Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume

Authors: Sami I Al-Majed, John E Thompson, Kenneth F Watson, Adrienne G Randolph

Published in: Critical Care | Issue 6/2004

Login to get access

Abstract

Introduction

The objective of this laboratory study was to measure the effect of decreased lung compliance and endotracheal tube (ETT) leakage on measured exhaled tidal volume at the airway and at the ventilator, in a research study with a test lung.

Methods

The subjects were infant, adult and pediatric test lungs. In the test lung model, lung compliances were set to normal and to levels seen in acute respiratory distress syndrome. Set tidal volume was 6 ml/kg across a range of simulated weights and ETT sizes. Data were recorded from both the ventilator light-emitting diode display and the CO2SMO Plus monitor display by a single observer. Effective tidal volume was calculated from a standard equation.

Results

In all test lung models, exhaled tidal volume measured at the airway decreased markedly with decreasing lung compliance, but measurement at the ventilator showed minimal change. In the absence of a simulated ETT leak, calculation of the effective tidal volume led to measurements very similar to exhaled tidal volume measured at the ETT. With a simulated ETT tube leak, the effective tidal volume markedly overestimated tidal volume measured at the airway.

Conclusion

Previous investigators have emphasized the need to measure tidal volume at the ETT for all children. When ETT leakage is minimal, it seems from our simulated lung models that calculation of effective tidal volume would give similar readings to tidal volume measured at the airway, even in small patients. Future studies of tidal volume measurement accuracy in mechanically ventilated children should control for the degree of ETT leakage.
Appendix
Available only for authorised users
Literature
1.
go back to reference Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN, Cheifetz IM: Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 2000, 162: 2109-2112.CrossRefPubMed Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN, Cheifetz IM: Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 2000, 162: 2109-2112.CrossRefPubMed
2.
go back to reference Castle RA, Dunne CJ, Mok Q, Wade AM, Stocks J: Accuracy of displayed values of tidal volume in the pediatric intensive care unit. Crit Care Med 2002, 30: 2566-2574. 10.1097/00003246-200211000-00027CrossRefPubMed Castle RA, Dunne CJ, Mok Q, Wade AM, Stocks J: Accuracy of displayed values of tidal volume in the pediatric intensive care unit. Crit Care Med 2002, 30: 2566-2574. 10.1097/00003246-200211000-00027CrossRefPubMed
3.
go back to reference Neve V, Vernox S, Forget P, Noizet O, Sadik A, Leteurtre S, Cremer R, Fourier C, Riou Y, Leclerc F: Comparison of measurement of flow, volume, and pressure at the Y piece to those displayed by the ventilator in children [abstract]. Crit Care Med 2001, 29: A142. Neve V, Vernox S, Forget P, Noizet O, Sadik A, Leteurtre S, Cremer R, Fourier C, Riou Y, Leclerc F: Comparison of measurement of flow, volume, and pressure at the Y piece to those displayed by the ventilator in children [abstract]. Crit Care Med 2001, 29: A142.
4.
go back to reference The Acute Respiratory Distress Syndrome Network: Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000, 342: 1301-1308. 10.1056/NEJM200005043421801CrossRef The Acute Respiratory Distress Syndrome Network: Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000, 342: 1301-1308. 10.1056/NEJM200005043421801CrossRef
5.
go back to reference Hess D, Kacmarek RM: Technical aspects of the patient-ventilator interface. In In Principles and Practice of Mechanical Ventilation. 1st edition. Edited by: Tobin MJ. New York: McGraw-Hill; 1994:1055-1056. Hess D, Kacmarek RM: Technical aspects of the patient-ventilator interface. In In Principles and Practice of Mechanical Ventilation. 1st edition. Edited by: Tobin MJ. New York: McGraw-Hill; 1994:1055-1056.
6.
go back to reference Kallet RH, Corral W, Silverman HJ, Luce JM: Implementation of a low tidal ventilation protocol for patients with acute lung injury or acute respiratory distress syndrome. Respir Care 2001, 45: 1024-1036. Kallet RH, Corral W, Silverman HJ, Luce JM: Implementation of a low tidal ventilation protocol for patients with acute lung injury or acute respiratory distress syndrome. Respir Care 2001, 45: 1024-1036.
7.
go back to reference Deakers TW, Reynolds G, Stretton M, Newth CJ: Cuffed endotracheal tubes in pediatric intensive care. J Pediatr 1994, 125: 57-62.CrossRefPubMed Deakers TW, Reynolds G, Stretton M, Newth CJ: Cuffed endotracheal tubes in pediatric intensive care. J Pediatr 1994, 125: 57-62.CrossRefPubMed
8.
go back to reference Main E, Castle R, Stocks J, James I, Hatch D: The influence of endotracheal tube leak on the assessment of respiratory function in ventilated children. Intensive Care Med 2001, 27: 1788-1797. 10.1007/s001340101105CrossRefPubMed Main E, Castle R, Stocks J, James I, Hatch D: The influence of endotracheal tube leak on the assessment of respiratory function in ventilated children. Intensive Care Med 2001, 27: 1788-1797. 10.1007/s001340101105CrossRefPubMed
Metadata
Title
Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
Authors
Sami I Al-Majed
John E Thompson
Kenneth F Watson
Adrienne G Randolph
Publication date
01-12-2004
Publisher
BioMed Central
Published in
Critical Care / Issue 6/2004
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc2954

Other articles of this Issue 6/2004

Critical Care 6/2004 Go to the issue