Published in: BMC Pulmonary Medicine 1/2013

Open Access 01-12-2013 | Research article

Effect of dynamic random leaks on the monitoring accuracy of home mechanical ventilators: a bench study

Authors: Ana Sogo, Jaume Montanyà, Eduard Monsó, Lluís Blanch, Xavier Pomares, Manel Lujàn

Published in: BMC Pulmonary Medicine | Issue 1/2013

Login to get access

Abstract

Background

So far, the accuracy of tidal volume (VT) and leak measures provided by the built-in software of commercial home ventilators has only been tested using bench linear models with fixed calibrated and continuous leaks. The objective was to assess the reliability of the estimation of tidal volume (VT) and unintentional leaks in a single tubing bench model which introduces random dynamic leaks during inspiratory or expiratory phases.

Methods

The built-in software of four commercial home ventilators and a fifth ventilator-independent ad hoc designed external software tool were tested with two levels of leaks and two different models with excess leaks (inspiration or expiration). The external software analyzed separately the inspiratory and expiratory unintentional leaks.

Results

In basal condition, all ventilators but one underestimated tidal volume with values ranging between -1.5 ± 3.3% to -8.7% ± 3.27%. In the model with excess of inspiratory leaks, VT was overestimated by all four commercial software tools, with values ranging from 18.27 ± 7.05% to 35.92 ± 17.7%, whereas the ventilator independent-software gave a smaller difference (3.03 ± 2.6%). Leaks were underestimated by two applications with values of -11.47 ± 6.32 and -5.9 ± 0.52 L/min. With expiratory leaks, VT was overestimated by the software of one ventilator and the ventilator-independent software and significantly underestimated by the other three, with deviations ranging from +10.94 ± 7.1 to -48 ± 23.08%. The four commercial tools tested overestimated unintentional leaks, with values between 2.19 ± 0.85 to 3.08 ± 0.43 L/min.

Conclusions

In a bench model, the presence of unintentional random leaks may be a source of error in the measurement of VT and leaks provided by the software of home ventilators. Analyzing leaks during inspiration and expiration separately may reduce this source of error.
Appendix
Available only for authorised users
Literature
1.
go back to reference Janssens JP, Borel JC, Pepin JL, SomnoNIV group: Nocturnal monitoring of home noninvasive ventilation: the contribution of simple tools such as pulse oximetry, capnography, built-in ventilator software and autonomic markers of sleep fragmentation. Thorax. 2011, 66: 438-445. 10.1136/thx.2010.139782.CrossRefPubMed Janssens JP, Borel JC, Pepin JL, SomnoNIV group: Nocturnal monitoring of home noninvasive ventilation: the contribution of simple tools such as pulse oximetry, capnography, built-in ventilator software and autonomic markers of sleep fragmentation. Thorax. 2011, 66: 438-445. 10.1136/thx.2010.139782.CrossRefPubMed
2.
go back to reference Pasquina P, Adler D, Farr P, Bourqui P, Bridevaux PO, Janssens JP: What does built-in software of home ventilators tell us? An observational study of 150 patients on home ventilation. Respiration. 2012, 83: 293-299. 10.1159/000330598.CrossRefPubMed Pasquina P, Adler D, Farr P, Bourqui P, Bridevaux PO, Janssens JP: What does built-in software of home ventilators tell us? An observational study of 150 patients on home ventilation. Respiration. 2012, 83: 293-299. 10.1159/000330598.CrossRefPubMed
3.
go back to reference Fernandez R, Rubinós G, Rodríguez F, García A, Rodríguez P, Casan P: Home mechanical ventilation through mask: monitoring leakage and nocturnal oxygenation at home. . Respiration 2013 in press. DOI: 10.1159/000341983 Fernandez R, Rubinós G, Rodríguez F, García A, Rodríguez P, Casan P: Home mechanical ventilation through mask: monitoring leakage and nocturnal oxygenation at home. . Respiration 2013 in press. DOI: 10.1159/000341983
4.
go back to reference Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC: Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation. Intensive Care Med. 2010, 36: 2074-2080. 10.1007/s00134-010-2044-9.CrossRefPubMed Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC: Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation. Intensive Care Med. 2010, 36: 2074-2080. 10.1007/s00134-010-2044-9.CrossRefPubMed
5.
go back to reference Rabec C, Georges M, Kabeya NK, Baudouin N, Massin F, Reybet-Degat O, Camus P: Evaluating non invasive ventilation using a monitoring system coupled to a ventilator: a bench to-bedside study. Eur Respir J. 2009, 34: 902-913. 10.1183/09031936.00170508.CrossRefPubMed Rabec C, Georges M, Kabeya NK, Baudouin N, Massin F, Reybet-Degat O, Camus P: Evaluating non invasive ventilation using a monitoring system coupled to a ventilator: a bench to-bedside study. Eur Respir J. 2009, 34: 902-913. 10.1183/09031936.00170508.CrossRefPubMed
6.
go back to reference Contal O, Vignaux L, Combescure C, Pepin JL, Jolliet P, Janssens JP: Monitoring of non-invasive ventilation by built-in software of home bi-level ventilators: a bench study. Chest. 2012, 141: 469-476. 10.1378/chest.11-0485.CrossRefPubMed Contal O, Vignaux L, Combescure C, Pepin JL, Jolliet P, Janssens JP: Monitoring of non-invasive ventilation by built-in software of home bi-level ventilators: a bench study. Chest. 2012, 141: 469-476. 10.1378/chest.11-0485.CrossRefPubMed
7.
go back to reference Luján M, Sogo A, Pomares X, Monsó E, Sales B, Blanch L: Effect of leak and breathing pattern on the accuracy of tidal volume estimation by commercial home ventilators: a bench study. Resp Care. 2013, 58: 770-777. 10.4187/respcare.02010.CrossRef Luján M, Sogo A, Pomares X, Monsó E, Sales B, Blanch L: Effect of leak and breathing pattern on the accuracy of tidal volume estimation by commercial home ventilators: a bench study. Resp Care. 2013, 58: 770-777. 10.4187/respcare.02010.CrossRef
8.
go back to reference Oliveri C, Costa R, Conti G, Navalesi P: Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives. Intensive Care Med. 2012, 38: 160-167. 10.1007/s00134-011-2416-9.CrossRef Oliveri C, Costa R, Conti G, Navalesi P: Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives. Intensive Care Med. 2012, 38: 160-167. 10.1007/s00134-011-2416-9.CrossRef
9.
go back to reference Blanch L, Sales B, Montanya J, Lucangelo U, Garcia-Esquirol O, Villagra A, Chacon E, Estruga A, Borelli M, Burgueño MJ, Oliva JC, Fernandez R, Villar J, Kacmarek R, Murias G: Validation of the Better Care® system to detect ineffective efforts during expiration in mechanically ventilated patients: a pilot study. Intensive Care Med. 2012, 38: 772-780. 10.1007/s00134-012-2493-4.CrossRefPubMed Blanch L, Sales B, Montanya J, Lucangelo U, Garcia-Esquirol O, Villagra A, Chacon E, Estruga A, Borelli M, Burgueño MJ, Oliva JC, Fernandez R, Villar J, Kacmarek R, Murias G: Validation of the Better Care® system to detect ineffective efforts during expiration in mechanically ventilated patients: a pilot study. Intensive Care Med. 2012, 38: 772-780. 10.1007/s00134-012-2493-4.CrossRefPubMed
10.
go back to reference Brown SP, Miller WC, Eason JM: Ergometry and Calorimetry: the measurement of work and energy. Exercise physiology: basis of human movement in health and disease. 2006, Baltimore: Lippincott Williams and Wilkins, 98-126. Brown SP, Miller WC, Eason JM: Ergometry and Calorimetry: the measurement of work and energy. Exercise physiology: basis of human movement in health and disease. 2006, Baltimore: Lippincott Williams and Wilkins, 98-126.
11.
go back to reference Prieto L, Lamarca L, Casado A, Alonso J: The evaluation of agreement on continuous variables by the intraclass correlation coefficient. J Epidemiol Community Health. 1997, 51: 579-581.CrossRefPubMedPubMedCentral Prieto L, Lamarca L, Casado A, Alonso J: The evaluation of agreement on continuous variables by the intraclass correlation coefficient. J Epidemiol Community Health. 1997, 51: 579-581.CrossRefPubMedPubMedCentral
12.
go back to reference Carteaux G, Lyazidi A, Cordoba-Izquierdo A, Vignaux L, Jolliet P, Thille AW, Richard M, Brochard L: Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study. Chest. 2012, 142: 367-376. 10.1378/chest.1389772.CrossRefPubMed Carteaux G, Lyazidi A, Cordoba-Izquierdo A, Vignaux L, Jolliet P, Thille AW, Richard M, Brochard L: Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study. Chest. 2012, 142: 367-376. 10.1378/chest.1389772.CrossRefPubMed
13.
go back to reference Battisti A, Tassaux D, Janssens JP, Michotte JB, Jaber S, Jolliet P: Performance characteristics of 10 home mechanical ventilators in pressure-support mode:a comparative bench study. Chest. 2005, 127: 1784-1792. 10.1378/chest.127.5.1784.CrossRefPubMed Battisti A, Tassaux D, Janssens JP, Michotte JB, Jaber S, Jolliet P: Performance characteristics of 10 home mechanical ventilators in pressure-support mode:a comparative bench study. Chest. 2005, 127: 1784-1792. 10.1378/chest.127.5.1784.CrossRefPubMed
14.
go back to reference Vignaux L, Tassaux D, Jolliet P: Performance of non-invasive ventilation modes on ICU ventilators during pressure support: a bench model study. Intensive Care Med. 2007, 33: 1444-1451. 10.1007/s00134-007-0713-0.CrossRefPubMed Vignaux L, Tassaux D, Jolliet P: Performance of non-invasive ventilation modes on ICU ventilators during pressure support: a bench model study. Intensive Care Med. 2007, 33: 1444-1451. 10.1007/s00134-007-0713-0.CrossRefPubMed
15.
go back to reference Rabec C, Rodenstein DO, Leger P, Rouault S, Perrin C, Gonzalez-Bermejo J, SomnoNIV group: Ventilator modes and settings during non-invasive ventilation: effects on respiratory events and implication for their identification. Thorax. 2011, 66: 170-178. 10.1136/thx.2010.142661.CrossRefPubMed Rabec C, Rodenstein DO, Leger P, Rouault S, Perrin C, Gonzalez-Bermejo J, SomnoNIV group: Ventilator modes and settings during non-invasive ventilation: effects on respiratory events and implication for their identification. Thorax. 2011, 66: 170-178. 10.1136/thx.2010.142661.CrossRefPubMed
Metadata
Title
Effect of dynamic random leaks on the monitoring accuracy of home mechanical ventilators: a bench study
Authors
Ana Sogo
Jaume Montanyà
Eduard Monsó
Lluís Blanch
Xavier Pomares
Manel Lujàn
Publication date
01-12-2013
Publisher
BioMed Central
Published in
BMC Pulmonary Medicine / Issue 1/2013
Electronic ISSN: 1471-2466
DOI
https://doi.org/10.1186/1471-2466-13-75

Other articles of this Issue 1/2013

BMC Pulmonary Medicine 1/2013 Go to the issue