Published in: International Orthopaedics 6/2012

Open Access 01-06-2012 | Original Paper

Polyethylene thickness is a risk factor for wear necessitating insert exchange

Authors: Bart G. Pijls, Henrica M. J. Van der Linden-Van der Zwaag, Rob G. H. H. Nelissen

Published in: International Orthopaedics | Issue 6/2012

Login to get access

Abstract

Purpose

The aim of this observational study was to investigate the optimal minimal polyethylene (PE) thickness in total knee arthroplasty (TKA) and identify other risk factors associated with revision of the insert due to wear.

Methods

A total of 84 TKA were followed for 11–16 years. All patients received the same prosthesis design (Interax; Howmedica/ Stryker) with halfbearings: separate PE-inserts medially and laterally. Statistical analysis comprised Cox-regression to correct for confounding.

Results

Eight knees (9.5%) had been revised due to thinning inserts and an additional patient is scheduled for revision. PE thickness, diagnosis, BMI and weight are risk factors for insert exchange. For each millimetre decrease in PE thickness, the risk of insert exchange increases 3.0 times, which remains after correction for age, gender, weight, diagnosis and femoral-tibial angle. Insert exchange was 4.73 times more likely in OA-patients compared to RA-patients. For every unit increase in BMI and weight the risk for insert exchange increases 1.40 times and 1.14 times, respectively.

Conclusions

In conclusion we therefore advise against the use of thin PE inserts in modular TKA and recommend PE inserts with a minimal 8-mm thickness.
Literature
1.
go back to reference Crowninshield RD, Rosenberg AG, Sporer SM (2006) Changing demographics of patients with total joint replacement. Clin Orthop Relat Res 443:266–272PubMedCrossRef Crowninshield RD, Rosenberg AG, Sporer SM (2006) Changing demographics of patients with total joint replacement. Clin Orthop Relat Res 443:266–272PubMedCrossRef
2.
go back to reference Kurtz SM, Ong KL, Lau E, Widmer M, Maravic M, Gomez-Barrena E, de Fatima de Pina M, Manno V, Torre M, Walter WL, de Steiger R, Geesink RG, Peltola M, Roder C (2011) International survey of primary and revision total knee replacement. Int Orthop. Mar 15 [Epub ahead of print] doi:10.1007/s00264-00011-01235-00265 Kurtz SM, Ong KL, Lau E, Widmer M, Maravic M, Gomez-Barrena E, de Fatima de Pina M, Manno V, Torre M, Walter WL, de Steiger R, Geesink RG, Peltola M, Roder C (2011) International survey of primary and revision total knee replacement. Int Orthop. Mar 15 [Epub ahead of print] doi:10.​1007/​s00264-00011-01235-00265
3.
go back to reference Tsao A, Mintz L, McRae CR, Stulberg SD, Wright T (1993) Failure of the porous-coated anatomic prosthesis in total knee arthroplasty due to severe polyethylene wear. J Bone Joint Surg Am 75(1):19–26PubMed Tsao A, Mintz L, McRae CR, Stulberg SD, Wright T (1993) Failure of the porous-coated anatomic prosthesis in total knee arthroplasty due to severe polyethylene wear. J Bone Joint Surg Am 75(1):19–26PubMed
4.
go back to reference Fisher J, Jennings LM, Galvin AL, Jin ZM, Stone MH, Ingham E (2010) 2009 Knee Society Presidential Guest Lecture: Polyethylene wear in total knees. Clin Orthop Relat Res 468(1):12–18PubMedCrossRef Fisher J, Jennings LM, Galvin AL, Jin ZM, Stone MH, Ingham E (2010) 2009 Knee Society Presidential Guest Lecture: Polyethylene wear in total knees. Clin Orthop Relat Res 468(1):12–18PubMedCrossRef
5.
go back to reference Bartel DL, Bicknell VL, Wright TM (1986) The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am 68(7):1041–1051PubMed Bartel DL, Bicknell VL, Wright TM (1986) The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am 68(7):1041–1051PubMed
6.
go back to reference Bugbee WD, Ammeen DJ, Parks NL, Engh GA (1998) 4- to 10-year results with the anatomic modular total knee. Clin Orthop Relat Res 348(348):158–165PubMed Bugbee WD, Ammeen DJ, Parks NL, Engh GA (1998) 4- to 10-year results with the anatomic modular total knee. Clin Orthop Relat Res 348(348):158–165PubMed
7.
go back to reference Kilgus DJ, Moreland JR, Finerman GA, Funahashi TT, Tipton JS (1991) Catastrophic wear of tibial polyethylene inserts. Clin Orthop Relat Res 273(273):223–231PubMed Kilgus DJ, Moreland JR, Finerman GA, Funahashi TT, Tipton JS (1991) Catastrophic wear of tibial polyethylene inserts. Clin Orthop Relat Res 273(273):223–231PubMed
8.
go back to reference Meding JB, Ritter MA, Faris PM (2001) Total knee arthroplasty with 4.4 mm of tibial polyethylene: 10-year followup. Clin Orthop Relat Res 388:112–117PubMedCrossRef Meding JB, Ritter MA, Faris PM (2001) Total knee arthroplasty with 4.4 mm of tibial polyethylene: 10-year followup. Clin Orthop Relat Res 388:112–117PubMedCrossRef
9.
go back to reference Nelissen RG, Valstar ER, Rozing PM (1998) The effect of hydroxyapatite on the micromotion of total knee prostheses. A prospective, randomized, double-blind study. J Bone Joint Surg Am 80(11):1665–1672PubMed Nelissen RG, Valstar ER, Rozing PM (1998) The effect of hydroxyapatite on the micromotion of total knee prostheses. A prospective, randomized, double-blind study. J Bone Joint Surg Am 80(11):1665–1672PubMed
10.
go back to reference Higuchi H, Hatayama K, Shimizu M, Kobayashi A, Kobayashi T, Takagishi K (2009) Relationship between joint gap difference and range of motion in total knee arthroplasty: a prospective randomised study between different platforms. Int Orthop 33(4):997–1000PubMedCrossRef Higuchi H, Hatayama K, Shimizu M, Kobayashi A, Kobayashi T, Takagishi K (2009) Relationship between joint gap difference and range of motion in total knee arthroplasty: a prospective randomised study between different platforms. Int Orthop 33(4):997–1000PubMedCrossRef
11.
go back to reference Rousseau MA, Lazennec JY-Y, Catonne Y (2008) Early mechanical failure in total knee arthroplasty. Int Orthop 32(1):53–56PubMedCrossRef Rousseau MA, Lazennec JY-Y, Catonne Y (2008) Early mechanical failure in total knee arthroplasty. Int Orthop 32(1):53–56PubMedCrossRef
12.
go back to reference Sugimoto K, Iwai M, Okahashi K, Kaneko K, Tanaka M, Takakura Y (2005) Premature failure of the polyethylene tibial bearing surface of the Interax knee arthroplasty. J Arthroplasty 20(5):658–663PubMedCrossRef Sugimoto K, Iwai M, Okahashi K, Kaneko K, Tanaka M, Takakura Y (2005) Premature failure of the polyethylene tibial bearing surface of the Interax knee arthroplasty. J Arthroplasty 20(5):658–663PubMedCrossRef
13.
go back to reference Atkins D, Best D, Briss PA, Eccles M, Falck-Ytter Y, Flottorp S, Guyatt GH, Harbour RT, Haugh MC, Henry D, Hill S, Jaeschke R, Leng G, Liberati A, Magrini N, Mason J, Middleton P, Mrukowicz J, O'Connell D, Oxman AD, Phillips B, Schunemann HJ, Edejer TT, Varonen H, Vist GE, Williams JW Jr, Zaza S (2004) Grading quality of evidence and strength of recommendations. BMJ 328(7454):1490PubMedCrossRef Atkins D, Best D, Briss PA, Eccles M, Falck-Ytter Y, Flottorp S, Guyatt GH, Harbour RT, Haugh MC, Henry D, Hill S, Jaeschke R, Leng G, Liberati A, Magrini N, Mason J, Middleton P, Mrukowicz J, O'Connell D, Oxman AD, Phillips B, Schunemann HJ, Edejer TT, Varonen H, Vist GE, Williams JW Jr, Zaza S (2004) Grading quality of evidence and strength of recommendations. BMJ 328(7454):1490PubMedCrossRef
14.
go back to reference Insall JN, Dorr LD, Scott RD, Scott WN (1989) Rationale of the Knee Society clinical rating system. Clin Orthop Relat Res 248:13–14PubMed Insall JN, Dorr LD, Scott RD, Scott WN (1989) Rationale of the Knee Society clinical rating system. Clin Orthop Relat Res 248:13–14PubMed
15.
go back to reference Spicer DD, Pomeroy DL, Badenhausen WE, Schaper LA, Curry JI, Suthers KE, Smith MW (2001) Body mass index as a predictor of outcome in total knee replacement. Int Orthop 25(14):246–249PubMed Spicer DD, Pomeroy DL, Badenhausen WE, Schaper LA, Curry JI, Suthers KE, Smith MW (2001) Body mass index as a predictor of outcome in total knee replacement. Int Orthop 25(14):246–249PubMed
16.
go back to reference Bert JM, Reuben J, Kelly F, Gross M, Elting J (1998) The incidence of modular tibial polyethylene insert exchange in total knee arthroplasty when polyethylene failure occurs. J Arthroplasty 13(6):609–614PubMedCrossRef Bert JM, Reuben J, Kelly F, Gross M, Elting J (1998) The incidence of modular tibial polyethylene insert exchange in total knee arthroplasty when polyethylene failure occurs. J Arthroplasty 13(6):609–614PubMedCrossRef
17.
19.
go back to reference Collier MB, Engh CA Jr, McAuley JP, Engh GA (2007) Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am 89(6):1306–1314PubMedCrossRef Collier MB, Engh CA Jr, McAuley JP, Engh GA (2007) Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am 89(6):1306–1314PubMedCrossRef
20.
go back to reference Griffin WL, Scott RD, Dalury DF, Mahoney OM, Chiavetta JB, Odum SM (2007) Modular insert exchange in knee arthroplasty for treatment of wear and osteolysis. Clin Orthop Relat Res 464:132–137PubMed Griffin WL, Scott RD, Dalury DF, Mahoney OM, Chiavetta JB, Odum SM (2007) Modular insert exchange in knee arthroplasty for treatment of wear and osteolysis. Clin Orthop Relat Res 464:132–137PubMed
21.
go back to reference Willson SE, Munro ML, Sandwell JC, Ezzet KA, Colwell CW Jr (2010) Isolated tibial polyethylene insert exchange outcomes after total knee arthroplasty. Clin Orthop Relat Res 468(1):96–101PubMedCrossRef Willson SE, Munro ML, Sandwell JC, Ezzet KA, Colwell CW Jr (2010) Isolated tibial polyethylene insert exchange outcomes after total knee arthroplasty. Clin Orthop Relat Res 468(1):96–101PubMedCrossRef
22.
go back to reference Chillag KJ, Barth E (1991) An analysis of polyethylene thickness in modular total knee components. Clin Orthop Relat Res 273:261–263PubMed Chillag KJ, Barth E (1991) An analysis of polyethylene thickness in modular total knee components. Clin Orthop Relat Res 273:261–263PubMed
24.
go back to reference Reay E, Wu J, Holland J, Deehan D (2009) Premature failure of Kinemax Plus total knee replacements. J Bone Joint Surg Br 91(5):604–611PubMedCrossRef Reay E, Wu J, Holland J, Deehan D (2009) Premature failure of Kinemax Plus total knee replacements. J Bone Joint Surg Br 91(5):604–611PubMedCrossRef
25.
go back to reference Feng EL, Stulberg SD, Wixson RL (1994) Progressive subluxation and polyethylene wear in total knee replacements with flat articular surfaces. Clin Orthop Relat Res 299:60–71PubMed Feng EL, Stulberg SD, Wixson RL (1994) Progressive subluxation and polyethylene wear in total knee replacements with flat articular surfaces. Clin Orthop Relat Res 299:60–71PubMed
26.
go back to reference Toksvig-Larsen S, Ryd L, Stentstrom A, Dansgard F, Jonsson K, Robertson O, Lindstrand A (1996) The porous-coated anatomic total knee experience. Special emphasis on complications and wear. J Arthroplasty 11(1):11–17PubMedCrossRef Toksvig-Larsen S, Ryd L, Stentstrom A, Dansgard F, Jonsson K, Robertson O, Lindstrand A (1996) The porous-coated anatomic total knee experience. Special emphasis on complications and wear. J Arthroplasty 11(1):11–17PubMedCrossRef
Metadata
Title
Polyethylene thickness is a risk factor for wear necessitating insert exchange
Authors
Bart G. Pijls
Henrica M. J. Van der Linden-Van der Zwaag
Rob G. H. H. Nelissen
Publication date
01-06-2012
Publisher
Springer-Verlag
Published in
International Orthopaedics / Issue 6/2012
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-011-1412-6

Other articles of this Issue 6/2012

International Orthopaedics 6/2012 Go to the issue