Published in: European Journal of Nuclear Medicine and Molecular Imaging 11/2023

Open Access 17-06-2023 | Original Article

No impact of attenuation and scatter correction on the interpretation of dopamine transporter SPECT in patients with clinically uncertain parkinsonian syndrome

Authors: Tassilo Schiebler, Ivayla Apostolova, Franziska Lara Mathies, Catharina Lange, Susanne Klutmann, Ralph Buchert

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 11/2023

Login to get access

Abstract  

Purpose

The benefit from attenuation and scatter correction (ASC) of dopamine transporter (DAT)-SPECT for the detection of nigrostriatal degeneration in clinical routine is still a matter of debate. The current study evaluated the impact of ASC on visual interpretation and semi-quantitative analysis of DAT-SPECT in a large patient sample.

Methods

One thousand seven hundred forty consecutive DAT-SPECT with 123I-FP-CIT from clinical routine were included retrospectively. SPECT images were reconstructed iteratively without and with ASC. Attenuation correction was based on uniform attenuation maps, scatter correction on simulation. All SPECT images were categorized with respect to the presence versus the absence of Parkinson-typical reduction of striatal 123I-FP-CIT uptake by three independent readers. Image reading was performed twice to assess intra-reader variability. The specific 123I-FP-CIT binding ratio (SBR) was used for automatic categorization, separately with and without ASC.

Results

The mean proportion of cases with discrepant categorization by the same reader between the two reading sessions was practically the same without and with ASC, about 2.2%. The proportion of DAT-SPECT with discrepant categorization without versus with ASC by the same reader was 1.66% ± 0.50% (1.09–1.95%), not exceeding the benchmark of 2.2% from intra-reader variability. This also applied to automatic categorization of the DAT-SPECT images based on the putamen SBR (1.78% discrepant cases between without versus with ASC).

Conclusion

Given the large sample size, the current findings provide strong evidence against a relevant impact of ASC with uniform attenuation and simulation-based scatter correction on the clinical utility of DAT-SPECT to detect nigrostriatal degeneration in patients with clinically uncertain parkinsonian syndrome.
Appendix
Available only for authorised users
Literature
2.
go back to reference Berardelli A, Wenning G, Antonini A, Berg D, Bloem B, Bonifati V, et al. EFNS/MDS-ES recommendations for the diagnosis of P arkinson’s disease. Eur J Neurol. 2013;20:16–34.CrossRefPubMed Berardelli A, Wenning G, Antonini A, Berg D, Bloem B, Bonifati V, et al. EFNS/MDS-ES recommendations for the diagnosis of P arkinson’s disease. Eur J Neurol. 2013;20:16–34.CrossRefPubMed
4.
go back to reference Albert NL, Unterrainer M, Diemling M, Xiong G, Bartenstein P, Koch W, et al. Implementation of the European multicentre database of healthy controls for [(123)I]FP-CIT SPECT increases diagnostic accuracy in patients with clinically uncertain parkinsonian syndromes. Eur J Nucl Med Mol Imaging. 2016;43:1315–22. https://doi.org/10.1007/s00259-015-3304-2.CrossRefPubMed Albert NL, Unterrainer M, Diemling M, Xiong G, Bartenstein P, Koch W, et al. Implementation of the European multicentre database of healthy controls for [(123)I]FP-CIT SPECT increases diagnostic accuracy in patients with clinically uncertain parkinsonian syndromes. Eur J Nucl Med Mol Imaging. 2016;43:1315–22. https://​doi.​org/​10.​1007/​s00259-015-3304-2.CrossRefPubMed
7.
go back to reference Kim KM, Varrone A, Watabe H, Shidahara M, Fujita M, Innis RB, et al. Contribution of scatter and attenuation compensation to SPECT images of nonuniformly distributed brain activities. J Nucl Med. 2003;44:512–9.PubMed Kim KM, Varrone A, Watabe H, Shidahara M, Fujita M, Innis RB, et al. Contribution of scatter and attenuation compensation to SPECT images of nonuniformly distributed brain activities. J Nucl Med. 2003;44:512–9.PubMed
9.
go back to reference Booij J, Sokole EB, Stabin MG, Janssen AGM, de Bruin K, van Royen EA. Human biodistribution and dosimetry of [I-123]FP-CIT: a potent radioligand for imaging of dopamine transporters. Eur J Nucl Med. 1998;25:24–30.CrossRefPubMed Booij J, Sokole EB, Stabin MG, Janssen AGM, de Bruin K, van Royen EA. Human biodistribution and dosimetry of [I-123]FP-CIT: a potent radioligand for imaging of dopamine transporters. Eur J Nucl Med. 1998;25:24–30.CrossRefPubMed
16.
go back to reference Varrone A, Dickson JC, Tossici-Bolt L, Sera T, Asenbaum S, Booij J, et al. European multicentre database of healthy controls for [123I]FP-CIT SPECT (ENC-DAT): age-related effects, gender differences and evaluation of different methods of analysis. Eur J Nucl Med Mol Imaging. 2013;40:213–27. https://doi.org/10.1007/s00259-012-2276-8.CrossRefPubMed Varrone A, Dickson JC, Tossici-Bolt L, Sera T, Asenbaum S, Booij J, et al. European multicentre database of healthy controls for [123I]FP-CIT SPECT (ENC-DAT): age-related effects, gender differences and evaluation of different methods of analysis. Eur J Nucl Med Mol Imaging. 2013;40:213–27. https://​doi.​org/​10.​1007/​s00259-012-2276-8.CrossRefPubMed
17.
go back to reference Crespo C, Gallego J, Cot A, Falcon C, Bullich S, Pareto D, et al. Quantification of dopaminergic neurotransmission SPECT studies with 123I-labelled radioligands. A comparison between different imaging systems and data acquisition protocols using Monte Carlo simulation. Eur J Nucl Med Mol Imaging. 2008;35:1334–42. https://doi.org/10.1007/s00259-007-0711-z.CrossRefPubMed Crespo C, Gallego J, Cot A, Falcon C, Bullich S, Pareto D, et al. Quantification of dopaminergic neurotransmission SPECT studies with 123I-labelled radioligands. A comparison between different imaging systems and data acquisition protocols using Monte Carlo simulation. Eur J Nucl Med Mol Imaging. 2008;35:1334–42. https://​doi.​org/​10.​1007/​s00259-007-0711-z.CrossRefPubMed
21.
go back to reference Bartenstein P, Grunwald F, Kuwert T, Tatsch K, Sabri O, Benkert O, et al. Clinical applications of single photon emission tomography in neuromedicine. 1. Neuro-oncology, epilepsy, movement disorders, cerebrovascular disease. Nuklearmedizin. 2000;39:180–95.CrossRefPubMed Bartenstein P, Grunwald F, Kuwert T, Tatsch K, Sabri O, Benkert O, et al. Clinical applications of single photon emission tomography in neuromedicine. 1. Neuro-oncology, epilepsy, movement disorders, cerebrovascular disease. Nuklearmedizin. 2000;39:180–95.CrossRefPubMed
25.
go back to reference Diemling M. HERMES camera correction for the ENCDAT database using DaTscan. Hermes Medical Solutions; 2021. Diemling M. HERMES camera correction for the ENCDAT database using DaTscan. Hermes Medical Solutions; 2021.
27.
go back to reference HybridRecon (White Paper). Hermes Medical Solutions. HybridRecon (White Paper). Hermes Medical Solutions.
35.
go back to reference Subramaniam RM, Frey KA, Hunt CH, Mercier GA Jr, Solnes LB, Colletti PM, et al. ACR–ACNM practice parameter for the performance of dopamine transporter (DAT) Single photon emission computed tomography (SPECT) imaging for movement disorders. Clin Nucl Med. 2017;42:847–52.CrossRefPubMed Subramaniam RM, Frey KA, Hunt CH, Mercier GA Jr, Solnes LB, Colletti PM, et al. ACR–ACNM practice parameter for the performance of dopamine transporter (DAT) Single photon emission computed tomography (SPECT) imaging for movement disorders. Clin Nucl Med. 2017;42:847–52.CrossRefPubMed
36.
39.
go back to reference Niznik HB, Fogel EF, Fassos FF, Seeman P. The dopamine transporter is absent in parkinsonian putamen and reduced in the caudate nucleus. J Neurochem. 1991;56:192–8.CrossRefPubMed Niznik HB, Fogel EF, Fassos FF, Seeman P. The dopamine transporter is absent in parkinsonian putamen and reduced in the caudate nucleus. J Neurochem. 1991;56:192–8.CrossRefPubMed
42.
go back to reference Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single-isotope and dual-isotope Spect. J Nucl Med. 1993;34:2216–21.PubMed Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single-isotope and dual-isotope Spect. J Nucl Med. 1993;34:2216–21.PubMed
45.
go back to reference Soret M, Koulibaly PM, Darcourt J, Hapdey S, Buvat I. Quantitative accuracy of dopaminergic neurotransmission imaging with (123)I SPECT. J Nucl Med. 2003;44:1184–93.PubMed Soret M, Koulibaly PM, Darcourt J, Hapdey S, Buvat I. Quantitative accuracy of dopaminergic neurotransmission imaging with (123)I SPECT. J Nucl Med. 2003;44:1184–93.PubMed
54.
go back to reference Lapa C, Spehl TS, Brumberg J, Isaias IU, Schlogl S, Lassmann M, et al. Influence of CT-based attenuation correction on dopamine transporter SPECT with [I-123] FP-CIT. Am J Nucl Med Molec. 2015;5:278–86. Lapa C, Spehl TS, Brumberg J, Isaias IU, Schlogl S, Lassmann M, et al. Influence of CT-based attenuation correction on dopamine transporter SPECT with [I-123] FP-CIT. Am J Nucl Med Molec. 2015;5:278–86.
55.
go back to reference Bienkiewicz M, Gorska-Chrzastek M, Siennicki J, Gajos A, Bogucki A, Mochecka-Thoelke A, et al. Impact of CT based attenuation correction on quantitative assessment of DaTSCAN ((123)I-ioflupane) imaging in diagnosis of extrapyramidal diseases. Nucl Med Rev Cent East Eur. 2008;11:53–8.PubMed Bienkiewicz M, Gorska-Chrzastek M, Siennicki J, Gajos A, Bogucki A, Mochecka-Thoelke A, et al. Impact of CT based attenuation correction on quantitative assessment of DaTSCAN ((123)I-ioflupane) imaging in diagnosis of extrapyramidal diseases. Nucl Med Rev Cent East Eur. 2008;11:53–8.PubMed
Metadata
Title
No impact of attenuation and scatter correction on the interpretation of dopamine transporter SPECT in patients with clinically uncertain parkinsonian syndrome
Authors
Tassilo Schiebler
Ivayla Apostolova
Franziska Lara Mathies
Catharina Lange
Susanne Klutmann
Ralph Buchert
Publication date
17-06-2023
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 11/2023
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-023-06293-2

Other articles of this Issue 11/2023

European Journal of Nuclear Medicine and Molecular Imaging 11/2023 Go to the issue