IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry

Authors: Yang Zhang, Lin Zheng, Ailing Bian, Qi Zhou

Published in: International Ophthalmology | Issue 2/2018

Login to get access

Abstract

Purpose

To compare the postoperative intraocular pressure (IOP) of eyes following pars plana vitrectomy (PPV) combined with intravitreal silicone oil (SO) tamponade by Corneal Visualization Scheimpflug Technology (CST), Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT).

Methods

Thirty-eight participants who had undergone PPV combined with SO tamponade to treat vitreoretinal diseases were enrolled. Postoperative IOP measurements were obtained using CST, NCT and GAT. Inter-device agreement was assessed by Bland–Altman analysis. The correlation coefficient was used to describe the potential postoperative factors affecting the postoperative IOP differences between each device.

Results

Bland–Altman analysis revealed the bias between CST and GAT, between CST and NCT, and between GAT and NCT to be −0.2, 2.1 and 2.4 mmHg, respectively. CST and GAT correlated well with each other. NCT values were lower than those of GAT and CST (all p < 0.05), whereas CST values did not differ from the GAT readings. Central corneal thickness, corneal biomechanical properties and age showed significant correlation with the differences of CST–NCT and GAT–NCT.

Conclusions

In SO tamponade eyes, NCT obtains lower IOP than other tonometry techniques, and CST is highly consistent with GAT. CST offers an optional non-contact method for measuring postoperative IOP in SO tamponade eyes.
Literature
1.
go back to reference Teke MY, Elgin U, Sen E et al (2014) Intravitreal silicone oil induced changes in corneal biomechanics. Int Ophthalmol 34:457–463CrossRefPubMed Teke MY, Elgin U, Sen E et al (2014) Intravitreal silicone oil induced changes in corneal biomechanics. Int Ophthalmol 34:457–463CrossRefPubMed
2.
go back to reference Reznicek L, Muth D, Kampik A et al (2013) Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma. Br J Ophthalmol 97:1410–1414CrossRefPubMed Reznicek L, Muth D, Kampik A et al (2013) Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma. Br J Ophthalmol 97:1410–1414CrossRefPubMed
3.
go back to reference Nemeth G, Hassan Z, Csutak A et al (2013) Repeatability of ocular biomechanical data measurements with a Scheimpflug-based noncontact device on normal corneas. J Refract Surg 29:558–563CrossRefPubMed Nemeth G, Hassan Z, Csutak A et al (2013) Repeatability of ocular biomechanical data measurements with a Scheimpflug-based noncontact device on normal corneas. J Refract Surg 29:558–563CrossRefPubMed
4.
go back to reference Eller AW, Gardner TW, D’Antonio JA (1992) Survey of intraocular silicone oil use in the United States. Ophthalmology 99:1174–1176CrossRefPubMed Eller AW, Gardner TW, D’Antonio JA (1992) Survey of intraocular silicone oil use in the United States. Ophthalmology 99:1174–1176CrossRefPubMed
5.
go back to reference Narayanan R, Tibra N, Mathai A et al (2012) Sutureless 23-gauge versus 20-gauge vitrectomy with silicone oil injection in rhegmatogenous retinal detachment. Retina 32:1013–1016CrossRefPubMed Narayanan R, Tibra N, Mathai A et al (2012) Sutureless 23-gauge versus 20-gauge vitrectomy with silicone oil injection in rhegmatogenous retinal detachment. Retina 32:1013–1016CrossRefPubMed
6.
go back to reference Ichhpujani P, Jindal A, Jay Katz L (2009) Silicone oil induced glaucoma: a review. Graefes Arch Clin Exp Ophthalmol 247:1585–1593CrossRefPubMed Ichhpujani P, Jindal A, Jay Katz L (2009) Silicone oil induced glaucoma: a review. Graefes Arch Clin Exp Ophthalmol 247:1585–1593CrossRefPubMed
7.
go back to reference Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 44:367–408CrossRefPubMed Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 44:367–408CrossRefPubMed
8.
go back to reference Gunvant P, O’Leary DJ, Baskaran M et al (2005) Evaluation of tonometric correction factors. J Glaucoma 14:337–343CrossRefPubMed Gunvant P, O’Leary DJ, Baskaran M et al (2005) Evaluation of tonometric correction factors. J Glaucoma 14:337–343CrossRefPubMed
9.
go back to reference Schroeder B, Hager A, Kutschan A et al (2008) Measurement of viscoelastic corneal parameters (corneal hysteresis) in patients with primary open angle glaucoma. Ophthalmologe 105:916–920CrossRefPubMed Schroeder B, Hager A, Kutschan A et al (2008) Measurement of viscoelastic corneal parameters (corneal hysteresis) in patients with primary open angle glaucoma. Ophthalmologe 105:916–920CrossRefPubMed
10.
go back to reference Kaushik S, Pandav SS, Banger A et al (2012) Relationship between corneal biomechanical properties, central corneal thickness, and intraocular pressure across the spectrum of glaucoma. Am J Ophthalmol 153(840–849):e2 Kaushik S, Pandav SS, Banger A et al (2012) Relationship between corneal biomechanical properties, central corneal thickness, and intraocular pressure across the spectrum of glaucoma. Am J Ophthalmol 153(840–849):e2
11.
go back to reference Hong J, Yu Z, Jiang C et al (2015) Corvis ST Tonometer for measuring postoperative IOP in LASIK patients. Optom Vis Sci 92:589–595CrossRefPubMed Hong J, Yu Z, Jiang C et al (2015) Corvis ST Tonometer for measuring postoperative IOP in LASIK patients. Optom Vis Sci 92:589–595CrossRefPubMed
12.
go back to reference Le Q, Wang X, Lv J et al (2012) In vivo laser scanning confocal microscopy of the cornea in patients with silicone oil tamponade after vitreoretinal surgery. Cornea 31:876–882CrossRefPubMed Le Q, Wang X, Lv J et al (2012) In vivo laser scanning confocal microscopy of the cornea in patients with silicone oil tamponade after vitreoretinal surgery. Cornea 31:876–882CrossRefPubMed
13.
go back to reference Kamiya K, Shimizu K, Ohmoto F et al (2010) Time course of corneal biomechanical parameters after phacoe-mulsification with intraocular lens implantation. Cornea 29:1256–1260CrossRefPubMed Kamiya K, Shimizu K, Ohmoto F et al (2010) Time course of corneal biomechanical parameters after phacoe-mulsification with intraocular lens implantation. Cornea 29:1256–1260CrossRefPubMed
14.
go back to reference Smedowski A, Weglarz B, Tarnawska D et al (2014) Comparison of three intraocular pressure measurement methods including biomechanical properties of the cornea. Investig Ophthalmol Vis Sci 55:666–673CrossRef Smedowski A, Weglarz B, Tarnawska D et al (2014) Comparison of three intraocular pressure measurement methods including biomechanical properties of the cornea. Investig Ophthalmol Vis Sci 55:666–673CrossRef
15.
go back to reference Hong J, Xu J, Wei A et al (2013) A new tonometer-the Corvis ST tonometer: clinical comparison with noncontact and Goldmann applanation tonometers. Investig Ophthalmol Vis Sci 54:659–665CrossRef Hong J, Xu J, Wei A et al (2013) A new tonometer-the Corvis ST tonometer: clinical comparison with noncontact and Goldmann applanation tonometers. Investig Ophthalmol Vis Sci 54:659–665CrossRef
16.
go back to reference Leung CK, Ye C, Weinreb RN (2013) An ultra-high speed Scheimpflug camera for evaluation for corneal deformation response and its impact on IOP measurements. Investig Ophthalmol Vis Sci 54:2885–2892CrossRef Leung CK, Ye C, Weinreb RN (2013) An ultra-high speed Scheimpflug camera for evaluation for corneal deformation response and its impact on IOP measurements. Investig Ophthalmol Vis Sci 54:2885–2892CrossRef
17.
go back to reference Kling S, Marcos S (2013) Contributing factors to corneal deformation in air puff measurements. Investig Ophthalmol Vis Sci 54:5078–5085CrossRef Kling S, Marcos S (2013) Contributing factors to corneal deformation in air puff measurements. Investig Ophthalmol Vis Sci 54:5078–5085CrossRef
18.
go back to reference Huseynova T, Waring GO 4th, Roberts C et al (2014) Corneal biomechanics as a function of intraocular pressure and pachymetry by dynamic infrared signal and Scheimpflug imaging analysis in normal eyes. Am J Ophthalmol 157:885–893CrossRefPubMed Huseynova T, Waring GO 4th, Roberts C et al (2014) Corneal biomechanics as a function of intraocular pressure and pachymetry by dynamic infrared signal and Scheimpflug imaging analysis in normal eyes. Am J Ophthalmol 157:885–893CrossRefPubMed
Metadata
Title
IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry
Authors
Yang Zhang
Lin Zheng
Ailing Bian
Qi Zhou
Publication date
01-04-2018
Publisher
Springer Netherlands
Published in
International Ophthalmology / Issue 2/2018
Print ISSN: 0165-5701
Electronic ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-017-0519-1

Other articles of this Issue 2/2018

International Ophthalmology 2/2018 Go to the issue