Published in: Surgical and Radiologic Anatomy 6/2013

01-08-2013 | Original Article

Cone-beam computed tomography and microtomography for alveolar bone measurements

Authors: Nathália Ferrare, André Ferreira Leite, Hugo César Pinto Marques Caracas, Ricardo Bentes de Azevedo, Nilce Santos de Melo, Paulo Tadeu de Souza Figueiredo

Published in: Surgical and Radiologic Anatomy | Issue 6/2013

Login to get access

Abstract

Purpose

To compare cone-beam computed tomography (CBCT) and microtomography (micro-CT) for alveolar bone measurements.

Methods

Forty teeth and alveolar bone blocks of five pigs were scanned on a micro-CT with a 9.05 μm pixel size, and on a CBCT device at 0.125 mm voxel size. One height and four thickness measurements were performed twice in standardized slices by two radiologists to verify reliability. Agreement between imaging methods was assessed by correlation coefficients, Bland–Altman plots, and the difference was tested by a Wilcoxon signed-rank test.

Results

Regarding intra- and interobserver agreements, all bone measurements presented excellent precision values for micro-CT, but interobserver agreement for CBCT presented good to moderate values. Bone height differed about 0.3 mm, but no statistically significant differences were found for the bone thickness measurements.

Conclusion

CBCT underestimated bone height. No statistically significant differences were found for bone thickness. Regions of thin bone tissue may not be visualized on CBCT images. There are risks of underestimating bone measurements with CBCT and assuming bone loss that does not exist clinically. Although the difference of the bone height measurement was small, the clinical relevance must be analyzed on how to interpret CBCT
Literature
1.
go back to reference Al-Ekrish AA, Ekram M (2011) A comparative study of the accuracy and reliability of multidetector computed tomography and cone beam computed tomography in the assessment of dental implant site dimensions. Dentomaxillofac Radiol 40:67–75. doi:10.1259/dmfr/27546065 PubMedCrossRef Al-Ekrish AA, Ekram M (2011) A comparative study of the accuracy and reliability of multidetector computed tomography and cone beam computed tomography in the assessment of dental implant site dimensions. Dentomaxillofac Radiol 40:67–75. doi:10.​1259/​dmfr/​27546065 PubMedCrossRef
4.
5.
go back to reference Hassan B, Souza PC, Jacobs R, Berti SA, van der Stelt P (2010) Influence of scanning and reconstruction parameters on quality of three-dimensional surface models of the dental arches from cone beam computed tomography. Clin Oral Investig 14:303–310. doi:10.1007/s00784-009-0291-3 PubMedCrossRef Hassan B, Souza PC, Jacobs R, Berti SA, van der Stelt P (2010) Influence of scanning and reconstruction parameters on quality of three-dimensional surface models of the dental arches from cone beam computed tomography. Clin Oral Investig 14:303–310. doi:10.​1007/​s00784-009-0291-3 PubMedCrossRef
9.
go back to reference Liang X, Jacobs R, Hassan B, Li L, Pauwels R, Corpas L, Souza PC, Martens W, Shahbazian M, Alonso A, Lambrichts I (2010) A comparative evaluation of cone beam computed tomography (cbct) and multi-slice Ct (Msct) Part I. on subjective image quality. Eur J Radiol 75:265–269. doi:10.1016/j.ejrad.2009.03.042 PubMedCrossRef Liang X, Jacobs R, Hassan B, Li L, Pauwels R, Corpas L, Souza PC, Martens W, Shahbazian M, Alonso A, Lambrichts I (2010) A comparative evaluation of cone beam computed tomography (cbct) and multi-slice Ct (Msct) Part I. on subjective image quality. Eur J Radiol 75:265–269. doi:10.​1016/​j.​ejrad.​2009.​03.​042 PubMedCrossRef
15.
go back to reference Momin MA, Matsumoto K, Ejima K, Asaumi R, Kawai T, Arai Y, Honda K, Yosue T (2012) Correlation of mandibular impacted tooth and bone morphology determined by cone beam computed topography on a premise oh third molar operation. Surg Radiol Anat. doi:10.1007/s00276-012-1031-y PubMed Momin MA, Matsumoto K, Ejima K, Asaumi R, Kawai T, Arai Y, Honda K, Yosue T (2012) Correlation of mandibular impacted tooth and bone morphology determined by cone beam computed topography on a premise oh third molar operation. Surg Radiol Anat. doi:10.​1007/​s00276-012-1031-y PubMed
16.
go back to reference Oth O, Louryan S, van Sint Jan S, Rooze M, Glineur R (2012) Impact of the mandibular divergence on the position of the inferior alveolar nerve and mylohyoid nerve: a computed tomography atudy and its relevance to bilateral sagittal aplit osteotomy. Surg Radiol Anat. doi:10.1007/s00276-012-1010-3 PubMed Oth O, Louryan S, van Sint Jan S, Rooze M, Glineur R (2012) Impact of the mandibular divergence on the position of the inferior alveolar nerve and mylohyoid nerve: a computed tomography atudy and its relevance to bilateral sagittal aplit osteotomy. Surg Radiol Anat. doi:10.​1007/​s00276-012-1010-3 PubMed
17.
18.
19.
23.
go back to reference Stratemann AS, Huang JC, Maki K, Miller AJ, Hatcher DC (2008) Comparison of cone beam computed tomography imaging with physical measures. Dentomaxillofac Radiol 37:1–14. doi:10.1259/dmfr/31349994 CrossRef Stratemann AS, Huang JC, Maki K, Miller AJ, Hatcher DC (2008) Comparison of cone beam computed tomography imaging with physical measures. Dentomaxillofac Radiol 37:1–14. doi:10.​1259/​dmfr/​31349994 CrossRef
24.
26.
go back to reference Thomsen JS, Laib A, Koller B, Prohaskas S, Mosekilde LI, Gowin W (2005) Sterelogical measures of trabecular bone structure: comparison of 3D micro computed tomography with 2D histological sections in human proximal tibial one biopsies. J Microsc 218:171–179. doi:10.111/j.1365-2818.2005.01469.x PubMedCrossRef Thomsen JS, Laib A, Koller B, Prohaskas S, Mosekilde LI, Gowin W (2005) Sterelogical measures of trabecular bone structure: comparison of 3D micro computed tomography with 2D histological sections in human proximal tibial one biopsies. J Microsc 218:171–179. doi:10.​111/​j.​1365-2818.​2005.​01469.​x PubMedCrossRef
27.
go back to reference Timock AM, Cook V, McDonald T, Leo MC, Crowe J, Benninger BL, Covell DA Jr (2011) Accuracy and reliability of buccal bone height and thickness measurements from cone-beam computed tomography imaging. Am J Orthod Dentofacial Orthop 140:734–744. doi:10.1016/j.ajodo.2011.06.021 PubMedCrossRef Timock AM, Cook V, McDonald T, Leo MC, Crowe J, Benninger BL, Covell DA Jr (2011) Accuracy and reliability of buccal bone height and thickness measurements from cone-beam computed tomography imaging. Am J Orthod Dentofacial Orthop 140:734–744. doi:10.​1016/​j.​ajodo.​2011.​06.​021 PubMedCrossRef
28.
go back to reference Vandenberghe B, Luchsinger S, Hostens J, Dhoore E, Jacobs R; The SEDENTEXCT Project Consortium (2012) The influence of exposure parameters on jawbone model accuracy using cone beam computed tomography and multi-slice computed tomography. Dentomaxillofac Radiol 41:466-474. doi: 10.1259/dmfr/81272805 Vandenberghe B, Luchsinger S, Hostens J, Dhoore E, Jacobs R; The SEDENTEXCT Project Consortium (2012) The influence of exposure parameters on jawbone model accuracy using cone beam computed tomography and multi-slice computed tomography. Dentomaxillofac Radiol 41:466-474. doi: 10.​1259/​dmfr/​81272805
29.
go back to reference Veyre-Goulet S, Fortin T, Thierry A (2008) Accuracy of linear measurement provided by cone beam computed tomography to assess bone quantity in the posterior maxilla: a human cadaver study. Clin Implant Dent Relat Res 10:226–230. doi:10.1111/j.1708-8208.2008.00083.x PubMed Veyre-Goulet S, Fortin T, Thierry A (2008) Accuracy of linear measurement provided by cone beam computed tomography to assess bone quantity in the posterior maxilla: a human cadaver study. Clin Implant Dent Relat Res 10:226–230. doi:10.​1111/​j.​1708-8208.​2008.​00083.​x PubMed
Metadata
Title
Cone-beam computed tomography and microtomography for alveolar bone measurements
Authors
Nathália Ferrare
André Ferreira Leite
Hugo César Pinto Marques Caracas
Ricardo Bentes de Azevedo
Nilce Santos de Melo
Paulo Tadeu de Souza Figueiredo
Publication date
01-08-2013
Publisher
Springer Paris
Published in
Surgical and Radiologic Anatomy / Issue 6/2013
Print ISSN: 0930-1038
Electronic ISSN: 1279-8517
DOI
https://doi.org/10.1007/s00276-013-1080-x

Other articles of this Issue 6/2013

Surgical and Radiologic Anatomy 6/2013 Go to the issue