Published in: Journal of Orthopaedic Surgery and Research 1/2020

01-12-2020 | Laser | Research article

Study of the tribological properties of surface structures using ultrashort laser pulses to reduce wear in endoprosthetics

Authors: Lea Theresa Backes, Paul Oldorf, Rigo Peters, Robert Wendlandt, Georg Schnell, Arndt-Peter Schulz

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2020

Login to get access

Abstract

Background

Loosening of prostheses and functional disorders represent a far-reaching problem in the clinic, and the long-term outcomes are essentially determined by wear. Despite all advances, up to 10% of prostheses still fail after 10 years. In particular, more active patients show increased revision rates.

Methods

The objective of this thesis is to examine whether the applied microstructures of the articulating surfaces can lead to a reduction in abrasion. Three different structural geometries (dimples, offset lines, grid lines) were defined. In an experimental test setup according to DIN ISO 6474 (Deutsches Institut für Normung, International Organization for Standardization), a tribological test of metal and ceramic pairings was performed using two-dimensional ring-on-disc (RoD) tests.

Results

In both material groups, the structuring had a positive effect on the wear behaviour. In the ceramic group, an abrasion reduction of 22.6% was achieved. However, it is important to take into account the limited informative value due to the hardness of the material. Two of the three Cobalt-Chrome-Molybdenum (CoCrMo) structure geometries (grids, offset lines) also showed a significant reduction in abrasion compared to the reference group, with a maximum wear reduction of 55.5%.

Conclusion

By reducing abrasion, surface structuring could be used to extend the life of prostheses and minimise the number of revisions.
Literature
1.
go back to reference Statistisches Bundesamt (2018). Gesundheit - Fallpauschalenbezogene Krankenhausstatistik (DRG-Statistik) Operationen und Prozeduren der vollstationären Patientinnen und Patienten in Krankenhäusern (4-Steller); 2017. p. 22. Statistisches Bundesamt (2018). Gesundheit - Fallpauschalenbezogene Krankenhausstatistik (DRG-Statistik) Operationen und Prozeduren der vollstationären Patientinnen und Patienten in Krankenhäusern (4-Steller); 2017. p. 22.
2.
go back to reference Pilz V, Hanstein T, Skripitz R. Projections of primary hip arthroplasty in Germany until 2040. Acta Orthop. 2018;89:308–13.CrossRef Pilz V, Hanstein T, Skripitz R. Projections of primary hip arthroplasty in Germany until 2040. Acta Orthop. 2018;89:308–13.CrossRef
3.
go back to reference Kircher J, Bergschmidt P, Bader R, Kluess D, Besser-Mahuzir E, Leder A, et al. The importance of wear couples for younger endoprosthesis. Orthopade. 2007;36:337–46.CrossRef Kircher J, Bergschmidt P, Bader R, Kluess D, Besser-Mahuzir E, Leder A, et al. The importance of wear couples for younger endoprosthesis. Orthopade. 2007;36:337–46.CrossRef
4.
go back to reference MacInnes SJ, Gordon A, Wilkinson JM. Risk Factors for aseptic loosening following total hip arthroplasty. In: Rijeka FS, editor. Recent advances in arthroplasty. 1st ed: IntechOpen; 2012. p. 275–95. MacInnes SJ, Gordon A, Wilkinson JM. Risk Factors for aseptic loosening following total hip arthroplasty. In: Rijeka FS, editor. Recent advances in arthroplasty. 1st ed: IntechOpen; 2012. p. 275–95.
5.
go back to reference Grimberg A, Jansson V, Liebs T, Melsheimer O, Steinbrück A. Jahresbericht 2017 – Mit Sicherheit mehr Qualität: Endoprothesenregister Deutschland (EPRD); 2018. p. 12–27. Grimberg A, Jansson V, Liebs T, Melsheimer O, Steinbrück A. Jahresbericht 2017 – Mit Sicherheit mehr Qualität: Endoprothesenregister Deutschland (EPRD); 2018. p. 12–27.
6.
go back to reference Lochner K, Fritsche A, Jonitz A, Hansmann D, Mueller P, Mueller-Hilke B, et al. The potential role of human osteoblasts for periprosthetic osteolysis following exposure to wear particles. Int J Mol Med. 2011;38:1055–63. Lochner K, Fritsche A, Jonitz A, Hansmann D, Mueller P, Mueller-Hilke B, et al. The potential role of human osteoblasts for periprosthetic osteolysis following exposure to wear particles. Int J Mol Med. 2011;38:1055–63.
8.
go back to reference Tarabolsi M, Klassen T, Mantwill F, Gärtner F, Siegel F, Schulz AP. Patterned CoCrMo and Al2O3 surfaces for reduced free wear debris in artificial joint arthroplasty. J Biomed Mater Res A. 2013;101(12):2337–3456.CrossRef Tarabolsi M, Klassen T, Mantwill F, Gärtner F, Siegel F, Schulz AP. Patterned CoCrMo and Al2O3 surfaces for reduced free wear debris in artificial joint arthroplasty. J Biomed Mater Res A. 2013;101(12):2337–3456.CrossRef
9.
go back to reference Implants for surgery – ceramic materials based on high purity alumina (ISO 6474: 1994). In: International Organisation for Standardization; 1994. p. 1–6. Implants for surgery – ceramic materials based on high purity alumina (ISO 6474: 1994). In: International Organisation for Standardization; 1994. p. 1–6.
10.
go back to reference Kärrholm J, Mohaddes M, Odin D, Vinblad J, Rogmark C, Rolfson O. Annual Report 2017. The Swedish Hip Arthroplasty Register; 2018. p. 62–76. Kärrholm J, Mohaddes M, Odin D, Vinblad J, Rogmark C, Rolfson O. Annual Report 2017. The Swedish Hip Arthroplasty Register; 2018. p. 62–76.
11.
go back to reference Chyr A, Qiu M, Speltz JW, Jacobsen RL, Sanders AP, Raeymaekers B. A patterned microtexture to reduce friction and increase longevity of prosthetic hip joints. Wear. 2014;315(1-2):51–7.CrossRef Chyr A, Qiu M, Speltz JW, Jacobsen RL, Sanders AP, Raeymaekers B. A patterned microtexture to reduce friction and increase longevity of prosthetic hip joints. Wear. 2014;315(1-2):51–7.CrossRef
12.
go back to reference Borjali A, Langhorn J, Monson K, Raeymaekers B. Using a patterned microtexture to reduce polyethylene wear in metal-on-polyethylene prosthetic bearing couples. Wear. 2017;392-393:77–83.CrossRef Borjali A, Langhorn J, Monson K, Raeymaekers B. Using a patterned microtexture to reduce polyethylene wear in metal-on-polyethylene prosthetic bearing couples. Wear. 2017;392-393:77–83.CrossRef
13.
go back to reference Myant C, Underwood R, Fan J, Cann PM. Lubrication of metal-on-metal hip joints: the effect of protein content and load on film formation and wear. J Mech Behav Biomed Mater. 2012;6:30–40.CrossRef Myant C, Underwood R, Fan J, Cann PM. Lubrication of metal-on-metal hip joints: the effect of protein content and load on film formation and wear. J Mech Behav Biomed Mater. 2012;6:30–40.CrossRef
14.
go back to reference Myant C, Cann P. On the matter of synovial fluid lubrication: implications for metal-on-metal hip tribology. J Mech Behav Biomed Mater. 2014;34:338–48.CrossRef Myant C, Cann P. On the matter of synovial fluid lubrication: implications for metal-on-metal hip tribology. J Mech Behav Biomed Mater. 2014;34:338–48.CrossRef
16.
go back to reference Schleicher I, Kilian O, Schnettler R. Die primäre Endoprothetik des Hüftgelenkes. In: Krukemeyer MG, Möllenhoff G, editors. Endoprothetik - Ein Leitfaden für den Praktiker. 3rd ed. Berlin; Boston: Walter de Gruyter GmbH; 2012. p. 99–127. Schleicher I, Kilian O, Schnettler R. Die primäre Endoprothetik des Hüftgelenkes. In: Krukemeyer MG, Möllenhoff G, editors. Endoprothetik - Ein Leitfaden für den Praktiker. 3rd ed. Berlin; Boston: Walter de Gruyter GmbH; 2012. p. 99–127.
17.
go back to reference Qiu ZY, Chen C, Wang XM, Lee IS. Advances in the surface modification techniques of bone-related implants for last 10 years. Regen Biomater. 2014;1(1):67–79.CrossRef Qiu ZY, Chen C, Wang XM, Lee IS. Advances in the surface modification techniques of bone-related implants for last 10 years. Regen Biomater. 2014;1(1):67–79.CrossRef
18.
go back to reference Graziani G, Boi M, Bianchi M. A review on ionic substitutions in hydroxyapatite thin films: towards complete biomimetism. Coatings. 2018;8(8):269.CrossRef Graziani G, Boi M, Bianchi M. A review on ionic substitutions in hydroxyapatite thin films: towards complete biomimetism. Coatings. 2018;8(8):269.CrossRef
21.
go back to reference Tsikandylakis G, Mohaddes M, Cnudde P, Eskelinen A, Kärrholm J, Rolfson O. Head size in primary total hip arthroplasty. EFORT Open Rev. 2018;3(5):225–31.CrossRef Tsikandylakis G, Mohaddes M, Cnudde P, Eskelinen A, Kärrholm J, Rolfson O. Head size in primary total hip arthroplasty. EFORT Open Rev. 2018;3(5):225–31.CrossRef
Metadata
Title
Study of the tribological properties of surface structures using ultrashort laser pulses to reduce wear in endoprosthetics
Authors
Lea Theresa Backes
Paul Oldorf
Rigo Peters
Robert Wendlandt
Georg Schnell
Arndt-Peter Schulz
Publication date
01-12-2020
Publisher
BioMed Central
Keyword
Laser
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2020
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-020-01719-1

Other articles of this Issue 1/2020

Journal of Orthopaedic Surgery and Research 1/2020 Go to the issue