Published in: Skeletal Radiology 5/2014

01-05-2014 | Scientific Article

Hip imaging of avascular necrosis at 7 Tesla compared with 3 Tesla

Authors: J. M. Theysohn, O. Kraff, N. Theysohn, S. Orzada, S. Landgraeber, M. E. Ladd, T. C. Lauenstein

Published in: Skeletal Radiology | Issue 5/2014

Login to get access

Abstract

Objectives

To compare ultra-high field, high-resolution bilateral magnetic resonance imaging (MRI) of the hips at 7 Tesla (T) with 3 T MRI in patients with avascular necrosis (AVN) of the femoral head by subjective image evaluations, contrast measurements, and evaluation of the appearance of imaging abnormalities.

Materials and Methods

Thirteen subjects with avascular necrosis treated using advanced core decompression underwent MRI at both 7 T and 3 T. Sequence parameters as well as resolution were kept identical for both field strengths. All MR images (MEDIC, DESS, PD/T2w TSE, T1w TSE, and STIR) were evaluated by two radiologists with regard to subjective image quality, soft tissue contrasts, B1 homogeneity (four-point scale, higher values indicating better image quality) and depiction of imaging abnormalities of the femoral heads (three-point scale, higher values indicating the superiority of 7 T). Contrast ratios of soft tissues were calculated and compared with subjective data.

Results

7-T imaging of the femoral joints, as well as 3-T imaging, achieved “good” to “very good” quality in all sequences. 7 T showed significantly higher soft tissue contrasts for T2w and MEDIC compared with 3 T (cartilage/fluid: 2.9 vs 2.2 and 3.6 vs 2.6), better detailed resolution for cartilage defects (PDw, T2w, T1w, MEDIC, DESS > 2.5) and better visibility of joint effusions (MEDIC 2.6; PDw/T2w 2.4; DESS 2.2). Image homogeneity compared with 3 T (3.9–4.0 for all sequences) was degraded, especially in TSE sequences at 7 T through signal variations (7 T: 2.1–2.9); to a lesser extent also GRE sequences (7 T: 2.9–3.5). Imaging findings related to untreated or treated AVN were better delineated at 3 T (≤1.8), while joint effusions (2.2–2.6) and cartilage defects (2.5–3.0) were better visualized at 7 T. STIR performed much more poorly at 7 T, generating large contrast variations (1.5).

Conclusions

7-T hip MRI showed comparable results in hip joint imaging compared with 3 T with slight advantages in contrast detail (cartilage defects) and fluid detection at 7 T when accepting image degradation medially.
Literature
1.
go back to reference Bolog N, Nanz D, Weishaupt D. Musculoskeletal MR imaging at 3.0 T: current status and future perspectives. Eur Radiol. 2006;16(6):1298–307.PubMedCrossRef Bolog N, Nanz D, Weishaupt D. Musculoskeletal MR imaging at 3.0 T: current status and future perspectives. Eur Radiol. 2006;16(6):1298–307.PubMedCrossRef
2.
go back to reference Kollia K, Maderwald S, Putzki N, Schlamann M, Theysohn JM, Kraff O, et al. First clinical study on ultra-high-field MR imaging in patients with multiple sclerosis: comparison of 1.5T and 7T. AJNR Am J Neuroradiol. 2009;30(4):699–702.PubMedCrossRef Kollia K, Maderwald S, Putzki N, Schlamann M, Theysohn JM, Kraff O, et al. First clinical study on ultra-high-field MR imaging in patients with multiple sclerosis: comparison of 1.5T and 7T. AJNR Am J Neuroradiol. 2009;30(4):699–702.PubMedCrossRef
3.
go back to reference Vaughan JT, Garwood M, Collins CM, Liu W, DelaBarre L, Adriany G, et al. 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images. Magn Reson Med. 2001;46(1):24–30.PubMedCrossRef Vaughan JT, Garwood M, Collins CM, Liu W, DelaBarre L, Adriany G, et al. 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images. Magn Reson Med. 2001;46(1):24–30.PubMedCrossRef
4.
go back to reference International Electrotechnical Commission. Medical electrical equipment—part 2–33: particular requirements for the safety of magnetic resonance equipment for medical diagnosis. 2002; 60601–2–33. International Electrotechnical Commission. Medical electrical equipment—part 2–33: particular requirements for the safety of magnetic resonance equipment for medical diagnosis. 2002; 60601–2–33.
5.
go back to reference Regatte RR, Schweitzer SME. Ultra-high-field MRI of the musculoskeletal system at 7.0T. J Magn Reson Imaging. 2007;25(2):262–9.PubMedCrossRef Regatte RR, Schweitzer SME. Ultra-high-field MRI of the musculoskeletal system at 7.0T. J Magn Reson Imaging. 2007;25(2):262–9.PubMedCrossRef
6.
go back to reference Trattnig S, Zbyn S, Schmitt B, Friedrich K, Juras V, Szomolanyi P, et al. Advanced MR methods at ultra-high field (7 Tesla) for clinical musculoskeletal applications. Eur Radiol. 2012;22(11):2338–46.PubMedCrossRef Trattnig S, Zbyn S, Schmitt B, Friedrich K, Juras V, Szomolanyi P, et al. Advanced MR methods at ultra-high field (7 Tesla) for clinical musculoskeletal applications. Eur Radiol. 2012;22(11):2338–46.PubMedCrossRef
7.
go back to reference Moser E, Stahlberg F, Ladd ME, Trattnig S. 7-T MR—from research to clinical applications? NMR Biomed. 2012;25(5):695–716.PubMedCrossRef Moser E, Stahlberg F, Ladd ME, Trattnig S. 7-T MR—from research to clinical applications? NMR Biomed. 2012;25(5):695–716.PubMedCrossRef
8.
go back to reference Franklin KM, Dale BM, Merkle EM. Improvement in B1-inhomogeneity artifacts in the abdomen at 3T MR imaging using a radiofrequency cushion. J Magn Reson Imaging. 2008;27(6):1443–7.PubMedCrossRef Franklin KM, Dale BM, Merkle EM. Improvement in B1-inhomogeneity artifacts in the abdomen at 3T MR imaging using a radiofrequency cushion. J Magn Reson Imaging. 2008;27(6):1443–7.PubMedCrossRef
9.
go back to reference Hecht EM, Lee RF, Taouli B, Sodickson DK. Perspectives on body MR imaging at ultrahigh field. Magn Reson Imaging Clin N Am. 2007;15(3):449–65, viii.PubMedCrossRef Hecht EM, Lee RF, Taouli B, Sodickson DK. Perspectives on body MR imaging at ultrahigh field. Magn Reson Imaging Clin N Am. 2007;15(3):449–65, viii.PubMedCrossRef
10.
go back to reference Winter L, Kellman P, Renz W, Grassl A, Hezel F, Thalhammer C, et al. Comparison of three multichannel transmit/receive radiofrequency coil configurations for anatomic and functional cardiac MRI at 7.0T: implications for clinical imaging. Eur Radiol. 2012;22(10):2211–20.PubMedCrossRef Winter L, Kellman P, Renz W, Grassl A, Hezel F, Thalhammer C, et al. Comparison of three multichannel transmit/receive radiofrequency coil configurations for anatomic and functional cardiac MRI at 7.0T: implications for clinical imaging. Eur Radiol. 2012;22(10):2211–20.PubMedCrossRef
11.
go back to reference Deniz CM, Brown R, Lattanzi R, Alon L, Sodickson DK, Zhu Y. Maximum efficiency radiofrequency shimming: theory and initial application for hip imaging at 7 tesla. Magn Reson Med. 2013;69(5):1379–88.PubMedCrossRef Deniz CM, Brown R, Lattanzi R, Alon L, Sodickson DK, Zhu Y. Maximum efficiency radiofrequency shimming: theory and initial application for hip imaging at 7 tesla. Magn Reson Med. 2013;69(5):1379–88.PubMedCrossRef
12.
go back to reference Snyder CJ, Delabarre L, Moeller S, Tian J, Akgun C, Van de Moortele PF, et al. Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7 T. Magn Reson Med. 2012;67(4):954–64.PubMedCentralPubMedCrossRef Snyder CJ, Delabarre L, Moeller S, Tian J, Akgun C, Van de Moortele PF, et al. Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7 T. Magn Reson Med. 2012;67(4):954–64.PubMedCentralPubMedCrossRef
13.
go back to reference Ellermann J, Goerke U, Morgan P, Ugurbil K, Tian J, Schmitter S, et al. Simultaneous bilateral hip joint imaging at 7 Tesla using fast transmit B(1) shimming methods and multichannel transmission—a feasibility study. NMR Biomed. 2012;25(10):1202–8.PubMedCentralPubMedCrossRef Ellermann J, Goerke U, Morgan P, Ugurbil K, Tian J, Schmitter S, et al. Simultaneous bilateral hip joint imaging at 7 Tesla using fast transmit B(1) shimming methods and multichannel transmission—a feasibility study. NMR Biomed. 2012;25(10):1202–8.PubMedCentralPubMedCrossRef
14.
go back to reference Theysohn JM, Kraff O, Orzada S, Theysohn N, Classen T, Landgraeber S, et al. Bilateral hip imaging at 7 Tesla using a multi-channel transmit technology: initial results presenting anatomical detail in healthy volunteers and pathological changes in patients with avascular necrosis of the femoral head. Skeletal Radiol. 2013;42(11):1555–63.PubMedCrossRef Theysohn JM, Kraff O, Orzada S, Theysohn N, Classen T, Landgraeber S, et al. Bilateral hip imaging at 7 Tesla using a multi-channel transmit technology: initial results presenting anatomical detail in healthy volunteers and pathological changes in patients with avascular necrosis of the femoral head. Skeletal Radiol. 2013;42(11):1555–63.PubMedCrossRef
15.
go back to reference Umutlu L, Kraff O, Fischer A, Kinner S, Maderwald S, Nassenstein K, et al. Seven-Tesla MRI of the female pelvis. Eur Radiol. 2013;23(9):2364–73.PubMedCrossRef Umutlu L, Kraff O, Fischer A, Kinner S, Maderwald S, Nassenstein K, et al. Seven-Tesla MRI of the female pelvis. Eur Radiol. 2013;23(9):2364–73.PubMedCrossRef
16.
go back to reference Umutlu L, Maderwald S, Kraff O, Kinner S, Schaefer LC, Wrede K, et al. New look at renal vasculature: 7 Tesla nonenhanced T1-weighted FLASH imaging. J Magn Reson Imaging. 2012;36(3):714–21.PubMedCrossRef Umutlu L, Maderwald S, Kraff O, Kinner S, Schaefer LC, Wrede K, et al. New look at renal vasculature: 7 Tesla nonenhanced T1-weighted FLASH imaging. J Magn Reson Imaging. 2012;36(3):714–21.PubMedCrossRef
17.
go back to reference Juras V, Welsch G, Bar P, Kronnerwetter C, Fujita H, Trattnig S. Comparison of 3T and 7T MRI clinical sequences for ankle imaging. Eur J Radiol. 2012;81(8):1846–50.PubMedCrossRef Juras V, Welsch G, Bar P, Kronnerwetter C, Fujita H, Trattnig S. Comparison of 3T and 7T MRI clinical sequences for ankle imaging. Eur J Radiol. 2012;81(8):1846–50.PubMedCrossRef
18.
go back to reference Nordmeyer-Massner JA, Wyss M, Andreisek G, Pruessmann KP, Hodler J. In vitro and in vivo comparison of wrist MR imaging at 3.0 and 7.0 tesla using a gradient echo sequence and identical eight-channel coil array designs. J Magn Reson Imaging. 2011;33(3):661–7.PubMedCrossRef Nordmeyer-Massner JA, Wyss M, Andreisek G, Pruessmann KP, Hodler J. In vitro and in vivo comparison of wrist MR imaging at 3.0 and 7.0 tesla using a gradient echo sequence and identical eight-channel coil array designs. J Magn Reson Imaging. 2011;33(3):661–7.PubMedCrossRef
19.
go back to reference Brown R, Storey P, Geppert C, McGorty K, Leite AP, Babb J, et al. Breast MRI at 7 Tesla with a bilateral coil and T1-weighted acquisition with robust fat suppression: image evaluation and comparison with 3 Tesla. Eur Radiol. 2013;23(11):2969–78.PubMedCrossRef Brown R, Storey P, Geppert C, McGorty K, Leite AP, Babb J, et al. Breast MRI at 7 Tesla with a bilateral coil and T1-weighted acquisition with robust fat suppression: image evaluation and comparison with 3 Tesla. Eur Radiol. 2013;23(11):2969–78.PubMedCrossRef
20.
go back to reference Deligianni X, Bar P, Scheffler K, Trattnig S, Bieri O. High-resolution Fourier-encoded sub-millisecond echo time musculoskeletal imaging at 3 Tesla and 7 Tesla. Magn Reson Med. 2013;70(5):1434–9.PubMedCrossRef Deligianni X, Bar P, Scheffler K, Trattnig S, Bieri O. High-resolution Fourier-encoded sub-millisecond echo time musculoskeletal imaging at 3 Tesla and 7 Tesla. Magn Reson Med. 2013;70(5):1434–9.PubMedCrossRef
21.
go back to reference Suttie JJ, Delabarre L, Pitcher A, van de Moortele PF, Dass S, Snyder CJ, et al. 7 Tesla (T) human cardiovascular magnetic resonance imaging using FLASH and SSFP to assess cardiac function: validation against 1.5 T and 3 T. NMR Biomed. 2012;25(1):27–34.PubMedCentralPubMedCrossRef Suttie JJ, Delabarre L, Pitcher A, van de Moortele PF, Dass S, Snyder CJ, et al. 7 Tesla (T) human cardiovascular magnetic resonance imaging using FLASH and SSFP to assess cardiac function: validation against 1.5 T and 3 T. NMR Biomed. 2012;25(1):27–34.PubMedCentralPubMedCrossRef
22.
go back to reference Ficat RP. Idiopathic bone necrosis of the femoral head. Early diagnosis and treatment. J Bone Joint Surg Br. 1985;67(1):3–9.PubMed Ficat RP. Idiopathic bone necrosis of the femoral head. Early diagnosis and treatment. J Bone Joint Surg Br. 1985;67(1):3–9.PubMed
23.
go back to reference Arnoldi CC, Lemperg K, Linderholm H. Intraosseous hypertension and pain in the knee. J Bone Joint Surg Br. 1975;57(3):360–3.PubMed Arnoldi CC, Lemperg K, Linderholm H. Intraosseous hypertension and pain in the knee. J Bone Joint Surg Br. 1975;57(3):360–3.PubMed
24.
go back to reference Saito S, Ohzono K, Ono K. Early arteriopathy and postulated pathogenesis of osteonecrosis of the femoral head. The intracapital arterioles. Clin Orthop Relat Res. 1992;277:98–110.PubMed Saito S, Ohzono K, Ono K. Early arteriopathy and postulated pathogenesis of osteonecrosis of the femoral head. The intracapital arterioles. Clin Orthop Relat Res. 1992;277:98–110.PubMed
25.
go back to reference Lavernia CJ, Sierra RJ, Grieco FR. Osteonecrosis of the femoral head. J Am Acad Orthop Surg. 1999;7(4):250–61.PubMed Lavernia CJ, Sierra RJ, Grieco FR. Osteonecrosis of the femoral head. J Am Acad Orthop Surg. 1999;7(4):250–61.PubMed
26.
go back to reference Plancher KD, Razi A. Management of osteonecrosis of the femoral head. Orthop Clin North Am. 1997;28(3):461–77.PubMedCrossRef Plancher KD, Razi A. Management of osteonecrosis of the femoral head. Orthop Clin North Am. 1997;28(3):461–77.PubMedCrossRef
27.
go back to reference Mont MA, Hungerford DS. Non-traumatic avascular necrosis of the femoral head. J Bone Joint Surg Am. 1995;77(3):459–74.PubMed Mont MA, Hungerford DS. Non-traumatic avascular necrosis of the femoral head. J Bone Joint Surg Am. 1995;77(3):459–74.PubMed
28.
go back to reference Kyung TW, Lee JE, Phan TV, Yu R, Choi HS. Osteoclastogenesis by bone marrow-derived macrophages is enhanced in obese mice. J Nutr. 2009;139(3):502–6.PubMedCrossRef Kyung TW, Lee JE, Phan TV, Yu R, Choi HS. Osteoclastogenesis by bone marrow-derived macrophages is enhanced in obese mice. J Nutr. 2009;139(3):502–6.PubMedCrossRef
29.
go back to reference Rajagopal M, Balch Samora J, Ellis TJ. Efficacy of core decompression as treatment for osteonecrosis of the hip: a systematic review. Hip Int. 2012;22(5):489–93.PubMedCrossRef Rajagopal M, Balch Samora J, Ellis TJ. Efficacy of core decompression as treatment for osteonecrosis of the hip: a systematic review. Hip Int. 2012;22(5):489–93.PubMedCrossRef
30.
go back to reference Landgraeber S, Theysohn JM, Classen T, Jager M, Warwas S, Hohn HP, et al. Advanced core decompression, a new treatment option of avascular necrosis of the femoral head - a first follow-up. J Tissue Eng Regen Med. 2013;7(11):893–900.PubMedCrossRef Landgraeber S, Theysohn JM, Classen T, Jager M, Warwas S, Hohn HP, et al. Advanced core decompression, a new treatment option of avascular necrosis of the femoral head - a first follow-up. J Tissue Eng Regen Med. 2013;7(11):893–900.PubMedCrossRef
31.
go back to reference Yeh LR, Chen CK, Huang YL, Pan HB, Yang CF. Diagnostic performance of MR imaging in the assessment of subchondral fractures in avascular necrosis of the femoral head. Skeletal Radiol. 2009;38(6):559–64.PubMedCrossRef Yeh LR, Chen CK, Huang YL, Pan HB, Yang CF. Diagnostic performance of MR imaging in the assessment of subchondral fractures in avascular necrosis of the femoral head. Skeletal Radiol. 2009;38(6):559–64.PubMedCrossRef
32.
go back to reference Ikemura S, Yamamoto T, Motomura G, Nakashima Y, Mawatari T, Iwamoto Y. MRI evaluation of collapsed femoral heads in patients 60 years old or older: Differentiation of subchondral insufficiency fracture from osteonecrosis of the femoral head. AJR Am J Roentgenol. 2010;195(1):W63–8.PubMedCrossRef Ikemura S, Yamamoto T, Motomura G, Nakashima Y, Mawatari T, Iwamoto Y. MRI evaluation of collapsed femoral heads in patients 60 years old or older: Differentiation of subchondral insufficiency fracture from osteonecrosis of the femoral head. AJR Am J Roentgenol. 2010;195(1):W63–8.PubMedCrossRef
33.
go back to reference Iwata T, Nozawa S, Dohjima T, Yamamoto T, Ishimaru D, Tsugita M, et al. The value of T1-weighted coronal MRI scans in diagnosing occult fracture of the hip. J Bone Joint Surg Br. 2012;94(7):969–73.PubMedCrossRef Iwata T, Nozawa S, Dohjima T, Yamamoto T, Ishimaru D, Tsugita M, et al. The value of T1-weighted coronal MRI scans in diagnosing occult fracture of the hip. J Bone Joint Surg Br. 2012;94(7):969–73.PubMedCrossRef
34.
go back to reference Bluemke DA, Zerhouni EA. MRI of avascular necrosis of bone. Top Magn Reson Imaging. 1996;8(4):231–46.PubMedCrossRef Bluemke DA, Zerhouni EA. MRI of avascular necrosis of bone. Top Magn Reson Imaging. 1996;8(4):231–46.PubMedCrossRef
35.
36.
go back to reference Kramer J, Breitenseher M, Imhof H, Urban M, Plenk Jr H, Hofmann S. Diagnostic imaging in femur head necrosis. Orthopade. 2000;29(5):380–8.PubMed Kramer J, Breitenseher M, Imhof H, Urban M, Plenk Jr H, Hofmann S. Diagnostic imaging in femur head necrosis. Orthopade. 2000;29(5):380–8.PubMed
38.
go back to reference Orzada S, Quick HH, Ladd ME, Bahr A, Bolz T, Yazdanbakhsh P, et al. A flexible 8-channel transmit/receive body coil for 7 T human imaging. 17th Scientific Meeting ISMRM; 2009; Hawaii, USA; 2009. p. 2999. Orzada S, Quick HH, Ladd ME, Bahr A, Bolz T, Yazdanbakhsh P, et al. A flexible 8-channel transmit/receive body coil for 7 T human imaging. 17th Scientific Meeting ISMRM; 2009; Hawaii, USA; 2009. p. 2999.
39.
go back to reference Kraff O, Theysohn JM, Maderwald S, Saylor C, Ladd SC, Ladd ME, et al. MRI of the knee at 7.0 Tesla. Rofo. 2007;179(12):1231–5.PubMedCrossRef Kraff O, Theysohn JM, Maderwald S, Saylor C, Ladd SC, Ladd ME, et al. MRI of the knee at 7.0 Tesla. Rofo. 2007;179(12):1231–5.PubMedCrossRef
40.
41.
go back to reference Dietrich O, Reiser MF, Schoenberg SO. Artifacts in 3-T MRI: physical background and reduction strategies. Eur J Radiol. 2008;65(1):29–35.PubMedCrossRef Dietrich O, Reiser MF, Schoenberg SO. Artifacts in 3-T MRI: physical background and reduction strategies. Eur J Radiol. 2008;65(1):29–35.PubMedCrossRef
42.
go back to reference Saekho S, Boada FE, Noll DC, Stenger VA. Small tip angle three-dimensional tailored radiofrequency slab-select pulse for reduced B1 inhomogeneity at 3 T. Magn Reson Med. 2005;53(2):479–84.PubMedCrossRef Saekho S, Boada FE, Noll DC, Stenger VA. Small tip angle three-dimensional tailored radiofrequency slab-select pulse for reduced B1 inhomogeneity at 3 T. Magn Reson Med. 2005;53(2):479–84.PubMedCrossRef
43.
go back to reference Edelman RR. MR imaging of the pancreas: 1.5T versus 3T. Magn Reson Imaging Clin N Am. 2007;15(3):349–53.PubMedCrossRef Edelman RR. MR imaging of the pancreas: 1.5T versus 3T. Magn Reson Imaging Clin N Am. 2007;15(3):349–53.PubMedCrossRef
Metadata
Title
Hip imaging of avascular necrosis at 7 Tesla compared with 3 Tesla
Authors
J. M. Theysohn
O. Kraff
N. Theysohn
S. Orzada
S. Landgraeber
M. E. Ladd
T. C. Lauenstein
Publication date
01-05-2014
Publisher
Springer Berlin Heidelberg
Published in
Skeletal Radiology / Issue 5/2014
Print ISSN: 0364-2348
Electronic ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-014-1818-5

Other articles of this Issue 5/2014

Skeletal Radiology 5/2014 Go to the issue