Abstract

Purpose

The 2010 American Heart Association or European Resuscitation Council guidelines for cardiopulmonary resuscitation emphasize that rescuers should minimize interruption of chest compressions, even for endotracheal intubation. Cervical stabilization should also be maintained during traumatic cardiac arrest. The utility of the Pentax-AWS Airwayscope (AWS) video laryngoscope and Airtraq (ATQ) optic laryngoscope for airway management has been reported under cervical stabilization. We first evaluated ATQ utility during chest compression with or without cervical stabilization and then compared the AWS, ATQ, and Macintosh laryngoscope (McL) during chest compressions under cervical stabilization in a manikin.

Methods

In the first trial, 19 novice doctors performed tracheal intubation with ATQ during chest compression with or without cervical stabilization. In the second trial, 21 novice doctors performed tracheal intubation on a manikin with cervical stabilization using AWS, ATQ, and McL with or without chest compression in a manikin. The rate of successful intubation, time to intubation, and subjective difficulty of use (visual analog scale) were recorded.

Results

In the first trial, intubation time during chest compression was significantly shortened under cervical stabilization compared to without cervical stabilization (P < 0.05). In the second trial, using McL, 3 participants failed to perform tracheal intubation without chest compression and 11 failed during chest compression (P < 0.05). Using ATQ, all intubations were successful without chest compression, but 5 failed during chest compression (P < 0.05). Intubation time was significantly prolonged by chest compression using McL or ATQ (P < 0.05). All participants successfully secured the airway with AWS regardless of chest compression, and chest compression did not prolong intubation time. Chest compression worsened the score on the visual analog scale of laryngoscopy in the McL trial (P < 0.05), but not in ATQ or AWS trials. Difficulty of tube passage through the glottis increased with chest compression with the McL and ATQ (P < 0.05) but not with AWS.

Conclusion

The AWS was superior to McL and ATQ for endotracheal intubation during simulated cervical stabilization and chest compression.
Literature
1.
go back to reference Nolan JP, Soar J, Zideman DA, Biarent D, Bossaert LL, Deakin C, Koster RW, Wyllie J, Böttiger B. European Resuscitation Council Guidelines for Resuscitation 2010: Section 1. Executive summary. Resuscitation. 2010;81:1219–76.PubMedCrossRef Nolan JP, Soar J, Zideman DA, Biarent D, Bossaert LL, Deakin C, Koster RW, Wyllie J, Böttiger B. European Resuscitation Council Guidelines for Resuscitation 2010: Section 1. Executive summary. Resuscitation. 2010;81:1219–76.PubMedCrossRef
2.
go back to reference Hazinski MF, Nolan JP, Billi JE, Böttiger BW, Bossaert L, de Caen AR, Deakin CD, Drajer S, Eigel B, Hickey RW, Jacobs I, Kleinman ME, Kloeck W, Koster RW, Lim SH, Mancini ME, Montgomery WH, Morley PT, Morrison LJ, Nadkarni VM, O’Connor RE, Okada K, Perlman JM, Sayre MR, Shuster M, Soar J, Sunde K, Travers AH, Wyllie J, Zideman D. Part 1: executive summary: 2010 international consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Circulation. 2010;122:S250–75.PubMedCrossRef Hazinski MF, Nolan JP, Billi JE, Böttiger BW, Bossaert L, de Caen AR, Deakin CD, Drajer S, Eigel B, Hickey RW, Jacobs I, Kleinman ME, Kloeck W, Koster RW, Lim SH, Mancini ME, Montgomery WH, Morley PT, Morrison LJ, Nadkarni VM, O’Connor RE, Okada K, Perlman JM, Sayre MR, Shuster M, Soar J, Sunde K, Travers AH, Wyllie J, Zideman D. Part 1: executive summary: 2010 international consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Circulation. 2010;122:S250–75.PubMedCrossRef
3.
go back to reference Langeron O, Birenbaum A, Amour J. Airway management in trauma. Minerva Anesthesiol. 2009;75:307–11. Langeron O, Birenbaum A, Amour J. Airway management in trauma. Minerva Anesthesiol. 2009;75:307–11.
4.
go back to reference Dupanovic M, Fox H, Kovac A. Management of the airway in multitrauma. Curr Opin Anaesthesiol. 2010;23:276–82.PubMedCrossRef Dupanovic M, Fox H, Kovac A. Management of the airway in multitrauma. Curr Opin Anaesthesiol. 2010;23:276–82.PubMedCrossRef
5.
go back to reference Davis DP, Peay J, Sise MJ, Kennedy F, Simon F, Tominaga G, Steele J, Coimbra R. Prehospital airway and ventilation management: a trauma score and injury severity score-based analysis. J Trauma. 2010;69:294–301.PubMedCrossRef Davis DP, Peay J, Sise MJ, Kennedy F, Simon F, Tominaga G, Steele J, Coimbra R. Prehospital airway and ventilation management: a trauma score and injury severity score-based analysis. J Trauma. 2010;69:294–301.PubMedCrossRef
6.
go back to reference Maharaj CH, Buckley E, Harte BH, Laffey JG. Endotracheal intubation in patients with cervical spine immobilization: a comparison of Macintosh and Airtraq™ laryngoscopes. Anesthesiology. 2007;107:53–9.PubMedCrossRef Maharaj CH, Buckley E, Harte BH, Laffey JG. Endotracheal intubation in patients with cervical spine immobilization: a comparison of Macintosh and Airtraq™ laryngoscopes. Anesthesiology. 2007;107:53–9.PubMedCrossRef
7.
go back to reference Enomoto Y, Asai T, Arai T, Kamishima K, Okuda Y. Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study. Br J Anaesth. 2008;100:544–8.PubMedCrossRef Enomoto Y, Asai T, Arai T, Kamishima K, Okuda Y. Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study. Br J Anaesth. 2008;100:544–8.PubMedCrossRef
8.
go back to reference Sadamori T, Kusunoki S, Ishida M, Otani T, Tanigawa K. Video laryngoscopy for emergency tracheal intubation during chest compression. Resuscitation. 2008;77:155–6.PubMedCrossRef Sadamori T, Kusunoki S, Ishida M, Otani T, Tanigawa K. Video laryngoscopy for emergency tracheal intubation during chest compression. Resuscitation. 2008;77:155–6.PubMedCrossRef
9.
go back to reference Corso RM, Piraccini E, Agnoletti V, Gambale G. The Airtraq laryngoscope for emergency tracheal intubation without interruption of chest compression. Am J Emerg Med. 2010;28:971–2.PubMedCrossRef Corso RM, Piraccini E, Agnoletti V, Gambale G. The Airtraq laryngoscope for emergency tracheal intubation without interruption of chest compression. Am J Emerg Med. 2010;28:971–2.PubMedCrossRef
10.
go back to reference Komasawa N, Ueki R, Nomura H, Itani M, Kaminoh Y. Comparison of tracheal intubation by Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study. J Anesth. 2010;24:306–8.PubMedCrossRef Komasawa N, Ueki R, Nomura H, Itani M, Kaminoh Y. Comparison of tracheal intubation by Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study. J Anesth. 2010;24:306–8.PubMedCrossRef
11.
go back to reference Maruyama K, Tsukamoto S, Ohno S, Kobayashi K, Nakagawa H, Kitamura A, Hayashida M. Effect of cardiopulmonary resuscitation on intubation using a Macintosh laryngoscope, the AirWay Scope, and the gum elastic bougie: a manikin study. Resuscitation. 2010;81:1014–8.PubMedCrossRef Maruyama K, Tsukamoto S, Ohno S, Kobayashi K, Nakagawa H, Kitamura A, Hayashida M. Effect of cardiopulmonary resuscitation on intubation using a Macintosh laryngoscope, the AirWay Scope, and the gum elastic bougie: a manikin study. Resuscitation. 2010;81:1014–8.PubMedCrossRef
12.
go back to reference Koyama J, Iwashita T, Okamoto K. Comparison of three types of laryngoscope for tracheal intubation during rhythmic chest compressions: a manikin study. Resuscitation. 2010;81:1172–4.PubMedCrossRef Koyama J, Iwashita T, Okamoto K. Comparison of three types of laryngoscope for tracheal intubation during rhythmic chest compressions: a manikin study. Resuscitation. 2010;81:1172–4.PubMedCrossRef
13.
go back to reference Komasawa N, Ueki R, Itani M, Nishi S, Kaminoh Y. Validation of Pentax-AWS Airwayscope utility for intubation device during cardiopulmonary resuscitation on the ground. J Anesth. 2010;24:582–6.PubMedCrossRef Komasawa N, Ueki R, Itani M, Nishi S, Kaminoh Y. Validation of Pentax-AWS Airwayscope utility for intubation device during cardiopulmonary resuscitation on the ground. J Anesth. 2010;24:582–6.PubMedCrossRef
14.
go back to reference Neumar RW, Otto CW, Link MS, Kronick SL, Shuster M, Callaway CW, Kudenchuk PJ, Ornato JP, McNally B, Silvers SM, Passman RS, White RD, Hess EP, Tang W, Davis D, Sinz E, Morrison LJ. Part 8: Adult advanced cardiovascular life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122:S729–67.PubMedCrossRef Neumar RW, Otto CW, Link MS, Kronick SL, Shuster M, Callaway CW, Kudenchuk PJ, Ornato JP, McNally B, Silvers SM, Passman RS, White RD, Hess EP, Tang W, Davis D, Sinz E, Morrison LJ. Part 8: Adult advanced cardiovascular life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122:S729–67.PubMedCrossRef
15.
go back to reference Katz SH, Falk JL. Misplaced endotracheal tubes by paramedics in an urban emergency medical services system. Ann Emerg Med. 2001;37:32–7.PubMedCrossRef Katz SH, Falk JL. Misplaced endotracheal tubes by paramedics in an urban emergency medical services system. Ann Emerg Med. 2001;37:32–7.PubMedCrossRef
16.
go back to reference Helm M, Hossfeld B, Schafer S, Hoitz J, Lampl L. Factors influencing emergency intubation in the pre-hospital setting: a multicentre study in the German Helicopter Emergency Medical Service. Br J Anaesth. 2006;96:67–71.PubMedCrossRef Helm M, Hossfeld B, Schafer S, Hoitz J, Lampl L. Factors influencing emergency intubation in the pre-hospital setting: a multicentre study in the German Helicopter Emergency Medical Service. Br J Anaesth. 2006;96:67–71.PubMedCrossRef
17.
go back to reference Cobas MA, De la Peña MA, Manning R, Candiotti K, Varon AJ. Prehospital intubations and mortality: a level 1 trauma center perspective. Anesth Analg. 2009;109:489–93.PubMedCrossRef Cobas MA, De la Peña MA, Manning R, Candiotti K, Varon AJ. Prehospital intubations and mortality: a level 1 trauma center perspective. Anesth Analg. 2009;109:489–93.PubMedCrossRef
18.
go back to reference Sawin PD, Todd MM, Traynelis VC, Farrell SB, Nader A, Sato Y, Clausen JD, Goel VK. Cervical spine motion with direct laryngoscopy and orotracheal intubation: an in vivo cinefluoroscopic study of subjects without cervical abnormality. Anesthesiology. 1996;85:26–36.PubMedCrossRef Sawin PD, Todd MM, Traynelis VC, Farrell SB, Nader A, Sato Y, Clausen JD, Goel VK. Cervical spine motion with direct laryngoscopy and orotracheal intubation: an in vivo cinefluoroscopic study of subjects without cervical abnormality. Anesthesiology. 1996;85:26–36.PubMedCrossRef
19.
go back to reference Maruyama K, Yamada T, Hara K, Nakagawa H, Kitamura A. Tracheal intubation using an AirWay Scope in a patient with Halo-Vest Fixation for upper cervical spine injury. Br J Anaesth. 2009;102:565–6.PubMedCrossRef Maruyama K, Yamada T, Hara K, Nakagawa H, Kitamura A. Tracheal intubation using an AirWay Scope in a patient with Halo-Vest Fixation for upper cervical spine injury. Br J Anaesth. 2009;102:565–6.PubMedCrossRef
20.
go back to reference Maharaj CH, Higgins B, Harte BH, Laffey JG. Evaluation of ease of intubation with the Airtraq or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy: a manikin study. Anaesthesia. 2006;61:469–77.PubMedCrossRef Maharaj CH, Higgins B, Harte BH, Laffey JG. Evaluation of ease of intubation with the Airtraq or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy: a manikin study. Anaesthesia. 2006;61:469–77.PubMedCrossRef
21.
go back to reference Suzuki A, Toyama Y, Katsumi N, Kunisawa T, Sasaki R, Hirota K, Henderson JJ, Iwasaki H. The Pentax-AWS rigid indirect video laryngoscope: clinical assessment of performance in 320 cases. Anaesthesia. 2008;63:641–7.PubMedCrossRef Suzuki A, Toyama Y, Katsumi N, Kunisawa T, Sasaki R, Hirota K, Henderson JJ, Iwasaki H. The Pentax-AWS rigid indirect video laryngoscope: clinical assessment of performance in 320 cases. Anaesthesia. 2008;63:641–7.PubMedCrossRef
22.
go back to reference Koyama J, Aoyama T, Kusano Y, Seguchi T, Kawagishi K, Iwashita T, Okamoto K, Okudera H, Takasuna H, Hongo K. Description and first clinical application of Airway Scope for tracheal intubation. J Neurosurg Anesthesiol. 2006;18:247–50.PubMedCrossRef Koyama J, Aoyama T, Kusano Y, Seguchi T, Kawagishi K, Iwashita T, Okamoto K, Okudera H, Takasuna H, Hongo K. Description and first clinical application of Airway Scope for tracheal intubation. J Neurosurg Anesthesiol. 2006;18:247–50.PubMedCrossRef
23.
go back to reference Asai T, Liu EH, Matsumoto S, Hirabayashi Y, Seo N, Suzuki A, Toi T, Yasumoto K, Okuda Y. Use of the Pentax-AWS in 293 patients with difficult airways. Anesthesiology. 2009;110:898–904.PubMedCrossRef Asai T, Liu EH, Matsumoto S, Hirabayashi Y, Seo N, Suzuki A, Toi T, Yasumoto K, Okuda Y. Use of the Pentax-AWS in 293 patients with difficult airways. Anesthesiology. 2009;110:898–904.PubMedCrossRef
24.
go back to reference Liu L, Tanigawa K, Kusunoki S, Tamura T, Ota K, Yamaga S, Kida Y, Otani T, Sadamori T, Takeda T, Iwasaki Y, Hirohashi N. Tracheal intubation of a difficult airway using Airway Scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel. Anesth Analg. 2010;110:1049–55.PubMedCrossRef Liu L, Tanigawa K, Kusunoki S, Tamura T, Ota K, Yamaga S, Kida Y, Otani T, Sadamori T, Takeda T, Iwasaki Y, Hirohashi N. Tracheal intubation of a difficult airway using Airway Scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel. Anesth Analg. 2010;110:1049–55.PubMedCrossRef
25.
go back to reference Hirabayashi Y, Seo N. Tracheal intubation by non-anesthesia residents using the Pentax-AWS airway scope and Macintosh laryngoscope. J Clin Anesth. 2009;21:268–71.PubMedCrossRef Hirabayashi Y, Seo N. Tracheal intubation by non-anesthesia residents using the Pentax-AWS airway scope and Macintosh laryngoscope. J Clin Anesth. 2009;21:268–71.PubMedCrossRef
26.
go back to reference Hirabayashi Y. Airway Scope versus Macintosh laryngoscope: a manikin study. Emerg Med J. 2007;24:357–8.PubMedCrossRef Hirabayashi Y. Airway Scope versus Macintosh laryngoscope: a manikin study. Emerg Med J. 2007;24:357–8.PubMedCrossRef
27.
go back to reference Jordan GM, Silsby J, Bayley G, Cook TM, Difficult Airway S. Evaluation of four manikins as simulators for teaching airway management procedures specified in the Difficult Airway Society guidelines, and other advanced airway skills. Anaesthesia. 2007;62:708–12.PubMedCrossRef Jordan GM, Silsby J, Bayley G, Cook TM, Difficult Airway S. Evaluation of four manikins as simulators for teaching airway management procedures specified in the Difficult Airway Society guidelines, and other advanced airway skills. Anaesthesia. 2007;62:708–12.PubMedCrossRef
Metadata
Title
Comparison of Pentax-AWS Airwayscope video laryngoscope, Airtraq optic laryngoscope, and Macintosh laryngoscope during cardiopulmonary resuscitation under cervical stabilization: a manikin study
Authors
Nobuyasu Komasawa
Ryusuke Ueki
Hanako Kohama
Shin-ichi Nishi
Yoshiroh Kaminoh
Publication date
01-12-2011
Publisher
Springer Japan
Published in
Journal of Anesthesia / Issue 6/2011
Print ISSN: 0913-8668
Electronic ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-011-1218-0

Other articles of this Issue 6/2011

Journal of Anesthesia 6/2011 Go to the issue