Published in: Skeletal Radiology 5/2013

01-05-2013 | Scientific Article

CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT)

Authors: Seppo K. Koskinen, Ville V. Haapamäki, Jari Salo, Nina C. Lindfors, Mika Kortesniemi, Lauri Seppälä, Kimmo T. Mattila

Published in: Skeletal Radiology | Issue 5/2013

Login to get access

Abstract

Purpose

To evaluate the feasibility and intra- and interobserver agreement of CBCT arthrography of wrist ligaments, triangular fibrocartilaginous complex (TFCC), and to assess the sensitivity (SE), specificity (SP), accuracy (ACC), and positive and negative predictive value (PPV, NPV) of CBCT arthrography in the diagnosis of scapholunate (SLL) and lunotriquetral (LTL) ligament tears, TFCC, and cartilage abnormalities of the scaphoid and lunate with their corresponding radial surfaces (scaphoid and lunate fossa) using a novel, mobile, dedicated extremity CBCT scanner.

Materials and methods

Fifty-two consecutively enrolled subjects (26 M, 26 F, mean age 38 years, range 18–66 years) with suspected wrist ligament tears underwent CBCT-arthrography before normally scheduled MR arthrography.An extremity CBCT was used for imaging with isotropic voxel size of 0.4 × 0.4 × 0.4 mm3. Subsequent routine 1.5 T MRI was performed using a dedicated wrist coil.Two observers reviewed the anonymized CBCT images twice for contrast enhancement (CE) and technical details (TD), for tears of the SLL, LTL, and TFCC. Also, cartilage abnormalities of the scaphoid and lunate with their corresponding radial surfaces (scaphoid and lunate fossa) were evaluated. Inter- and intraobserver agreement was determined using weighted kappa statistics. Since no surgery was performed, MRI served as a reference standard, and SE and SP, ACC, PPV, and NPV were calculated.

Results

Intra- and interobserver kappa values for both readers (reader 1/reader 2; first reading/second reading) with 95 % confidence limits were: CE 0.54 (0.08–1.00)/ 0.75 (0.46–1.00); 0.73 (0.29–1.00)/ 0.45 (0.07–0.83), TD 0.53 (0.30–0.88)/ 0.86 (0.60–1.00); 0.56 (0.22–0.91)/ 0.67 (0.37–0.98), SLL 0.59 (0.25–0.93)/ 0.66 (0.42–0.91); 0.31 (0.06–0.56)/ 0.49 (0.26–0.73), LTL 0.83 (0.66–1.00)/ 0.68 (0.46–0.91); 0.90 (0.79–1.00)/ 0.48 (0.22–0.74); TFCC (0.72–1.00)/ (0.79–1.00); 0.65 (0.43–0.87)/ 0.59 (0.35–0.83), radius (scaphoid fossa) 0.45 (0.12–0.77)/ 0.64 (0.31–0.96); 0.58 (0.19–0.96)/ 0.38 (0.09–0.66), scaphoid 0.43 (0.12–0.74)/ 0.76 (0.55–0.96); 0.37 (0.00–0.75)/ 0.32 (0.04–0.59), radius (lunate fossa) 0.68 (0.36–1.00)/ 0.42 (0.00–0.86); 0.62 (0.29–0.96)/ 0.51 (0.12–0.91), and lunate 0.53 (0.16–0.90)/ 0.68 (0.44–0.91); 0.59 (0.29–0.88)/ 0.42 (0.00–0.84), respectively.
The overall mean accuracy was 82–92 % and specificity was 81–94 %. Sensitivity for LTL and TFCC tears was 76–83, but for SLL tears it was 58 %. For cartilage abnormalities, the accuracy and negative predictive value were high, 90–98 %.

Conclusions

A dedicated CBCT extremity scanner is a new method for evaluating the wrist ligaments and radiocarpal cartilage. The method has an overall accuracy of 82–86 % and specificity 81–91 %. For cartilage abnormalities, the accuracy and negative predictive value were high.
Literature
1.
go back to reference Zlatkin MB, Chao PC, Osterman AL, Schnall MD, Dalinka MK, Kressel HY. Chronic wrist pain: evaluation with high-resolution imaging. Radiology. 1989;173:723–9.PubMed Zlatkin MB, Chao PC, Osterman AL, Schnall MD, Dalinka MK, Kressel HY. Chronic wrist pain: evaluation with high-resolution imaging. Radiology. 1989;173:723–9.PubMed
2.
go back to reference Smith DK. MR imaging of normal and injured wrist ligaments. Magn Reson Imaging Clin N Am. 1995;3:229–48.PubMed Smith DK. MR imaging of normal and injured wrist ligaments. Magn Reson Imaging Clin N Am. 1995;3:229–48.PubMed
3.
go back to reference Golimbu CN, Firooznia H, Melone CP, Rafii M, Weinreb J, Leber C. Tears of the triangular fibrocartilage of the wrist: MR imaging. Radiology. 1989;173:731–3.PubMed Golimbu CN, Firooznia H, Melone CP, Rafii M, Weinreb J, Leber C. Tears of the triangular fibrocartilage of the wrist: MR imaging. Radiology. 1989;173:731–3.PubMed
4.
go back to reference Hobby JL, Tom BD, Bearcroft PW, Dixon AK. Magnetic resonance imaging of the wrist: diagnostic performance statistics. Clin Radiol. 2001;56:50–7.PubMedCrossRef Hobby JL, Tom BD, Bearcroft PW, Dixon AK. Magnetic resonance imaging of the wrist: diagnostic performance statistics. Clin Radiol. 2001;56:50–7.PubMedCrossRef
5.
go back to reference Haims AH, Moore AE, Schweitzer ME, Morrison WB, Deely D, Culp RW, et al. MRI in the diagnosis of cartilage injury in the wrist. Am J Roentgenol. 2004;182:1267–70.CrossRef Haims AH, Moore AE, Schweitzer ME, Morrison WB, Deely D, Culp RW, et al. MRI in the diagnosis of cartilage injury in the wrist. Am J Roentgenol. 2004;182:1267–70.CrossRef
6.
go back to reference Moser T, Dosch JC, Moussaoui A, Dietemann JL. Wrist ligament tears: evaluation of MRI and combined MDCT and MR arthrography. Am J Roentgenol. 2007;188:1278–86.CrossRef Moser T, Dosch JC, Moussaoui A, Dietemann JL. Wrist ligament tears: evaluation of MRI and combined MDCT and MR arthrography. Am J Roentgenol. 2007;188:1278–86.CrossRef
7.
go back to reference Cerezal L, de Dios Berná-Mestre J, Canga A, et al. MR and CT arthrography of the wrist. Semin Musculoskelet Radiol. 2012;16:27–41.PubMedCrossRef Cerezal L, de Dios Berná-Mestre J, Canga A, et al. MR and CT arthrography of the wrist. Semin Musculoskelet Radiol. 2012;16:27–41.PubMedCrossRef
8.
go back to reference Mozzo P, Procacci C, Tacconi A, et al. A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results. Eur Radiol. 1998;8:1558–64.PubMedCrossRef Mozzo P, Procacci C, Tacconi A, et al. A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results. Eur Radiol. 1998;8:1558–64.PubMedCrossRef
9.
go back to reference Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of compact computed tomographic apparatus for dental use. Dentomaxillofac Radiol. 1999;28:245–8.PubMedCrossRef Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of compact computed tomographic apparatus for dental use. Dentomaxillofac Radiol. 1999;28:245–8.PubMedCrossRef
10.
go back to reference De Vos W, Casselman J, Swennen GRJ. Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: a systematic review of the literature. Int J Oral Maxillofac Surg. 2009;38:609–25.PubMedCrossRef De Vos W, Casselman J, Swennen GRJ. Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: a systematic review of the literature. Int J Oral Maxillofac Surg. 2009;38:609–25.PubMedCrossRef
11.
go back to reference Faccioli N, Foti G, Barillari M, Atzei A, Mucelli RP. Finger fractures imaging: accuracy of cone-beam computed tomography and multislice computed tomography. Skeletal Radiol. 2010;39:1087–95.PubMedCrossRef Faccioli N, Foti G, Barillari M, Atzei A, Mucelli RP. Finger fractures imaging: accuracy of cone-beam computed tomography and multislice computed tomography. Skeletal Radiol. 2010;39:1087–95.PubMedCrossRef
12.
go back to reference De Cock J, Mermuys K, Goubau J, van Petegem S, Houthoofdt B, Casselman JW. Cone-beam computed tomography: a new low-dose, high-resolution imaging technique of the wrist, presentation of three cases with technique. Skeletal Radiol. 2012;41:93–6.PubMedCrossRef De Cock J, Mermuys K, Goubau J, van Petegem S, Houthoofdt B, Casselman JW. Cone-beam computed tomography: a new low-dose, high-resolution imaging technique of the wrist, presentation of three cases with technique. Skeletal Radiol. 2012;41:93–6.PubMedCrossRef
13.
go back to reference Zbijewski W, De Jean P, Prakash P, et al. A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization. Med Phys. 2011;38:4700–13.PubMedCrossRef Zbijewski W, De Jean P, Prakash P, et al. A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization. Med Phys. 2011;38:4700–13.PubMedCrossRef
14.
15.
go back to reference Biswas D, Bible JE, Bohan M, Simpson AK, Whang PG, Grauer JN. Radiation exposure from musculoskeletal computerized tomographic scans. J Bone Joint Surg Am. 2009;91:1882–9.PubMedCrossRef Biswas D, Bible JE, Bohan M, Simpson AK, Whang PG, Grauer JN. Radiation exposure from musculoskeletal computerized tomographic scans. J Bone Joint Surg Am. 2009;91:1882–9.PubMedCrossRef
16.
go back to reference Landis JR, Koch GG. The measurement of observer agreement for categorial data. Biometrics. 1997;33:159–74.CrossRef Landis JR, Koch GG. The measurement of observer agreement for categorial data. Biometrics. 1997;33:159–74.CrossRef
17.
go back to reference Guggenberger R, Fischer MA, Hodler J, Pfammatter T, Andreisek G. Flat-panel CT-arthrography. Feasibility study and comparison to multidetector CT arthrography. Invest Radiol. 2012;47:312–8.PubMedCrossRef Guggenberger R, Fischer MA, Hodler J, Pfammatter T, Andreisek G. Flat-panel CT-arthrography. Feasibility study and comparison to multidetector CT arthrography. Invest Radiol. 2012;47:312–8.PubMedCrossRef
18.
go back to reference Zheng Z-Z, Shan H, Li X. Fat-suppressed 3D T1-weighted gradient-echo imaging of the cartilage with a volumetric interpolated breath-hold examination. Am J Roentgenol. 2010;194:W414–9.CrossRef Zheng Z-Z, Shan H, Li X. Fat-suppressed 3D T1-weighted gradient-echo imaging of the cartilage with a volumetric interpolated breath-hold examination. Am J Roentgenol. 2010;194:W414–9.CrossRef
19.
go back to reference Heigl B, Kowarschik M. High-Speed Reconstruction for C-Arm Computed Tomography. In Proceedings of the 9th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, pp. 25–28, Lindau, Germany, July 2007. Heigl B, Kowarschik M. High-Speed Reconstruction for C-Arm Computed Tomography. In Proceedings of the 9th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, pp. 25–28, Lindau, Germany, July 2007.
20.
go back to reference Tsiklakis K, Donta C, Gavala S, Karayianni K, Kamenopoulou V, Hourdakis CJ. Dose reduction in maxillofacial imaging using low-dose cone-beam CT. Eur J Radiol. 2005;56:413–7.PubMedCrossRef Tsiklakis K, Donta C, Gavala S, Karayianni K, Kamenopoulou V, Hourdakis CJ. Dose reduction in maxillofacial imaging using low-dose cone-beam CT. Eur J Radiol. 2005;56:413–7.PubMedCrossRef
21.
go back to reference Roberts JA, Drage NA, Davies J, Thomas DW. Effective dose from cone-beam CT examinations in dentistry. Br J Radiol. 2009;82:35–40.PubMedCrossRef Roberts JA, Drage NA, Davies J, Thomas DW. Effective dose from cone-beam CT examinations in dentistry. Br J Radiol. 2009;82:35–40.PubMedCrossRef
22.
go back to reference Koong B. Cone-beam imaging: is this the ultimate imaging modality? Clin Oral Implants Res. 2010;21:1201–8.PubMedCrossRef Koong B. Cone-beam imaging: is this the ultimate imaging modality? Clin Oral Implants Res. 2010;21:1201–8.PubMedCrossRef
23.
go back to reference Hart D, Wall BF. Radiation Exposure of the UK population from Medical and Dental X-ray Examinations; Report NRPB-W4; National Radiological Protection Board, UK; 2002. Hart D, Wall BF. Radiation Exposure of the UK population from Medical and Dental X-ray Examinations; Report NRPB-W4; National Radiological Protection Board, UK; 2002.
24.
go back to reference Donker DK, Hasman A, van Geijn HP. Interpretation of low kappa values. Int J Biomed Comp. 1993;33:55–64.CrossRef Donker DK, Hasman A, van Geijn HP. Interpretation of low kappa values. Int J Biomed Comp. 1993;33:55–64.CrossRef
25.
go back to reference Scheck RJ, Romagnolo A, Hierner R, Pfluger T, Wilhelm K, Hahn K. The carpal ligaments in MR arthrography of the wrist: correlation with standard MRI and wrist arthroscopy. J Magn Reson Imaging. 1999;9:468–74.PubMedCrossRef Scheck RJ, Romagnolo A, Hierner R, Pfluger T, Wilhelm K, Hahn K. The carpal ligaments in MR arthrography of the wrist: correlation with standard MRI and wrist arthroscopy. J Magn Reson Imaging. 1999;9:468–74.PubMedCrossRef
26.
go back to reference Omlor G, Jung M, Grieser T, Ludwig K. Depiction of the triangular fibro-cartilage in patients with ulnar-sided wrist pain: comparison of direct multi-slice CT arthrography and direct MR arthrography. Eur Radiol. 2009;19:147–51.PubMedCrossRef Omlor G, Jung M, Grieser T, Ludwig K. Depiction of the triangular fibro-cartilage in patients with ulnar-sided wrist pain: comparison of direct multi-slice CT arthrography and direct MR arthrography. Eur Radiol. 2009;19:147–51.PubMedCrossRef
Metadata
Title
CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT)
Authors
Seppo K. Koskinen
Ville V. Haapamäki
Jari Salo
Nina C. Lindfors
Mika Kortesniemi
Lauri Seppälä
Kimmo T. Mattila
Publication date
01-05-2013
Publisher
Springer-Verlag
Published in
Skeletal Radiology / Issue 5/2013
Print ISSN: 0364-2348
Electronic ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-012-1516-0

Other articles of this Issue 5/2013

Skeletal Radiology 5/2013 Go to the issue