Skip to main content
Top
Published in: Clinical Neuroradiology 3/2014

01-09-2014 | Original Aricle

Flat Panel Detector Computed Tomography for the Interaction Between Contrast-Enhanced Thrombi and Stent Retrievers in Stroke Therapy: A Pilot Study

Authors: K. J. Wenger, cand med, J. Berkefeld, M. Wagner

Published in: Clinical Neuroradiology | Issue 3/2014

Login to get access

Abstract

Objectives

The purpose of this study was to create virtual 3D views of the interaction between contrast-enhanced thrombi and three different types of stent retrievers.

Materials and Methods

Artificial thrombi with and without contrast agent were created and introduced into a silicone tube with saline solution. The stent retrievers (Aperio (A), Solitaire FR (B), Revive (C)) were released around the thrombi. For each retriever, two sets (0 min, 5 min) of flat panel computed tomography (CT) data were acquired on a Siemens Axiom Artis Zee biplane angiography system and reconstructed using syngo InSpace 3D software.

Results

The filaments of all three stent retrievers were displaced by the thrombus immediately after deployment. Another study series after 5 min showed further expansion of the stents into the thrombus. A tends to appose best around obstacles. A and B allow cushion-like bulging of clot material into the stent lumen. C showed only moderate curvilinear bulging and a minor degree of interaction in between stent filaments and thrombus.

Conclusions

Flexible design allows better apposition of the device to the surface of the thrombus. A relatively strong longitudinal structure combined with large gaps in between the stent filaments seems to favor migration into the clot. The main capture mechanism seems to be engagement of the clot between the crossings of stent struts.
Literature
1.
go back to reference Rohde S, Bösel J, Hacke W, Bendszus M. Stent retriever technology: concept, application and initial results. J Neurointerv Surg. 2011 Nov 23;4(6)455–8.PubMedCrossRef Rohde S, Bösel J, Hacke W, Bendszus M. Stent retriever technology: concept, application and initial results. J Neurointerv Surg. 2011 Nov 23;4(6)455–8.PubMedCrossRef
2.
go back to reference Mordasini P, Frabetti N, Gralla J, Schroth G, Fischer U, Arnold M et al. In vivo evaluation of the first dedicated combined flow-restoration and mechanical thrombectomy device in a swine model of acute vessel occlusion. AJNR Am J Neuroradiol. 2011 Feb;32(2):294–300.PubMedCrossRef Mordasini P, Frabetti N, Gralla J, Schroth G, Fischer U, Arnold M et al. In vivo evaluation of the first dedicated combined flow-restoration and mechanical thrombectomy device in a swine model of acute vessel occlusion. AJNR Am J Neuroradiol. 2011 Feb;32(2):294–300.PubMedCrossRef
3.
go back to reference Roth C, Papanagiotou P, Behnke S, Walter S, Haass A, Becker C et al. Stent-assisted mechanical recanalization for treatment of acute intracerebral artery occlusions. Stroke. 2010 Nov;41(11):2559–67.PubMedCrossRef Roth C, Papanagiotou P, Behnke S, Walter S, Haass A, Becker C et al. Stent-assisted mechanical recanalization for treatment of acute intracerebral artery occlusions. Stroke. 2010 Nov;41(11):2559–67.PubMedCrossRef
4.
go back to reference Mordasini P, Brekenfeld C, Byrne JV, Fischer U, Arnold M, Jung S et al. Experimental evaluation of immediate recanalization effect and recanalization efficacy of a new thrombus retriever for acute stroke treatment in vivo. AJNR Am J Neuroradiol. 2013 Jan;34(1):153–8.PubMedCrossRef Mordasini P, Brekenfeld C, Byrne JV, Fischer U, Arnold M, Jung S et al. Experimental evaluation of immediate recanalization effect and recanalization efficacy of a new thrombus retriever for acute stroke treatment in vivo. AJNR Am J Neuroradiol. 2013 Jan;34(1):153–8.PubMedCrossRef
5.
go back to reference Chandler AB. In vitro thrombotic coagulation of the blood; a method for producing a thrombus. Lab Invest. 1958 Mar–Apr;7(2):110–4.PubMed Chandler AB. In vitro thrombotic coagulation of the blood; a method for producing a thrombus. Lab Invest. 1958 Mar–Apr;7(2):110–4.PubMed
6.
go back to reference Wenger K, Nagl F, Wagner M, Berkefeld J. Improvement of stent retriever design and efficacy of mechanical thrombectomy in a flow model. Cardiovasc Intervent Radiol. 2013 Feb;36(1):192–7.PubMedCrossRef Wenger K, Nagl F, Wagner M, Berkefeld J. Improvement of stent retriever design and efficacy of mechanical thrombectomy in a flow model. Cardiovasc Intervent Radiol. 2013 Feb;36(1):192–7.PubMedCrossRef
Metadata
Title
Flat Panel Detector Computed Tomography for the Interaction Between Contrast-Enhanced Thrombi and Stent Retrievers in Stroke Therapy: A Pilot Study
Authors
K. J. Wenger, cand med
J. Berkefeld
M. Wagner
Publication date
01-09-2014
Publisher
Springer Berlin Heidelberg
Published in
Clinical Neuroradiology / Issue 3/2014
Print ISSN: 1869-1439
Electronic ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-013-0246-6

Other articles of this Issue 3/2014

Clinical Neuroradiology 3/2014 Go to the issue