Published in: Current Diabetes Reports 9/2022

07-07-2022 | COVID-19 | Hospital Management of Diabetes (A Wallia and JJ Seley, Section Editors)

CGM in the Hospital: Is It Ready for Prime Time?

Authors: Elizabeth O. Buschur, Eileen Faulds, Kathleen Dungan

Published in: Current Diabetes Reports | Issue 9/2022

Login to get access

Abstract

Purpose of Review

The use of continuous glucose monitoring (CGM) in the hospital setting is growing with more patients using these devices at home and when admitted to the hospital, especially during the COVID-19 pandemic.

Recent Findings

Historically, most evidence for CGM use in the inpatient setting was limited to small studies utilizing outdated CGM technology and analyzing accuracy of sensor measurements. Previous studies have shown reduced sensor accuracy during extreme hypo- or hyperglycemia, rapid fluctuations of glucose, compression of the sensor itself, and in those who are critically ill. Studies that are more recent have shown CGM to have adequate accuracy and may be effective in reducing hypoglycemia in hospitalized patients; some studies have also showed improvement in time in target glycemic range. Furthermore, CGM may reduce nursing workload, cost of inpatient care, and use of personal protective equipment and face-to-face patient care especially for patients during the COVID-19 pandemic.

Summary

This review will describe the evidence for use of CGM in hospitalized critically ill or non-critically ill patients, address accuracy and safety considerations, and outline paths for future implementation.
Literature
2.
go back to reference American DA. 14. Diabetes care in the hospital: standards of medical care in diabetes-2018. Diabetes Care. 2018;41(Suppl 1):S144–S51.CrossRef American DA. 14. Diabetes care in the hospital: standards of medical care in diabetes-2018. Diabetes Care. 2018;41(Suppl 1):S144–S51.CrossRef
3.
go back to reference van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, et al. Intensive insulin therapy in critically ill patients. N Engl J Med. 2001;345(19):1359–67.PubMedCrossRef van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, et al. Intensive insulin therapy in critically ill patients. N Engl J Med. 2001;345(19):1359–67.PubMedCrossRef
4.
go back to reference American DA. 15. Diabetes care in the hospital: standards of medical care in diabetes-2021. Diabetes Care. 2021;44(Suppl 1):S211–S20.CrossRef American DA. 15. Diabetes care in the hospital: standards of medical care in diabetes-2021. Diabetes Care. 2021;44(Suppl 1):S211–S20.CrossRef
5.
go back to reference Clement S, Braithwaite SS, Magee MF, Ahmann A, Smith EP, Schafer RG, et al. Management of diabetes and hyperglycemia in hospitals. Diabetes Care. 2004;27(2):553–91.PubMedCrossRef Clement S, Braithwaite SS, Magee MF, Ahmann A, Smith EP, Schafer RG, et al. Management of diabetes and hyperglycemia in hospitals. Diabetes Care. 2004;27(2):553–91.PubMedCrossRef
6.
go back to reference Moghissi ES, Korytkowski MT, DiNardo M, Einhorn D, Hellman R, Hirsch IB, et al. American Association of Clinical Endocrinologists and American Diabetes Association consensus statement on inpatient glycemic control. Diabetes Care. 2009;32(6):1119–31.PubMedPubMedCentralCrossRef Moghissi ES, Korytkowski MT, DiNardo M, Einhorn D, Hellman R, Hirsch IB, et al. American Association of Clinical Endocrinologists and American Diabetes Association consensus statement on inpatient glycemic control. Diabetes Care. 2009;32(6):1119–31.PubMedPubMedCentralCrossRef
7.
go back to reference Bogun M, Inzucchi SE. Inpatient management of diabetes and hyperglycemia. Clin Ther. 2013;35(5):724–33.PubMedCrossRef Bogun M, Inzucchi SE. Inpatient management of diabetes and hyperglycemia. Clin Ther. 2013;35(5):724–33.PubMedCrossRef
8.
go back to reference Cardona S, Gomez PC, Vellanki P, Anzola I, Ramos C, Urrutia MA, et al. Clinical characteristics and outcomes of symptomatic and asymptomatic hypoglycemia in hospitalized patients with diabetes. BMJ Open Diabetes Res Care. 2018;6(1):e000607.PubMedPubMedCentralCrossRef Cardona S, Gomez PC, Vellanki P, Anzola I, Ramos C, Urrutia MA, et al. Clinical characteristics and outcomes of symptomatic and asymptomatic hypoglycemia in hospitalized patients with diabetes. BMJ Open Diabetes Res Care. 2018;6(1):e000607.PubMedPubMedCentralCrossRef
9.
go back to reference Brod M, Christensen T, Thomsen TL, Bushnell DM. The impact of non-severe hypoglycemic events on work productivity and diabetes management. Value Health. 2011;14(5):665–71.PubMedCrossRef Brod M, Christensen T, Thomsen TL, Bushnell DM. The impact of non-severe hypoglycemic events on work productivity and diabetes management. Value Health. 2011;14(5):665–71.PubMedCrossRef
10.
go back to reference Umpierrez GE, Klonoff DC. Diabetes technology update: use of insulin pumps and continuous glucose monitoring in the hospital. Diabetes Care. 2018;41(8):1579–89.PubMedPubMedCentralCrossRef Umpierrez GE, Klonoff DC. Diabetes technology update: use of insulin pumps and continuous glucose monitoring in the hospital. Diabetes Care. 2018;41(8):1579–89.PubMedPubMedCentralCrossRef
11.
go back to reference Wollersheim T, Engelhardt LJ, Pachulla J, Moergeli R, Koch S, Spies C, et al. Accuracy, reliability, feasibility and nurse acceptance of a subcutaneous continuous glucose management system in critically ill patients: a prospective clinical trial. Ann Intensive Care. 2016;6(1):70.PubMedPubMedCentralCrossRef Wollersheim T, Engelhardt LJ, Pachulla J, Moergeli R, Koch S, Spies C, et al. Accuracy, reliability, feasibility and nurse acceptance of a subcutaneous continuous glucose management system in critically ill patients: a prospective clinical trial. Ann Intensive Care. 2016;6(1):70.PubMedPubMedCentralCrossRef
17.
go back to reference Moghissi ES, Korytkowski MT, DiNardo M, Einhorn D, Hellman R, Hirsch IB, et al. American Association of Clinical Endocrinologists and American Diabetes Association consensus statement on inpatient glycemic control. Endocr Pract. 2009;15(4):353–69.PubMedCrossRef Moghissi ES, Korytkowski MT, DiNardo M, Einhorn D, Hellman R, Hirsch IB, et al. American Association of Clinical Endocrinologists and American Diabetes Association consensus statement on inpatient glycemic control. Endocr Pract. 2009;15(4):353–69.PubMedCrossRef
18.
go back to reference Inoue S, Egi M, Kotani J, Morita K. Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review. Crit Care. 2013;17(2):R48.PubMedPubMedCentralCrossRef Inoue S, Egi M, Kotani J, Morita K. Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review. Crit Care. 2013;17(2):R48.PubMedPubMedCentralCrossRef
19.•
go back to reference Galindo RJ, Umpierrez GE, Rushakoff RJ, Basu A, Lohnes S, Nichols JH, et al. Continuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline. J Diabetes Sci Technol. 2020;14(6):1035–64. This guideline includes the most up-to-date reccomendations for use of CGM in the inpatient settingPubMedPubMedCentralCrossRef Galindo RJ, Umpierrez GE, Rushakoff RJ, Basu A, Lohnes S, Nichols JH, et al. Continuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline. J Diabetes Sci Technol. 2020;14(6):1035–64. This guideline includes the most up-to-date reccomendations for use of CGM in the inpatient settingPubMedPubMedCentralCrossRef
20.
go back to reference Bochicchio GV, Nasraway S, Moore L, Furnary A, Nohra E, Bochicchio K. Results of a multicenter prospective pivotal trial of the first inline continuous glucose monitor in critically ill patients. J Trauma Acute Care Surg. 2017;82(6):1049–54.PubMedCrossRef Bochicchio GV, Nasraway S, Moore L, Furnary A, Nohra E, Bochicchio K. Results of a multicenter prospective pivotal trial of the first inline continuous glucose monitor in critically ill patients. J Trauma Acute Care Surg. 2017;82(6):1049–54.PubMedCrossRef
21.
go back to reference Rijkenberg S, van Steen SC, DeVries JH, van der Voort PHJ. Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients. J Clin Monit Comput. 2018;32(5):953–64.PubMedCrossRef Rijkenberg S, van Steen SC, DeVries JH, van der Voort PHJ. Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients. J Clin Monit Comput. 2018;32(5):953–64.PubMedCrossRef
22.
go back to reference Holzinger U, Warszawska J, Kitzberger R, Wewalka M, Miehsler W, Herkner H, et al. Real-time continuous glucose monitoring in critically ill patients: a prospective randomized trial. Diabetes Care. 2010;33(3):467–72.PubMedCrossRef Holzinger U, Warszawska J, Kitzberger R, Wewalka M, Miehsler W, Herkner H, et al. Real-time continuous glucose monitoring in critically ill patients: a prospective randomized trial. Diabetes Care. 2010;33(3):467–72.PubMedCrossRef
23.
go back to reference Boom DT, Sechterberger MK, Rijkenberg S, Kreder S, Bosman RJ, Wester JP, et al. Insulin treatment guided by subcutaneous continuous glucose monitoring compared to frequent point-of-care measurement in critically ill patients: a randomized controlled trial. Crit Care. 2014;18(4):453.PubMedPubMedCentralCrossRef Boom DT, Sechterberger MK, Rijkenberg S, Kreder S, Bosman RJ, Wester JP, et al. Insulin treatment guided by subcutaneous continuous glucose monitoring compared to frequent point-of-care measurement in critically ill patients: a randomized controlled trial. Crit Care. 2014;18(4):453.PubMedPubMedCentralCrossRef
24.
go back to reference De Block CE, Gios J, Verheyen N, Manuel-y-Keenoy B, Rogiers P, Jorens PG, et al. Randomized evaluation of glycemic control in the medical intensive care unit using real-time continuous glucose monitoring (REGIMEN Trial). Diabetes Technol Ther. 2015;17(12):889–98.PubMedCrossRef De Block CE, Gios J, Verheyen N, Manuel-y-Keenoy B, Rogiers P, Jorens PG, et al. Randomized evaluation of glycemic control in the medical intensive care unit using real-time continuous glucose monitoring (REGIMEN Trial). Diabetes Technol Ther. 2015;17(12):889–98.PubMedCrossRef
25.
go back to reference Krinsley JS, Bruns DE, Boyd JC. The impact of measurement frequency on the domains of glycemic control in the critically ill--a Monte Carlo simulation. J Diabetes Sci Technol. 2015;9(2):237–45.PubMedPubMedCentralCrossRef Krinsley JS, Bruns DE, Boyd JC. The impact of measurement frequency on the domains of glycemic control in the critically ill--a Monte Carlo simulation. J Diabetes Sci Technol. 2015;9(2):237–45.PubMedPubMedCentralCrossRef
26.
go back to reference Rodriguez LM, Knight RJ, Heptulla RA. Continuous glucose monitoring in subjects after simultaneous pancreas-kidney and kidney-alone transplantation. Diabetes Technol Ther. 2010;12(5):347–51.PubMedPubMedCentralCrossRef Rodriguez LM, Knight RJ, Heptulla RA. Continuous glucose monitoring in subjects after simultaneous pancreas-kidney and kidney-alone transplantation. Diabetes Technol Ther. 2010;12(5):347–51.PubMedPubMedCentralCrossRef
27.
go back to reference Burt MG, Roberts GW, Aguilar-Loza NR, Stranks SN. Brief report: comparison of continuous glucose monitoring and finger-prick blood glucose levels in hospitalized patients administered basal-bolus insulin. Diabetes Technol Ther. 2013;15(3):241–5.PubMedCrossRef Burt MG, Roberts GW, Aguilar-Loza NR, Stranks SN. Brief report: comparison of continuous glucose monitoring and finger-prick blood glucose levels in hospitalized patients administered basal-bolus insulin. Diabetes Technol Ther. 2013;15(3):241–5.PubMedCrossRef
28.
go back to reference Schaupp L, Donsa K, Neubauer KM, Mader JK, Aberer F, Holl B, et al. Taking a closer look--continuous glucose monitoring in non-critically ill hospitalized patients with type 2 diabetes mellitus under basal-bolus insulin therapy. Diabetes Technol Ther. 2015;17(9):611–8.PubMedCrossRef Schaupp L, Donsa K, Neubauer KM, Mader JK, Aberer F, Holl B, et al. Taking a closer look--continuous glucose monitoring in non-critically ill hospitalized patients with type 2 diabetes mellitus under basal-bolus insulin therapy. Diabetes Technol Ther. 2015;17(9):611–8.PubMedCrossRef
29.
go back to reference Gomez AM, Umpierrez GE, Munoz OM, Herrera F, Rubio C, Aschner P, et al. Continuous glucose monitoring versus capillary point-of-care testing for inpatient glycemic control in type 2 diabetes patients hospitalized in the general ward and treated with a basal bolus insulin regimen. J Diabetes Sci Technol. 2015;10(2):325–9.PubMedPubMedCentralCrossRef Gomez AM, Umpierrez GE, Munoz OM, Herrera F, Rubio C, Aschner P, et al. Continuous glucose monitoring versus capillary point-of-care testing for inpatient glycemic control in type 2 diabetes patients hospitalized in the general ward and treated with a basal bolus insulin regimen. J Diabetes Sci Technol. 2015;10(2):325–9.PubMedPubMedCentralCrossRef
30.
go back to reference Gu W, Liu Y, Chen Y, Deng W, Ran X, Chen L, et al. Multicentre randomized controlled trial with sensor-augmented pump vs multiple daily injections in hospitalized patients with type 2 diabetes in China: time to reach target glucose. Diabetes Metab. 2017;43(4):359–63.PubMedCrossRef Gu W, Liu Y, Chen Y, Deng W, Ran X, Chen L, et al. Multicentre randomized controlled trial with sensor-augmented pump vs multiple daily injections in hospitalized patients with type 2 diabetes in China: time to reach target glucose. Diabetes Metab. 2017;43(4):359–63.PubMedCrossRef
31.
go back to reference Galindo RJ, Migdal AL, Davis GM, Urrutia MA, Albury B, Zambrano C, et al. Comparison of the freestyle libre pro flash continuous glucose monitoring (CGM) system and point-of-care capillary glucose testing in hospitalized patients with type 2 diabetes treated with basal-bolus insulin regimen. Diabetes Care. 2020;43(11):2730–5.PubMedPubMedCentralCrossRef Galindo RJ, Migdal AL, Davis GM, Urrutia MA, Albury B, Zambrano C, et al. Comparison of the freestyle libre pro flash continuous glucose monitoring (CGM) system and point-of-care capillary glucose testing in hospitalized patients with type 2 diabetes treated with basal-bolus insulin regimen. Diabetes Care. 2020;43(11):2730–5.PubMedPubMedCentralCrossRef
32.
go back to reference Davis GM, Spanakis EK, Migdal AL, Singh LG, Albury B, Urrutia MA, et al. Accuracy of Dexcom G6 continuous glucose monitoring in non-critically ill hospitalized patients with diabetes. Diabetes Care. 2021; Davis GM, Spanakis EK, Migdal AL, Singh LG, Albury B, Urrutia MA, et al. Accuracy of Dexcom G6 continuous glucose monitoring in non-critically ill hospitalized patients with diabetes. Diabetes Care. 2021;
33.
go back to reference Singh LG, Satyarengga M, Marcano I, Scott WH, Pinault LF, Feng Z, et al. Reducing Inpatient hypoglycemia in the general wards using real-time continuous glucose monitoring: the glucose telemetry system, a randomized clinical trial. Diabetes Care. 2020;43(11):2736–43.PubMedPubMedCentralCrossRef Singh LG, Satyarengga M, Marcano I, Scott WH, Pinault LF, Feng Z, et al. Reducing Inpatient hypoglycemia in the general wards using real-time continuous glucose monitoring: the glucose telemetry system, a randomized clinical trial. Diabetes Care. 2020;43(11):2736–43.PubMedPubMedCentralCrossRef
34.
go back to reference Thabit H, Hartnell S, Allen JM, Lake A, Wilinska ME, Ruan Y, et al. Closed-loop insulin delivery in inpatients with type 2 diabetes: a randomised, parallel-group trial. Lancet Diabetes Endocrinol. 2017;5(2):117–24.PubMedCrossRef Thabit H, Hartnell S, Allen JM, Lake A, Wilinska ME, Ruan Y, et al. Closed-loop insulin delivery in inpatients with type 2 diabetes: a randomised, parallel-group trial. Lancet Diabetes Endocrinol. 2017;5(2):117–24.PubMedCrossRef
35.
go back to reference Fortmann AL, Spierling Bagsic SR, Talavera L, Garcia IM, Sandoval H, Hottinger A, et al. Glucose as the fifth vital sign: a randomized controlled trial of continuous glucose monitoring in a non-ICU hospital setting. Diabetes Care. 2020;43(11):2873–7.PubMedPubMedCentralCrossRef Fortmann AL, Spierling Bagsic SR, Talavera L, Garcia IM, Sandoval H, Hottinger A, et al. Glucose as the fifth vital sign: a randomized controlled trial of continuous glucose monitoring in a non-ICU hospital setting. Diabetes Care. 2020;43(11):2873–7.PubMedPubMedCentralCrossRef
36.
go back to reference Perez-Guzman MC, Duggan E, Gibanica S, Cardona S, Corujo-Rodriguez A, Faloye A, et al. Continuous glucose monitoring in the operating room and cardiac intensive care unit. Diabetes Care. 2021;44(3):e50–e2.PubMedPubMedCentralCrossRef Perez-Guzman MC, Duggan E, Gibanica S, Cardona S, Corujo-Rodriguez A, Faloye A, et al. Continuous glucose monitoring in the operating room and cardiac intensive care unit. Diabetes Care. 2021;44(3):e50–e2.PubMedPubMedCentralCrossRef
37.
go back to reference Nair BG, Dellinger EP, Flum DR, Rooke GA, Hirsch IB. A pilot study of the feasibility and accuracy of inpatient continuous glucose monitoring. Diabetes Care. 2020;43(11):e168–e9.PubMedCrossRef Nair BG, Dellinger EP, Flum DR, Rooke GA, Hirsch IB. A pilot study of the feasibility and accuracy of inpatient continuous glucose monitoring. Diabetes Care. 2020;43(11):e168–e9.PubMedCrossRef
38.
go back to reference Richardson S, Hirsch JS, Narasimhan M, Crawford JM, McGinn T, Davidson KW, et al. Presenting characteristics, comorbidities, and outcomes among 5700 patients hospitalized with COVID-19 in the New York City Area. JAMA. 2020;323(20):2052–9.PubMedPubMedCentralCrossRef Richardson S, Hirsch JS, Narasimhan M, Crawford JM, McGinn T, Davidson KW, et al. Presenting characteristics, comorbidities, and outcomes among 5700 patients hospitalized with COVID-19 in the New York City Area. JAMA. 2020;323(20):2052–9.PubMedPubMedCentralCrossRef
39.
go back to reference Guan WJ, Ni ZY, Hu Y, Liang WH, Ou CQ, He JX, et al. Clinical characteristics of coronavirus disease 2019 in China. N Engl J Med. 2020;382(18):1708–20.PubMedCrossRef Guan WJ, Ni ZY, Hu Y, Liang WH, Ou CQ, He JX, et al. Clinical characteristics of coronavirus disease 2019 in China. N Engl J Med. 2020;382(18):1708–20.PubMedCrossRef
40.
go back to reference Guo W, Li M, Dong Y, Zhou H, Zhang Z, Tian C, Qin R, Wang H, Shen W, Du K, Zhao L, Fan H, Luo S, Hu D. Diabetes is a risk factor for the progression and prognosis of COVID-19. Diabetes Metab Res Rev. 2020:e3319. Guo W, Li M, Dong Y, Zhou H, Zhang Z, Tian C, Qin R, Wang H, Shen W, Du K, Zhao L, Fan H, Luo S, Hu D. Diabetes is a risk factor for the progression and prognosis of COVID-19. Diabetes Metab Res Rev. 2020:e3319.
43.
go back to reference Faulds ER BA, Sumner L, Jones L, McNett M, Smetana KS, May CC, Buschur E, Exline MC, Ringel MD, Dungan K. Use of continuous glucose monitor in critically ill COVID-19 patients requiring insulin infusion: an observational study. J Clin Endocrinol Metab 2021;106(10):e4007–16. https://doi.org/10.1210/clinem/dgab409. Faulds ER BA, Sumner L, Jones L, McNett M, Smetana KS, May CC, Buschur E, Exline MC, Ringel MD, Dungan K. Use of continuous glucose monitor in critically ill COVID-19 patients requiring insulin infusion: an observational study. J Clin Endocrinol Metab 2021;106(10):e4007–16. https://​doi.​org/​10.​1210/​clinem/​dgab409.
44.
go back to reference Davis GM, Faulds E, Walker T, Vigliotti D, Rabinovich M, Hester J, et al. Remote continuous glucose monitoring with a computerized insulin infusion protocol for critically ill patients in a COVID-19 medical ICU: proof of concept. Diabetes Care. 2021;44(4):1055–8.PubMedPubMedCentralCrossRef Davis GM, Faulds E, Walker T, Vigliotti D, Rabinovich M, Hester J, et al. Remote continuous glucose monitoring with a computerized insulin infusion protocol for critically ill patients in a COVID-19 medical ICU: proof of concept. Diabetes Care. 2021;44(4):1055–8.PubMedPubMedCentralCrossRef
45.
go back to reference Faulds ER, Jones L, McNett M, Smetana KS, May CC, Sumner L, et al. Facilitators and barriers to nursing implementation of continuous glucose monitoring (CGM) in critically ill patients with COVID-19. Endocr Pract. 2021;27(4):354–61.PubMedPubMedCentralCrossRef Faulds ER, Jones L, McNett M, Smetana KS, May CC, Sumner L, et al. Facilitators and barriers to nursing implementation of continuous glucose monitoring (CGM) in critically ill patients with COVID-19. Endocr Pract. 2021;27(4):354–61.PubMedPubMedCentralCrossRef
46.
go back to reference Chow KW, Kelly DJ, Rieff MC, Skala PA, Kravets I, Charitou MM, et al. Outcomes and healthcare provider perceptions of real-time continuous glucose monitoring (rtCGM) in patients with diabetes and COVID-19 admitted to the ICU. J Diabetes Sci Technol. 2021;15(3):607–14.PubMedPubMedCentralCrossRef Chow KW, Kelly DJ, Rieff MC, Skala PA, Kravets I, Charitou MM, et al. Outcomes and healthcare provider perceptions of real-time continuous glucose monitoring (rtCGM) in patients with diabetes and COVID-19 admitted to the ICU. J Diabetes Sci Technol. 2021;15(3):607–14.PubMedPubMedCentralCrossRef
47.
go back to reference Agarwal S, Mathew J, Davis GM, Shephardson A, Levine A, Louard R, et al. Continuous glucose monitoring in the intensive care unit during the COVID-19 pandemic. Diabetes Care. 2021;44(3):847–9.PubMedCrossRef Agarwal S, Mathew J, Davis GM, Shephardson A, Levine A, Louard R, et al. Continuous glucose monitoring in the intensive care unit during the COVID-19 pandemic. Diabetes Care. 2021;44(3):847–9.PubMedCrossRef
48.
go back to reference Sadhu AR, Serrano IA, Xu J, Nisar T, Lucier J, Pandya AR, et al. Continuous glucose monitoring in critically ill patients with COVID-19: results of an emergent pilot study. J Diabetes Sci Technol. 2020;14(6):1065–73.PubMedPubMedCentralCrossRef Sadhu AR, Serrano IA, Xu J, Nisar T, Lucier J, Pandya AR, et al. Continuous glucose monitoring in critically ill patients with COVID-19: results of an emergent pilot study. J Diabetes Sci Technol. 2020;14(6):1065–73.PubMedPubMedCentralCrossRef
49.
go back to reference Murray-Bachmann R, Leung TM, Myers AK, Murthi S, Sarbanes M, Ziskovich K, et al. Reliability of continuous glucose monitoring system in the inpatient setting. J Clin Transl Endocrinol. 2021;25:100262.PubMedPubMedCentral Murray-Bachmann R, Leung TM, Myers AK, Murthi S, Sarbanes M, Ziskovich K, et al. Reliability of continuous glucose monitoring system in the inpatient setting. J Clin Transl Endocrinol. 2021;25:100262.PubMedPubMedCentral
50.
go back to reference Klarskov CK, Windum NA, Olsen MT, Dungu AM, Jensen AK, Lindegaard B, Pedersen-Bjergaard U, Krisensen PL. Telemetric continuous glucose monitoring during the COVID-19 pandemic in isolated hospitalized patients in Denmark - a randomized controlled exploratory trial. Diabetes Technol Ther. 2022;24(2):102–12. https://doi.org/10.1089/dia.2021.0291. Klarskov CK, Windum NA, Olsen MT, Dungu AM, Jensen AK, Lindegaard B, Pedersen-Bjergaard U, Krisensen PL. Telemetric continuous glucose monitoring during the COVID-19 pandemic in isolated hospitalized patients in Denmark - a randomized controlled exploratory trial. Diabetes Technol Ther. 2022;24(2):102–12. https://​doi.​org/​10.​1089/​dia.​2021.​0291.
51.
go back to reference Gomez AM, Henao DC, Munoz OM, Aschner P, Yepes CA, Jojoa R, et al. Glycemic control metrics using flash glucose monitoring and hospital complications in patients with COVID-19. Diabetes Metab Syndr. 2021;15(2):499–503.PubMedPubMedCentralCrossRef Gomez AM, Henao DC, Munoz OM, Aschner P, Yepes CA, Jojoa R, et al. Glycemic control metrics using flash glucose monitoring and hospital complications in patients with COVID-19. Diabetes Metab Syndr. 2021;15(2):499–503.PubMedPubMedCentralCrossRef
52.
go back to reference Tripyla A, Herzig D, Joachim D, Nakas CT, Amiet F, Andreou A, et al. Performance of a factory-calibrated, real-time continuous glucose monitoring system during elective abdominal surgery. Diabetes Obes Metab. 2020;22(9):1678–82.PubMedCrossRef Tripyla A, Herzig D, Joachim D, Nakas CT, Amiet F, Andreou A, et al. Performance of a factory-calibrated, real-time continuous glucose monitoring system during elective abdominal surgery. Diabetes Obes Metab. 2020;22(9):1678–82.PubMedCrossRef
53.
go back to reference Migdal AL, Spanakis EK, Galindo RJ, Davis G, Singh LG, Satyarengga M, et al. Accuracy and precision of continuous glucose monitoring in hospitalized patients undergoing radiology procedures. J Diabetes Sci Technol. 2020;14(6):1135–6.PubMedPubMedCentralCrossRef Migdal AL, Spanakis EK, Galindo RJ, Davis G, Singh LG, Satyarengga M, et al. Accuracy and precision of continuous glucose monitoring in hospitalized patients undergoing radiology procedures. J Diabetes Sci Technol. 2020;14(6):1135–6.PubMedPubMedCentralCrossRef
55.
go back to reference Takatsu Y, Shiozaki T, Miyati T, Asahara M, Tani Y. Are the recorded data of flash glucose monitoring systems influenced by radiological examinations? Radiol Phys Technol. 2019;12(2):224–9.PubMedCrossRef Takatsu Y, Shiozaki T, Miyati T, Asahara M, Tani Y. Are the recorded data of flash glucose monitoring systems influenced by radiological examinations? Radiol Phys Technol. 2019;12(2):224–9.PubMedCrossRef
56.
go back to reference Rajendran R, Rayman G. Point-of-care blood glucose testing for diabetes care in hospitalized patients: an evidence-based review. J Diabetes Sci Technol. 2014;8(6):1081–90.PubMedPubMedCentralCrossRef Rajendran R, Rayman G. Point-of-care blood glucose testing for diabetes care in hospitalized patients: an evidence-based review. J Diabetes Sci Technol. 2014;8(6):1081–90.PubMedPubMedCentralCrossRef
58.
go back to reference Wadwa RP, Laffel LM, Shah VN, Garg SK. Accuracy of a factory-calibrated, real-time continuous glucose monitoring system during 10 days of use in youth and adults with diabetes. Diabetes Technol Ther. 2018;20(6):395–402.PubMedPubMedCentralCrossRef Wadwa RP, Laffel LM, Shah VN, Garg SK. Accuracy of a factory-calibrated, real-time continuous glucose monitoring system during 10 days of use in youth and adults with diabetes. Diabetes Technol Ther. 2018;20(6):395–402.PubMedPubMedCentralCrossRef
59.
go back to reference Freckmann G, Pleus S, Grady M, Setford S, Levy B. Measures of accuracy for continuous glucose monitoring and blood glucose monitoring devices. J Diabetes Sci Technol. 2019;13(3):575–83.PubMedCrossRef Freckmann G, Pleus S, Grady M, Setford S, Levy B. Measures of accuracy for continuous glucose monitoring and blood glucose monitoring devices. J Diabetes Sci Technol. 2019;13(3):575–83.PubMedCrossRef
60.
go back to reference Faulds ER, Boutsicaris A, Sumner L, Jones L, McNett M, Smetana KS, et al. Use of continuous glucose monitor in critically ill covid-19 patients requiring insulin infusion: an observational study. J Clin Endocrinol Metab. 2021;106(10):e4007–e16.PubMedCrossRef Faulds ER, Boutsicaris A, Sumner L, Jones L, McNett M, Smetana KS, et al. Use of continuous glucose monitor in critically ill covid-19 patients requiring insulin infusion: an observational study. J Clin Endocrinol Metab. 2021;106(10):e4007–e16.PubMedCrossRef
61.
go back to reference Davis GM, Spanakis EK, Migdal AL, Singh LG, Albury B, Urrutia MA, et al. Accuracy of Dexcom G6 continuous glucose monitoring in non-critically ill hospitalized patients with diabetes. Diabetes Care. 2021;44(7):1641–6.PubMedPubMedCentralCrossRef Davis GM, Spanakis EK, Migdal AL, Singh LG, Albury B, Urrutia MA, et al. Accuracy of Dexcom G6 continuous glucose monitoring in non-critically ill hospitalized patients with diabetes. Diabetes Care. 2021;44(7):1641–6.PubMedPubMedCentralCrossRef
62.
go back to reference Spanakis EK, Levitt DL, Siddiqui T, Singh LG, Pinault L, Sorkin J, et al. The effect of continuous glucose monitoring in preventing inpatient hypoglycemia in general wards: the glucose telemetry system. J Diabetes Sci Technol. 2018;12(1):20–5.PubMedCrossRef Spanakis EK, Levitt DL, Siddiqui T, Singh LG, Pinault L, Sorkin J, et al. The effect of continuous glucose monitoring in preventing inpatient hypoglycemia in general wards: the glucose telemetry system. J Diabetes Sci Technol. 2018;12(1):20–5.PubMedCrossRef
63.
go back to reference Curran GM, Bauer M, Mittman B, Pyne JM, Stetler C. Effectiveness-implementation hybrid designs: combining elements of clinical effectiveness and implementation research to enhance public health impact. Med Care. 2012;50(3):217–26.PubMedPubMedCentralCrossRef Curran GM, Bauer M, Mittman B, Pyne JM, Stetler C. Effectiveness-implementation hybrid designs: combining elements of clinical effectiveness and implementation research to enhance public health impact. Med Care. 2012;50(3):217–26.PubMedPubMedCentralCrossRef
64.
go back to reference American Diabetes A. 7. Diabetes technology: standards of medical care in diabetes-2021. Diabetes Care. 2021;44(Suppl 1):S85–99.CrossRef American Diabetes A. 7. Diabetes technology: standards of medical care in diabetes-2021. Diabetes Care. 2021;44(Suppl 1):S85–99.CrossRef
65.
go back to reference Grunberger G, Sherr J, Allende M, Blevins T, Bode B, Handelsman Y, et al. American Association of Clinical Endocrinology clinical practice guideline: the use of advanced technology in the management of persons with diabetes mellitus. Endocr Pract. 2021;27(6):505–37.PubMedCrossRef Grunberger G, Sherr J, Allende M, Blevins T, Bode B, Handelsman Y, et al. American Association of Clinical Endocrinology clinical practice guideline: the use of advanced technology in the management of persons with diabetes mellitus. Endocr Pract. 2021;27(6):505–37.PubMedCrossRef
Metadata
Title
CGM in the Hospital: Is It Ready for Prime Time?
Authors
Elizabeth O. Buschur
Eileen Faulds
Kathleen Dungan
Publication date
07-07-2022
Publisher
Springer US
Published in
Current Diabetes Reports / Issue 9/2022
Print ISSN: 1534-4827
Electronic ISSN: 1539-0829
DOI
https://doi.org/10.1007/s11892-022-01484-x

Other articles of this Issue 9/2022

Current Diabetes Reports 9/2022 Go to the issue

Pathogenesis of Type 2 Diabetes and Insulin Resistance (M-E Patti, Section Editor)

Lipodystrophy for the Diabetologist—What to Look For

Health Care Delivery Systems and Implementation in Diabetes (ME McDonnell and AR Sadhu, Section Editor)

Electronic Health Record–Based Decision-Making Support in Inpatient Diabetes Management

Hospital Management of Diabetes (A Wallia and J Jeffrie Seley, Section Editors)

Pharmacist Role in Providing Inpatient Diabetes Management

Pediatric Type 2 and Monogenic Diabetes (O Pinhas-Hamiel , Section Editor)

Diabetes Out-of-the-Box: Diabetes Mellitus and Impairment in Hearing and Vision

Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.