Skip to main content
Top
Published in: International Orthopaedics 12/2020

01-12-2020 | Original Paper

Does patient-specific instrumentation increase the risk of notching in the anterior femoral cortex in total knee arthroplasty? A comparative prospective trial

Authors: Song Ke, Tianfei Ran, Yongqin He, Mingrui Lv, Xin Song, Yuanyuan Zhou, Yuan Xu, Min Wang

Published in: International Orthopaedics | Issue 12/2020

Login to get access

Abstract

Purpose

Patient-specific instrumentation (PSI) was usually applied in total knee arthroplasty (TKA) to acquire a favourable alignment. We hypothesized that using PSI had a potential risk of notching in the anterior femoral cortex, because the femoral component may be placed in an overextension position due to the distal femoral sagittal anteversion. The aim of this study was to figure out the relationship between the notch and the distal femoral sagittal anteversion in PSI-assisted TKA.

Methods

One hundred thirty-one patients who were to undergo total knee arthroplasty (TKA) were randomly divided into conventional instrumentation (CI) group and PSI group. The computed tomography (CT) data of lower extremities was collected and imported to the Mimics software to reconstruct the three-dimensional (3D) bone image of the femur. The angle between distal femoral anatomic axis (DFAA) and femoral mechanical axis (FMA) on sagittal plane was defined as distal femoral sagittal anteverted angle (DFSAA) and measured. The number of notch intra-operative and post-operative was recorded. Then, we calculated the incidence of the notch and analyzed its relationship with DFSAA.

Results

The average DFSAA of 262 femurs is 2.5° ± 1.5° (range, 0.0°–5.7°). When DFSAA ≥ 3°, the incidence of notch was 7.10% in CI group and 33.30% in PSI group, respectively, which shows significant statistical difference in the two groups (P = 0.016 < 0.05). When DFSAA < 3°, the incidence of notch was 6.50% in CI group and 5.30% in PSI group, respectively, which shows no significant statistical difference in the two groups (P = 0.667 > 0.05).

Conclusion

DFSAA could be taken as an indicator to predict the notch when performing TKA assisted with PSI. Especially when the DFSAA ≥ 3°, the risk of notch could be markedly increased.
Literature
5.
17.
go back to reference Maus U, Marques CJ, Scheunemann D, Lampe F, Lazovic D, Hommel H, Vogel D, Haunschild M, Pfitzner T (2018) No improvement in reducing outliers in coronal axis alignment with patient-specific instrumentation. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 26(9):2788–2796. https://doi.org/10.1007/s00167-017-4741-1CrossRef Maus U, Marques CJ, Scheunemann D, Lampe F, Lazovic D, Hommel H, Vogel D, Haunschild M, Pfitzner T (2018) No improvement in reducing outliers in coronal axis alignment with patient-specific instrumentation. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 26(9):2788–2796. https://​doi.​org/​10.​1007/​s00167-017-4741-1CrossRef
22.
go back to reference Ko JH, Han CD, Shin KH, Nguku L, Yang IH, Lee WS, Kim KI, Park KK (2016) Femur bowing could be a risk factor for implant flexion in conventional total knee arthroplasty and notching in navigated total knee arthroplasty. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 24(8):2476–2482. https://doi.org/10.1007/s00167-015-3863-6CrossRef Ko JH, Han CD, Shin KH, Nguku L, Yang IH, Lee WS, Kim KI, Park KK (2016) Femur bowing could be a risk factor for implant flexion in conventional total knee arthroplasty and notching in navigated total knee arthroplasty. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 24(8):2476–2482. https://​doi.​org/​10.​1007/​s00167-015-3863-6CrossRef
27.
go back to reference Minoda Y, Kobayashi A, Iwaki H, Mitsuhiko I, Kadoya Y, Ohashi H, Takaoka K, Nakamura H (2010) The risk of notching the anterior femoral cortex with the use of navigation systems in total knee arthroplasty. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 18(6):718–722. https://doi.org/10.1007/s00167-009-0927-5CrossRef Minoda Y, Kobayashi A, Iwaki H, Mitsuhiko I, Kadoya Y, Ohashi H, Takaoka K, Nakamura H (2010) The risk of notching the anterior femoral cortex with the use of navigation systems in total knee arthroplasty. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 18(6):718–722. https://​doi.​org/​10.​1007/​s00167-009-0927-5CrossRef
30.
go back to reference Boonen B, Schotanus MG, Kerens B, van der Weegen W, van Drumpt RA, Kort NP (2013) Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 21(10):2206–2212. https://doi.org/10.1007/s00167-013-2620-yCrossRef Boonen B, Schotanus MG, Kerens B, van der Weegen W, van Drumpt RA, Kort NP (2013) Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 21(10):2206–2212. https://​doi.​org/​10.​1007/​s00167-013-2620-yCrossRef
Metadata
Title
Does patient-specific instrumentation increase the risk of notching in the anterior femoral cortex in total knee arthroplasty? A comparative prospective trial
Authors
Song Ke
Tianfei Ran
Yongqin He
Mingrui Lv
Xin Song
Yuanyuan Zhou
Yuan Xu
Min Wang
Publication date
01-12-2020
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 12/2020
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-020-04779-4

Other articles of this Issue 12/2020

International Orthopaedics 12/2020 Go to the issue