Efficacy of automatic mode switching in DDDR mode pacemakers: The most 2 study

Authors: Maurizio Santomauro, Luca Ottaviano, Alessio Borrelli, Gennaro Galasso, Carlo Duilio, Nicola Monteforte, Luigi Padeletti, Annibale Sandro Montenero, Peter Andrew, Massimo Chiariello

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 1/2008

Login to get access

Abstract

Background

Effective automatic mode switching (AMS) algorithms capable of detecting a range of supraventricular tachyarrhythmias is important given evidence of atrial fibrillation (AF), atrial flutter (AFL), and atrial tachycardia (AT) post-implantation of pacemakers.

Objectives

The aim of the study was to assess the efficacy, defined as ability to detect a specific atrial rate and activate AMS, of five different AMS mechanisms during simulation of AF, AFL, and AT.

Materials and methods

A total of 48 subjects (35 men, 13 women; mean age: 69 ± 8 years) implanted with DDDR pacemakers utilizing five different AMS mechanisms (mean atrial rate, rate cut-off, complex ‘fallback’ algorithm, retriggerable atrial refractory period, and physiological band ‘beat-to-beat’) were tested using an external electronic device that simulated the occurrence of supraventricular tachyarrhythmias. AF, AFL, and AT were simulated by delivering low voltage pulse trains at 350, 250 and 160 beats/min, respectively.

Results

Mean efficacy for all AMS mechanisms was 81% [range: 57% to 100%] at 350 beats/min, 81% [range: 57–100%] at 250 beats/min, and 79% [range: 57–100%] at 160 beats/min. The AMS mechanisms that yielded 100% efficacy were the rate cut-off and physiological band ‘beat-to-beat.’

Conclusion

Not all AMS algorithms are equally efficacious at detecting atrial arrhythmias and subsequently activating AMS. Our results suggest that the most efficacious AMS algorithms are those that use rate cut-off and physiological band ‘beat-to-beat’ to detect supraventricular tachyarrhythmias.
Literature
1.
go back to reference Kamalvand, K., Tan, K., Kotsakis, A., Bucknall, C., & Sulke, N. (1997). Is mode switching beneficial? A randomized study in patients with paroxysmal atrial tachyarrhythmias. Journal of the American College of Cardiology, 30(2), 496–504.PubMedCrossRef Kamalvand, K., Tan, K., Kotsakis, A., Bucknall, C., & Sulke, N. (1997). Is mode switching beneficial? A randomized study in patients with paroxysmal atrial tachyarrhythmias. Journal of the American College of Cardiology, 30(2), 496–504.PubMedCrossRef
2.
go back to reference Brignole, M., Gianfranchi, L., Menozzi, C., Alboni, P., Musso, G., Bongiorni, M. G., et al. (1997). Assessment of atrioventricular junction ablation and DDDR mode-switching pacemaker versus pharmacological treatment in patients with severely symptomatic paroxysmal atrial fibrillation: A randomized controlled study. Circulation, 96(8), 2617–2624.PubMed Brignole, M., Gianfranchi, L., Menozzi, C., Alboni, P., Musso, G., Bongiorni, M. G., et al. (1997). Assessment of atrioventricular junction ablation and DDDR mode-switching pacemaker versus pharmacological treatment in patients with severely symptomatic paroxysmal atrial fibrillation: A randomized controlled study. Circulation, 96(8), 2617–2624.PubMed
3.
go back to reference Marshall, H. J., Harris, Z. I., Griffith, M. J., Holder, R. L., & Gammage, M. D. (1999). Prospective randomized study of ablation and pacing versus medical therapy for paroxysmal atrial fibrillation: Effects of pacing mode and mode-switch algorithm. Circulation, 99(12), 1587–1592.PubMed Marshall, H. J., Harris, Z. I., Griffith, M. J., Holder, R. L., & Gammage, M. D. (1999). Prospective randomized study of ablation and pacing versus medical therapy for paroxysmal atrial fibrillation: Effects of pacing mode and mode-switch algorithm. Circulation, 99(12), 1587–1592.PubMed
4.
go back to reference Garrigue, S., Cazeau, S., Ritter, P., Lazarus, A., Gras, D., & Mugica, J. (1996). Incidence of atrial arrhythmia in patients with long term dual-chamber pacemakers. Contribution of the Holter function of pacemakers. Archives des maladies du coeur et des vaisseaux, 89(7), 873–881.PubMed Garrigue, S., Cazeau, S., Ritter, P., Lazarus, A., Gras, D., & Mugica, J. (1996). Incidence of atrial arrhythmia in patients with long term dual-chamber pacemakers. Contribution of the Holter function of pacemakers. Archives des maladies du coeur et des vaisseaux, 89(7), 873–881.PubMed
5.
go back to reference Defaye, P., Dournaux, F., & Mouton, E. (1998). Prevalence of supraventricular arrhythmias from the automated analysis of data stored in the DDD pacemakers of 617 patients: The AIDA study. The AIDA Multicenter Study Group. Automatic interpretation for diagnosis assistance. Pacing and Clinical Electrophysiology, 21(1 Pt 2), 250–255.PubMedCrossRef Defaye, P., Dournaux, F., & Mouton, E. (1998). Prevalence of supraventricular arrhythmias from the automated analysis of data stored in the DDD pacemakers of 617 patients: The AIDA study. The AIDA Multicenter Study Group. Automatic interpretation for diagnosis assistance. Pacing and Clinical Electrophysiology, 21(1 Pt 2), 250–255.PubMedCrossRef
6.
go back to reference Stabile, G., De Simone, A., & Romano, E. (2005). Automatic mode switching in atrial fibrillation. Indian Pacing and Electrophysiology Journal, 5(3), 186–196.PubMed Stabile, G., De Simone, A., & Romano, E. (2005). Automatic mode switching in atrial fibrillation. Indian Pacing and Electrophysiology Journal, 5(3), 186–196.PubMed
7.
go back to reference Lau, C. P., Leung, S. K., Tse, H. F., & Barold, S. S. (2002). Automatic mode switching of implantable pacemakers: I. Principles of instrumentation, clinical, and hemodynamic considerations. Pacing and Clinical Electrophysiology, 25(6):967–983.PubMedCrossRef Lau, C. P., Leung, S. K., Tse, H. F., & Barold, S. S. (2002). Automatic mode switching of implantable pacemakers: I. Principles of instrumentation, clinical, and hemodynamic considerations. Pacing and Clinical Electrophysiology, 25(6):967–983.PubMedCrossRef
8.
go back to reference Ishikawa, T., Sumita, S., Kikuchi, M., Nakagawa, T., Kosuge, M., Kuji, N., et al. (2000). Incidence of atrial flutter and atrial fibrillation in patients with implanted physiological pacemakers. Japanese Circulation Journal, 64(7), 505–509.PubMedCrossRef Ishikawa, T., Sumita, S., Kikuchi, M., Nakagawa, T., Kosuge, M., Kuji, N., et al. (2000). Incidence of atrial flutter and atrial fibrillation in patients with implanted physiological pacemakers. Japanese Circulation Journal, 64(7), 505–509.PubMedCrossRef
9.
go back to reference Cheung, J. W., Keating, R. J., Stein, K. M., Markowitz, S. M., Iwai, S., Shah, B. K., et al. (2006). Newly detected atrial fibrillation following dual chamber pacemaker implantation. Journal of Cardiovascular Electrophysiology, 17(12), 1323–1328.PubMedCrossRef Cheung, J. W., Keating, R. J., Stein, K. M., Markowitz, S. M., Iwai, S., Shah, B. K., et al. (2006). Newly detected atrial fibrillation following dual chamber pacemaker implantation. Journal of Cardiovascular Electrophysiology, 17(12), 1323–1328.PubMedCrossRef
10.
go back to reference Padeletti, L., Gasparini, M., Porciani, M. C., Pieragnoli, P., Colella, A., Michelucci, A., et al. (2002). How to test mode switching in pacemakers implanted in patients: the MOST study. Pacing and Clinical Electrophysiology, 25(2), 156–160.PubMedCrossRef Padeletti, L., Gasparini, M., Porciani, M. C., Pieragnoli, P., Colella, A., Michelucci, A., et al. (2002). How to test mode switching in pacemakers implanted in patients: the MOST study. Pacing and Clinical Electrophysiology, 25(2), 156–160.PubMedCrossRef
Metadata
Title
Efficacy of automatic mode switching in DDDR mode pacemakers: The most 2 study
Authors
Maurizio Santomauro
Luca Ottaviano
Alessio Borrelli
Gennaro Galasso
Carlo Duilio
Nicola Monteforte
Luigi Padeletti
Annibale Sandro Montenero
Peter Andrew
Massimo Chiariello
Publication date
01-01-2008
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2008
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-007-9173-2

Other articles of this Issue 1/2008

Journal of Interventional Cardiac Electrophysiology 1/2008 Go to the issue