Sensitivity and reliability of cerebral oxygenation responses to postural changes measured with near-infrared spectroscopy

Authors: Arjen Mol, Jeffrey H. H. Woltering, Willy N. J. M. Colier, Andrea B. Maier, Carel G. M. Meskers, Richard J. A. van Wezel

Published in: European Journal of Applied Physiology | Issue 5/2019

Login to get access

Abstract

Purpose

Cerebral oxygenation as measured by near-infrared spectroscopy (NIRS) might be useful to discriminate between physiological and pathological responses after standing up in individuals with orthostatic hypotension. This study addressed the physiological sensitivity of the cerebral oxygenation responses as measured by NIRS to different types and speeds of postural changes in healthy adults and assessed the reliability of these responses.

Methods

Cerebral oxygenated hemoglobin (O2Hb), deoxygenated hemoglobin (HHb) and tissue saturation index (TSI) were measured bilaterally on the forehead of 15 healthy individuals (12 male, age range 18–27) using NIRS. Participants performed three repeats of sit to stand, and slow and rapid supine to stand movements. Responses were defined as the difference between mean, minimum and maximum O2Hb, HHb and TSI values after standing up and baseline. Test–retest, interobserver and intersensor reliabilities were addressed using intraclass correlation coefficients (ICCs).

Results

The minimum O2Hb response was most sensitive to postural changes and showed significant differences (− 4.09 µmol/L, p < 0.001) between standing up from sitting and supine position, but not between standing up at different speeds (− 0.31 µmol/L, p = 0.70). The minimum O2Hb response was the most reliable parameter (ICC > 0.6).

Conclusions

In healthy individuals, NIRS-based cerebral oxygenation parameters are sensitive to postural change and discriminate between standing up from supine and sitting position with minimum O2Hb response as the most sensitive and reliable parameter. The results underpin the potential value for future clinical use of NIRS in individuals with orthostatic hypotension.
Appendix
Available only for authorised users
Literature
go back to reference Hiraoka M, Firbank M, Essenpreis M et al (1993) A Monte Carlo investigation of optical pathlength in inhomogeneous tissue and its application to near- infrared spectroscopy. Phys Med Biol 38:1859–1876CrossRefPubMed Hiraoka M, Firbank M, Essenpreis M et al (1993) A Monte Carlo investigation of optical pathlength in inhomogeneous tissue and its application to near- infrared spectroscopy. Phys Med Biol 38:1859–1876CrossRefPubMed
go back to reference Kawaguchi T, Uyama O, Konishi M et al (2001) Orthostatic Hypotension in Elderly Persons During Passive Standing: A Comparison With Young Persons. J Gerontol 56:273–280CrossRef Kawaguchi T, Uyama O, Konishi M et al (2001) Orthostatic Hypotension in Elderly Persons During Passive Standing: A Comparison With Young Persons. J Gerontol 56:273–280CrossRef
go back to reference Smielewski P, Kirkpatrick PJ, Minhas P et al (1995) Can cerebrovascular reactivity be measured with near-infrared spectroscopy? Stroke 26:2285–2292CrossRefPubMed Smielewski P, Kirkpatrick PJ, Minhas P et al (1995) Can cerebrovascular reactivity be measured with near-infrared spectroscopy? Stroke 26:2285–2292CrossRefPubMed
Metadata
Title
Sensitivity and reliability of cerebral oxygenation responses to postural changes measured with near-infrared spectroscopy
Authors
Arjen Mol
Jeffrey H. H. Woltering
Willy N. J. M. Colier
Andrea B. Maier
Carel G. M. Meskers
Richard J. A. van Wezel
Publication date
01-05-2019
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Applied Physiology / Issue 5/2019
Print ISSN: 1439-6319
Electronic ISSN: 1439-6327
DOI
https://doi.org/10.1007/s00421-019-04101-0

Other articles of this Issue 5/2019

European Journal of Applied Physiology 5/2019 Go to the issue