Published in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 5/2018

01-09-2018 | Original Article

Effects of reducing light-curing time of a high-power LED device on shear bond strength of brackets

Authors: Layene Figueiredo Almeida, DDS, MS, Prof. Lidia Parsekian Martins, DDS, MS, PhD, Prof. Dr. Renato Parsekian Martins, DDS, MS, PhD

Published in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie | Issue 5/2018

Login to get access

Abstract

Purpose

To assess the effects of reducing the curing time of a high-power light-emitting diode (LED) unit (Valo, Ultradent, South Jordan, UT, USA) on shear bond strength (SBS) of metal brackets and on the amount of adhesive remnant of two orthodontic composites.

Methods

Eighty human premolars were divided into four groups (G1–4) according to curing time and composite: G1 (Transbond XT, 6 s), G2 (Opal Bond MV, 6 s), G3 (Transbond XT, 3 s), and G4 (Opal Bond MV, 3 s). Twenty-four hours after bonding, brackets were subject to a SBS test performed with a universal testing machine. Enamel surface was analyzed by SEM and the amount of adhesive remnant was assessed by the Image J software area calculation tool. Two-way analysis of variance was used for statistical analysis of SBS data, while Friedman and Mann–Whitney post hoc tests were used to analyze data on the amount of adhesive remnant.

Results

Time and composite significantly affected SBS (p < 0.001). The 6 s curing showed a higher SBS value (21.56 MPa) in comparison to 3 s curing (15.79 MPa). Transbond XT composite showed a significantly higher SBS value (21.06 MPa) compared to Opal Bond MV (16.29 MPa). After the SBS test, Opal Bond MV showed a significantly greater amount of composite adhered to enamel (p < 0.001).

Conclusion

Reducing exposure time from 6 to 3 s significantly decreased mean values of SBS, even with the use of a high-power LED unit. Reduction in time did not affect the amount of adhesive remnant.
Literature
1.
go back to reference Amato PAF, Martins RP, dos Santos Cruz CA et al (2014) Time reduction of light curing: Influence on conversion degree and microhardness of orthodontic composites. Am J Orthod Dentofacial Orthop 146(1):40–46CrossRefPubMed Amato PAF, Martins RP, dos Santos Cruz CA et al (2014) Time reduction of light curing: Influence on conversion degree and microhardness of orthodontic composites. Am J Orthod Dentofacial Orthop 146(1):40–46CrossRefPubMed
2.
go back to reference Artun J, Bergland S (1984) Clinical trials with crystal growth conditioning as an alternative to acid etch enamel pretreatment. Am J Orthod 85(4):333–340CrossRefPubMed Artun J, Bergland S (1984) Clinical trials with crystal growth conditioning as an alternative to acid etch enamel pretreatment. Am J Orthod 85(4):333–340CrossRefPubMed
3.
go back to reference Bishara SE (2003) Ceramic brackets: a clinical perspective. World J Orthod 4(1):61–66 Bishara SE (2003) Ceramic brackets: a clinical perspective. World J Orthod 4(1):61–66
4.
go back to reference Bishara SE, Fonseca JM, Fehr DE et al (1994) Debonding forces applied to ceramic brackets simulating clinical conditions. Angle Orthod 64(4):277–282PubMed Bishara SE, Fonseca JM, Fehr DE et al (1994) Debonding forces applied to ceramic brackets simulating clinical conditions. Angle Orthod 64(4):277–282PubMed
5.
go back to reference Bland J, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310CrossRefPubMed Bland J, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310CrossRefPubMed
6.
go back to reference Blankenau R, Erickson RL, Rueggeberg F (1999) New light curing options for composite restorations. Compend Contin Educ Dent 20(2):122–125PubMed Blankenau R, Erickson RL, Rueggeberg F (1999) New light curing options for composite restorations. Compend Contin Educ Dent 20(2):122–125PubMed
7.
go back to reference Bowen RL, Rodriguez MS (1962) Tensile strength and modulus of elasticity of tooth structure and several restorative materials. J Am Dent Assoc 64:378–387CrossRefPubMed Bowen RL, Rodriguez MS (1962) Tensile strength and modulus of elasticity of tooth structure and several restorative materials. J Am Dent Assoc 64:378–387CrossRefPubMed
9.
go back to reference Cerekja E, Cakirer B (2011) Effect of short curing times with a high-intensity light-emitting diode or high-power halogen on shear bond strength of metal brackets before and after thermocycling. Angle Orthod 81(3):510–516CrossRefPubMed Cerekja E, Cakirer B (2011) Effect of short curing times with a high-intensity light-emitting diode or high-power halogen on shear bond strength of metal brackets before and after thermocycling. Angle Orthod 81(3):510–516CrossRefPubMed
10.
go back to reference Dall’Igna CM, Marchioro EM, Spohr AM et al (2011) Effect of curing time on the bond strength of a bracket-bonding system cured with a light-emitting diode or plasma arc light. Eur J Orthod 33(1):55–59CrossRefPubMed Dall’Igna CM, Marchioro EM, Spohr AM et al (2011) Effect of curing time on the bond strength of a bracket-bonding system cured with a light-emitting diode or plasma arc light. Eur J Orthod 33(1):55–59CrossRefPubMed
11.
go back to reference De Munck J, Van Landuyt K, Coutinho E et al (2005) Micro-tensile bond strength of adhesives bonded to class-I cavity-bottom dentin after thermo-cycling. Dent Mater 21(11):999–1007CrossRefPubMed De Munck J, Van Landuyt K, Coutinho E et al (2005) Micro-tensile bond strength of adhesives bonded to class-I cavity-bottom dentin after thermo-cycling. Dent Mater 21(11):999–1007CrossRefPubMed
12.
go back to reference De Munck J, Van Landuyt K, Peumans M et al (2005) A critical review of the durability of adhesion of tooth tissue: methods and results. J Dent Res 84(2):118–132CrossRefPubMed De Munck J, Van Landuyt K, Peumans M et al (2005) A critical review of the durability of adhesion of tooth tissue: methods and results. J Dent Res 84(2):118–132CrossRefPubMed
13.
go back to reference Donatelli RE, Lee SJ (2013) How to report reliability in orthodontic research: part 1. Am J Orthod Dentofacial Orthop 144(1):156–161CrossRefPubMed Donatelli RE, Lee SJ (2013) How to report reliability in orthodontic research: part 1. Am J Orthod Dentofacial Orthop 144(1):156–161CrossRefPubMed
14.
go back to reference Eliades T, Eliades G, Brantley WA et al (1995) Residual monomer leaching from chemically cured and visible light-cured orthodontics adhesives. Am J Orthod Dentofacial Orthop 108(3):316–321CrossRefPubMed Eliades T, Eliades G, Brantley WA et al (1995) Residual monomer leaching from chemically cured and visible light-cured orthodontics adhesives. Am J Orthod Dentofacial Orthop 108(3):316–321CrossRefPubMed
15.
go back to reference Ferracane J, Greener E (1984) The effect of composite formulation on the degree of conversion and mechanical properties of dental restorative composites. J Biomed Mater Res 20(1):121–131CrossRef Ferracane J, Greener E (1984) The effect of composite formulation on the degree of conversion and mechanical properties of dental restorative composites. J Biomed Mater Res 20(1):121–131CrossRef
16.
go back to reference Finnema KJ, Ozcan M, Post WJ et al (2010) In-vitro orthodontic bond strength testing: a systematic review and meta-analysis. Am J Orthod Dentofacial Orthop 137(5):615–622CrossRefPubMed Finnema KJ, Ozcan M, Post WJ et al (2010) In-vitro orthodontic bond strength testing: a systematic review and meta-analysis. Am J Orthod Dentofacial Orthop 137(5):615–622CrossRefPubMed
17.
18.
go back to reference Lamper T, Steinhäuser-Andresen S, Huth KC et al (2012) Does a reduction of polymerization time and bonding steps affect the bond strength of brackets? Clin Oral Investig 16(2):665–671CrossRefPubMed Lamper T, Steinhäuser-Andresen S, Huth KC et al (2012) Does a reduction of polymerization time and bonding steps affect the bond strength of brackets? Clin Oral Investig 16(2):665–671CrossRefPubMed
19.
go back to reference Mavropoulos A, Staudt CB, Kiliaridis S et al (2005) Light curing time reduction. Eur J Orthod 27(4):408–412CrossRefPubMed Mavropoulos A, Staudt CB, Kiliaridis S et al (2005) Light curing time reduction. Eur J Orthod 27(4):408–412CrossRefPubMed
20.
go back to reference Ogaard B, Fjeld M (2010) The enamel surface and bonding in orthodontics. Semin Orthod 16(1):37–48CrossRef Ogaard B, Fjeld M (2010) The enamel surface and bonding in orthodontics. Semin Orthod 16(1):37–48CrossRef
21.
go back to reference Retief DH (1974) Failure at the dental adhesive-etched enamel interface. J Oral Rehabil 1(13):265–284CrossRefPubMed Retief DH (1974) Failure at the dental adhesive-etched enamel interface. J Oral Rehabil 1(13):265–284CrossRefPubMed
22.
go back to reference Reynolds JR (1975) A review of direct orthodontic bonding. Br J Orthod 2(3):171–178CrossRef Reynolds JR (1975) A review of direct orthodontic bonding. Br J Orthod 2(3):171–178CrossRef
24.
go back to reference Rueggeberg F (2011) State-of-the-art: dental photocuring—a review. Dent Mater J 27(1):39–52CrossRef Rueggeberg F (2011) State-of-the-art: dental photocuring—a review. Dent Mater J 27(1):39–52CrossRef
25.
go back to reference Rueggeberg F, Jordan D (1993) Effect of light-tip distance on polymerization of composite. Int J Prosthodont 6(4):364–370PubMed Rueggeberg F, Jordan D (1993) Effect of light-tip distance on polymerization of composite. Int J Prosthodont 6(4):364–370PubMed
26.
go back to reference Sokucu O, Siso SH, Ozturk F, Nalcaci R (2010) Shear bond strength of orthodontic brackets cured with different light sources under thermocycling. Eur J Dent 4(3):257–262PubMedPubMedCentral Sokucu O, Siso SH, Ozturk F, Nalcaci R (2010) Shear bond strength of orthodontic brackets cured with different light sources under thermocycling. Eur J Dent 4(3):257–262PubMedPubMedCentral
27.
go back to reference Swanson T, Dunn WJ, Childers DE et al (2004) Shear bond strength of orthodontic brackets bonded with light-emitting diode curing units at various polymerization times. Am J Orthod Dentofacial Orthop 125(3):337–341CrossRefPubMed Swanson T, Dunn WJ, Childers DE et al (2004) Shear bond strength of orthodontic brackets bonded with light-emitting diode curing units at various polymerization times. Am J Orthod Dentofacial Orthop 125(3):337–341CrossRefPubMed
29.
go back to reference Thind BS, Stirrups DR, Lloyd CH (2006) A comparison of tungsten-quartz-halogen, plasma arc and light-emitting diode light sources for the polymerization of an orthodontic adhesive. Eur J Orthod 28(1):78–82CrossRefPubMed Thind BS, Stirrups DR, Lloyd CH (2006) A comparison of tungsten-quartz-halogen, plasma arc and light-emitting diode light sources for the polymerization of an orthodontic adhesive. Eur J Orthod 28(1):78–82CrossRefPubMed
30.
31.
go back to reference Yu HS, Lee KJ, Jin GC et al (2007) Comparison of the shear bond strength of brackets using the led curing light and plasma arc curing light: polymerization time. World J Orthod 8(2):129–135PubMed Yu HS, Lee KJ, Jin GC et al (2007) Comparison of the shear bond strength of brackets using the led curing light and plasma arc curing light: polymerization time. World J Orthod 8(2):129–135PubMed
32.
go back to reference Yuasa T, Iijima M, Ito S, Muguruma T, Saito T, Mizoguchi I (2010) Effects of long-term storage and thermocycling on bond strength of two self-etching primer adhesive systems. Eur J Orthod 32(3):285–290CrossRefPubMed Yuasa T, Iijima M, Ito S, Muguruma T, Saito T, Mizoguchi I (2010) Effects of long-term storage and thermocycling on bond strength of two self-etching primer adhesive systems. Eur J Orthod 32(3):285–290CrossRefPubMed
Metadata
Title
Effects of reducing light-curing time of a high-power LED device on shear bond strength of brackets
Authors
Layene Figueiredo Almeida, DDS, MS
Prof. Lidia Parsekian Martins, DDS, MS, PhD
Prof. Dr. Renato Parsekian Martins, DDS, MS, PhD
Publication date
01-09-2018
Publisher
Springer Medizin
Published in
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie / Issue 5/2018
Print ISSN: 1434-5293
Electronic ISSN: 1615-6714
DOI
https://doi.org/10.1007/s00056-018-0146-0

Other articles of this Issue 5/2018

Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 5/2018 Go to the issue

Mitteilungen DGKFO

Mitteilungen DGKFO