Published in: European Radiology 9/2008

01-09-2008 | Computer Tomography

Image quality on dual-source computed-tomographic coronary angiography

Authors: Johannes Rixe, Andreas Rolf, Guido Conradi, Albrecht Elsaesser, Helge Moellmann, Holger M. Nef, Georg Bachmann, Christian W. Hamm, Thorsten Dill

Published in: European Radiology | Issue 9/2008

Login to get access

Abstract

Multi-detector CT reliably permits visualization of coronary arteries, but due to the occurrence of motion artefacts at heart rates >65 bpm caused by a temporal resolution of 165 ms, its utilisation has so far been limited to patients with a preferably low heart rate. We investigated the assessment of image quality on computed tomography of coronary arteries in a large series of patients without additional heart rate control using dual-source computed tomography (DSCT). DSCT (Siemens Somatom Definition, 83-ms temporal resolution) was performed in 165 consecutive patients (mean age 64 ± 11.4 years) after injection of 60–80 ml of contrast. Data sets were reconstructed in 5% intervals of the cardiac cycle and evaluated by two readers in consensus concerning evaluability of the coronary arteries and presence of motion and beam-hardening artefacts using the AHA 16-segment coronary model. Mean heart rate during CT was 65 ± 10.5 bpm; visualisation without artefacts was possible in 98.7% of 2,541 coronary segments. Only two segments were considered unevaluable due to cardiac motion; 30 segments were unassessable due to poor signal-to-noise ratio or coronary calcifications (both n = 15). Data reconstruction at 65-70% of the cardiac cycle provided for the best image quality. For heart rates >85 bpm, a systolic reconstruction at 45% revealed satisfactory results. Compared with earlier CT generations, DSCT provides for non-invasive coronary angiography with diagnostic image quality even at heart rates >65 bpm and thus may broaden the spectrum of patients that can be investigated non-invasively.
Literature
1.
2.
go back to reference Ropers D, Pohle FK, Kuettner A et al (2006) Diagnostic accuracy of noninvasive coronary angiography in patients after bypass surgery using 64-slice spiral computed tomography with 330-ms gantry rotation. Circulation 114:2334–2341PubMedCrossRef Ropers D, Pohle FK, Kuettner A et al (2006) Diagnostic accuracy of noninvasive coronary angiography in patients after bypass surgery using 64-slice spiral computed tomography with 330-ms gantry rotation. Circulation 114:2334–2341PubMedCrossRef
3.
go back to reference Sun Z, Jiang W (2006) Diagnostic value of multislice computed tomography angiography in coronary artery disease: a meta-analysis. Eur J Radiol 60:279–286PubMedCrossRef Sun Z, Jiang W (2006) Diagnostic value of multislice computed tomography angiography in coronary artery disease: a meta-analysis. Eur J Radiol 60:279–286PubMedCrossRef
4.
go back to reference Sun Z, Lin C, Davidson R, Dong C, Liao Y (2007) Diagnostic value of 64-slice CT angiography in coronary artery disease: A systematic review. Eur J Radiol (Epub ahead of print) Sun Z, Lin C, Davidson R, Dong C, Liao Y (2007) Diagnostic value of 64-slice CT angiography in coronary artery disease: A systematic review. Eur J Radiol (Epub ahead of print)
5.
go back to reference Vanhoenacker PK, Heijenbrok-Kal MH, Van Heste R et al (2007) Diagnostic performance of multidetector CT angiography for assessment of coronary artery disease: meta-analysis. Radiology 244:419–428PubMedCrossRef Vanhoenacker PK, Heijenbrok-Kal MH, Van Heste R et al (2007) Diagnostic performance of multidetector CT angiography for assessment of coronary artery disease: meta-analysis. Radiology 244:419–428PubMedCrossRef
6.
go back to reference Fox K, Garcia MA, Ardissino D et al (2006) Guidelines on the management of stable angina pectoris: executive summary: the Task Force on the Management of Stable Angina Pectoris of the European Society of Cardiology. Eur Heart J 27:1341–1381PubMedCrossRef Fox K, Garcia MA, Ardissino D et al (2006) Guidelines on the management of stable angina pectoris: executive summary: the Task Force on the Management of Stable Angina Pectoris of the European Society of Cardiology. Eur Heart J 27:1341–1381PubMedCrossRef
7.
go back to reference Budoff MJ, Achenbach S, Blumenthal RS et al (2006) Assessment of coronary artery disease by cardiac computed tomography: a scientific statement from the American Heart Association Committee on Cardiovascular Imaging and Intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology. Circulation 114:1761–1791PubMedCrossRef Budoff MJ, Achenbach S, Blumenthal RS et al (2006) Assessment of coronary artery disease by cardiac computed tomography: a scientific statement from the American Heart Association Committee on Cardiovascular Imaging and Intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology. Circulation 114:1761–1791PubMedCrossRef
8.
go back to reference Cordeiro MA, Miller JM, Schmidt A et al (2006) Non-invasive half millimetre 32 detector row computed tomography angiography accurately excludes significant stenoses in patients with advanced coronary artery disease and high calcium scores. Heart 92:589–597PubMedCrossRef Cordeiro MA, Miller JM, Schmidt A et al (2006) Non-invasive half millimetre 32 detector row computed tomography angiography accurately excludes significant stenoses in patients with advanced coronary artery disease and high calcium scores. Heart 92:589–597PubMedCrossRef
9.
go back to reference Herzog C, Arning-Erb M, Zangos S et al (2006) Multi-detector row CT coronary angiography: influence of reconstruction technique and heart rate on image quality. Radiology 238:75–86PubMedCrossRef Herzog C, Arning-Erb M, Zangos S et al (2006) Multi-detector row CT coronary angiography: influence of reconstruction technique and heart rate on image quality. Radiology 238:75–86PubMedCrossRef
10.
go back to reference Hoffmann MH, Shi H, Schmitz BL et al (2005) Noninvasive coronary angiography with multislice computed tomography. Jama 293:2471–2478PubMedCrossRef Hoffmann MH, Shi H, Schmitz BL et al (2005) Noninvasive coronary angiography with multislice computed tomography. Jama 293:2471–2478PubMedCrossRef
11.
go back to reference Giesler T, Baum U, Ropers D et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR 179:911–916PubMed Giesler T, Baum U, Ropers D et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR 179:911–916PubMed
12.
go back to reference Knez A, Becker CR, Leber A et al (2001) Usefulness of multislice spiral computed tomography angiography for determination of coronary artery stenoses. Am J Cardiol 88:1191–1194PubMedCrossRef Knez A, Becker CR, Leber A et al (2001) Usefulness of multislice spiral computed tomography angiography for determination of coronary artery stenoses. Am J Cardiol 88:1191–1194PubMedCrossRef
13.
go back to reference Nieman K, Oudkerk M, Rensing BJ et al (2001) Coronary angiography with multi-slice computed tomography. Lancet 357:599–603PubMedCrossRef Nieman K, Oudkerk M, Rensing BJ et al (2001) Coronary angiography with multi-slice computed tomography. Lancet 357:599–603PubMedCrossRef
14.
go back to reference Schroeder S, Kopp AF, Kuettner A et al (2002) Influence of heart rate on vessel visibility in noninvasive coronary angiography using new multislice computed tomography: experience in 94 patients. Clin Imaging 26:106–111PubMedCrossRef Schroeder S, Kopp AF, Kuettner A et al (2002) Influence of heart rate on vessel visibility in noninvasive coronary angiography using new multislice computed tomography: experience in 94 patients. Clin Imaging 26:106–111PubMedCrossRef
15.
go back to reference Leber AW, Knez A, von Ziegler F et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef Leber AW, Knez A, von Ziegler F et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef
16.
go back to reference Nikolaou K, Knez A, Rist C et al (2006) Accuracy of 64-MDCT in the diagnosis of ischemic heart disease. AJR 187:111–117PubMedCrossRef Nikolaou K, Knez A, Rist C et al (2006) Accuracy of 64-MDCT in the diagnosis of ischemic heart disease. AJR 187:111–117PubMedCrossRef
17.
go back to reference Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557PubMedCrossRef Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557PubMedCrossRef
18.
go back to reference Mollet NR, Cademartiri F, van Mieghem CA et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323PubMedCrossRef Mollet NR, Cademartiri F, van Mieghem CA et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323PubMedCrossRef
19.
go back to reference Pugliese F, Mollet NR, Runza G et al (2006) Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris. Eur Radiol 16:575–582PubMedCrossRef Pugliese F, Mollet NR, Runza G et al (2006) Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris. Eur Radiol 16:575–582PubMedCrossRef
20.
go back to reference Ropers D, Baum U, Pohle K et al (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666PubMedCrossRef Ropers D, Baum U, Pohle K et al (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666PubMedCrossRef
21.
go back to reference Ropers D, Regenfus M, Wasmeier G, Achenbach S (2004) Non-interventional cardiac diagnostics: computed tomography, magnetic resonance and real-time three-dimensional echocardiography. Techniques and clinical applications. Minerva Cardioangiol 52:407–417PubMed Ropers D, Regenfus M, Wasmeier G, Achenbach S (2004) Non-interventional cardiac diagnostics: computed tomography, magnetic resonance and real-time three-dimensional echocardiography. Techniques and clinical applications. Minerva Cardioangiol 52:407–417PubMed
22.
go back to reference Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268PubMedCrossRef Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268PubMedCrossRef
23.
go back to reference Flohr TG, Schaller S, Stierstorfer K et al (2005) Multi-detector row CT systems and image-reconstruction techniques. Radiology 235:756–773PubMedCrossRef Flohr TG, Schaller S, Stierstorfer K et al (2005) Multi-detector row CT systems and image-reconstruction techniques. Radiology 235:756–773PubMedCrossRef
24.
go back to reference Achenbach S, Ropers D, Kuettner A et al (2006) Contrast-enhanced coronary artery visualization by dual-source computed tomography–initial experience. Eur J Radiol 57:331–335PubMedCrossRef Achenbach S, Ropers D, Kuettner A et al (2006) Contrast-enhanced coronary artery visualization by dual-source computed tomography–initial experience. Eur J Radiol 57:331–335PubMedCrossRef
25.
go back to reference Johnson TR, Nikolaou K, Wintersperger BJ et al (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415PubMedCrossRef Johnson TR, Nikolaou K, Wintersperger BJ et al (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415PubMedCrossRef
26.
go back to reference Scheffel H, Alkadhi H, Plass A et al (2006) Accuracy of dual-source CT coronary angiography: First experience in a high pre-test probability population without heart rate control. Eur Radiol 16:2739–2747PubMedCrossRef Scheffel H, Alkadhi H, Plass A et al (2006) Accuracy of dual-source CT coronary angiography: First experience in a high pre-test probability population without heart rate control. Eur Radiol 16:2739–2747PubMedCrossRef
27.
go back to reference Austen WG, Edwards JE, Frye RL et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40PubMed Austen WG, Edwards JE, Frye RL et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40PubMed
28.
go back to reference Wintersperger BJ, Nikolaou K, von Ziegler F et al (2006) Image quality, motion artifacts, and reconstruction timing of 64-slice coronary computed tomography angiography with 0.33-second rotation speed. Invest Radiol 41:436–442PubMedCrossRef Wintersperger BJ, Nikolaou K, von Ziegler F et al (2006) Image quality, motion artifacts, and reconstruction timing of 64-slice coronary computed tomography angiography with 0.33-second rotation speed. Invest Radiol 41:436–442PubMedCrossRef
29.
go back to reference Leschka S, Scheffel H, Desbiolles L et al (2007) Image quality and reconstruction intervals of dual-source CT coronary angiography: recommendations for ECG-pulsing windowing. Invest Radiol 42:543–549PubMedCrossRef Leschka S, Scheffel H, Desbiolles L et al (2007) Image quality and reconstruction intervals of dual-source CT coronary angiography: recommendations for ECG-pulsing windowing. Invest Radiol 42:543–549PubMedCrossRef
30.
go back to reference Husmann L, Valenta I, Gaemperli O et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197PubMedCrossRef Husmann L, Valenta I, Gaemperli O et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197PubMedCrossRef
31.
go back to reference Hoffmann MH, Lessick J, Manzke R et al (2006) Automatic determination of minimal cardiac motion phases for computed tomography imaging: initial experience. Eur Radiol 16:365–373PubMedCrossRef Hoffmann MH, Lessick J, Manzke R et al (2006) Automatic determination of minimal cardiac motion phases for computed tomography imaging: initial experience. Eur Radiol 16:365–373PubMedCrossRef
Metadata
Title
Image quality on dual-source computed-tomographic coronary angiography
Authors
Johannes Rixe
Andreas Rolf
Guido Conradi
Albrecht Elsaesser
Helge Moellmann
Holger M. Nef
Georg Bachmann
Christian W. Hamm
Thorsten Dill
Publication date
01-09-2008
Publisher
Springer-Verlag
Published in
European Radiology / Issue 9/2008
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-008-0947-4

Other articles of this Issue 9/2008

European Radiology 9/2008 Go to the issue