Published in: Journal of Orthopaedic Surgery and Research 1/2016

Open Access 01-12-2016 | Research article

Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates

Authors: Christopher Ioannou, Matthew Knight, Luca Daniele, Lee Flueckiger, Ezekiel S. L. Tan

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2016

Login to get access

Abstract

Background

The objective of this study is to analyse the effectiveness of the surgical torque limiter during operative use. The study also investigates the potential differences in torque between hand and drill-based screw insertion into locking plates using a standardised torque limiter.

Methods

Torque for both hand and power screw insertion was measured through a load cell, registering 6.66 points per second. This was performed in a controlled environment using synthetic bone, a locking plate and locking screws to simulate plate fixation. Screws were inserted by hand and by drill with torque values measured.

Results

The surgical torque limiter (1.5 Nm) was effective as the highest recorded reading in the study was 1.409 Nm. Comparatively, there is a statistically significant difference between screw insertion methods. Torque produced for manually driven screw insertion into locking plates was 1.289 Nm (95 % CI 1.269–1.308) with drill-powered screw insertion at 0.740 Nm (95 % CI 0.723–0.757).

Conclusions

The surgical torque limiter proved to be effective as per product specifications. Screws inserted under power produce significantly less torque when compared to manual insertion by hand. This is likely related to the mechanism of the torque limiter when being used at higher speeds for which it was designed. We conclude that screws may be inserted using power to the plate with the addition of a torque limiter. It is recommended that all screws inserted by drill be hand tightened to achieve adequate torque values.
Literature
1.
go back to reference Wagner M. General principles for the clinical use of the LCP. Injury. 2003;34:31–42. Wagner M. General principles for the clinical use of the LCP. Injury. 2003;34:31–42.
2.
3.
go back to reference LCP locking compression plate—combine without compromise. Synthes instruction manual. 2006. LCP locking compression plate—combine without compromise. Synthes instruction manual. 2006.
4.
go back to reference LCP distal tibia plate. Combine without compromise. Synthes instruction manual. 2015. LCP distal tibia plate. Combine without compromise. Synthes instruction manual. 2015.
5.
go back to reference Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent. 1999;81(2):159–166. Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent. 1999;81(2):159–166.
6.
go back to reference Boero Baroncelli A, et al. Effect of screw insertion torque on push-out and cantilever bending properties of five different angle-stable systems. Vet Surg. 2013;42(3):308–15.CrossRefPubMed Boero Baroncelli A, et al. Effect of screw insertion torque on push-out and cantilever bending properties of five different angle-stable systems. Vet Surg. 2013;42(3):308–15.CrossRefPubMed
Metadata
Title
Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
Authors
Christopher Ioannou
Matthew Knight
Luca Daniele
Lee Flueckiger
Ezekiel S. L. Tan
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2016
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-016-0458-y

Other articles of this Issue 1/2016

Journal of Orthopaedic Surgery and Research 1/2016 Go to the issue