Skip to main content
Top
Published in: European Radiology 8/2005

01-08-2005 | Magnetic Resonance

Comparison of volume, four- and eight-channel head coils using standard and parallel imaging

Authors: Elke R. Gizewski, Stefan Maderwald, Isabel Wanke, Susanne Goehde, Michael Forsting, Mark E. Ladd

Published in: European Radiology | Issue 8/2005

Login to get access

Abstract

Array coils can potentially offer increased signal-to-noise ratio (SNR) over standard coils adjacent to the array elements, while preserving the SNR at the center of the volume. The SNR advantage should theoretically increase with the number of array elements. Parallel acquisition techniques (PAT), on the other hand, can benefit acquisition times or spatial resolution at a cost to SNR as well as image quality. This study examines the question of whether SNR and image quality are still acceptable with two different array coils (four and eight channels) in conjunction with PAT when compared to standard imaging with a volume coil. All imaging was on a 1.5 T MR scanner. T2-weighted, FLAIR, diffusion-weighted, and time of flight (TOF) angiography images were performed with and without PAT in a phantom and in ten healthy volunteers. The phantom measurements demonstrated superior SNR for the eight-channel coil versus the four-channel and standard head coils. Using the eight-channel head coil for in vivo imaging, image quality with PAT (acceleration factor=2) was scored similar to images without PAT using the volume coil. The four-channel head coil suffered from inhomogeneity, lower SNR and poorer image quality when using PAT compared to standard imaging with the volume head coil. Both the in vivo and the phantom results indicate that the eight-channel head coil should be used for the highest quality brain images; this coil can be combined with PAT sequences for shorter acquisition time without a significant decrease in image quality relative to a volume coil without PAT.
Literature
1.
go back to reference Porter JR, Wright SM, Reykowski A (1998) A 16-element phased-array head coil. Magn Reson Med 40:272–279 Porter JR, Wright SM, Reykowski A (1998) A 16-element phased-array head coil. Magn Reson Med 40:272–279
2.
go back to reference Henry RG, Fischbein NJ, Dillon WP, Vigneron DB, Nelson SJ (2001) High-sensitivity coil array for head and neck imaging: technical note. Am J Neuroradiol 22:1881–1886 Henry RG, Fischbein NJ, Dillon WP, Vigneron DB, Nelson SJ (2001) High-sensitivity coil array for head and neck imaging: technical note. Am J Neuroradiol 22:1881–1886
3.
go back to reference Hadley JR, Chapman BE, Roberts JA, Chapman DC, Goodrich KC, Buswell HR, Alexander AL, Tsuruda JS, Parker DL (2000) A three-coil comparison for MR angiography. J Magn Reson Imaging 11:458–468 Hadley JR, Chapman BE, Roberts JA, Chapman DC, Goodrich KC, Buswell HR, Alexander AL, Tsuruda JS, Parker DL (2000) A three-coil comparison for MR angiography. J Magn Reson Imaging 11:458–468
4.
go back to reference Ozsarlak O, Van Goethem JW, Parizel PM (2004) 3D time-of-flight MR angiography of the intracranial vessels: optimization of the technique with water excitation, parallel acquisition, eight-channel phased-array head coil and low-dose contrast administration. Eur Radiol 14:2067–2071 Ozsarlak O, Van Goethem JW, Parizel PM (2004) 3D time-of-flight MR angiography of the intracranial vessels: optimization of the technique with water excitation, parallel acquisition, eight-channel phased-array head coil and low-dose contrast administration. Eur Radiol 14:2067–2071
5.
go back to reference Tintera J, Gawehn J, Bauermann T, Vucurevic G, Stoeter P (2004) New partially parallel acquisition technique in cerebral imaging: preliminary findings. Eur Radiol 14:2273–2281 Tintera J, Gawehn J, Bauermann T, Vucurevic G, Stoeter P (2004) New partially parallel acquisition technique in cerebral imaging: preliminary findings. Eur Radiol 14:2273–2281
6.
go back to reference Tsuchiya K, Aoki C, Fujikawa A, Hachiya J (2004) Three-dimensional MR digital subtraction angiography using parallel imaging and keyhole data sampling in cerebrovascular diseases: initial experience. Eur Radiol 14:1494–1497 Tsuchiya K, Aoki C, Fujikawa A, Hachiya J (2004) Three-dimensional MR digital subtraction angiography using parallel imaging and keyhole data sampling in cerebrovascular diseases: initial experience. Eur Radiol 14:1494–1497
7.
go back to reference Sodickson DK, McKenzie CA, Ohliger MA, Yeh EN, Price MD (2002) Recent advances in image reconstruction, coil sensitivity calibration, and coil array design for SMASH and generalized parallel MRI. Magma 13:158–163 Sodickson DK, McKenzie CA, Ohliger MA, Yeh EN, Price MD (2002) Recent advances in image reconstruction, coil sensitivity calibration, and coil array design for SMASH and generalized parallel MRI. Magma 13:158–163
8.
go back to reference Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A (2002) Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 47:1202–1210 Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A (2002) Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 47:1202–1210
9.
go back to reference Maki JH, Wilson GJ, Eubank WB, Hoogeveen RM (2002) Utilizing SENSE to achieve lower station sub-millimeter isotropic resolution and minimal venous enhancement in peripheral MR angiography. J Magn Reson Imaging 15:484–491 Maki JH, Wilson GJ, Eubank WB, Hoogeveen RM (2002) Utilizing SENSE to achieve lower station sub-millimeter isotropic resolution and minimal venous enhancement in peripheral MR angiography. J Magn Reson Imaging 15:484–491
10.
go back to reference Wang J, Kluge T, Nittka M, Jellus V, Kühn B, Kiefer B (2001) Parallel acquisition techniques with modified sense reconstruction mSense. Proceedings of the Parallel Imaging Basics and Clinical Application, Würzburg, Gemany, 1, pp 89 Wang J, Kluge T, Nittka M, Jellus V, Kühn B, Kiefer B (2001) Parallel acquisition techniques with modified sense reconstruction mSense. Proceedings of the Parallel Imaging Basics and Clinical Application, Würzburg, Gemany, 1, pp 89
11.
go back to reference Firbank MJ, Coulthard A, Harrison RM, Williams ED (1999) A comparison of two methods for measuring the signal to noise ratio on MR images. Phys Med Biol 44:N261–N264 Firbank MJ, Coulthard A, Harrison RM, Williams ED (1999) A comparison of two methods for measuring the signal to noise ratio on MR images. Phys Med Biol 44:N261–N264
12.
go back to reference Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P (1999) Sense: sensitivity encoding for fast MRI. Magn Reson Med 42:952–962 Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P (1999) Sense: sensitivity encoding for fast MRI. Magn Reson Med 42:952–962
13.
go back to reference Griswold MA, Kannengiesser S, Heidemann RM, Wang J, Jakob PM (2004) Field-of-view limitations in parallel imaging. Magn Reson Med 52:1118–1126 Griswold MA, Kannengiesser S, Heidemann RM, Wang J, Jakob PM (2004) Field-of-view limitations in parallel imaging. Magn Reson Med 52:1118–1126
14.
go back to reference Sodickson DK, Griswold MA, Jakob PM (1999) Smash imaging. Magn Reson Imaging Clin N Am 7:237–254, (vii–viii) Sodickson DK, Griswold MA, Jakob PM (1999) Smash imaging. Magn Reson Imaging Clin N Am 7:237–254, (vii–viii)
15.
go back to reference Wintersperger BJ, Nikolaou K, Dietrich O, Rieber J, Nittka M, Reiser MF, Schoenberg SO (2003) Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm. Eur Radiol 13:1931–1936 Wintersperger BJ, Nikolaou K, Dietrich O, Rieber J, Nittka M, Reiser MF, Schoenberg SO (2003) Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm. Eur Radiol 13:1931–1936
Metadata
Title
Comparison of volume, four- and eight-channel head coils using standard and parallel imaging
Authors
Elke R. Gizewski
Stefan Maderwald
Isabel Wanke
Susanne Goehde
Michael Forsting
Mark E. Ladd
Publication date
01-08-2005
Publisher
Springer-Verlag
Published in
European Radiology / Issue 8/2005
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-005-2730-0

Other articles of this Issue 8/2005

European Radiology 8/2005 Go to the issue