Open Access 07-10-2023 | Computed Tomography | Scientific Article

Superior metal artifact reduction of tin-filtered low-dose CT in imaging of lumbar spinal instrumentation compared to conventional computed tomography

Authors: Christoph Stern, Florian Wanivenhaus, Andrea B. Rosskopf, Mazda Farshad, Reto Sutter

Published in: Skeletal Radiology | Issue 4/2024

Login to get access

Abstract

Objective

To compare the image quality of low-dose CT (LD-CT) with tin filtration of the lumbar spine after metal implants to standard clinical CT, and to evaluate the potential for metal artifact and dose reduction.

Materials and methods

CT protocols were optimized in a cadaver torso. Seventy-four prospectively included patients with metallic lumbar implants were scanned with both standard CT (120 kV) and tin-filtered LD-CT (Sn140kV). CT dose parameters and qualitative measures (1 = worst,4 = best) were compared. Quantitative measures included noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and the width and attenuation of the most prominent hypodense metal artifact. Standard CT and LD-CT were assessed for imaging findings.

Results

Tin-filtered LD-CT was performed with 60% dose saving compared to standard CT (median effective dose 3.22 mSv (quartile 1–3: 2.73–3.49 mSv) versus 8.02 mSv (6.42–9.27 mSv; p < .001). Image quality of CT and tin-filtered low-dose CT was good with excellent depiction of anatomy, while image noise was lower for CT and artifacts were weaker for tin-filtered LD-CT. Quantitative measures also revealed increased noise for tin-filtered low-dose CT (41.5HU), lower SNR (2) and CNR (0.6) compared to CT (32HU,3.55,1.03, respectively) (all p < .001). However, tin-filtered LD-CT performed superior regarding the width and attenuation of hypodense metal artifacts (2.9 mm and -767.5HU for LD-CT vs. 4.1 mm and -937HU for CT; all p < .001). No difference between methods was observed in detection of imaging findings.

Conclusion

Tin-filtered LD-CT with 60% dose saving performs comparable to standard CT in detection of pathology and surgery related complications after lumbar spinal instrumentation, and shows superior metal artifact reduction.
Appendix
Available only for authorised users
Literature
1.
go back to reference Yuan C, Zhou J, Wang L, Deng Z. Adjacent segment disease after minimally invasive transforaminal lumbar interbody fusion for degenerative lumbar diseases: incidence and risk factors. BMC Musculoskelet Disord. 2022;23(1):982.CrossRefPubMedPubMedCentral Yuan C, Zhou J, Wang L, Deng Z. Adjacent segment disease after minimally invasive transforaminal lumbar interbody fusion for degenerative lumbar diseases: incidence and risk factors. BMC Musculoskelet Disord. 2022;23(1):982.CrossRefPubMedPubMedCentral
2.
go back to reference Sarwahi V, Ayan S, Amaral T, Wendolowski S, Gecelter R, Lo Y, et al. Can Postoperative Radiographs Accurately Identify Screw Misplacements? Spine Deform. 2017;5(2):109–16.CrossRefPubMed Sarwahi V, Ayan S, Amaral T, Wendolowski S, Gecelter R, Lo Y, et al. Can Postoperative Radiographs Accurately Identify Screw Misplacements? Spine Deform. 2017;5(2):109–16.CrossRefPubMed
3.
go back to reference Wanivenhaus F, Bauer DE, Laux C, Stern C, Cornaz F, Wetzel O, et al. Risk factors for L5 pedicle fractures after single-level posterior spinal fusion. Spine J. 2022;22(6):927–33.CrossRefPubMed Wanivenhaus F, Bauer DE, Laux C, Stern C, Cornaz F, Wetzel O, et al. Risk factors for L5 pedicle fractures after single-level posterior spinal fusion. Spine J. 2022;22(6):927–33.CrossRefPubMed
4.
go back to reference Biswas D, Bible JE, Bohan M, Simpson AK, Whang PG, Grauer JN. Radiation exposure from musculoskeletal computerized tomographic scans. J Bone Joint Surg Am. 2009;91(8):1882–9.CrossRefPubMed Biswas D, Bible JE, Bohan M, Simpson AK, Whang PG, Grauer JN. Radiation exposure from musculoskeletal computerized tomographic scans. J Bone Joint Surg Am. 2009;91(8):1882–9.CrossRefPubMed
5.
go back to reference Simpson AK, Whang PG, Jonisch A, Haims A, Grauer JN. The radiation exposure associated with cervical and lumbar spine radiographs. J Spinal Disord Tech. 2008;21(6):409–12.CrossRefPubMed Simpson AK, Whang PG, Jonisch A, Haims A, Grauer JN. The radiation exposure associated with cervical and lumbar spine radiographs. J Spinal Disord Tech. 2008;21(6):409–12.CrossRefPubMed
6.
go back to reference Gordic S, Morsbach F, Schmidt B, Allmendinger T, Flohr T, Husarik D, et al. Ultralow-dose chest computed tomography for pulmonary nodule detection: first performance evaluation of single energy scanning with spectral shaping. Invest Radiol. 2014;49(7):465–73.CrossRefPubMed Gordic S, Morsbach F, Schmidt B, Allmendinger T, Flohr T, Husarik D, et al. Ultralow-dose chest computed tomography for pulmonary nodule detection: first performance evaluation of single energy scanning with spectral shaping. Invest Radiol. 2014;49(7):465–73.CrossRefPubMed
7.
go back to reference Euler A, Obmann MM, Szucs-Farkas Z, Mileto A, Zaehringer C, Falkowski AL, et al. Comparison of image quality and radiation dose between split-filter dual-energy images and single-energy images in single-source abdominal CT. Eur Radiol. 2018;28(8):3405–12.CrossRefPubMed Euler A, Obmann MM, Szucs-Farkas Z, Mileto A, Zaehringer C, Falkowski AL, et al. Comparison of image quality and radiation dose between split-filter dual-energy images and single-energy images in single-source abdominal CT. Eur Radiol. 2018;28(8):3405–12.CrossRefPubMed
8.
go back to reference Braun FM, Johnson TR, Sommer WH, Thierfelder KM, Meinel FG. Chest CT using spectral filtration: radiation dose, image quality, and spectrum of clinical utility. Eur Radiol. 2015;25(6):1598–606.CrossRefPubMed Braun FM, Johnson TR, Sommer WH, Thierfelder KM, Meinel FG. Chest CT using spectral filtration: radiation dose, image quality, and spectrum of clinical utility. Eur Radiol. 2015;25(6):1598–606.CrossRefPubMed
9.
go back to reference Messerli M, Kluckert T, Knitel M, Rengier F, Warschkow R, Alkadhi H, et al. Computer-aided detection (CAD) of solid pulmonary nodules in chest x-ray equivalent ultralow dose chest CT - first in-vivo results at dose levels of 0.13mSv. Eur J Radiol. 2016;85(12):2217–24.CrossRefPubMed Messerli M, Kluckert T, Knitel M, Rengier F, Warschkow R, Alkadhi H, et al. Computer-aided detection (CAD) of solid pulmonary nodules in chest x-ray equivalent ultralow dose chest CT - first in-vivo results at dose levels of 0.13mSv. Eur J Radiol. 2016;85(12):2217–24.CrossRefPubMed
10.
go back to reference Messerli M, Ottilinger T, Warschkow R, Leschka S, Alkadhi H, Wildermuth S, et al. Emphysema quantification and lung volumetry in chest X-ray equivalent ultralow dose CT - Intra-individual comparison with standard dose CT. Eur J Radiol. 2017;91:1–9.CrossRefPubMed Messerli M, Ottilinger T, Warschkow R, Leschka S, Alkadhi H, Wildermuth S, et al. Emphysema quantification and lung volumetry in chest X-ray equivalent ultralow dose CT - Intra-individual comparison with standard dose CT. Eur J Radiol. 2017;91:1–9.CrossRefPubMed
11.
go back to reference Leyendecker P, Faucher V, Labani A, Noblet V, Lefebvre F, Magotteaux P, et al. Prospective evaluation of ultra-low-dose contrast-enhanced 100-kV abdominal computed tomography with tin filter: effect on radiation dose reduction and image quality with a third-generation dual-source CT system. Eur Radiol 2018. Leyendecker P, Faucher V, Labani A, Noblet V, Lefebvre F, Magotteaux P, et al. Prospective evaluation of ultra-low-dose contrast-enhanced 100-kV abdominal computed tomography with tin filter: effect on radiation dose reduction and image quality with a third-generation dual-source CT system. Eur Radiol 2018.
12.
go back to reference Huflage H, Grunz JP, Hackenbroch C, Halt D, Luetkens KS, Alfred Schmidt AM, et al. Metal artefact reduction in low-dose computed tomography: Benefits of tin prefiltration versus postprocessing of dual-energy datasets over conventional CT imaging. Radiography (Lond). 2022;28(3):690–6.CrossRefPubMed Huflage H, Grunz JP, Hackenbroch C, Halt D, Luetkens KS, Alfred Schmidt AM, et al. Metal artefact reduction in low-dose computed tomography: Benefits of tin prefiltration versus postprocessing of dual-energy datasets over conventional CT imaging. Radiography (Lond). 2022;28(3):690–6.CrossRefPubMed
13.
go back to reference Hackenbroch C, Schule S, Halt D, Zengerle L, Beer M. Metal artifact reduction with tin prefiltration in computed tomography: A cadaver study for comparison with other novel techniques. Invest Radiol. 2022;57(3):194–203.CrossRefPubMed Hackenbroch C, Schule S, Halt D, Zengerle L, Beer M. Metal artifact reduction with tin prefiltration in computed tomography: A cadaver study for comparison with other novel techniques. Invest Radiol. 2022;57(3):194–203.CrossRefPubMed
14.
go back to reference Zhou W, Bartlett DJ, Diehn FE, Glazebrook KN, Kotsenas AL, Carter RE, et al. Reduction of metal artifacts and improvement in dose efficiency using photon-counting detector computed tomography and tin filtration. Invest Radiol. 2019;54(4):204–11.CrossRefPubMedPubMedCentral Zhou W, Bartlett DJ, Diehn FE, Glazebrook KN, Kotsenas AL, Carter RE, et al. Reduction of metal artifacts and improvement in dose efficiency using photon-counting detector computed tomography and tin filtration. Invest Radiol. 2019;54(4):204–11.CrossRefPubMedPubMedCentral
15.
go back to reference Lee SK, Kim JS, Yoon SW, Kim JM. Development of CT Effective Dose Conversion Factors from Clinical CT Examinations in the Republic of Korea. Diagnostics (Basel) 2020; 10(9). Lee SK, Kim JS, Yoon SW, Kim JM. Development of CT Effective Dose Conversion Factors from Clinical CT Examinations in the Republic of Korea. Diagnostics (Basel) 2020; 10(9).
16.
Metadata
Title
Superior metal artifact reduction of tin-filtered low-dose CT in imaging of lumbar spinal instrumentation compared to conventional computed tomography
Authors
Christoph Stern
Florian Wanivenhaus
Andrea B. Rosskopf
Mazda Farshad
Reto Sutter
Publication date
07-10-2023
Publisher
Springer Berlin Heidelberg
Published in
Skeletal Radiology / Issue 4/2024
Print ISSN: 0364-2348
Electronic ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-023-04467-5

Other articles of this Issue 4/2024

Skeletal Radiology 4/2024 Go to the issue