Published in: The International Journal of Cardiovascular Imaging 1/2013

01-06-2013 | Original Paper

Image quality and radiation dose of 128-slice dual-source CT venography using low kilovoltage combined with high-pitch scanning and automatic tube current modulation

Authors: Chan Kue Park, Ki Seok Choo, Ung Bae Jeon, Seung Kug Baik, Yong Woo Kim, Tae Un Kim, Chang Won Kim, Yeon Ju Jeong, Dong Wook Jeong, Soo Jin Lim

Published in: The International Journal of Cardiovascular Imaging | Special Issue 1/2013

Login to get access

Abstract

To compare vascular enhancement, image quality, and radiation dose of 128-slice dual-source CT venography (CTV) between an imaging setting of 120 kVp with low pitch, and a setting of 100 kVp combined with high pitch and automatic tube current modulation. A total of 100 patients with suspected deep vein thrombosis and varicose veins were divided into two groups: Group 1 [50 patients, 120 kVp, low pitch (0.6), and fixed 120 mA) and Group 2 (50 patients, 100 kVp, high pitch (3.0), and automatic tube current modulation]. Two radiologists, who were blinded to the image protocol, assessed vascular enhancement and image noise in the inferior vena cava (IVC), femoral vein, and popliteal vein. They also assigned an image quality score independently using a 5-point visual scale. Effective dose was estimated using the dose-length product (DLP). Group demographics, radiation dose, vascular enhancement, image noise, and image quality in the two groups were analyzed. Mean vascular enhancement of the IVC, femoral vein, and popliteal vein was significantly higher in group 2 than in group 1, and images in group 2 had significantly higher image noise. However, there were no significant differences in subjective image quality score of the IVC, femoral vein, and popliteal vein. The mean DLP in group 2 (402.10 ± 94.29 mGy cm) was significantly lower than that in group 1 (973.36 ± 63.20 mGy cm) (P < 0.001). Lower extremity CTV using 100 kVp, high pitch (3.0), and automatic tube current modulation improved vascular enhancement with acceptable image quality and low radiation dose.
Literature
1.
go back to reference Byrne JJ, O’Neil EE (1952) Fatal pulmonary emboli; a study of 130 autopsy proven fatal emboli. Am J Surg 83(1):47–49PubMedCrossRef Byrne JJ, O’Neil EE (1952) Fatal pulmonary emboli; a study of 130 autopsy proven fatal emboli. Am J Surg 83(1):47–49PubMedCrossRef
2.
go back to reference Kalra MK, Maher MM, Rizzo S, Kanarek D, Shepard JA (2004) Radiation exposure from chest CT: issues and strategies. J Korean Med Sci 19(2):159–166PubMedCrossRef Kalra MK, Maher MM, Rizzo S, Kanarek D, Shepard JA (2004) Radiation exposure from chest CT: issues and strategies. J Korean Med Sci 19(2):159–166PubMedCrossRef
3.
go back to reference Tack D, Gevenois PA (2004) Radiation dose in computed tomography of the chest. JBR-BTR 87(6):281–288PubMed Tack D, Gevenois PA (2004) Radiation dose in computed tomography of the chest. JBR-BTR 87(6):281–288PubMed
4.
go back to reference Maher MM, Kalra MK, Toth TL, Wittram C, Saini S, Shepard J (2004) Application of rational practice and technical advances for optimizing radiation dose for chest CT. J Thorac Imaging 19(1):16–23PubMedCrossRef Maher MM, Kalra MK, Toth TL, Wittram C, Saini S, Shepard J (2004) Application of rational practice and technical advances for optimizing radiation dose for chest CT. J Thorac Imaging 19(1):16–23PubMedCrossRef
5.
go back to reference Mayo JR, Aldrich J, Muller NL, Fleischner S (2003) Radiation exposure at chest CT: a statement of the Fleischner Society. Radiology 228(1):15–21PubMedCrossRef Mayo JR, Aldrich J, Muller NL, Fleischner S (2003) Radiation exposure at chest CT: a statement of the Fleischner Society. Radiology 228(1):15–21PubMedCrossRef
6.
go back to reference McNitt-Gray MF (2002) AAPM/RSNA Physics Tutorial for Residents: topics in CT Radiation dose in CT. Radiographics 22(6):1541–1553PubMedCrossRef McNitt-Gray MF (2002) AAPM/RSNA Physics Tutorial for Residents: topics in CT Radiation dose in CT. Radiographics 22(6):1541–1553PubMedCrossRef
7.
go back to reference Hamberg LM, Rhea JT, Hunter GJ, Thrall JH (2003) Multi-detector row CT: radiation dose characteristics. Radiology 226(3):762–772PubMedCrossRef Hamberg LM, Rhea JT, Hunter GJ, Thrall JH (2003) Multi-detector row CT: radiation dose characteristics. Radiology 226(3):762–772PubMedCrossRef
8.
go back to reference Silverman PM, Kalender WA, Hazle JD (2001) Common terminology for single and multislice helical CT. AJR Am J Roentgenol 176(5):1135–1136PubMedCrossRef Silverman PM, Kalender WA, Hazle JD (2001) Common terminology for single and multislice helical CT. AJR Am J Roentgenol 176(5):1135–1136PubMedCrossRef
10.
go back to reference Kalra MK, Maher MM, Toth TL, Schmidt B, Westerman BL, Morgan HT, Saini S (2004) Techniques and applications of automatic tube current modulation for CT. Radiology 233(3):649–657PubMedCrossRef Kalra MK, Maher MM, Toth TL, Schmidt B, Westerman BL, Morgan HT, Saini S (2004) Techniques and applications of automatic tube current modulation for CT. Radiology 233(3):649–657PubMedCrossRef
11.
go back to reference Kubo T, Lin PJ, Stiller W, Takahashi M, Kauczor HU, Ohno Y, Hatabu H (2008) Radiation dose reduction in chest CT: a review. AJR Am J Roentgenol 190(2):335–343PubMedCrossRef Kubo T, Lin PJ, Stiller W, Takahashi M, Kauczor HU, Ohno Y, Hatabu H (2008) Radiation dose reduction in chest CT: a review. AJR Am J Roentgenol 190(2):335–343PubMedCrossRef
12.
go back to reference Fujikawa A, Matsuoka S, Kuramochi K, Yoshikawa T, Yagihashi K, Kurihara Y, Nakajima Y (2011) Vascular enhancement and image quality of CT venography: comparison of standard and low kilovoltage settings. AJR Am J Roentgenol 197(4):838–843PubMedCrossRef Fujikawa A, Matsuoka S, Kuramochi K, Yoshikawa T, Yagihashi K, Kurihara Y, Nakajima Y (2011) Vascular enhancement and image quality of CT venography: comparison of standard and low kilovoltage settings. AJR Am J Roentgenol 197(4):838–843PubMedCrossRef
13.
go back to reference Kulkarni NM, Sahani DV, Desai GS, Kalva SP (2012) Indirect computed tomography venography of the lower extremities using single-source dual-energy computed tomography: advantage of low-kiloelectron volt monochromatic images. J Vasc Interv Radiol 23(7):879–886PubMedCrossRef Kulkarni NM, Sahani DV, Desai GS, Kalva SP (2012) Indirect computed tomography venography of the lower extremities using single-source dual-energy computed tomography: advantage of low-kiloelectron volt monochromatic images. J Vasc Interv Radiol 23(7):879–886PubMedCrossRef
14.
go back to reference Oda S, Utsunomiya D, Awai K, Takaoka H, Nakaura T, Katahira K, Morishita S, Namimoto T, Yamashita Y (2011) Indirect computed tomography venography with a low-tube-voltage technique: reduction in the radiation and contrast material dose–a prospective randomized study. J Comput Assist Tomogr 35(5):631–636PubMedCrossRef Oda S, Utsunomiya D, Awai K, Takaoka H, Nakaura T, Katahira K, Morishita S, Namimoto T, Yamashita Y (2011) Indirect computed tomography venography with a low-tube-voltage technique: reduction in the radiation and contrast material dose–a prospective randomized study. J Comput Assist Tomogr 35(5):631–636PubMedCrossRef
15.
go back to reference Nakaura T, Awai K, Oda S, Yanaga Y, Namimoto T, Harada K, Uemura S, Yamashita Y (2011) A low-kilovolt (peak) high-tube current technique improves venous enhancement and reduces the radiation dose at indirect multidetector-row CT venography: initial experience. J Comput Assist Tomogr 35(1):141–147PubMedCrossRef Nakaura T, Awai K, Oda S, Yanaga Y, Namimoto T, Harada K, Uemura S, Yamashita Y (2011) A low-kilovolt (peak) high-tube current technique improves venous enhancement and reduces the radiation dose at indirect multidetector-row CT venography: initial experience. J Comput Assist Tomogr 35(1):141–147PubMedCrossRef
16.
go back to reference Wintermark M, Maeder P, Verdun FR, Thiran JP, Valley JF, Schnyder P, Meuli R (2000) Using 80 kVp versus 120 kVp in perfusion CT measurement of regional cerebral blood flow. AJNR Am J Neuroradiol 21(10):1881–1884PubMed Wintermark M, Maeder P, Verdun FR, Thiran JP, Valley JF, Schnyder P, Meuli R (2000) Using 80 kVp versus 120 kVp in perfusion CT measurement of regional cerebral blood flow. AJNR Am J Neuroradiol 21(10):1881–1884PubMed
17.
go back to reference Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217(2):430–435PubMed Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217(2):430–435PubMed
18.
go back to reference Heyer CM, Mohr PS, Lemburg SP, Peters SA, Nicolas V (2007) Image quality and radiation exposure at pulmonary CT angiography with 100- or 120-kVp protocol: prospective randomized study. Radiology 245(2):577–583PubMedCrossRef Heyer CM, Mohr PS, Lemburg SP, Peters SA, Nicolas V (2007) Image quality and radiation exposure at pulmonary CT angiography with 100- or 120-kVp protocol: prospective randomized study. Radiology 245(2):577–583PubMedCrossRef
19.
go back to reference Schueller-Weidekamm C, Schaefer-Prokop CM, Weber M, Herold CJ, Prokop M (2006) CT angiography of pulmonary arteries to detect pulmonary embolism: improvement of vascular enhancement with low kilovoltage settings. Radiology 241(3):899–907PubMedCrossRef Schueller-Weidekamm C, Schaefer-Prokop CM, Weber M, Herold CJ, Prokop M (2006) CT angiography of pulmonary arteries to detect pulmonary embolism: improvement of vascular enhancement with low kilovoltage settings. Radiology 241(3):899–907PubMedCrossRef
20.
go back to reference Sigal-Cinqualbre AB, Hennequin R, Abada HT, Chen X, Paul JF (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231(1):169–174PubMedCrossRef Sigal-Cinqualbre AB, Hennequin R, Abada HT, Chen X, Paul JF (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231(1):169–174PubMedCrossRef
21.
go back to reference Matsuoka S, Hunsaker AR, Gill RR, Oliva IB, Trotman-Dickenson B, Jacobson FL, Hatabu H (2009) Vascular enhancement and image quality of MDCT pulmonary angiography in 400 cases: comparison of standard and low kilovoltage settings. AJR Am J Roentgenol 192(6):1651–1656PubMedCrossRef Matsuoka S, Hunsaker AR, Gill RR, Oliva IB, Trotman-Dickenson B, Jacobson FL, Hatabu H (2009) Vascular enhancement and image quality of MDCT pulmonary angiography in 400 cases: comparison of standard and low kilovoltage settings. AJR Am J Roentgenol 192(6):1651–1656PubMedCrossRef
22.
go back to reference Boone JM, Geraghty EM, Seibert JA, Wootton-Gorges SL (2003) Dose reduction in pediatric CT: a rational approach. Radiology 228(2):352–360PubMedCrossRef Boone JM, Geraghty EM, Seibert JA, Wootton-Gorges SL (2003) Dose reduction in pediatric CT: a rational approach. Radiology 228(2):352–360PubMedCrossRef
23.
go back to reference Coche E, Vynckier S, Octave-Prignot M (2006) Pulmonary embolism: radiation dose with multi-detector row CT and digital angiography for diagnosis. Radiology 240(3):690–697PubMedCrossRef Coche E, Vynckier S, Octave-Prignot M (2006) Pulmonary embolism: radiation dose with multi-detector row CT and digital angiography for diagnosis. Radiology 240(3):690–697PubMedCrossRef
24.
go back to reference Garg K, Kemp JL, Russ PD, Baron AE (2001) Thromboembolic disease: variability of interobserver agreement in the interpretation of CT venography with CT pulmonary angiography. AJR Am J Roentgenol 176(4):1043–1047PubMedCrossRef Garg K, Kemp JL, Russ PD, Baron AE (2001) Thromboembolic disease: variability of interobserver agreement in the interpretation of CT venography with CT pulmonary angiography. AJR Am J Roentgenol 176(4):1043–1047PubMedCrossRef
Metadata
Title
Image quality and radiation dose of 128-slice dual-source CT venography using low kilovoltage combined with high-pitch scanning and automatic tube current modulation
Authors
Chan Kue Park
Ki Seok Choo
Ung Bae Jeon
Seung Kug Baik
Yong Woo Kim
Tae Un Kim
Chang Won Kim
Yeon Ju Jeong
Dong Wook Jeong
Soo Jin Lim
Publication date
01-06-2013
Publisher
Springer Netherlands
Published in
The International Journal of Cardiovascular Imaging / Issue Special Issue 1/2013
Print ISSN: 1569-5794
Electronic ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-013-0252-4

Other articles of this Special Issue 1/2013

The International Journal of Cardiovascular Imaging 1/2013 Go to the issue