Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 2/2017

01-02-2017 | Cornea

Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery

Authors: Alberto López-Miguel, María Sanchidrián, Itziar Fernández, Alfredo Holgueras, Miguel J. Maldonado

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 2/2017

Login to get access

Abstract

Purpose

To compare specular microscopy (Topcon SP-3000P, Topcon Corporation, Tokyo, Japan) and ultrasound (US) technology when evaluating central corneal thickness (CCT) prior to and after phacoemulsification.

Methods

Corneal edema was assessed in phacoemulsification patients due to senile cataract by measuring CCT preoperatively and 1 day, 1 month, and 3 months postoperatively. Bland–Altman analysis was performed to assess interchangeability between pachymetry techniques for each visit. Repeated measures analysis of variance was performed to evaluate variation in CCT agreement depending on the degree of corneal edema.

Results

One hundred and eighteen patients aged 73.9 ± 10.1 years were recruited. Topcon SP-3000P provided significant (p < 0.0001) lower CCT values than US with and without induced corneal edema. Mean differences between CCT techniques for baseline, 1 day, 1 month and 3 months after cataract surgery were −28.9 ± 22.6, −30.5 ± 41.4, −32.3 ± 16.2 and −33.0 ± 16.9 μm, respectively. The systematic bias observed was not significantly different among the 4 visits (p = 0.59). The estimated limits of agreement (based on 1.96 standard deviation) were substantial, being 90.5, 165.8, 64.9 and 67.5 um at baseline, 1 day and 1 and 3 months.

Conclusions

Topcon SP-3000P provides similar systematic bias in comparison with US technique for CCT measurements regardless of the degree of corneal edema after phacoemulsification, which should be reduced after applying a constant calibration adjustment of ∼30 μm. However, interchangeability between techniques may be still limited by the notable random measurement error.
Appendix
Available only for authorised users
Literature
1.
go back to reference Viswanathan D, Goldberg I, Graham SL (2013) Relationship of change in central corneal thickness to visual field progression in eyes with glaucoma. Graefes Arch Clin Exp Ophthalmol 251:1593–1599CrossRefPubMed Viswanathan D, Goldberg I, Graham SL (2013) Relationship of change in central corneal thickness to visual field progression in eyes with glaucoma. Graefes Arch Clin Exp Ophthalmol 251:1593–1599CrossRefPubMed
2.
go back to reference Leccisotti A (2007) Corneal ectasia after photorefractive keratectomy. Graefes Arch Clin Exp Ophthalmol 245:869–875CrossRefPubMed Leccisotti A (2007) Corneal ectasia after photorefractive keratectomy. Graefes Arch Clin Exp Ophthalmol 245:869–875CrossRefPubMed
3.
go back to reference Kopplin LJ, Przepyszny K, Schmotzer B et al (2012) Relationship of Fuchs endothelial corneal dystrophy severity to central corneal thickness. Arch Ophthalmol 130:433–439CrossRefPubMed Kopplin LJ, Przepyszny K, Schmotzer B et al (2012) Relationship of Fuchs endothelial corneal dystrophy severity to central corneal thickness. Arch Ophthalmol 130:433–439CrossRefPubMed
4.
go back to reference Claesson M, Armitage WJ, Stenevi U (2009) Corneal oedema after cataract surgery: predisposing factors and corneal graft outcome. Acta Ophthalmol 87:154–159CrossRefPubMed Claesson M, Armitage WJ, Stenevi U (2009) Corneal oedema after cataract surgery: predisposing factors and corneal graft outcome. Acta Ophthalmol 87:154–159CrossRefPubMed
5.
go back to reference Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 44:367–408CrossRefPubMed Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 44:367–408CrossRefPubMed
6.
go back to reference Zhao MH, Zou J, Wang WQ et al (2007) Comparison of central corneal thickness as measured by non-contact specular microscopy and ultrasound pachymetry before and post LASIK. Clin Exp Ophthalmol 35:818–823CrossRefPubMed Zhao MH, Zou J, Wang WQ et al (2007) Comparison of central corneal thickness as measured by non-contact specular microscopy and ultrasound pachymetry before and post LASIK. Clin Exp Ophthalmol 35:818–823CrossRefPubMed
7.
go back to reference Osuagwu UL, Ogbuehi KC (2013) Evaluation of the comparative effect of tetracaine on central corneal thickness measured by a contact and noncontact pachymeter. J Ocul Pharmacol Ther 29:68–74CrossRefPubMed Osuagwu UL, Ogbuehi KC (2013) Evaluation of the comparative effect of tetracaine on central corneal thickness measured by a contact and noncontact pachymeter. J Ocul Pharmacol Ther 29:68–74CrossRefPubMed
8.
go back to reference Bao F, Wang Q, Cheng S et al (2014) Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices. Cornea 33:576–581CrossRefPubMed Bao F, Wang Q, Cheng S et al (2014) Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices. Cornea 33:576–581CrossRefPubMed
9.
go back to reference Kuerten D, Plange N, Koch EC et al (2015) Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT. Graefes Arch Clin Exp Ophthalmol 253:1105–1109CrossRefPubMed Kuerten D, Plange N, Koch EC et al (2015) Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT. Graefes Arch Clin Exp Ophthalmol 253:1105–1109CrossRefPubMed
10.
go back to reference Price MO, Gorovoy M, Price FW Jr et al (2013) Descemet’s stripping automated endothelial keratoplasty: three-year graft and endothelial cell survival compared with penetrating keratoplasty. Ophthalmology 120:246–251CrossRefPubMed Price MO, Gorovoy M, Price FW Jr et al (2013) Descemet’s stripping automated endothelial keratoplasty: three-year graft and endothelial cell survival compared with penetrating keratoplasty. Ophthalmology 120:246–251CrossRefPubMed
11.
go back to reference Módis L Jr, Langenbucher A, Seitz B (2001) Scanning-slit and specular microscopic pachymetry in comparison with ultrasonic determination of corneal thickness. Cornea 20:711–714CrossRefPubMed Módis L Jr, Langenbucher A, Seitz B (2001) Scanning-slit and specular microscopic pachymetry in comparison with ultrasonic determination of corneal thickness. Cornea 20:711–714CrossRefPubMed
12.
go back to reference Bovelle R, Kaufman SC, Thompson HW et al (1999) Corneal thickness measurements with the Topcon SP-2000P specular microscope and an ultrasound pachymeter. Arch Ophthalmol 117:868–870CrossRefPubMed Bovelle R, Kaufman SC, Thompson HW et al (1999) Corneal thickness measurements with the Topcon SP-2000P specular microscope and an ultrasound pachymeter. Arch Ophthalmol 117:868–870CrossRefPubMed
13.
go back to reference Almubrad TM, Osuagwu UL, Alabbadi I et al (2011) Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter. Clin Ophthalmol 5:871–876CrossRefPubMedPubMedCentral Almubrad TM, Osuagwu UL, Alabbadi I et al (2011) Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter. Clin Ophthalmol 5:871–876CrossRefPubMedPubMedCentral
14.
go back to reference Tai LY, Khaw KW, Ng CM et al (2013) Central corneal thickness measurements with different imaging devices and ultrasound pachymetry. Cornea 32:766–771CrossRefPubMed Tai LY, Khaw KW, Ng CM et al (2013) Central corneal thickness measurements with different imaging devices and ultrasound pachymetry. Cornea 32:766–771CrossRefPubMed
15.
go back to reference Khaja WA, Grover S, Kelmenson AT et al (2015) Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan. Clin Ophthalmol 9:1065–1070PubMedPubMedCentral Khaja WA, Grover S, Kelmenson AT et al (2015) Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan. Clin Ophthalmol 9:1065–1070PubMedPubMedCentral
16.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed
17.
go back to reference McAlinden C, Khadka J, Pesudovs K (2011) Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Opt 31:330–338CrossRefPubMed McAlinden C, Khadka J, Pesudovs K (2011) Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Opt 31:330–338CrossRefPubMed
18.
go back to reference López-Miguel A, Martínez-Almeída L, Coco-Martín MB et al (2014) Clinical utility of combined Placido-scanning-slit midperipheral and thinnest point pachymetry after corneal ablation for myopia. Cornea 33:266–270CrossRefPubMed López-Miguel A, Martínez-Almeída L, Coco-Martín MB et al (2014) Clinical utility of combined Placido-scanning-slit midperipheral and thinnest point pachymetry after corneal ablation for myopia. Cornea 33:266–270CrossRefPubMed
19.
go back to reference Gordon MO, Beiser JA, Brandt JD et al (2002) The ocular hypertension treatment study; baseline factors that predict the onset of primary open-angle glaucoma; for the Ocular Hypertension Treatment Study Group. Arch Ophthalmol 120:714–720CrossRefPubMed Gordon MO, Beiser JA, Brandt JD et al (2002) The ocular hypertension treatment study; baseline factors that predict the onset of primary open-angle glaucoma; for the Ocular Hypertension Treatment Study Group. Arch Ophthalmol 120:714–720CrossRefPubMed
20.
go back to reference Ogbuehi KC, Almubrad TM (2005) Repeatability of central corneal thickness measurements measured with the Topcon SP2000P specular microscope. Graefes Arch Clin Exp Ophthalmol 243:798–802CrossRefPubMed Ogbuehi KC, Almubrad TM (2005) Repeatability of central corneal thickness measurements measured with the Topcon SP2000P specular microscope. Graefes Arch Clin Exp Ophthalmol 243:798–802CrossRefPubMed
21.
go back to reference López-Miguel A, Martínez-Almeida L, Mateo ME et al (2013) Dependability of pachymetry measurements after myopic advanced surface ablation using scanning-slit topography and specular microscopy. Invest Ophthalmol Vis Sci 54:1054–1060CrossRefPubMed López-Miguel A, Martínez-Almeida L, Mateo ME et al (2013) Dependability of pachymetry measurements after myopic advanced surface ablation using scanning-slit topography and specular microscopy. Invest Ophthalmol Vis Sci 54:1054–1060CrossRefPubMed
22.
go back to reference Ogbuehi KC, Osuagwu UL (2012) Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness. Clin Ophthalmol 6:761–769PubMedPubMedCentral Ogbuehi KC, Osuagwu UL (2012) Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness. Clin Ophthalmol 6:761–769PubMedPubMedCentral
23.
go back to reference Wong MM, Shukla AN, Munir WM (2014) Correlation of corneal thickness and volume with intraoperative phacoemulsification parameters using Scheimpflug imaging and optical coherence tomography. J Cataract Refract Surg 40:2067–2075CrossRefPubMed Wong MM, Shukla AN, Munir WM (2014) Correlation of corneal thickness and volume with intraoperative phacoemulsification parameters using Scheimpflug imaging and optical coherence tomography. J Cataract Refract Surg 40:2067–2075CrossRefPubMed
24.
go back to reference Ljubimov AV, Atilano SR, Garner MH et al (2002) Extracellular matrix and Na+, K + −ATPase in human corneas following cataract surgery: comparison with bullous keratopathy and Fuchs’ dystrophy corneas. Cornea 21:74–80CrossRefPubMed Ljubimov AV, Atilano SR, Garner MH et al (2002) Extracellular matrix and Na+, K + −ATPase in human corneas following cataract surgery: comparison with bullous keratopathy and Fuchs’ dystrophy corneas. Cornea 21:74–80CrossRefPubMed
Metadata
Title
Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery
Authors
Alberto López-Miguel
María Sanchidrián
Itziar Fernández
Alfredo Holgueras
Miguel J. Maldonado
Publication date
01-02-2017
Publisher
Springer Berlin Heidelberg
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 2/2017
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-016-3537-1

Other articles of this Issue 2/2017

Graefe's Archive for Clinical and Experimental Ophthalmology 2/2017 Go to the issue