01-07-2015 | Cornea

Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT

Authors: David Kuerten, Niklas Plange, Eva Charlotte Koch, Antonis Koutsonas, Peter Walter, Matthias Fuest

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 7/2015

Login to get access

Abstract

Purpose

The purpose of this study is to determine the influence of post-surgical corneal edema on the reliability and reproducibility of central corneal thickness (CCT) measurements by a Scheimpflug camera (Pentacam), ultrasound pachymetry (USP), and anterior-segment spectral-domain optical coherence tomography (AS-OCT).

Methods

Thirty-two patients planned for cataract surgery (n = 16) or vitrectomy (n =  6) were included in a prospective study. The non-surgery eye was used as control. Two investigators acquired two measurements each, with the Pentacam (Oculus, Germany) and the AS-OCT (Heidelberg Engineering, Germany) in a randomized order, followed by USP (Tomey SP-100, Germany). CCT was evaluated using the apex value for Pentacam, the corneal apex cut in AS–OCT and averaging eight single measurements for USP. Coefficients of variation (COV) and intra-class correlation coefficients (ICC) were determined.

Results

Post-surgery corneas showed a thickness of (investigators 1 and 2): Pentacam (615.9 ± 58.02 μm and 615.1 ± 60.17 μm), USP (601.4 ± 63.77 μm and 614.5 ± 70.91 μm), AS-OCT (608.8 ± 65.67 μm and 606.9 ± 64.41 μm) ,with no significant difference (ANOVA p > 0.99). The COVs (investigators 1 and 2) for control eyes were: Pentacam (0.78 ± 0.52 and 0.70 ± 0.76), USP (0.66 ± 0.29 and 0.98 ± 0.44), AS-OCT (0.59 ± 0.61 and 0.59 ± 0.40). The COVs (investigators 1 and 2) for post-surgical eyes were: Pentacam (0.98 ± 1.25 and 0.97 ± 0.73), USP (0.73 ± 0.64 and 1.35 ± 0.85), AS-OCT (1.34 ± 1.57 and 1.19 ± 1.18).The ICC was determined in post-surgery corneas (ICC > 0.96) and control corneas (ICC > 0.95).

Conclusion

USP measurements have the highest user dependence. Post-surgical corneal edema leads to higher intraobserver variability. All methods reached a high level of agreement in CCT determination in edematous as well as healthy corneas.
Literature
1.
go back to reference Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK 2nd, Wilson MR, Kass MA (2002) The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 120(6):714–720, discussion 829–730PubMedCrossRef Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK 2nd, Wilson MR, Kass MA (2002) The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 120(6):714–720, discussion 829–730PubMedCrossRef
3.
4.
go back to reference Martin R, de Juan V, Rodriguez G, Cuadrado R, Fernandez I (2007) Measurement of corneal swelling variations without removal of the contact lens during extended wear. Invest Ophthalmol Vis Sci 48(7):3043–3050. doi:10.1167/iovs.06-1372 PubMedCrossRef Martin R, de Juan V, Rodriguez G, Cuadrado R, Fernandez I (2007) Measurement of corneal swelling variations without removal of the contact lens during extended wear. Invest Ophthalmol Vis Sci 48(7):3043–3050. doi:10.​1167/​iovs.​06-1372 PubMedCrossRef
9.
go back to reference Miglior S, Albe E, Guareschi M, Mandelli G, Gomarasca S, Orzalesi N (2004) Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness. Br J Ophthalmol 88(2):174–177PubMedCentralPubMedCrossRef Miglior S, Albe E, Guareschi M, Mandelli G, Gomarasca S, Orzalesi N (2004) Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness. Br J Ophthalmol 88(2):174–177PubMedCentralPubMedCrossRef
11.
go back to reference de Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM (2007) Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 144(5):712–718. doi:10.1016/j.ajo.2007.07.021 PubMedCrossRef de Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM (2007) Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 144(5):712–718. doi:10.​1016/​j.​ajo.​2007.​07.​021 PubMedCrossRef
14.
go back to reference Hiraoka M, Amano S, Oshika T, Kato S, Hori S (2001) Factors contributing to corneal complications after vitrectomy in diabetic patients. Jpn J Ophthalmol 45(5):492–495PubMedCrossRef Hiraoka M, Amano S, Oshika T, Kato S, Hori S (2001) Factors contributing to corneal complications after vitrectomy in diabetic patients. Jpn J Ophthalmol 45(5):492–495PubMedCrossRef
15.
go back to reference du Toit R, Vega JA, Fonn D, Simpson T (2003) Diurnal variation of corneal sensitivity and thickness. Cornea 22(3):205–209PubMedCrossRef du Toit R, Vega JA, Fonn D, Simpson T (2003) Diurnal variation of corneal sensitivity and thickness. Cornea 22(3):205–209PubMedCrossRef
16.
go back to reference Zadnik K, Mutti DO, Adams AJ (1992) The repeatability of measurement of the ocular components. Invest Ophthalmol Vis Sci 33(7):2325–2333PubMed Zadnik K, Mutti DO, Adams AJ (1992) The repeatability of measurement of the ocular components. Invest Ophthalmol Vis Sci 33(7):2325–2333PubMed
18.
go back to reference Li EY, Mohamed S, Leung CK, Rao SK, Cheng AC, Cheung CY, Lam DS (2007) Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography. Ophthalmology 114(10):1842–1847. doi:10.1016/j.ophtha.2007.02.017 PubMedCrossRef Li EY, Mohamed S, Leung CK, Rao SK, Cheng AC, Cheung CY, Lam DS (2007) Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography. Ophthalmology 114(10):1842–1847. doi:10.​1016/​j.​ophtha.​2007.​02.​017 PubMedCrossRef
19.
22.
go back to reference Wu W, Wang Y, Xu L (2014) Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes. Graefes Arch Clin Exp Ophthalmol 252(1):91–99. doi:10.1007/s00417-013-2502-5 PubMedCrossRef Wu W, Wang Y, Xu L (2014) Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes. Graefes Arch Clin Exp Ophthalmol 252(1):91–99. doi:10.​1007/​s00417-013-2502-5 PubMedCrossRef
23.
go back to reference Pierro L, Conforto E, Resti AG, Lattanzio R (1998) High-frequency ultrasound biomicroscopy versus ultrasound and optical pachymetry for the measurement of corneal thickness. Ophthalmologica 212(1):1–3PubMedCrossRef Pierro L, Conforto E, Resti AG, Lattanzio R (1998) High-frequency ultrasound biomicroscopy versus ultrasound and optical pachymetry for the measurement of corneal thickness. Ophthalmologica 212(1):1–3PubMedCrossRef
24.
go back to reference Marsich MW, Bullimore MA (2000) The repeatability of corneal thickness measures. Cornea 19(6):792–795PubMedCrossRef Marsich MW, Bullimore MA (2000) The repeatability of corneal thickness measures. Cornea 19(6):792–795PubMedCrossRef
25.
go back to reference Mertz GW (1980) Overnight swelling of the living human cornea. J Am Optom Assoc 51(3):211–214PubMed Mertz GW (1980) Overnight swelling of the living human cornea. J Am Optom Assoc 51(3):211–214PubMed
26.
go back to reference Kotecha A, Crabb DP, Spratt A, Garway-Heath DF (2009) The relationship between diurnal variations in intraocular pressure measurements and central corneal thickness and corneal hysteresis. Invest Ophthalmol Vis Sci 50(9):4229–4236. doi:10.1167/iovs.08-2955 PubMedCrossRef Kotecha A, Crabb DP, Spratt A, Garway-Heath DF (2009) The relationship between diurnal variations in intraocular pressure measurements and central corneal thickness and corneal hysteresis. Invest Ophthalmol Vis Sci 50(9):4229–4236. doi:10.​1167/​iovs.​08-2955 PubMedCrossRef
27.
go back to reference Wheeler NC, Morantes CM, Kristensen RM, Pettit TH, Lee DA (1992) Reliability coefficients of three corneal pachymeters. Am J Ophthalmol 113(6):645–651PubMedCrossRef Wheeler NC, Morantes CM, Kristensen RM, Pettit TH, Lee DA (1992) Reliability coefficients of three corneal pachymeters. Am J Ophthalmol 113(6):645–651PubMedCrossRef
28.
go back to reference Iester M, Telani S, Frezzotti P, Manni G, Uva M, Figus M, Perdicchi A (2012) Differences in central corneal thickness between the paired eyes and the severity of the glaucomatous damage. Eye (Lond) 26(11):1424–1430. doi:10.1038/eye.2012.179 CrossRef Iester M, Telani S, Frezzotti P, Manni G, Uva M, Figus M, Perdicchi A (2012) Differences in central corneal thickness between the paired eyes and the severity of the glaucomatous damage. Eye (Lond) 26(11):1424–1430. doi:10.​1038/​eye.​2012.​179 CrossRef
Metadata
Title
Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT
Authors
David Kuerten
Niklas Plange
Eva Charlotte Koch
Antonis Koutsonas
Peter Walter
Matthias Fuest
Publication date
01-07-2015
Publisher
Springer Berlin Heidelberg
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 7/2015
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-015-2998-y

Other articles of this Issue 7/2015

Graefe's Archive for Clinical and Experimental Ophthalmology 7/2015 Go to the issue