01-09-2020 | Computed Tomography | Vascular-Interventional

Effective dose to patient measurements for flat-detector computed tomography protocols in acute stroke care

Authors: Alex Brehm, Georg Stamm, Matthias Lüpke, Christian Riedel, Bram Stieltjes, Marios-Nikos Psychogios

Published in: European Radiology | Issue 9/2020

Login to get access

Abstract

Objectives

The aim was to measure the effective dose of flat-detector CT (FDCT) whole-brain imaging, biphasic FDCT angiography (FDCT-A), and FDCT perfusion (FDCT-P) protocols and compare it to previously reported effective dose values of multidetector CT (MDCT) applications.

Materials

We measured effective dose according to the IRCP 103 using an anthropomorphic phantom equipped with thermoluminescent dosimeters (TLDs). Placement was according to anatomical positions of each organ. In total, 60 TLDs (≥ 4 TLDs/organ) were placed into and onto the phantom to account for all relevant organs. Organs within the primary beam were covered with more TLDs. Additionally, we measured dose to the eye lens with two TLDs per eye. Protocols which we routinely use in clinical practice were measured on a biplane angiography system.

Results

The effective dose of the 20-s protocol/7-s protocol for whole-brain imaging was 2.6 mSv/2.4 mSv. The radiation dose to the eye lens was 24/23 mGy. For the biphasic high-/low-dose FDCT-A protocol, the effective dose was 8.9/2.8 mSv respectively. The eye lens dose was 60/14 mGy. The contribution of bolus tracking to the effective dose was 0.66 mSv (assuming average duration of 14 s). The multisweep FDCT-P protocol had an effective dose of 5.9 mSv and an eye lens dose of 46 mGy.

Conclusion

Except for the high-dose biphasic FDCT-A protocol, FDCT applications used in neuroradiology have effective doses, which do not deviate more than 1 mSv from previously reported values for MDCT applications. However, the effective dose to the eye lens in commonly used stroke paradigms exceeds the recommended annual dose twofold.

Key Points

• Flat-detector computed tomography (FDCT) can be used for acute and periinterventional imaging of acute stroke patients and in neurointerventions.
• Except for the high-dose FDCT angiography protocol, the effective doses do not deviate more than 1 mSv from previously reported values for multidetector CT applications.
• Strategies to decrease the effective lens dose especially in younger patients should be evaluated in the future.
Literature
1.
go back to reference Psychogios MN, Behme D, Schregel K et al (2017) One-stop management of acute stroke patients: minimizing door-to-reperfusion times. Stroke 48:3152–3155PubMed Psychogios MN, Behme D, Schregel K et al (2017) One-stop management of acute stroke patients: minimizing door-to-reperfusion times. Stroke 48:3152–3155PubMed
2.
go back to reference Hoelter P, Goelitz P, Lang S et al (2019) Visualization of large vessel occlusion, clot extent, and collateral supply using volume perfusion flat detector computed tomography in acute stroke patients. Acta Radiol 60(11):1504–1511PubMed Hoelter P, Goelitz P, Lang S et al (2019) Visualization of large vessel occlusion, clot extent, and collateral supply using volume perfusion flat detector computed tomography in acute stroke patients. Acta Radiol 60(11):1504–1511PubMed
3.
go back to reference Brehm A, Tsogkas I, Maier IL et al (2019) One-stop management with perfusion for transfer patients with stroke due to a large-vessel occlusion: feasibility and effects on in-hospital times. AJNR Am J Neuroradiol 40(8):1330–1334PubMedPubMedCentral Brehm A, Tsogkas I, Maier IL et al (2019) One-stop management with perfusion for transfer patients with stroke due to a large-vessel occlusion: feasibility and effects on in-hospital times. AJNR Am J Neuroradiol 40(8):1330–1334PubMedPubMedCentral
5.
go back to reference Lian Y, Xiao J, Ji X et al (2015) Protracted low-dose radiation exposure and cataract in a cohort of Chinese industry radiographers. Occup Environ Med 72(9):640–647PubMed Lian Y, Xiao J, Ji X et al (2015) Protracted low-dose radiation exposure and cataract in a cohort of Chinese industry radiographers. Occup Environ Med 72(9):640–647PubMed
6.
go back to reference Elmaraezy A, Morra ME, Mohammed AT et al (2019) Risk of cataract among interventional cardiologists and catheterization lab staff : a systematic review. Catheter Cardiovasc Interv 90(1):1–9 Elmaraezy A, Morra ME, Mohammed AT et al (2019) Risk of cataract among interventional cardiologists and catheterization lab staff : a systematic review. Catheter Cardiovasc Interv 90(1):1–9
7.
go back to reference Internationalen Strahlenschutzkommission (2007) Veröffentlichungen der Internationalen Strahlenschutzkommission Deutsche Ausgabe herausgegeben vom Bundesamt für Strahlenschutz Internationalen Strahlenschutzkommission (2007) Veröffentlichungen der Internationalen Strahlenschutzkommission Deutsche Ausgabe herausgegeben vom Bundesamt für Strahlenschutz
8.
go back to reference Yang P, Niu K, Wu Y et al (2017) Evaluation of collaterals and clot burden using time-resolved C-arm conebeam CT angiography in the angiography suite: a feasibility study. AJNR Am J Neuroradiol 38(4):747–752PubMed Yang P, Niu K, Wu Y et al (2017) Evaluation of collaterals and clot burden using time-resolved C-arm conebeam CT angiography in the angiography suite: a feasibility study. AJNR Am J Neuroradiol 38(4):747–752PubMed
9.
go back to reference Maier XIL, Leyhe XJR, Tsogkas XI et al (2018) Diagnosing early ischemic changes with the latest-generation flat detector CT: a comparative study with multidetector CT. AJNR Am J Neuroradiol 39(5):881–886PubMedPubMedCentral Maier XIL, Leyhe XJR, Tsogkas XI et al (2018) Diagnosing early ischemic changes with the latest-generation flat detector CT: a comparative study with multidetector CT. AJNR Am J Neuroradiol 39(5):881–886PubMedPubMedCentral
10.
go back to reference Struffert T, Hauer M, Banckwitz R, Köhler C, Royalty K, Doerfler A (2014) Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology. Eur Radiol 24(6):1257–1265PubMed Struffert T, Hauer M, Banckwitz R, Köhler C, Royalty K, Doerfler A (2014) Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology. Eur Radiol 24(6):1257–1265PubMed
11.
go back to reference Lin HC, Lai TJ, Tseng HC, Wang CH, Tseng YL, Chen CY (2019) Radiation doses with various body weights of phantoms in brain 128-slice MDCT examination. J Radiat Res 60(4):466–475PubMedPubMedCentral Lin HC, Lai TJ, Tseng HC, Wang CH, Tseng YL, Chen CY (2019) Radiation doses with various body weights of phantoms in brain 128-slice MDCT examination. J Radiat Res 60(4):466–475PubMedPubMedCentral
12.
go back to reference Fiorella D, Turk A, Chaudry I et al (2014) A prospective, multicenter pilot study investigating the utility of flat detector derived parenchymal blood volume maps to estimate cerebral blood volume in stroke patients. J Neurointerv Surg 6(6):451–456PubMed Fiorella D, Turk A, Chaudry I et al (2014) A prospective, multicenter pilot study investigating the utility of flat detector derived parenchymal blood volume maps to estimate cerebral blood volume in stroke patients. J Neurointerv Surg 6(6):451–456PubMed
13.
go back to reference Yang P, Niu K, Wu Y et al (2015) New capability for one-stop-shop acute ischemic. Stroke 46(12):3383–3389PubMed Yang P, Niu K, Wu Y et al (2015) New capability for one-stop-shop acute ischemic. Stroke 46(12):3383–3389PubMed
14.
go back to reference Psychogios M, Schramm P, Frölich AM et al (2013) Alberta stroke program early CT scale evaluation of multimodal computed tomography in predicting clinical outcomes of stroke patients treated with aspiration thrombectomy. Stroke 44(8):2188–2193PubMed Psychogios M, Schramm P, Frölich AM et al (2013) Alberta stroke program early CT scale evaluation of multimodal computed tomography in predicting clinical outcomes of stroke patients treated with aspiration thrombectomy. Stroke 44(8):2188–2193PubMed
15.
go back to reference Psychogios M-N, Schramm P, Buhk J et al (2010) Angiographic CT after intravenous contrast agent application: a noninvasive follow-up tool after intracranial angioplasty and stenting. AJNR Am J Neuroradiol 31(10):1886–1891PubMed Psychogios M-N, Schramm P, Buhk J et al (2010) Angiographic CT after intravenous contrast agent application: a noninvasive follow-up tool after intracranial angioplasty and stenting. AJNR Am J Neuroradiol 31(10):1886–1891PubMed
16.
go back to reference Schregel K, Tsogkas I, Peter C et al (2018) Outcome prediction using perfusion parameters and collateral scores of multi-phase and single-phase CT angiography in acute stroke: need for one, two, three, or thirty scans? J Stroke 20(3):362–372PubMedPubMedCentral Schregel K, Tsogkas I, Peter C et al (2018) Outcome prediction using perfusion parameters and collateral scores of multi-phase and single-phase CT angiography in acute stroke: need for one, two, three, or thirty scans? J Stroke 20(3):362–372PubMedPubMedCentral
17.
go back to reference Ainsbury EA, Barnard S, Bright S et al (2016) Ionizing radiation induced cataracts: recent biological and mechanistic developments and perspectives for future research. Mutat Res 770(Pt B):238–261PubMed Ainsbury EA, Barnard S, Bright S et al (2016) Ionizing radiation induced cataracts: recent biological and mechanistic developments and perspectives for future research. Mutat Res 770(Pt B):238–261PubMed
18.
go back to reference Dauer LT, Ainsbury EA, Dynlacht J et al (2017) Guidance on radiation dose limits for the lens of the eye: overview of the recommendations in NCRP Commentary No. 26. Int J Radiat Biol 93(10):1015–1023PubMed Dauer LT, Ainsbury EA, Dynlacht J et al (2017) Guidance on radiation dose limits for the lens of the eye: overview of the recommendations in NCRP Commentary No. 26. Int J Radiat Biol 93(10):1015–1023PubMed
19.
go back to reference Bolch WE, Dietze G, Petoussi-henss N, Zankl M (2015) Dosimetric models of the eye and lens of the eye and their use in assessing dose coefficients for ocular exposures. Ann ICRP 44(1 Suppl):91–111 Bolch WE, Dietze G, Petoussi-henss N, Zankl M (2015) Dosimetric models of the eye and lens of the eye and their use in assessing dose coefficients for ocular exposures. Ann ICRP 44(1 Suppl):91–111
20.
go back to reference Fahrig R, Dixon R, Payne T, Morin RL, Ganguly A, Strobel N (2006) Dose and image quality for a cone-beam C-arm CT system. Med Phys 33(12):4541–4550PubMed Fahrig R, Dixon R, Payne T, Morin RL, Ganguly A, Strobel N (2006) Dose and image quality for a cone-beam C-arm CT system. Med Phys 33(12):4541–4550PubMed
21.
go back to reference Kalender WA, Kyriakou Y (2007) Flat-detector computed tomography (FD-CT). Eur Radiol 17(11):2767–2779PubMed Kalender WA, Kyriakou Y (2007) Flat-detector computed tomography (FD-CT). Eur Radiol 17(11):2767–2779PubMed
Metadata
Title
Effective dose to patient measurements for flat-detector computed tomography protocols in acute stroke care
Authors
Alex Brehm
Georg Stamm
Matthias Lüpke
Christian Riedel
Bram Stieltjes
Marios-Nikos Psychogios
Publication date
01-09-2020
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 9/2020
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-06891-w

Other articles of this Issue 9/2020

European Radiology 9/2020 Go to the issue