Published in: European Radiology 6/2015

01-06-2015 | Musculoskeletal

Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system

Authors: S. Demehri, A. Muhit, W. Zbijewski, J. W. Stayman, J. Yorkston, N. Packard, R. Senn, D. Yang, D. Foos, G. K. Thawait, L. M. Fayad, A. Chhabra, J. A. Carrino, J. H. Siewerdsen

Published in: European Radiology | Issue 6/2015

Login to get access

Abstract

Objective

To assess visualization tasks using cone-beam CT (CBCT) compared to multi-detector CT (MDCT) for musculoskeletal extremity imaging.

Methods

Ten cadaveric hands and ten knees were examined using a dedicated CBCT prototype and a clinical multi-detector CT using nominal protocols (80kVp-108mAs for CBCT; 120kVp- 300mAs for MDCT). Soft tissue and bone visualization tasks were assessed by four radiologists using five-point satisfaction (for CBCT and MDCT individually) and five-point preference (side-by-side CBCT versus MDCT image quality comparison) rating tests. Ratings were analyzed using Kruskal–Wallis and Wilcoxon signed-rank tests, and observer agreement was assessed using the Kappa-statistic.

Results

Knee CBCT images were rated “excellent” or “good” (median scores 5 and 4) for “bone” and “soft tissue” visualization tasks. Hand CBCT images were rated “excellent” or “adequate” (median scores 5 and 3) for “bone” and “soft tissue” visualization tasks. Preference tests rated CBCT equivalent or superior to MDCT for bone visualization and favoured the MDCT for soft tissue visualization tasks. Intraobserver agreement for CBCT satisfaction tests was fair to almost perfect (κ ~ 0.26–0.92), and interobserver agreement was fair to moderate (κ ~ 0.27–0.54).

Conclusion

CBCT provided excellent image quality for bone visualization and adequate image quality for soft tissue visualization tasks.

Key Points

CBCT provided adequate image quality for diagnostic tasks in extremity imaging.
CBCT images were “excellent” for “bone” and “good/adequate” for “soft tissue” visualization tasks.
CBCT image quality was equivalent/superior to MDCT for bone visualization tasks.
Appendix
Available only for authorised users
Literature
1.
go back to reference Fayad LM, Corl F, Fishman EK (2009) Pediatric skeletal trauma: use of multiplanar reformatted and three-dimensional 64-row multidetector CT in the emergency department. Radiographics 29(1):135–50CrossRefPubMed Fayad LM, Corl F, Fishman EK (2009) Pediatric skeletal trauma: use of multiplanar reformatted and three-dimensional 64-row multidetector CT in the emergency department. Radiographics 29(1):135–50CrossRefPubMed
2.
go back to reference Demehri S, Chalian M, Farahani SJ, Fishman EK, Fayad LM (2014) Detection and characterization of tendon abnormalities with multidetector computed tomography. J Comput Assist Tomogr 38(2):299–307CrossRefPubMed Demehri S, Chalian M, Farahani SJ, Fishman EK, Fayad LM (2014) Detection and characterization of tendon abnormalities with multidetector computed tomography. J Comput Assist Tomogr 38(2):299–307CrossRefPubMed
3.
go back to reference Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 72(1):75–80PubMed Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 72(1):75–80PubMed
4.
go back to reference O”Connell A, Conover DL, Zhang Y et al (2010) Cone-beam CT for breast imaging: Radiation dose, breast coverage, and image quality. AJR Am J Roentgenol 195(2):496–509CrossRef O”Connell A, Conover DL, Zhang Y et al (2010) Cone-beam CT for breast imaging: Radiation dose, breast coverage, and image quality. AJR Am J Roentgenol 195(2):496–509CrossRef
5.
go back to reference Ingrosso G, Miceli R, Fedele D et al (2012) Cone-beam computed tomography in hypofractionated stereotactic radiotherapy for brain metastases. Radiat Oncol 7(1):54CrossRefPubMedCentralPubMed Ingrosso G, Miceli R, Fedele D et al (2012) Cone-beam computed tomography in hypofractionated stereotactic radiotherapy for brain metastases. Radiat Oncol 7(1):54CrossRefPubMedCentralPubMed
6.
go back to reference Yu JJ, Kim GT, Choi YS et al (2012) Accuracy of a cone beam computed tomography-guided surgical stent for orthodontic mini-implant placement. Angle Orthod 82(2):275–83CrossRefPubMed Yu JJ, Kim GT, Choi YS et al (2012) Accuracy of a cone beam computed tomography-guided surgical stent for orthodontic mini-implant placement. Angle Orthod 82(2):275–83CrossRefPubMed
7.
go back to reference Carrino JA, Al Muhit A, Zbijewski W, Thawait GK, Stayman JW, Packard N, Senn R, Yang D, Foos DH, Yorkston J, Siewerdsen JH (2014) Dedicated cone-beam CT system for extremity imaging. Radiology 270(3):816–24CrossRefPubMedCentralPubMed Carrino JA, Al Muhit A, Zbijewski W, Thawait GK, Stayman JW, Packard N, Senn R, Yang D, Foos DH, Yorkston J, Siewerdsen JH (2014) Dedicated cone-beam CT system for extremity imaging. Radiology 270(3):816–24CrossRefPubMedCentralPubMed
8.
go back to reference Hirschmann A, Pfirrmann CW, Klammer G, Espinosa N, Buck FM (2014) Upright Cone CT of the hindfoot: Comparison of the non-weight-bearing with the upright weight-bearing position. Eur Radiol 24(3):553–8CrossRefPubMed Hirschmann A, Pfirrmann CW, Klammer G, Espinosa N, Buck FM (2014) Upright Cone CT of the hindfoot: Comparison of the non-weight-bearing with the upright weight-bearing position. Eur Radiol 24(3):553–8CrossRefPubMed
9.
go back to reference Tuominen EK, Kankare J, Koskinen SK, Mattila KT (2013) Weight-bearing CT imaging of the lower extremity. AJR Am J Roentgenol 200(1):146–8CrossRefPubMed Tuominen EK, Kankare J, Koskinen SK, Mattila KT (2013) Weight-bearing CT imaging of the lower extremity. AJR Am J Roentgenol 200(1):146–8CrossRefPubMed
10.
go back to reference Suzuki T, Ito S, Handa S et al (2010) New low-field extremity MRI, compacTscan: comparison with whole-body 1.5 T conventional MRI. Mod Rheumatol 20(4):331–6CrossRefPubMed Suzuki T, Ito S, Handa S et al (2010) New low-field extremity MRI, compacTscan: comparison with whole-body 1.5 T conventional MRI. Mod Rheumatol 20(4):331–6CrossRefPubMed
11.
go back to reference Choi JH, Fahrig R, Keil A et al (2013) Fiducial marker-based correction for involuntary motion in weight-bearing C-arm CT scanning of knees. Part I. Numerical model-based optimization. Med Phys 40(9):091905CrossRefPubMedCentralPubMed Choi JH, Fahrig R, Keil A et al (2013) Fiducial marker-based correction for involuntary motion in weight-bearing C-arm CT scanning of knees. Part I. Numerical model-based optimization. Med Phys 40(9):091905CrossRefPubMedCentralPubMed
12.
go back to reference Koskinen SK, Haapamäki VV, Salo J et al (2013) CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT). Skelet Radiol 42(5):649–57CrossRef Koskinen SK, Haapamäki VV, Salo J et al (2013) CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT). Skelet Radiol 42(5):649–57CrossRef
13.
go back to reference Guggenberger R, Winklhofer S, Spiczak JV, Andreisek G, Alkadhi H (2013) In vitro high-resolution flat-panel computed tomographic arthrography for artificial cartilage defect detection: comparison with multidetector computed tomography. Invest Radiol 48(8):614–21CrossRefPubMed Guggenberger R, Winklhofer S, Spiczak JV, Andreisek G, Alkadhi H (2013) In vitro high-resolution flat-panel computed tomographic arthrography for artificial cartilage defect detection: comparison with multidetector computed tomography. Invest Radiol 48(8):614–21CrossRefPubMed
14.
go back to reference Finkenstaedt T, Morsbach F, Calcagni M, Vich M, Pfirrmann CW, Alkadhi H, Runge VM, Andreisek G, Guggenberger R (2014) Metallic artifacts from internal scaphoid fracture fixation screws: comparison between C-arm flat-panel, cone-beam, and multidetector computed tomography. Invest Radiol 49(8):532–9CrossRefPubMed Finkenstaedt T, Morsbach F, Calcagni M, Vich M, Pfirrmann CW, Alkadhi H, Runge VM, Andreisek G, Guggenberger R (2014) Metallic artifacts from internal scaphoid fracture fixation screws: comparison between C-arm flat-panel, cone-beam, and multidetector computed tomography. Invest Radiol 49(8):532–9CrossRefPubMed
15.
go back to reference Zbijewski W, De Jean P, Prakash P et al (2011) A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization. Med Phys 38(8):4700–4713CrossRefPubMedCentralPubMed Zbijewski W, De Jean P, Prakash P et al (2011) A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization. Med Phys 38(8):4700–4713CrossRefPubMedCentralPubMed
16.
go back to reference Dixon RL, Anderson JA, Bakalyar DM, et al (2010) Comprehensive methodology for the evaluation of radiation dose in x-ray computed tomography. AAPM Task Group Report 111 Dixon RL, Anderson JA, Bakalyar DM, et al (2010) Comprehensive methodology for the evaluation of radiation dose in x-ray computed tomography. AAPM Task Group Report 111
17.
go back to reference Boone JM, Strauss KJ, Cody DD, et al (2011) Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations. AAPM Task Group Report 204 Boone JM, Strauss KJ, Cody DD, et al (2011) Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations. AAPM Task Group Report 204
18.
go back to reference Kashani H, Gang JG, Shkumat NA et al (2009) Development of a high-performance dual-energy chest imaging system: initial investigation of diagnostic performance. Acad Radiol 16(4):464–476CrossRefPubMedCentralPubMed Kashani H, Gang JG, Shkumat NA et al (2009) Development of a high-performance dual-energy chest imaging system: initial investigation of diagnostic performance. Acad Radiol 16(4):464–476CrossRefPubMedCentralPubMed
19.
go back to reference Williams DB, Siewerdsen JH, Tward DJ et al (2007) Optimal kvp selection for dual-energy imaging of the chest: evaluation by task-specific observer preference tests. Med Phys 34(10):3916–3925CrossRefPubMed Williams DB, Siewerdsen JH, Tward DJ et al (2007) Optimal kvp selection for dual-energy imaging of the chest: evaluation by task-specific observer preference tests. Med Phys 34(10):3916–3925CrossRefPubMed
20.
go back to reference Bidgood WD Jr, Horii SC (1992) Introduction to the ACR-NEMA DICOM standard. Radiographics 12(2):345–355CrossRefPubMed Bidgood WD Jr, Horii SC (1992) Introduction to the ACR-NEMA DICOM standard. Radiographics 12(2):345–355CrossRefPubMed
21.
go back to reference Kruskal WH, Wallis WA (1952) Use of ranks in One-criterion variance analysis. J Am Stat Assoc 47(260):583–621CrossRef Kruskal WH, Wallis WA (1952) Use of ranks in One-criterion variance analysis. J Am Stat Assoc 47(260):583–621CrossRef
22.
go back to reference Zou KH, Carrino JA (2002) Comparison of accuracy and interreader agreement in side-by-side versus independent evaluations of MR imaging of the medial collateral ligament of the elbow. Acad Radiol 9(5):520–525CrossRefPubMed Zou KH, Carrino JA (2002) Comparison of accuracy and interreader agreement in side-by-side versus independent evaluations of MR imaging of the medial collateral ligament of the elbow. Acad Radiol 9(5):520–525CrossRefPubMed
23.
go back to reference Wilcoxon F (1945) Individual Comparisons by Ranking Methods. Biom Bull 1(6):80–83CrossRef Wilcoxon F (1945) Individual Comparisons by Ranking Methods. Biom Bull 1(6):80–83CrossRef
24.
go back to reference Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20(1):37–46CrossRef Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20(1):37–46CrossRef
25.
go back to reference Cohen J (1968) Weighted kappa: nominal scale agreement with provision for scaled disagreement or partial credit. Psychol Bull 70(4):213–20CrossRefPubMed Cohen J (1968) Weighted kappa: nominal scale agreement with provision for scaled disagreement or partial credit. Psychol Bull 70(4):213–20CrossRefPubMed
26.
go back to reference Mac Niocaill RF, Quinlan JF, Stapleton RD et al (2011) Inter- and intra-observer variability associated with the use of the Mirels” scoring system for metastatic bone lesions. Int Orthop 35(1):83–86CrossRefPubMedCentralPubMed Mac Niocaill RF, Quinlan JF, Stapleton RD et al (2011) Inter- and intra-observer variability associated with the use of the Mirels” scoring system for metastatic bone lesions. Int Orthop 35(1):83–86CrossRefPubMedCentralPubMed
27.
go back to reference Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33(1):159–174CrossRefPubMed Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33(1):159–174CrossRefPubMed
29.
go back to reference Sisniega A, Zbijewski W, Badal A et al (2013) Monte Carlo study of the effects of system geometry and antiscatter grids on cone-beam CT scatter distributions. Med Phys 40(5):051915CrossRefPubMedCentralPubMed Sisniega A, Zbijewski W, Badal A et al (2013) Monte Carlo study of the effects of system geometry and antiscatter grids on cone-beam CT scatter distributions. Med Phys 40(5):051915CrossRefPubMedCentralPubMed
30.
go back to reference Erdogan H, Fessler JA (1999) Monotonic algorithms for transmission tomography. IEEE Trans Med Imaging 18(9):801–814CrossRefPubMed Erdogan H, Fessler JA (1999) Monotonic algorithms for transmission tomography. IEEE Trans Med Imaging 18(9):801–814CrossRefPubMed
31.
go back to reference Muhit AA, Arora S, Ogawa M et al (2013) Peripheral quantitative CT (pQCT) using a dedicated extremity cone-beam CT scanner. SPIE Med Imaging: Biomed Appl Mol, Struct Func Imaging 29:867203–867207 Muhit AA, Arora S, Ogawa M et al (2013) Peripheral quantitative CT (pQCT) using a dedicated extremity cone-beam CT scanner. SPIE Med Imaging: Biomed Appl Mol, Struct Func Imaging 29:867203–867207
33.
go back to reference Odgaard A (1997) Three-dimensional methods for quantification of cancellous bone architecture. Bone 20(4):315–28CrossRefPubMed Odgaard A (1997) Three-dimensional methods for quantification of cancellous bone architecture. Bone 20(4):315–28CrossRefPubMed
34.
go back to reference Roos PG, Colbeth RE, Mollov I et al (2004) Multiple-gain-ranging readout method to extend the dynamic range of amorphous silicon flat-panel imagers. ProcSPIE Phys Med Imaging 5368(1):139–149 Roos PG, Colbeth RE, Mollov I et al (2004) Multiple-gain-ranging readout method to extend the dynamic range of amorphous silicon flat-panel imagers. ProcSPIE Phys Med Imaging 5368(1):139–149
Metadata
Title
Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system
Authors
S. Demehri
A. Muhit
W. Zbijewski
J. W. Stayman
J. Yorkston
N. Packard
R. Senn
D. Yang
D. Foos
G. K. Thawait
L. M. Fayad
A. Chhabra
J. A. Carrino
J. H. Siewerdsen
Publication date
01-06-2015
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 6/2015
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-014-3546-6

Other articles of this Issue 6/2015

European Radiology 6/2015 Go to the issue