01-06-2021 | Implantable Cardioverter Defibrillator | Original Article

Investigating the cause of hemolysis in patients supported by a pulsatile ventricular assist device

Authors: Takeaki Harada, Yoshihisa Tanoue, Yasuhisa Oishi, Hiromichi Sonoda, Satoshi Kimura, Satoshi Fujita, Tomoki Ushijima, Ryo Kosaka, Koichi Kojima, Akira Shiose

Published in: Heart and Vessels | Issue 6/2021

Login to get access

Abstract

A survey conducted by Abiomed, Inc. revealed that 10 of 60 patients who received ventricular assistance via the AB5000 ventricular assist device (VAD) experienced hemolysis. The present study was conducted to investigate which factors influence hemolysis under pulsatile-flow VADs such as the AB5000. We compared the specificity of the AB5000 and its driving console with those of the NIPRO-VAD and VCT50χ under severe heart failure conditions using a mock circulatory system with a glycerol water solution. We used the mock circuit with bovine blood to confirm which pump conditions were most likely to cause hemolysis. In addition, we measured the shear velocity using particle image velocimetry by analyzing the seeding particle motion for both the AB5000 and NIPRO-VAD under the same conditions as those indicated in the initial experiment. Finally, we analyzed the correlation between negative pressure, exposure time, and hemolysis by continuously exposing fixed vacuum pressures for fixed times in a sealed device injected with bovine blood. Applying higher vacuum pressure to the AB5000 pump yielded a larger minimum inlet pressure and a longer exposure time when the negative pressure was under − 10 mmHg. The plasma-free hemoglobin increased as more negative pressure was driven into the AB5000 pump. Moreover, the negative pressure interacted with the exposure time, inducing hemolysis. This study revealed that negative pressure and exposure time were both associated with hemolysis.
Literature
1.
go back to reference Friedman JA (2020) Experiences of left ventricular assist device-destination therapy recipients: a systematic review and meta-synthesis. Heart Lung 49:463–474CrossRef Friedman JA (2020) Experiences of left ventricular assist device-destination therapy recipients: a systematic review and meta-synthesis. Heart Lung 49:463–474CrossRef
2.
go back to reference Woodburn JL, Staley LL, Wordingham SE, Spadafore J, Boldea E, Williamson S, Hollenbach S, Ross HM, Steidley DE, Pajaro OE (2019) Destination therapy: standardizing the role of palliative medicine and delineating the DT-LVAD journey. J Pain Symptom Manag 57(2):330–340CrossRef Woodburn JL, Staley LL, Wordingham SE, Spadafore J, Boldea E, Williamson S, Hollenbach S, Ross HM, Steidley DE, Pajaro OE (2019) Destination therapy: standardizing the role of palliative medicine and delineating the DT-LVAD journey. J Pain Symptom Manag 57(2):330–340CrossRef
3.
go back to reference Samuels LE, Holmes EC, Garwood P, Ferdinand F (2005) Initial experience with the Abiomed AB5000 ventricular assist device system. Ann Thorac Surg 80:309–312CrossRef Samuels LE, Holmes EC, Garwood P, Ferdinand F (2005) Initial experience with the Abiomed AB5000 ventricular assist device system. Ann Thorac Surg 80:309–312CrossRef
4.
go back to reference Yasuda T, Shimokasa K, Funakubo A, Higami T, Kawamura T, Fukui Y (2000) An investigation of blood flow behavior and hemolysis in artificial organs. ASAIO J 46:527–531CrossRef Yasuda T, Shimokasa K, Funakubo A, Higami T, Kawamura T, Fukui Y (2000) An investigation of blood flow behavior and hemolysis in artificial organs. ASAIO J 46:527–531CrossRef
5.
go back to reference Hernandez AF, Grab JD, Gammie JS, O’Brien SM, Hammill BG, Rogers JG, Camacho MT, Dullum MK, Ferguson TB, Peterson ED (2007) A decade of short-term outcomes in post cardiac surgery ventricular assist device implantation: data from the Society of Thoracic Surgeons’ National Cardiac Database. Circulation 116:606–612CrossRef Hernandez AF, Grab JD, Gammie JS, O’Brien SM, Hammill BG, Rogers JG, Camacho MT, Dullum MK, Ferguson TB, Peterson ED (2007) A decade of short-term outcomes in post cardiac surgery ventricular assist device implantation: data from the Society of Thoracic Surgeons’ National Cardiac Database. Circulation 116:606–612CrossRef
6.
go back to reference Geisen U, Heilmann C, Beyersdorf F, Benk C, Berchtold-Herz M, Schlensak U, Zieger B (2008) Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease. Eur J Cardiothorac Surg 33:679–684CrossRef Geisen U, Heilmann C, Beyersdorf F, Benk C, Berchtold-Herz M, Schlensak U, Zieger B (2008) Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease. Eur J Cardiothorac Surg 33:679–684CrossRef
7.
go back to reference Inge K (2020) Haemolysis induced by mechanical circulatory support devices: unsolved problems. Perfusion 35(6):474–483CrossRef Inge K (2020) Haemolysis induced by mechanical circulatory support devices: unsolved problems. Perfusion 35(6):474–483CrossRef
8.
go back to reference Qi Q, Karl AN, Yiming W, Tarek MD, Sanjeev S (2010) Hemolysis and acute kidney failure. Am J Kidney Dis 56(4):780–784CrossRef Qi Q, Karl AN, Yiming W, Tarek MD, Sanjeev S (2010) Hemolysis and acute kidney failure. Am J Kidney Dis 56(4):780–784CrossRef
9.
go back to reference Giersiepen M, Wurzinger LJ, Opitz R, Reul H (1990) Estimation of shear stress-related blood damage in heart valve prostheses—in vitro comparison of 25 aortic valves. Int J Artif Organs 13:300–306CrossRef Giersiepen M, Wurzinger LJ, Opitz R, Reul H (1990) Estimation of shear stress-related blood damage in heart valve prostheses—in vitro comparison of 25 aortic valves. Int J Artif Organs 13:300–306CrossRef
10.
go back to reference Zhang L, Kapetanakis EI, Cooke RH, Sweet LC, Boyce SW (2007) Bi-ventricular circulatory support with the Abiomed AB5000 system in a patient with idiopathic refractory ventricular fibrillation. Ann Thorac Surg 83:298–300CrossRef Zhang L, Kapetanakis EI, Cooke RH, Sweet LC, Boyce SW (2007) Bi-ventricular circulatory support with the Abiomed AB5000 system in a patient with idiopathic refractory ventricular fibrillation. Ann Thorac Surg 83:298–300CrossRef
11.
go back to reference Takano H, Nakatani T, Taenaka Y, Kitoh Y, Hiramori K, Haze K, Itoh A, Fujita T, Manabe H (1992) Treatment of acute profound heart failure by ventricular assist device. Jpn Circ J 56:100–110CrossRef Takano H, Nakatani T, Taenaka Y, Kitoh Y, Hiramori K, Haze K, Itoh A, Fujita T, Manabe H (1992) Treatment of acute profound heart failure by ventricular assist device. Jpn Circ J 56:100–110CrossRef
12.
go back to reference Kimura M, Nishimura T, Kinoshita O, Kashiwa K, Kyo S, Ono M (2012) Hemodynamic influence of tilting disc value type on pump performance with the NIPRO-ventricular assist device. J Artif Organs 15:134–139CrossRef Kimura M, Nishimura T, Kinoshita O, Kashiwa K, Kyo S, Ono M (2012) Hemodynamic influence of tilting disc value type on pump performance with the NIPRO-ventricular assist device. J Artif Organs 15:134–139CrossRef
13.
go back to reference Takano H, Taenaka Y, Noda H, Kinoshita M, Yagura A, Tatsumi E, Sekii H, Sasaki E, Umezu M, Nakatani T (1989) Multi-institutional studies of the National Cardiovascular Center ventricular assist system: use in 92 patients. ASAIO Trans 35(3):541–544CrossRef Takano H, Taenaka Y, Noda H, Kinoshita M, Yagura A, Tatsumi E, Sekii H, Sasaki E, Umezu M, Nakatani T (1989) Multi-institutional studies of the National Cardiovascular Center ventricular assist system: use in 92 patients. ASAIO Trans 35(3):541–544CrossRef
14.
go back to reference Chambers SD, Laberteaux KR, Merz SI, Montoya JP, Bartlett RH (1996) Effects of static pressure on red blood cells on removal of the air interface. ASAIO J 42:947–950CrossRef Chambers SD, Laberteaux KR, Merz SI, Montoya JP, Bartlett RH (1996) Effects of static pressure on red blood cells on removal of the air interface. ASAIO J 42:947–950CrossRef
15.
go back to reference Nevaril CG, Hellums JD, Alfrey CP Jr, Lynch EC (1969) Physical effects in red blood cell trauma. AIChE J 15(5):707–711CrossRef Nevaril CG, Hellums JD, Alfrey CP Jr, Lynch EC (1969) Physical effects in red blood cell trauma. AIChE J 15(5):707–711CrossRef
16.
go back to reference Pohlmann JR, Toomasian JM, Hampton CE, Cook KE, Annich GM, Bartlett RH (2009) The relationships between air exposure, negative pressure, and hemolysis. ASAIO J 55:469–473CrossRef Pohlmann JR, Toomasian JM, Hampton CE, Cook KE, Annich GM, Bartlett RH (2009) The relationships between air exposure, negative pressure, and hemolysis. ASAIO J 55:469–473CrossRef
Metadata
Title
Investigating the cause of hemolysis in patients supported by a pulsatile ventricular assist device
Authors
Takeaki Harada
Yoshihisa Tanoue
Yasuhisa Oishi
Hiromichi Sonoda
Satoshi Kimura
Satoshi Fujita
Tomoki Ushijima
Ryo Kosaka
Koichi Kojima
Akira Shiose
Publication date
01-06-2021
Publisher
Springer Japan
Published in
Heart and Vessels / Issue 6/2021
Print ISSN: 0910-8327
Electronic ISSN: 1615-2573
DOI
https://doi.org/10.1007/s00380-021-01809-y

Other articles of this Issue 6/2021

Heart and Vessels 6/2021 Go to the issue