Improved design decreases wear in total knee arthroplasty with varus malalignment

Authors: Kazutaka Nishikawa, Ken Okazaki, Shuichi Matsuda, Yasutaka Tashiro, Shinya Kawahara, Hiroyuki Nakahara, Shigetoshi Okamoto, Takeshi Shimoto, Hidehiko Higaki, Yukihide Iwamoto

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 11/2014

Login to get access

Abstract

Purpose

Controversy still exists whether coronal malalignment would influence the long-term survival of total knee arthroplasty (TKA). The hypothesis was that an improved design of the articular surface of modern TKA would prevent the increase in contact stresses and thus decrease the wear even when the implant was placed in a varus position. Two different designs of TKA were compared biomechanically and clinically.

Methods

The patients whose prosthesis was initially placed in a varus alignment by the postoperative long-leg radiographs were selected. Seventeen knees using the NexGen LPS and 16 knees using the MG I were examined. Changes in postoperative alignment and the thickness of the polyethylene insert in a follow-up period of approximately 7 years were evaluated. Additionally, an in vitro biomechanical testing was conducted to measure the contact stresses and the contact area at the tibiofemoral joint of the NexGen LPS and the MG I components mounted on a servohydraulic testing device.

Results

Although the long-leg alignment did not change in NexGen LPS, the varus alignment significantly progressed in MG I. The thickness of polyethylene insert in MG I decreased a significantly greater amount compared with that in NexGen LPS. Biomechanical test showed that the NexGen LPS had a larger contact area and lower mean and peak contact stresses than the MG I significantly.

Conclusion

These results suggest that comprehensive factors of modern prosthesis including improved implant designs could improve the durability of polyethylene insert and decrease implant failures due to component malalignment.
Literature
1.
go back to reference Berend ME, Ritter MA, Meding JB, Faris PM, Keating EM, Redelman R, Faris GW, Davis KE (2004) Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res 428:26–34PubMedCrossRef Berend ME, Ritter MA, Meding JB, Faris PM, Keating EM, Redelman R, Faris GW, Davis KE (2004) Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res 428:26–34PubMedCrossRef
2.
go back to reference Bonner TJ, Eardley WG, Patterson P, Gregg PJ (2011) The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 years. J Bone Joint Surg Br 93(9):1217–1222PubMedCrossRef Bonner TJ, Eardley WG, Patterson P, Gregg PJ (2011) The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 years. J Bone Joint Surg Br 93(9):1217–1222PubMedCrossRef
3.
go back to reference Bugbee WD, Ammeen DJ, Parks NL, Engh GA (1998) 4- to 10-year results with the anatomic modular total knee. Clin Orthop Relat Res 348:158–165PubMedCrossRef Bugbee WD, Ammeen DJ, Parks NL, Engh GA (1998) 4- to 10-year results with the anatomic modular total knee. Clin Orthop Relat Res 348:158–165PubMedCrossRef
4.
go back to reference Chauhan SK, Clark GW, Lloyd S, Scott RG, Breidahl W, Sikorski JM (2004) Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol). J Bone Joint Surg Br 86(6):818–823PubMedCrossRef Chauhan SK, Clark GW, Lloyd S, Scott RG, Breidahl W, Sikorski JM (2004) Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol). J Bone Joint Surg Br 86(6):818–823PubMedCrossRef
5.
go back to reference D’Lima DD, Hermida JC, Chen PC, Colwell CW Jr (2001) Polyethylene wear and variations in knee kinematics. Clin Orthop Relat Res 392:124–130PubMedCrossRef D’Lima DD, Hermida JC, Chen PC, Colwell CW Jr (2001) Polyethylene wear and variations in knee kinematics. Clin Orthop Relat Res 392:124–130PubMedCrossRef
6.
go back to reference Ezzet KA, Hermida JC, Steklov N, D’Lima DD (2012) Wear of polyethylene against oxidized zirconium femoral components effect of aggressive kinematic conditions and malalignment in total knee arthroplasty. J Arthroplasty 27(1):116–121PubMedCrossRef Ezzet KA, Hermida JC, Steklov N, D’Lima DD (2012) Wear of polyethylene against oxidized zirconium femoral components effect of aggressive kinematic conditions and malalignment in total knee arthroplasty. J Arthroplasty 27(1):116–121PubMedCrossRef
7.
go back to reference Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 24:39–43PubMedCrossRef Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 24:39–43PubMedCrossRef
8.
go back to reference Kuster MS, Stachowiak GW (2002) Factors affecting polyethylene wear in total knee arthroplasty. Orthopedics 25:s235–s242PubMed Kuster MS, Stachowiak GW (2002) Factors affecting polyethylene wear in total knee arthroplasty. Orthopedics 25:s235–s242PubMed
9.
go back to reference Liau JJ, Cheng CK, Huang CH, Lo WH (2002) The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol, Avon) 17(2):140–146CrossRef Liau JJ, Cheng CK, Huang CH, Lo WH (2002) The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol, Avon) 17(2):140–146CrossRef
10.
go back to reference Lombardi AV Jr, Berend KR, Ng VY (2011) Neutral mechanical alignment: a requirement for successful TKA: affirms. Lombardi Orthop 34(9):e504–e506 Lombardi AV Jr, Berend KR, Ng VY (2011) Neutral mechanical alignment: a requirement for successful TKA: affirms. Lombardi Orthop 34(9):e504–e506
11.
go back to reference Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R (2009) Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty 24(4):570–578PubMedCrossRef Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R (2009) Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty 24(4):570–578PubMedCrossRef
12.
go back to reference Matsuda S, Miura H, Nagamine R, Urabe K, Harimaya K, Matsunobu T, Iwamoto Y (1999) Changes in knee alignment after total knee arthroplasty. J Arthroplasty 14(5):566–570PubMedCrossRef Matsuda S, Miura H, Nagamine R, Urabe K, Harimaya K, Matsunobu T, Iwamoto Y (1999) Changes in knee alignment after total knee arthroplasty. J Arthroplasty 14(5):566–570PubMedCrossRef
13.
go back to reference Matsuda S, Whiteside LA, White SE (1999) The effect of varus tilt on contact stresses in total knee arthroplasty: a biomechanical study. Orthopedics 22(3):303–307PubMed Matsuda S, Whiteside LA, White SE (1999) The effect of varus tilt on contact stresses in total knee arthroplasty: a biomechanical study. Orthopedics 22(3):303–307PubMed
14.
go back to reference Matziolis G, Adam J, Perka C (2010) Varus malalignment has no influence on clinical outcome in midterm follow-up after total knee replacement. Arch Orthop Trauma Surg 130(12):1487–1491PubMedCrossRef Matziolis G, Adam J, Perka C (2010) Varus malalignment has no influence on clinical outcome in midterm follow-up after total knee replacement. Arch Orthop Trauma Surg 130(12):1487–1491PubMedCrossRef
15.
go back to reference Pang HN, Yeo SJ, Chong HC, Chin PL, Ong J, Lo NN (2011) Computer-assisted gap balancing technique improves outcome in total knee arthroplasty, compared with conventional measured resection technique. Knee Surg Sports Traumatol Arthrosc 19(9):1496–1503PubMedCrossRef Pang HN, Yeo SJ, Chong HC, Chin PL, Ong J, Lo NN (2011) Computer-assisted gap balancing technique improves outcome in total knee arthroplasty, compared with conventional measured resection technique. Knee Surg Sports Traumatol Arthrosc 19(9):1496–1503PubMedCrossRef
16.
go back to reference Parratte S, Pagnano MW, Trousdale RT, Berry DJ (2010) Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surg Am 92(12):2143–2149PubMedCrossRef Parratte S, Pagnano MW, Trousdale RT, Berry DJ (2010) Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surg Am 92(12):2143–2149PubMedCrossRef
17.
go back to reference Perillo-Marcone A, Barrett DS, Taylor M (2000) The importance of tibial alignment: finite element analysis of tibial malalignment. J Arthroplasty 15(8):1020–1027PubMedCrossRef Perillo-Marcone A, Barrett DS, Taylor M (2000) The importance of tibial alignment: finite element analysis of tibial malalignment. J Arthroplasty 15(8):1020–1027PubMedCrossRef
18.
go back to reference Pijls BG, Van der Linden-Van der Zwaag HM, Nelissen RG (2012) Polyethylene thickness is a risk factor for wear necessitating insert exchange. Int Orthop 36(6):1175–1180PubMedCrossRefPubMedCentral Pijls BG, Van der Linden-Van der Zwaag HM, Nelissen RG (2012) Polyethylene thickness is a risk factor for wear necessitating insert exchange. Int Orthop 36(6):1175–1180PubMedCrossRefPubMedCentral
19.
go back to reference Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res 299:153–156PubMed Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res 299:153–156PubMed
20.
go back to reference Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA (2011) The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am 93(17):1588–1596PubMedCrossRef Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA (2011) The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am 93(17):1588–1596PubMedCrossRef
21.
go back to reference Taylor M, Tanner KE (1997) Fatigue failure of cancellous bone: a possible cause of implant migration and loosening. J Bone Joint Surg Br 79(2):181–182PubMedCrossRef Taylor M, Tanner KE (1997) Fatigue failure of cancellous bone: a possible cause of implant migration and loosening. J Bone Joint Surg Br 79(2):181–182PubMedCrossRef
Metadata
Title
Improved design decreases wear in total knee arthroplasty with varus malalignment
Authors
Kazutaka Nishikawa
Ken Okazaki
Shuichi Matsuda
Yasutaka Tashiro
Shinya Kawahara
Hiroyuki Nakahara
Shigetoshi Okamoto
Takeshi Shimoto
Hidehiko Higaki
Yukihide Iwamoto
Publication date
01-11-2014
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 11/2014
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-013-2506-z

Other articles of this Issue 11/2014

Knee Surgery, Sports Traumatology, Arthroscopy 11/2014 Go to the issue