Published in: Intensive Care Medicine 4/2007

01-04-2007 | Original

Automatic adjustment of noninvasive pressure support with a bilevel home ventilator in patients with acute respiratory failure: a feasibility study

Authors: Anne Battisti, Didier Tassaux, David Bassin, Philippe Jolliet

Published in: Intensive Care Medicine | Issue 4/2007

Login to get access

Abstract

Objective

To test the feasibility of applying noninvasive ventilation (NIV) using a prototype algorithm implemented in a bilevel ventilation device designed to adjust pressure support (PS) to maintain a clinician-set alveolar ventilation in patients with acute respiratory failure after initial stabilization.

Design and setting

Prospective crossover interventional study in an intensive care unit, university hospital.

Patients

19 patients receiving NIV for acute hypercapnic respiratory failure (13 men, 6 women; mean age 70 ± 11 years).

Methods

The same bilevel ventilator was used with manually adjusted PS and with the automated algorithm (autoPS), set to maintain the same alveolar ventilation as in PS. Sequence (measurements at end of each period): (a) prior to initiating NIV (baseline 1); (b) 45 min with manually set PS; (c) 60 min without NIV; (d) 45 min with autoPS; (e) 60 min without NIV; (f) 45 min with manually set PS.

Results

The magnitude of decrease in PaCO2 and increase in pH with autoPS was comparable to that of conventional PS, with the same alveolar ventilation and level of PS. No technical problem occurred in autoPS mode, and no NIV trial had to be discontinued because of patient discomfort.

Conclusions

These results suggest that the alveolar ventilation based automatic control of PS during NIV with a bilevel device is feasible and leads to beneficial effects in patients with acute respiratory failure comparable to those of manually set PS. Further studies should now explore the potential of this system over longer periods in patients with acute and chronic respiratory failure.
Literature
1.
go back to reference Antonelli M, Conti G et al (2001) Predictors of failure of noninvasive positive pressure ventilation in patients with acute hypoxemic respiratory failure: a multi-center study. Intensive Care Med 27:1718–1728PubMedCrossRef Antonelli M, Conti G et al (2001) Predictors of failure of noninvasive positive pressure ventilation in patients with acute hypoxemic respiratory failure: a multi-center study. Intensive Care Med 27:1718–1728PubMedCrossRef
2.
go back to reference Carlucci A, Richard J, Wysocki M, Lepage E, Brochard L (2001) Noninvasive versus conventional mechanical ventilation. An epidemiologic survey. Am J Respir Crit Care Med 163:874–880PubMed Carlucci A, Richard J, Wysocki M, Lepage E, Brochard L (2001) Noninvasive versus conventional mechanical ventilation. An epidemiologic survey. Am J Respir Crit Care Med 163:874–880PubMed
3.
go back to reference Tobin MJ, Jubran A, Laghi F (2001) Patient-ventilator interaction. Am J Respir Crit Care Med 163:1059–1063PubMed Tobin MJ, Jubran A, Laghi F (2001) Patient-ventilator interaction. Am J Respir Crit Care Med 163:1059–1063PubMed
4.
go back to reference Calderini E, Confalonieri M, Puccio P, Francavilla N, Stella L, Gregoretti C (1999) Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger. Intensive Care Med 25:662–667PubMedCrossRef Calderini E, Confalonieri M, Puccio P, Francavilla N, Stella L, Gregoretti C (1999) Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger. Intensive Care Med 25:662–667PubMedCrossRef
5.
go back to reference Prinianakis G, Delmastro M, Carlucci A, Ceriana P, Nava S (2004) Effect of varying the pressurisation rate during noninvasive pressure support ventilation. Eur Respir J 23:314–320PubMedCrossRef Prinianakis G, Delmastro M, Carlucci A, Ceriana P, Nava S (2004) Effect of varying the pressurisation rate during noninvasive pressure support ventilation. Eur Respir J 23:314–320PubMedCrossRef
6.
go back to reference Kondili E, Prinianakis G, Georgopoulos D (2003) Patient-ventilator interaction. Br J Anaesth 91:106–119PubMedCrossRef Kondili E, Prinianakis G, Georgopoulos D (2003) Patient-ventilator interaction. Br J Anaesth 91:106–119PubMedCrossRef
7.
go back to reference Vincent JL (2005) Give your patient a fast hug (at least) once a day. Crit Care Med 33:1225–1229PubMedCrossRef Vincent JL (2005) Give your patient a fast hug (at least) once a day. Crit Care Med 33:1225–1229PubMedCrossRef
8.
go back to reference Dojat M, Harf A, Touchard D, Lemaire F, Brochard L (2000) Clinical evaluation of a computer-controlled pressure support mode. Am J Respir Crit Care Med 161:1161–1166PubMed Dojat M, Harf A, Touchard D, Lemaire F, Brochard L (2000) Clinical evaluation of a computer-controlled pressure support mode. Am J Respir Crit Care Med 161:1161–1166PubMed
9.
go back to reference Plant P, Owen J, Elliott M (2000) Early use of non-invasive ventilation for acute exacerbations of chronic obstructive pulmonary disease on general respiratory wards: a multicentre randomised controlled trial. Lancet 355:1931–1935PubMedCrossRef Plant P, Owen J, Elliott M (2000) Early use of non-invasive ventilation for acute exacerbations of chronic obstructive pulmonary disease on general respiratory wards: a multicentre randomised controlled trial. Lancet 355:1931–1935PubMedCrossRef
10.
go back to reference Poponick JM, Renston JP, Bennett RP, Emerman CL (1999) Use of a ventilatory support system (BiPAP) for acute respiratory failure in the emergency department. Chest 116:166–171PubMedCrossRef Poponick JM, Renston JP, Bennett RP, Emerman CL (1999) Use of a ventilatory support system (BiPAP) for acute respiratory failure in the emergency department. Chest 116:166–171PubMedCrossRef
11.
go back to reference Cotes J (1979) Distribution of ventilation and perfusion. In: Cotes J (ed) Lung function. Assessment and application in medicine. Blackwell, Oxford, pp 131–159 Cotes J (1979) Distribution of ventilation and perfusion. In: Cotes J (ed) Lung function. Assessment and application in medicine. Blackwell, Oxford, pp 131–159
12.
go back to reference Brochard L, Mancebo J et al (1995) Noninvasive ventilation for acute exacerbations of chronic obstructive pulmonary disease. N Engl J Med 333:817–822PubMedCrossRef Brochard L, Mancebo J et al (1995) Noninvasive ventilation for acute exacerbations of chronic obstructive pulmonary disease. N Engl J Med 333:817–822PubMedCrossRef
13.
go back to reference Brochard L, Isabey D et al (1990) Reversal of acute exacerbations of chronic obstructive lung disease by inspiratory assistance with a face mask. N Engl J Med 323:1523–1530PubMedCrossRef Brochard L, Isabey D et al (1990) Reversal of acute exacerbations of chronic obstructive lung disease by inspiratory assistance with a face mask. N Engl J Med 323:1523–1530PubMedCrossRef
14.
go back to reference Hilbert G, Gruson D, Gbiki-Benissan G, Cardinaud J (1997) Sequential use of noninvasive pressure support ventilation for acute exacerbations of COPD. Intensive Care Med 23:955–961PubMedCrossRef Hilbert G, Gruson D, Gbiki-Benissan G, Cardinaud J (1997) Sequential use of noninvasive pressure support ventilation for acute exacerbations of COPD. Intensive Care Med 23:955–961PubMedCrossRef
15.
go back to reference Nava S, Gregoretti C et al (2005) Noninvasive ventilation to prevent respiratory failure after extubation in high-risk patients. Crit Care Med 33:2465–2470PubMedCrossRef Nava S, Gregoretti C et al (2005) Noninvasive ventilation to prevent respiratory failure after extubation in high-risk patients. Crit Care Med 33:2465–2470PubMedCrossRef
16.
go back to reference Battisti A, Tassaux D, Janssens J, Michotte J, Jaber S, Jolliet P (2005) Performance characteristics of ten recent bilevel ventilators: a comparative bench study. Chest 127:1784–1792PubMedCrossRef Battisti A, Tassaux D, Janssens J, Michotte J, Jaber S, Jolliet P (2005) Performance characteristics of ten recent bilevel ventilators: a comparative bench study. Chest 127:1784–1792PubMedCrossRef
17.
go back to reference Richard JC, Carlucci A et al (2002) Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 28:1049–1057PubMedCrossRef Richard JC, Carlucci A et al (2002) Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 28:1049–1057PubMedCrossRef
18.
go back to reference Tassaux D, Strasser S, Fonseca S, Dalmas E, Jolliet P (2002) Comparative bench study of triggering, pressurization and cycling between the home ventilator VPAPII and three ICU ventilators. Intensive Care Med 28:1254–1261PubMedCrossRef Tassaux D, Strasser S, Fonseca S, Dalmas E, Jolliet P (2002) Comparative bench study of triggering, pressurization and cycling between the home ventilator VPAPII and three ICU ventilators. Intensive Care Med 28:1254–1261PubMedCrossRef
19.
go back to reference Thys F, Roeseler J, Reynaert M, Liistro G, Rodenstein DO (2002) Noninvasive ventilation for acute respiratory failure: a prospective randomised placebo-controlled trial. Eur Respir J 20:545–255PubMedCrossRef Thys F, Roeseler J, Reynaert M, Liistro G, Rodenstein DO (2002) Noninvasive ventilation for acute respiratory failure: a prospective randomised placebo-controlled trial. Eur Respir J 20:545–255PubMedCrossRef
20.
go back to reference British Thoracic Society Standards of Care Committee (2002) Non-invasive ventilation in acute respiratory failure. Thorax 57:192–211CrossRef British Thoracic Society Standards of Care Committee (2002) Non-invasive ventilation in acute respiratory failure. Thorax 57:192–211CrossRef
21.
go back to reference Liesching T, Kwok H, Hill N (2003) Acute applications of noninvasive positive pressure ventilation. Chest 124:699–713PubMedCrossRef Liesching T, Kwok H, Hill N (2003) Acute applications of noninvasive positive pressure ventilation. Chest 124:699–713PubMedCrossRef
22.
24.
go back to reference Leung P, Jubran A, Tobin M (1997) Comparison of assisted ventilator modes on triggering, patient effort, and dyspnea. Am J Respir Crit Care Med 155:1940–1948PubMed Leung P, Jubran A, Tobin M (1997) Comparison of assisted ventilator modes on triggering, patient effort, and dyspnea. Am J Respir Crit Care Med 155:1940–1948PubMed
25.
go back to reference Schettino GP, Tucci MR, Sousa R, Valente Barbas CS, Passos Amato MB, Carvalho CR (2001) Mask mechanics and leak dynamics during noninvasive pressure support ventilation: a bench study. Intensive Care Med 27:1887–1891PubMedCrossRef Schettino GP, Tucci MR, Sousa R, Valente Barbas CS, Passos Amato MB, Carvalho CR (2001) Mask mechanics and leak dynamics during noninvasive pressure support ventilation: a bench study. Intensive Care Med 27:1887–1891PubMedCrossRef
26.
go back to reference Tassaux D, Gainnier M, Battisti A, Jolliet P (2005) Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med 172:1283–1289PubMedCrossRef Tassaux D, Gainnier M, Battisti A, Jolliet P (2005) Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med 172:1283–1289PubMedCrossRef
27.
go back to reference Brochard L, Harf A, Lorino H, Lemaire F (1989) Inspiratory pressure support prevents diaphragmatic fatigue during weaning from mechanical ventilation. Am Rev Respir Dis 139:513–521PubMed Brochard L, Harf A, Lorino H, Lemaire F (1989) Inspiratory pressure support prevents diaphragmatic fatigue during weaning from mechanical ventilation. Am Rev Respir Dis 139:513–521PubMed
Metadata
Title
Automatic adjustment of noninvasive pressure support with a bilevel home ventilator in patients with acute respiratory failure: a feasibility study
Authors
Anne Battisti
Didier Tassaux
David Bassin
Philippe Jolliet
Publication date
01-04-2007
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 4/2007
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-007-0550-1

Other articles of this Issue 4/2007

Intensive Care Medicine 4/2007 Go to the issue