In vitro investigation of the influence of tibial slope on quadriceps extension force after total knee arthroplasty

Authors: Sven Ostermeier, Christof Hurschler, Henning Windhagen, Christina Stukenborg-Colsman

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 10/2006

Login to get access

Abstract

The purpose of this study was to investigate the influence of tibial base plate angulation on knee kinematics and kinetics during knee arthroplasty. The amount of quadriceps force required to extend the knee and the anteroposterior displacement of a mobile bearing insert as well as tibiofemoral position were measured during an in vitro simulation of an isokinetic knee extension cycle. Human knee specimens (= 7, mean age 62, range 52–75 years, all male) were tested in a kinematic knee simulating machine after total knee arthroplasty (TKA) with a mobile bearing insert prosthesis (Interax®, Stryker/Howmedica). During simulation, a hydraulic cylinder applied sufficient force to the quadriceps tendon to produce an extension moment of 31 N m about the knee. The quadriceps load was measured using a load cell attached to the quadriceps tendon, the anteroposterior displacement of the mobile bearing insert as well as the relative tibiofemoral position was measured using an ultrasound base motion analysis system (CMS 100®, Zebris). Quadriceps load, insert and tibial displacement were first investigated with the tibial base plate implanted with a neutral tibial base plate orientation, and subsequently after 10° posterior angulation. The quadriceps forces needed to produce a 31 N m knee extension moment after TKA with neutral slope reached levels as high as 1,391 N (SD 82 N). After applying a posterior slope of 10°, maximum quadriceps force was measured to be up to 1,303 N (SD 34 N, P = 0.04). The mobile bearing insert was observed to move up to 0.1 mm (SD 4.2 mm) anteriorly relative to the tibial base plate with neutral tibial slope, and up to 1.0 mm (SD 4.5 mm, P = 0.47) with tibial slope. Femoral position relative to the tibia moved from a posterior position of 13.1 mm (SD 4.0 mm) anteriorly up to 0.5 mm (SD 6.3 mm), and from 16.0 mm (SD 6.4 mm, = 0.67) to 9.5 mm (SD 9.9 mm, P = 0.33) with a 10° tibial slope. Posterior slope of the tibial base plate resulted in a more physiologic insert movement with a more posterior position of the femur and reduced quadriceps force especially in knee flexion angles above 60° compared to TKA with a neutral slope of the tibial base plate. Thus, the data suggest that the quadriceps lever arm was improved, which might have positive effect in mobilization of patients after TKA.
Literature
1.
go back to reference Andriacchi TP (1988) Biomechanics and gait analysis in total knee arthroplasty. Orthop Rev 17:470–473PubMed Andriacchi TP (1988) Biomechanics and gait analysis in total knee arthroplasty. Orthop Rev 17:470–473PubMed
2.
go back to reference Andriacchi TP, Alexander EJ, Godman SB (1999) Understanding the role of functional adaptions in patients with total knee replacements. 1, 50–53. International Conference on Knee Replacement 1974–2024, IMechE Headquarters. Westminster, London, UK Andriacchi TP, Alexander EJ, Godman SB (1999) Understanding the role of functional adaptions in patients with total knee replacements. 1, 50–53. International Conference on Knee Replacement 1974–2024, IMechE Headquarters. Westminster, London, UK
3.
go back to reference Andriacchi TP, Galante JO, Fermier RW (1982) The influence of total knee-replacement design on walking and stair-climbing. J Bone Joint Surg [Am] 64:1328–1335 Andriacchi TP, Galante JO, Fermier RW (1982) The influence of total knee-replacement design on walking and stair-climbing. J Bone Joint Surg [Am] 64:1328–1335
4.
go back to reference Berman AT, Bosacco SJ, Israelite C (1991) Evaluation of total knee arthroplasty using isokinetic testing. Clin Orthop 271:106–113PubMed Berman AT, Bosacco SJ, Israelite C (1991) Evaluation of total knee arthroplasty using isokinetic testing. Clin Orthop 271:106–113PubMed
5.
go back to reference Dennis DA (1998) Proposed advantages of posterior cruciate ligament retention orthopaedics today 1[September/October], 10. New Jersey, Slack Incorporated Dennis DA (1998) Proposed advantages of posterior cruciate ligament retention orthopaedics today 1[September/October], 10. New Jersey, Slack Incorporated
6.
go back to reference Dorr L, Scott R, Ranawat CS (1985) Controversies of total knee arthroplasty: importance of retention of the posterior cruciate ligament. In: Ranawat C (ed) Total knee arthroplasty: techniques, results and complications. Springer, Berlin, Heidelberg New York pp 197–202 Dorr L, Scott R, Ranawat CS (1985) Controversies of total knee arthroplasty: importance of retention of the posterior cruciate ligament. In: Ranawat C (ed) Total knee arthroplasty: techniques, results and complications. Springer, Berlin, Heidelberg New York pp 197–202
7.
go back to reference Dorr LD (1988) Functional comparison of posterior of posterior cruciate retaining versus cruciate sacrificed total knee arthroplasty. Clin Orthop 236:36–41PubMed Dorr LD (1988) Functional comparison of posterior of posterior cruciate retaining versus cruciate sacrificed total knee arthroplasty. Clin Orthop 236:36–41PubMed
8.
go back to reference Draganich LF (1984) The influence of the cruciate ligaments, knee musculature and anatomy on knee joint loading. University of Illinois at Chicago Draganich LF (1984) The influence of the cruciate ligaments, knee musculature and anatomy on knee joint loading. University of Illinois at Chicago
9.
go back to reference Garg A, Walker PS (1990) Prediction of total knee motion using a three-dimensional computer-graphics model. J Biomech 23:45–58PubMedCrossRef Garg A, Walker PS (1990) Prediction of total knee motion using a three-dimensional computer-graphics model. J Biomech 23:45–58PubMedCrossRef
10.
go back to reference Giffin JR, Vogrin TM, Zantop T, Woo SL, Harner CD (2004) Effects of increasing tibial slope on the biomechanics of the knee. Am J Sports Med 32:376–382PubMedCrossRef Giffin JR, Vogrin TM, Zantop T, Woo SL, Harner CD (2004) Effects of increasing tibial slope on the biomechanics of the knee. Am J Sports Med 32:376–382PubMedCrossRef
11.
go back to reference Godest AC, de Cloke CS, Taylor M, Gregson PJ, Keane AJ, Sathasivan S, Walker PS (2000) A computational model for the prediction of total knee replacement kinematics in the sagittal plane. J Biomech 33:435–442PubMedCrossRef Godest AC, de Cloke CS, Taylor M, Gregson PJ, Keane AJ, Sathasivan S, Walker PS (2000) A computational model for the prediction of total knee replacement kinematics in the sagittal plane. J Biomech 33:435–442PubMedCrossRef
12.
go back to reference Hernigou P, Deschamps G (2004) Posterior slope of the tibial implant and the outcome of unicompartmental knee arthroplasty. J Bone Joint Surg Am 86-A:506–511PubMed Hernigou P, Deschamps G (2004) Posterior slope of the tibial implant and the outcome of unicompartmental knee arthroplasty. J Bone Joint Surg Am 86-A:506–511PubMed
13.
go back to reference Komistek RD, Dennis DA, Haas B (1999) Motion in LCS Rotating Platform. 1, 246–249. London, UK, Imech E. International conference on knee replacement 1974–2024. 22-4-0099 Komistek RD, Dennis DA, Haas B (1999) Motion in LCS Rotating Platform. 1, 246–249. London, UK, Imech E. International conference on knee replacement 1974–2024. 22-4-0099
14.
go back to reference Komistek RD, Dennis DA, Walker SA, Anderson DT (1998) A multicenter analysis of A/P translation in TKA. 11th conference of the ESB, Toulouse Komistek RD, Dennis DA, Walker SA, Anderson DT (1998) A multicenter analysis of A/P translation in TKA. 11th conference of the ESB, Toulouse
15.
go back to reference Komistek RD, Walker SA, Crossett LS, Running D, Dennis DA (1998) In vivo kinematics of PS implanted knees during gait. 11th Conference of the ESB, Toulouse Komistek RD, Walker SA, Crossett LS, Running D, Dennis DA (1998) In vivo kinematics of PS implanted knees during gait. 11th Conference of the ESB, Toulouse
16.
go back to reference Lewandowski PJ, Askew MJ, Lin DF, Hurst FW, Melby A (1997) Kinematics of posterior cruciate ligament-retaining and -sacrificing mobile bearing total knee arthroplasties. J Arthroplasty 12:777–784PubMedCrossRef Lewandowski PJ, Askew MJ, Lin DF, Hurst FW, Melby A (1997) Kinematics of posterior cruciate ligament-retaining and -sacrificing mobile bearing total knee arthroplasties. J Arthroplasty 12:777–784PubMedCrossRef
17.
go back to reference Nisell R, Ekholm J (1985) Mechanics of the knee. Acta Orthop Scand Suppl 216:1–42PubMed Nisell R, Ekholm J (1985) Mechanics of the knee. Acta Orthop Scand Suppl 216:1–42PubMed
18.
go back to reference Ostermeier S, Hurschler C, Stukenborg-Colsman C (2004) Quadriceps function after TKA—an in vitro study in a knee kinematic simulator. Clin Biomech (Bristol, Avon) 19:270–276CrossRef Ostermeier S, Hurschler C, Stukenborg-Colsman C (2004) Quadriceps function after TKA—an in vitro study in a knee kinematic simulator. Clin Biomech (Bristol, Avon) 19:270–276CrossRef
19.
go back to reference Rand JA (1994) Current concepts review: the patellofemoral joint in total knee arhtroplasty. JBJS 76-A:612–620 Rand JA (1994) Current concepts review: the patellofemoral joint in total knee arhtroplasty. JBJS 76-A:612–620
20.
go back to reference Stukenborg-Colsman C, Ostermeier S, Hurschler C, Wirth CJ (2002) Tibiofemoral contact stress after total knee arthroplasty comparison of fixed and mobile-bearing insert designs. Acta Orthop Scand 73:638–646PubMedCrossRef Stukenborg-Colsman C, Ostermeier S, Hurschler C, Wirth CJ (2002) Tibiofemoral contact stress after total knee arthroplasty comparison of fixed and mobile-bearing insert designs. Acta Orthop Scand 73:638–646PubMedCrossRef
21.
go back to reference Stukenborg-Colsman C, Ostermeier S, Wenger K, Wirth CJ (2002) Relative motion of a mobile bearing insert after total knee arthroplasty—a dynamic in vitro study. Clin Biomech 17:49–55CrossRef Stukenborg-Colsman C, Ostermeier S, Wenger K, Wirth CJ (2002) Relative motion of a mobile bearing insert after total knee arthroplasty—a dynamic in vitro study. Clin Biomech 17:49–55CrossRef
22.
go back to reference Walker PS, Sathasivam S (1999) The design of guide surfaces for fixed-bearing and mobile-bearing knee replacements. J Biomech 32:27–34PubMedCrossRef Walker PS, Sathasivam S (1999) The design of guide surfaces for fixed-bearing and mobile-bearing knee replacements. J Biomech 32:27–34PubMedCrossRef
23.
go back to reference Wimmer MA (1999) Wear of the polyethylene component created by rolling motion of the artificial knee joint. Shaker, Aachen Wimmer MA (1999) Wear of the polyethylene component created by rolling motion of the artificial knee joint. Shaker, Aachen
24.
go back to reference Wimmer MA, Andriacchi TP (1997) Tractive forces during rolling motion of the knee: implications for wear in total knee replacement. J Biomech 30:131–137PubMedCrossRef Wimmer MA, Andriacchi TP (1997) Tractive forces during rolling motion of the knee: implications for wear in total knee replacement. J Biomech 30:131–137PubMedCrossRef
Metadata
Title
In vitro investigation of the influence of tibial slope on quadriceps extension force after total knee arthroplasty
Authors
Sven Ostermeier
Christof Hurschler
Henning Windhagen
Christina Stukenborg-Colsman
Publication date
01-10-2006
Publisher
Springer-Verlag
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 10/2006
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-006-0078-x

Other articles of this Issue 10/2006

Knee Surgery, Sports Traumatology, Arthroscopy 10/2006 Go to the issue