop Trauma 18:503–508PubMedCrossRef Weight M, Collinge C (2004) Early results of the less invasive stabilization system for mechanically unstable fractures of the distal femur (AO/OTA types A2, A3, C2, and C3). J Orthop Trauma 18:503–508PubMedCrossRef
8.
go back to reference Frigg R, Appenzeller A, Christensen R, Frenk A, Gilbert S, Schavan R (2001) The development of the distal femur Less Invasive Stabilization System (LISS). Injury 32(Suppl 3):SC24–SC31PubMed Frigg R, Appenzeller A, Christensen R, Frenk A, Gilbert S, Schavan R (2001) The development of the distal femur Less Invasive Stabilization System (LISS). Injury 32(Suppl 3):SC24–SC31PubMed
9.
go back to reference Leung KS, Procter P, Robioneck B, Behrens K (1996) Geometric mismatch of the Gamma nail to the Chinese femur. Clin Orthop Relat Res:42-48 Leung KS, Procter P, Robioneck B, Behrens K (1996) Geometric mismatch of the Gamma nail to the Chinese femur. Clin Orthop Relat Res:42-48
10.
go back to reference Hak DJ, Stewart RL, Lee M (2004) Preliminary stabilization of the less invasive stabilization system. J Orthop Trauma 18:559–561PubMedCrossRef Hak DJ, Stewart RL, Lee M (2004) Preliminary stabilization of the less invasive stabilization system. J Orthop Trauma 18:559–561PubMedCrossRef
11.
go back to reference Krettek C, Schandelmaier P, Miclau T, Bertram R, Holmes W, Tscherne H (1997) Transarticular joint reconstruction and indirect plate osteosynthesis for complex distal supracondylar femoral fractures. Injury 28(Suppl 1):A31–A41PubMedCrossRef Krettek C, Schandelmaier P, Miclau T, Bertram R, Holmes W, Tscherne H (1997) Transarticular joint reconstruction and indirect plate osteosynthesis for complex distal supracondylar femoral fractures. Injury 28(Suppl 1):A31–A41PubMedCrossRef
Metadata
Title
Mismatch of anatomically pre-shaped locking plate on asian femurs could lead to malalignment in the minimally invasive plating of distal femoral fractures: a cadaveric study
Authors
Jin-Ho Hwang
Jong-Keon Oh
Chang-Wug Oh
Yong-Cheol Yoon
Hyeuk Woo Choi
Publication date
01-01-2012
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 1/2012
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-011-1375-5

Other articles of this Issue 1/2012

Archives of Orthopaedic and Trauma Surgery 1/2012 Go to the issue