Skip to main content
Top
Published in: Pediatric Surgery International 7/2017

01-07-2017 | Original Article

Incidence and outcome of retained Port-A-Cath fragments during removal

Authors: Olugbenga Michael Aworanti, Niall Linnane, Farhan Tareen, Alan Mortell

Published in: Pediatric Surgery International | Issue 7/2017

Login to get access

Abstract

Purpose

Port-A-Cath devices are frequently used for long-term venous access. We postulate that long-term use predisposes them to getting stuck and retained in a central vein at the time of removal. We aim to report the incidence and outcome of this complication.

Methods

Between January 2006 and July 2016, a retrospective review of all Port-A-Cath removals that were performed at our centre was conducted. At the time of removal, catheters that could not be removed from the vein were considered retained.

Results

During the study period, 107 children had 174 episodes of silicone Sitimplant (Vygon, Ecouen, France) Port-A-Cath insertions. These children required 135 removal episodes and there were 3 (2.2%) instances whereby the catheter fragment was retained. These episodes of retained catheters only occurred in children with factor VIII deficiency (4.1% incidence in this cohort). For each episode of catheter insertion and removal, the catheters had been left in situ for a mean duration of 43 months in children with factor deficiency and no retained fragments, and the mean duration was 91 months in children with factor deficiency and retained catheter fragments (p = 0.0011).

Conclusions

Port-A-Caths that are retained after attempted removal is a complication encountered predominantly in catheters that have been in use for a prolonged duration. Furthermore, factor replacement therapy in haemophiliacs may be a risk factor for this complication.
Literature
8.
go back to reference Quaretti P, Galli F, Fiorina I et al (2014) A refinement of Hong’s technique for the removal of stuck dialysis catheters: an easy solution to a complex problem. J Vasc Access 15:183–188. doi:10.5301/jva.5000186 CrossRefPubMed Quaretti P, Galli F, Fiorina I et al (2014) A refinement of Hong’s technique for the removal of stuck dialysis catheters: an easy solution to a complex problem. J Vasc Access 15:183–188. doi:10.​5301/​jva.​5000186 CrossRefPubMed
11.
go back to reference Hong J (2010) An easy technique for removal of a hemodialysis catheter stuck in central veins. J Vasc Access 11:59–62PubMed Hong J (2010) An easy technique for removal of a hemodialysis catheter stuck in central veins. J Vasc Access 11:59–62PubMed
12.
go back to reference Bessoud B, de Baere T, Kuoch V et al (2003) Experience at a single institution with endovascular treatment of mechanical complications caused by implanted central venous access devices in pediatric and adult patients. AJR Am J Roentgenol 180:527–532. doi:10.2214/ajr.180.2.1800527 CrossRefPubMed Bessoud B, de Baere T, Kuoch V et al (2003) Experience at a single institution with endovascular treatment of mechanical complications caused by implanted central venous access devices in pediatric and adult patients. AJR Am J Roentgenol 180:527–532. doi:10.​2214/​ajr.​180.​2.​1800527 CrossRefPubMed
15.
go back to reference Surov A, Wienke A, Carter JM et al (2009) Intravascular embolization of venous catheter-causes, clinical signs, and management: a systematic review. J Parenter Enter Nutr 33:677–685. doi:10.1177/0148607109335121 CrossRef Surov A, Wienke A, Carter JM et al (2009) Intravascular embolization of venous catheter-causes, clinical signs, and management: a systematic review. J Parenter Enter Nutr 33:677–685. doi:10.​1177/​0148607109335121​ CrossRef
Metadata
Title
Incidence and outcome of retained Port-A-Cath fragments during removal
Authors
Olugbenga Michael Aworanti
Niall Linnane
Farhan Tareen
Alan Mortell
Publication date
01-07-2017
Publisher
Springer Berlin Heidelberg
Published in
Pediatric Surgery International / Issue 7/2017
Print ISSN: 0179-0358
Electronic ISSN: 1437-9813
DOI
https://doi.org/10.1007/s00383-017-4103-6

Other articles of this Issue 7/2017

Pediatric Surgery International 7/2017 Go to the issue