Abstract

Objectives

To investigate the influence of orthodontic materials, field of view (FOV), and artifact reduction (AR) on the assessment of approximal caries using cone beam computed tomography.

Materials and methods

Forty non-cavitated and restoration-free human premolars and molars ranging from sound to various grades of lesions without cavitations were assigned to 13 groups with different combination of fix appliance equipment. CBCT (cone beam computed tomography) (Planmeca ProMax 3D Mid, Helsinki, Finland) images were obtained using combinations of three orthodontic bracket materials and two orthodontic archwire with small and large FOVs and with and without AR activation. Receiver operating characteristic (ROC) analysis was used to calculate the area under the ROC curve (AUC).

Results

Interobserver agreement ranged from 0.44 to 0.92 and intraobserver agreement ranged from 0.50 to 0.99. Teeth lacking orthodontic materials had the highest Az values at 0.84. FOV and AR activation did not significantly affect AUC values (P > 0.05). The AUC data were significantly reduced by the addition of stainless steel wire, NT wire, or a combination of a stainless steel bracket with stainless steel wire (P < 0.05).

Conclusions

The addition of stainless steel wire, NT wire, or a stainless steel bracket with stainless steel wire combination prevented the diagnosis of non-cavitated interproximal tooth caries by CBCT. With and without AR modes and different FOVs did not influence the diagnosis of interproximal caries lesions with different types of orthodontic equipment.

Clinical relevance

A wide variety of brackets and wire combinations are used in the clinic; however, the extent to which these combinations impact the diagnosis of caries by CBCT as the effects of FOV and AR algorithms are unknown.
Literature
5.
go back to reference Sanders MA, Hoyjberg C, Chu CB, Leggitt VL, Kim JS (2007) Common orthodontic appliances cause artifacts that degrade the diagnostic quality of CBCT images. J Calif Dent Assoc 35:850–857PubMed Sanders MA, Hoyjberg C, Chu CB, Leggitt VL, Kim JS (2007) Common orthodontic appliances cause artifacts that degrade the diagnostic quality of CBCT images. J Calif Dent Assoc 35:850–857PubMed
10.
go back to reference Gorelick L, Geiger AM, Gwinnett AJ (1982) Incidence of white spot formation after bonding and banding. Am J Orthod 81:93–98CrossRef Gorelick L, Geiger AM, Gwinnett AJ (1982) Incidence of white spot formation after bonding and banding. Am J Orthod 81:93–98CrossRef
11.
go back to reference Mizrahi E (1982) Enamel demineralization following orthodontic treatment. Am J Orthod 82:62–67CrossRef Mizrahi E (1982) Enamel demineralization following orthodontic treatment. Am J Orthod 82:62–67CrossRef
14.
go back to reference Prell D, Kyriakou Y, Struffert T, Dörfler A, Kalender W (2010) Metal artifact reduction for clipping and coiling in interventional C-arm CT. Am J Neuroradiol 31:634–639CrossRef Prell D, Kyriakou Y, Struffert T, Dörfler A, Kalender W (2010) Metal artifact reduction for clipping and coiling in interventional C-arm CT. Am J Neuroradiol 31:634–639CrossRef
15.
go back to reference Obuchowski NA (2003) Receiver operating characteristic curves and their use in radiology 1. Radiology 229:3–8CrossRef Obuchowski NA (2003) Receiver operating characteristic curves and their use in radiology 1. Radiology 229:3–8CrossRef
19.
go back to reference Rino Neto J, Silva FP, Chilvarquer I, Paiva JB, Hernandez AM (2012) Hausdorff distance evaluation of orthodontic accessories’ streaking artifacts in 3D model superimposition. Braz Oral Res 26:450–456CrossRef Rino Neto J, Silva FP, Chilvarquer I, Paiva JB, Hernandez AM (2012) Hausdorff distance evaluation of orthodontic accessories’ streaking artifacts in 3D model superimposition. Braz Oral Res 26:450–456CrossRef
23.
go back to reference de Rezende Barbosa GL, Sousa Melo SL, Alencar PN, Nascimento MC, Almeida SM (2016) Performance of an artefact reduction algorithm in the diagnosis of in vitro vertical root fracture in four different root filling conditions on CBCT images. Int Endod J 49:500–508. https://doi.org/10.1111/iej.12477 CrossRefPubMed de Rezende Barbosa GL, Sousa Melo SL, Alencar PN, Nascimento MC, Almeida SM (2016) Performance of an artefact reduction algorithm in the diagnosis of in vitro vertical root fracture in four different root filling conditions on CBCT images. Int Endod J 49:500–508. https://​doi.​org/​10.​1111/​iej.​12477 CrossRefPubMed
Metadata
Title
Evaluating the effects of orthodontic materials, field of view, and artifact reduction mode on accuracy of CBCT-based caries detection
Authors
Ozlem Isman, DDS, PhD
Ali Murat Aktan, DDS, PhD
Elif Tarim Ertas, DDS, PhD
Publication date
01-07-2020
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 7/2020
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-019-03112-7

Other articles of this Issue 7/2020

Clinical Oral Investigations 7/2020 Go to the issue