Skip to main content
Top
Published in: BMC Oral Health 1/2021

Open Access 01-12-2021 | Computed Tomography | Research article

Endodontic length measurements using 3D Endo, cone-beam computed tomography, and electronic apex locator

Author: Khoa Van Pham

Published in: BMC Oral Health | Issue 1/2021

Login to get access

Abstract

Background

The objective of this study is to investigate the accuracy of the 3D Endo software, cone-beam computed tomography (CBCT) software, and the electronic apex locator (EAL) in endodontic length determination.

Methods

302 root canals in 111 human extracted molars were chosen. Access cavity was performed, and root canal lengths were measured with a digital caliper for actual length (AL) and EAL for electronic length. Teeth were then scanned using CBCT device at voxel size of 0.10 mm. It measured root canal lengths using the CBCT (Romexis Viewer), 3D Endo for proposed length (3D-PL) and correct length (3D-CL). Mean differences between the four methods with the AL were calculated and compared. Fisher’s exact test, paired t-test, Bland-Altman plot were used to test the differences among the experimental modalities in working length determination at the significance of 0.05.

Results

The accuracy in the range of ± 0.5 mm of the EAL ProPex II was highest among the experimental modalities, however this method disagreed with the actual length.

Conclusions

The correct working length after adjustment from the semi-automatically length by the 3D Endo software and Romexis Viewer measurements agreed with the AL.
Literature
1.
go back to reference Pham KV, Khuc NK. The accuracy of endodontic length measurement using cone-beam computed tomography in comparison with electronic apex locators. Iran Endod J. 2020;15(1):12–7. Pham KV, Khuc NK. The accuracy of endodontic length measurement using cone-beam computed tomography in comparison with electronic apex locators. Iran Endod J. 2020;15(1):12–7.
2.
go back to reference Ricucci D, Pascon EA, Siqueira JF. The complexity of the apical anatomy. In: Versiani MA, Basrani B, Sousa-Neto MD, editors. The root canal anatomy in permanent dentition. Cham: Springer International Publishing; 2019. p. 241–254. Ricucci D, Pascon EA, Siqueira JF. The complexity of the apical anatomy. In: Versiani MA, Basrani B, Sousa-Neto MD, editors. The root canal anatomy in permanent dentition. Cham: Springer International Publishing; 2019. p. 241–254.
3.
go back to reference Nekoofar MH, Ghandi MM, Hayes SJ, Dummer PMH. The fundamental operating principles of electronic root canal length measurement devices. Int Endod J. 2006;39(8):595–609.CrossRef Nekoofar MH, Ghandi MM, Hayes SJ, Dummer PMH. The fundamental operating principles of electronic root canal length measurement devices. Int Endod J. 2006;39(8):595–609.CrossRef
4.
go back to reference Martins JNR, Marques D, Mata A, Caramês J. Clinical efficacy of electronic apex locators: systematic review. J Endod. 2014;40(6):759–77.CrossRef Martins JNR, Marques D, Mata A, Caramês J. Clinical efficacy of electronic apex locators: systematic review. J Endod. 2014;40(6):759–77.CrossRef
5.
go back to reference Nudera WJ. Three-dimensional evaluation of internal tooth anatomy. In: Fayad M, Johnson BR, editors. 3D imaging in endodontics: a new era in diagnosis and treatment. Cham: Springer International Publishing; 2016. p. 53–73. Nudera WJ. Three-dimensional evaluation of internal tooth anatomy. In: Fayad M, Johnson BR, editors. 3D imaging in endodontics: a new era in diagnosis and treatment. Cham: Springer International Publishing; 2016. p. 53–73.
6.
go back to reference Khademi J. Advanced CBCT for endodontics: technical considerations, perception, and decision-making. 1st ed. Batavia: Quintessence Publishing; 2017. Khademi J. Advanced CBCT for endodontics: technical considerations, perception, and decision-making. 1st ed. Batavia: Quintessence Publishing; 2017.
7.
go back to reference Yılmaz F, Kamburoğlu K, Şenel B. Endodontic working length measurement using cone-beam computed tomographic images obtained at different voxel sizes and field of views, periapical radiography, and apex locator: a comparative vivo study. J Endod. 2017;43(1):152–6.CrossRef Yılmaz F, Kamburoğlu K, Şenel B. Endodontic working length measurement using cone-beam computed tomographic images obtained at different voxel sizes and field of views, periapical radiography, and apex locator: a comparative vivo study. J Endod. 2017;43(1):152–6.CrossRef
8.
go back to reference de Morais ALG, de Alencar AHG, Estrela CRdA, Decurcio DA, Estrela C. Working length determination using cone-beam computed tomography, periapical radiography and electronic apex locator in teeth with apical periodontitis: a clinical study. Iran Endod J. 2016;11(3):5. de Morais ALG, de Alencar AHG, Estrela CRdA, Decurcio DA, Estrela C. Working length determination using cone-beam computed tomography, periapical radiography and electronic apex locator in teeth with apical periodontitis: a clinical study. Iran Endod J. 2016;11(3):5.
9.
go back to reference Tchorz J. 3D endo: three-dimensional endodontic treatment planning. Int J Comput Dent. 2017;20:87–92.PubMed Tchorz J. 3D endo: three-dimensional endodontic treatment planning. Int J Comput Dent. 2017;20:87–92.PubMed
10.
go back to reference Ludbrook J. Statistical techniques for comparing measurers and methods of measurement: a critical review. Clin Exp Pharmacol Physiol. 2002;29(7):527–36.CrossRef Ludbrook J. Statistical techniques for comparing measurers and methods of measurement: a critical review. Clin Exp Pharmacol Physiol. 2002;29(7):527–36.CrossRef
11.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Int J Nurs Stud. 2010;47(8):931–6.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Int J Nurs Stud. 2010;47(8):931–6.CrossRef
12.
go back to reference Segato AVK, Piasecki L, Felipe Iparraguirre Nuñovero M, da Silva Neto UX, Westphalen VPD, Gambarini G, Carneiro E. The accuracy of a new cone-beam computed tomographic software in the preoperative working length determination ex vivo. J Endod. 2018;44(6):1024–9.CrossRef Segato AVK, Piasecki L, Felipe Iparraguirre Nuñovero M, da Silva Neto UX, Westphalen VPD, Gambarini G, Carneiro E. The accuracy of a new cone-beam computed tomographic software in the preoperative working length determination ex vivo. J Endod. 2018;44(6):1024–9.CrossRef
13.
go back to reference Liang Y-H, Jiang L, Chen C, Gao X-J, Wesselink PR, Wu M-K, Shemesh H. The validity of cone-beam computed tomography in measuring root canal length using a gold standard. J Endod. 2013;39(12):1607–10.CrossRef Liang Y-H, Jiang L, Chen C, Gao X-J, Wesselink PR, Wu M-K, Shemesh H. The validity of cone-beam computed tomography in measuring root canal length using a gold standard. J Endod. 2013;39(12):1607–10.CrossRef
14.
go back to reference Connert T, Hülber-J M, Godt A, Löst C, ElAyouti A. Accuracy of endodontic working length determination using cone beam computed tomography. Int Endod J. 2014;47(7):698–703.CrossRef Connert T, Hülber-J M, Godt A, Löst C, ElAyouti A. Accuracy of endodontic working length determination using cone beam computed tomography. Int Endod J. 2014;47(7):698–703.CrossRef
15.
go back to reference Tchorz JP, Wrbas KT, Von See C, Vach K, Patzelt SBM. Accuracy of software-based three-dimensional root canal length measurements using cone-beam computed tomography. Eur Endod J. 2019;4(1):28–32.PubMed Tchorz JP, Wrbas KT, Von See C, Vach K, Patzelt SBM. Accuracy of software-based three-dimensional root canal length measurements using cone-beam computed tomography. Eur Endod J. 2019;4(1):28–32.PubMed
16.
go back to reference Pham K, Phan T. Evaluation of root canal preparation using two nickel-titanium instrument systems via cone-beam computed tomography. Saudi Endod J. 2019;9(3):210–5. Pham K, Phan T. Evaluation of root canal preparation using two nickel-titanium instrument systems via cone-beam computed tomography. Saudi Endod J. 2019;9(3):210–5.
17.
go back to reference Pham K, Nguyen N. Cutting efficiency and dentinal defects using two single-file continuous rotary nickel-titanium instruments. Saudi Endod J. 2020;10(1):56–60. Pham K, Nguyen N. Cutting efficiency and dentinal defects using two single-file continuous rotary nickel-titanium instruments. Saudi Endod J. 2020;10(1):56–60.
18.
go back to reference Pham K, Le A. Evaluation of roots and canal systems of mandibular first molars in a vietnamese subpopulation using cone-beam computed tomography. J Int Soc Prevent Commun Dent. 2019;9(4):356–62.CrossRef Pham K, Le A. Evaluation of roots and canal systems of mandibular first molars in a vietnamese subpopulation using cone-beam computed tomography. J Int Soc Prevent Commun Dent. 2019;9(4):356–62.CrossRef
19.
go back to reference Gambarini G, Ropini P, Piasecki L, Costantini R, Carneiro E, Testarelli L, Dummer PMH. A preliminary assessment of a new dedicated endodontic software for use with CBCT images to evaluate the canal complexity of mandibular molars. Int Endod J. 2018;51(3):259–68.CrossRef Gambarini G, Ropini P, Piasecki L, Costantini R, Carneiro E, Testarelli L, Dummer PMH. A preliminary assessment of a new dedicated endodontic software for use with CBCT images to evaluate the canal complexity of mandibular molars. Int Endod J. 2018;51(3):259–68.CrossRef
20.
go back to reference Ludlow JB, Timothy R, Walker C, Hunter R, Benavides E, Samuelson DB, Scheske MJ. Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol. 2014;44(1):20140197.CrossRef Ludlow JB, Timothy R, Walker C, Hunter R, Benavides E, Samuelson DB, Scheske MJ. Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol. 2014;44(1):20140197.CrossRef
21.
go back to reference Fayad MI, Nair M, Levin MD, Benavides E, Rubinstein RA, Barghan S, Hirschberg CS, Ruprecht A. AAE and AAOMR joint position statement: use of cone beam computed tomography in endodontics 2015 update. Oral Surg Oral Med Oral Pathol Oral Radiol. 2015;120(4):508–12.CrossRef Fayad MI, Nair M, Levin MD, Benavides E, Rubinstein RA, Barghan S, Hirschberg CS, Ruprecht A. AAE and AAOMR joint position statement: use of cone beam computed tomography in endodontics 2015 update.  Oral Surg Oral Med Oral Pathol Oral Radiol. 2015;120(4):508–12.CrossRef
Metadata
Title
Endodontic length measurements using 3D Endo, cone-beam computed tomography, and electronic apex locator
Author
Khoa Van Pham
Publication date
01-12-2021
Publisher
BioMed Central
Published in
BMC Oral Health / Issue 1/2021
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-021-01625-w

Other articles of this Issue 1/2021

BMC Oral Health 1/2021 Go to the issue